IAEA SAFETY STANDARDS
for protecting people and the environment
Draft version 2.5
Step 8. For Member States’ review and comments
Deadine for Comments: 30 April 2015
Radiation Protection and Safety
in Medical Uses of Ionizing Radiation
Proposed Joint Sponsors:
International Atomic Energy Agency, World Health Organization,
Pan American Health Organization and International Labour Office
DRAFT SAFETY GUIDE
DS399
FOREWORD
By Yukiya Amano
Director General
[standard text to be added]
PREFACE
In 2006, the Agency published the Fundamental Safety Principles (IAEA Safety
Standards Series No.
SF-1), jointly sponsored by the European Atomic Energy Community
(EURATOM), the Food and
Agriculture Organization of the United Nations (FAO), the IAEA,
the International Labour
Organization (ILO), the International Maritime Organization, the OECD
Nuclear Energy Agency
(OECD/NEA), the Pan American Health Organization (PAHO), the United
Nations Environment
Programme (UNEP) and the World Health Organization (WHO). That
publication sets out the
fundamental safety objective and the principles of protection and safety.
Requirements designed to
meet these are established in Radiation Protection and Safety of
Radiation Sources: International
Basic Safety Standards (the BSS), jointly sponsored by the European
Commission (EC/Euratom),
Food and Agriculture Organization of the United Nations (FAO), IAEA,
International Labour
Organization (ILO), OECD Nuclear Energy Agency (OECD/NEA), Pan
American Health
Organization (PAHO), United Nations Environment Programme (UNEP)
and World Health
Organization (WHO).
This Safety Guide, prepared jointly by the IAEA, the WHO, the PAHO and the
International Labour
Office, provides guidance on fulfilling the requirements of GSR Part 3 with
respect to medical uses of
ionizing radiation. It is aimed primarily at end-users in medical radiation
facilities where radiological
procedures are performed, including management, radiological
medical practitioners, medical
radiation technologists, medical physicists, radiation protection
officers and other health
professionals. It also provides recommendations and guidance to health
professionals who refer
patients for radiological procedures; to manufacturers and
suppliers of medical radiological
equipment; and to ethics committees with responsibilities for biomedical
research. This publication
provides recommendations and guidance on appropriate regulatory activities and
infrastructure, and is
therefore also applicable to regulatory bodies, health authorities, government
agencies in general, and
professional bodies.
The Safety Guide addresses all three categories of exposure:
occupational exposure for health
professionals performing radiological procedures; medical exposure,
primarily for the patients
undergoing the radiological procedures but also for carers and comforters and
for volunteers subject
to exposure as part of a programme of medical research; and public
exposure for members of the
public. A systematic approach should be applied to ensure that there is a
balance between being able
to utilize the benefits from medical uses of ionizing radiation and
minimizing the risk of radiation
effects to patients, workers and members of the public.
This Safety Guide supersedes Safety Guide Radiological Protection for Medical
Exposure to Ionizing
Radiation, RS-G-1.5, Safety Reports Series No. 38 Applying
Radiation Safety Standards in
Radiotherapy, Safety Reports Series No. 39 Applying Radiation Safety
Standards in Diagnostic
Radiology and Interventional Procedures Using X Rays, and Safety Reports Series
No. 40 Applying
Radiation Safety Standards in Nuclear Medicine.
CONTENTS
1. INTRODUCTION
10
BACKGROUND
10
SCOPE
11
STRUCTURE
11
2. GENERAL RECOMMENDATIONS FOR RADIATION PROTECTION AND SAFETY IN
MEDICAL USES OF RADIATION
12
GENERAL ASPECTS
12
TYPES OF EXPOSURE SITUATIONS AND CATEGORIES OF EXPOSURE
12
APPLICATION OF THE RADIATION PROTECTION REQUIREMENTS
13
Justification
14
Optimization of protection and safety
15
Dose limits
16
GRADED APPROACH
17
ROLES AND RESPONSIBILITIES
17
Government
17
Health authority
22
Professional bodies
24
Regulatory body
25
Medical radiation facility
29
Ethics committees
31
Suppliers of sources, equipment and software
32
Maintenance and servicing organizations
33
Referring medical practitioners
34
Patients
34
EDUCATION, TRAINING, QUALIFICATION AND COMPETENCE
35
Radiological medical practitioners
36
Medical radiation technologists
37
Medical physicists
37
Radiopharmacists
38
Other health professionals in the medical radiation facility
38
Referring medical practitioners
38
Radiation protection officers
38
Suppliers, installation, maintenance and servicing personnel
39
Maintaining competence (CME and CPD)
39
Equipment and software specific training
39
MANAGEMENT SYSTEM FOR RADIATION PROTECTION AND SAFETY
40
SAFETY ASSESSMENTS
42
3. SPECIFIC RECOMMENDATIONS FOR RADIATION PROTECTION AND SAFETY IN
DIAGNOSTIC RADIOLOGY AND IMAGE GUIDED INTERVENTIONAL PROCEDURES 43
INTRODUCTION
43
SAFETY OF MEDICAL RADIATION FACILITIES AND MEDICAL RADIOLOGICAL
EQUIPMENT
44
Radiology facilities
44
Medical radiological equipment, software and ancillary equipment
48
OCCUPATIONAL RADIATION PROTECTION
54
Introduction
54
Arrangements under the radiation protection programme
54
Assessment of occupational exposure and workers’ health surveillance
61
Information, instruction and training
65
Conditions of service and special arrangements
66
RADIATION PROTECTION OF PATIENTS, CARERS AND COMFORTERS, AND
VOLUNTEERS IN BIOMEDICAL RESEARCH
66
Introduction
66
Justification of medical exposures
67
Optimization of protection and safety
69
Pregnant female patients
83
Unintended and accidental medical exposures
83
Records and review
86
RADIATION PROTECTION OF THE PUBLIC
87
Introduction
87
Non-occupationally exposed workers and visitors
87
Monitoring and reporting
88
PREVENTION AND MITIGATION OF ACCIDENTS
88
Safety assessments
88
Accident prevention
89
Mitigation and contingency plans
90
4. SPECIFIC RECOMMENDATIONS FOR RADIATION PROTECTION AND SAFETY IN
NUCLEAR MEDICINE
91
INTRODUCTION
91
SAFETY OF MEDICAL RADIATION FACILITIES AND MEDICAL RADIOLOGICAL
EQUIPMENT
92
Nuclear medicine facilities
92
Radiopharmaceuticals
96
Medical radiological equipment, software and ancillary equipment
97
OCCUPATIONAL RADIATION PROTECTION
100
Introduction
100
Assessment of occupational exposure and workers’ health surveillance
108
Information, instruction and training
112
Conditions of service and special arrangements
112
Protection of workers responding to incidents in the nuclear medicine facility
113
RADIATION PROTECTION OF PATIENTS, CARERS AND COMFORTERS, AND
VOLUNTEERS IN BIOMEDICAL RESEARCH
113
Introduction
114
Justification of medical exposures
114
Optimization of protection and safety
117
Pregnant female patients
127
Breast-feeding patients
128
Release of patients after radiopharmaceutical therapy
128
Unintended and accidental medical exposures
129
Records and review
131
RADIATION PROTECTION OF THE PUBLIC
132
Introduction
132
Non-occupationally exposed workers and visitors
132
Members of the public in the wider public domain
133
Death of a patient who has undergone a nuclear medicine procedure
134
Radioactive waste
134
Monitoring and reporting
136
PREVENTION AND MITIGATION OF ACCIDENTS
136
Safety assessments
136
Accident prevention
137
Mitigation and contingency plans
137
SAFETY IN THE TRANSPORT OF RADIOACTIVE MATERIALS
140
5. SPECIFIC RECOMMENDATIONS FOR RADIATION PROTECTION AND SAFETY IN
RADIATION THERAPY
141
INTRODUCTION
141
SAFETY OF MEDICAL RADIATION FACILITIES AND MEDICAL RADIOLOGICAL
EQUIPMENT
143
Radiation therapy facilities
143
Medical radiological equipment, software and ancillary equipment
148
OCCUPATIONAL RADIATION PROTECTION
155
Introduction
155
Arrangements under the radiation protection programme
155
Assessment of occupational exposure and workers’ health surveillance
162
Information, instruction and training
165
Conditions of service and special arrangements
165
Protection of workers responding to incidents in the radiation therapy facility
166
RADIATION PROTECTION OF PATIENTS, CARERS AND COMFORTERS, AND
VOLUNTEERS IN BIOMEDICAL RESEARCH
167
Introduction
167
Justification of medical exposures
167
Optimization of protection and safety
168
Pregnant females patients
177
Release of patients after permanent brachytherapy implants
178
Unintended and accidental medical exposures
178
Records and review
181
RADIATION PROTECTION OF THE PUBLIC
182
Introduction
182
Non-occupationally exposed workers and visitors
182
Radioactive sources no longer in use
184
Monitoring and reporting
184
PREVENTION AND MITIGATION OF ACCIDENTS
184
Safety assessments
184
Accident prevention
185
Mitigation and contingency plans
186
SAFETY IN THE TRANSPORT OF RADIOACTIVE MATERIALS
188
APPENDIX I SUMMARY OF TYPICAL CAUSES OF AND CONTRIBUTING FACTORS TO
ACCIDENTAL EXPOSURES IN MEDICAL USES OF IONIZING RADIATION
190
DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES
190
NUCLEAR MEDICINE
190
RADIATION THERAPY
191
APPENDIX II AVOIDANCE OF PREGNANCY FOLLOWING RADIOPHARMACEUTICAL
THERAPY
193
APPENDIX III CESSATION OF BREAST FEEDING
194
REFERENCES
196
1. INTRODUCTION
BACKGROUND
1.1. Medical uses of ionizing radiation are amongst the longest established
applications of ionizing
radiation. Current estimates put the worldwide annual number of
diagnostic and interventional
radiological procedures at over 3000 million and at over 5 million radiation
therapy treatments [1].
These medical uses bring considerable public health benefits.
1.2. However, ionizing radiation can cause harm and a systematic
approach should be applied to
ensure that there is a balance between being able to utilize the benefits from
medical uses of ionizing
radiation and minimizing the risk of radiation effects to patients, workers and
members of the public.
1.3. Medical uses of ionizing radiation only have a place in the
context of medical practice. The
system for ensuring radiation protection and safety should fit in with the
larger system for ensuring
good medical practice. This Safety Guide focuses on the system of radiation
protection and safety.
1.4. The International Atomic Energy Agency (IAEA) Fundamental Safety
Principles [2] present the
fundamental safety objectives and principles of protection and safety.
Requirements designed to meet
these are established in Radiation Protection and Safety of Radiation
Sources: International Basic
Safety Standards (the BSS), jointly sponsored by the European Commission
(EC/Euratom), Food and
Agriculture Organization of the United Nations (FAO), IAEA,
International Labour Organization
(ILO), OECD Nuclear Energy Agency (OECD/NEA), Pan American Health Organization
(PAHO),
United Nations Environment Programme (UNEP) and World Health Organization (WHO)
[3].
1.5. This Safety Guide, prepared jointly by the IAEA, the WHO, the PAHO and the
ILO, provides
guidance on fulfilling the requirements of GSR Part 3 [3] (the BSS) with
respect to medical uses of
ionizing radiation.
1.6. The International Commission on Radiological Protection
(ICRP) has developed
recommendations for a system of radiation protection [4]. These and other
current recommendations
of the ICRP and the International Commission on Radiation Units and
Measurements (ICRU) have
been taken into account in preparing this Safety Guide.
1.7. It is assumed in this Safety Guide that the individual State has in place
an effective governmental,
legal and regulatory infrastructure for radiation safety that covers medical
uses of ionizing radiation.
1.8. This Safety Guide supersedes Safety Guide Radiological Protection
for Medical Exposure to
Ionizing Radiation, RS-G-1.5, Safety Reports Series No. 38 Applying Radiation
Safety Standards in
Radiotherapy, Safety Reports Series No. 39 Applying Radiation Safety
Standards in Diagnostic
Radiology and Interventional Procedures Using X Rays, and Safety Reports Series
No. 40 Applying
Radiation Safety Standards in Nuclear Medicine.
1.9. Unless otherwise stated, terms are used with the meanings ascribed to them
in the BSS [3] or the
IAEA Safety Glossary (2007 Edition) [5].
OBJECTIVE
1.10. The BSS [3] establish requirements for protection of people from harmful
effects of exposure to
ionizing radiation, for the safety of radiation sources and for
protection of the environment. This
Safety Guide recommends how medical uses of ionizing radiation should be
carried out safely within
the framework of the BSS.
1.11. The purpose of this publication is to provide recommendations
and guidance on meeting the
requirements for the safe use of radiation in medicine. It is aimed primarily
at end-users in medical
radiation facilities where radiological procedures are performed, including
management, radiological
medical practitioners, medical radiation technologists, medical physicists,
radiation protection officers
and other health professionals. It also provides recommendations and guidance
to health professionals
who refer patients for radiological procedures; to manufacturers and suppliers
of medical radiological
equipment; and to ethics committees with responsibilities for biomedical
research.
1.12. This publication provides recommendations and guidance on
appropriate regulatory activities
and infrastructure, and is therefore also applicable to regulatory bodies,
health authorities, government
agencies in general, and professional bodies.
SCOPE
1.13. This Safety Guide provides recommendations for ensuring radiation
protection and safety of
radiation sources with regard to patients, workers, carers and
comforters, volunteers in biomedical
research, and the public in medical uses of ionizing radiation. It
covers radiological procedures in
diagnostic radiology (including dentistry), image guided interventional
procedures, nuclear medicine,
and radiation therapy. Some of these radiological procedures may be
carried out in other medical
specialties, including, but not limited to, cardiology, vascular surgery,
urology, orthopaedic surgery,
obstetrics and gynaecology, emergency medicine,
gastroenterology, anaesthetics and pain
management.
1.14. Depending on the laws and regulations in a State, medical uses of
ionizing radiation may include
the use of ionizing radiation in other health care practices, such
as chiropractic, osteopathy and
podiatry. These uses are also within the scope of this Safety Guide.
1.15. This Safety Guide does not include recommendations or guidance
on human imaging using
ionizing radiation for purposes other than medical diagnosis, medical
treatment or biomedical
research. Such human imaging using ionizing radiation for other purposes
includes exposing people to
radiation for employment related, legal or health insurance purposes
without reference to clinical
indications, and human imaging using ionizing radiation for the
detection of concealed objects for
anti-smuggling purposes or for the detection of concealed objects that could be
used for criminal acts
that pose a national security threat. These applications will be covered in
Safety Guide DS471 [6].
STRUCTURE
1.16. Following this introductory section, Section 2 gives general
recommendations for radiation
protection and safety in medical uses of ionizing radiation. This
includes the application of the
principles of protection and safety; the use of the graded approach;
roles and responsibilities;
education, training, qualification and competence; management systems for
protection and safety; and
safety assessments.
1.17. Sections 3 – 5 give recommendations for specific areas of medical uses of
ionizing radiation –
Section 3 – diagnostic radiology and image guided interventional
procedures; Section 4 – nuclear
medicine; and Section 5 – radiation therapy. Guidance for hybrid
modalities is addressed in the
relevant sections, as appropriate.
1.18. Appendix I gives summary guidance on typical causes of and contributing
factors to accidental
exposures in medical uses of radiation. Appendices II and III give
recommendations on the avoidance
of pregnancy following radiopharmaceutical therapy and the cessation of
breast feeding following
administration of radiopharmaceuticals for diagnostic examinations,
respectively.
2. GENERAL RECOMMENDATIONS FOR RADIATION PROTECTION AND
SAFETY IN MEDICAL USES OF RADIATION
GENERAL ASPECTS
2.1. Medical uses of ionizing radiation take place in a variety of settings,
including hospitals, medical
centres, health clinics, specialist clinics, and dental practices.
Medical radiation facility is the term
used in the BSS to cover all the possible settings. Many medical radiation
facilities provide services
for more than one medical use of radiation. For example, a large hospital
typically may have facilities
for diagnostic radiology, image guided interventional procedures,
nuclear medicine and radiation
therapy. The authorization process for medical uses of ionizing radiation
varies from State to State. In
some States a single authorization may cover all specialties and activities
within the facility, whereas
others may authorize each specialty or application separately. For
example, in one State a hospital
may have a single authorization covering all of diagnostic radiology,
image guided interventional
procedures, nuclear medicine, and radiation therapy, whereas in another State
each of these areas or
applications may be authorized separately. Despite differences in
authorization, the guidance in this
Safety Guide remains applicable.
2.2. Traditionally each of the areas of diagnostic radiology, nuclear medicine,
and radiation therapy
were separate, with little or no combined usage. This has changed, with the
so-called hybrid imaging
systems involving both diagnostic radiology and nuclear medicine expertise, and
with the planning,
guidance and verification stages of radiation therapy increasingly
involving both imaging and
radiation therapy expertise. Cross references are given where appropriate.
2.3. As already noted above in para. 1.3, the setting for this Safety Guide is
the practice of medicine
(including dentistry, chiropractic, osteopathy and podiatry). The requirements
of the BSS for radiation
protection and safety of radiation sources apply for the uses of radiation in
medicine as elsewhere.
The requirements must be met and fitted in with medical structures and
processes and in health care
pathways, with the objective of improved patient care and patient outcomes.
TYPES OF EXPOSURE SITUATIONS AND CATEGORIES OF EXPOSURE
2.4. The requirements of the BSS are based around the three types of exposure
situations: planned
exposure situations, existing exposure situations, and emergency exposure
situations. Medical uses of
ionizing radiation are a planned exposure situation and the requirements of
Sections 2 and 3 of the
BSS apply, as appropriate. This includes situations of potential exposure that
is an exposure that is not
expected to occur with certainty, but could result from an accident or from an
event or a sequence of
events that may occur but is not certain to occur (BSS [3], paragraph 1.20
(a)]. Potential exposure can
be applicable to any of occupational, public and medical exposure,
where the event, if it occurs,
results in an exposure over and above what would be expected
normally. Situations when the
radiological procedures do not go as planned – that is, unintended and
accidental medical exposures
should be treated as planned exposure situations (BSS [3], paragraph 3.145].
Chapters 2, 4 and 5 of
this safety guide discuss the prevention and mitigation of events leading to a
potential exposure. In
extreme situations in medical settings when dangerous source is
involved (such as a radiotherapy
source), emergency exposure situation may occur affecting either workers or
member of the public.
For preparedness and response for emergency exposure situations requirements of
Sections 4 of the
BSS and specific safety standard GSR Part 7 apply [7].
2.5. Medical uses of ionizing radiation involve all three categories of
exposure: occupational exposure
for those involved in the performance of radiological procedures; medical
exposure, primarily for the
patients undergoing the radiological procedures but also for carers and
comforters and for volunteers
subject to exposure as part of a programme of medical research; and public
exposure for members of
the public, such as in waiting rooms. The requirements for radiation
protection and safety differ
according to the category of exposure, so it is important that the exposure of
persons is categorized
correctly. For example, a nurse assisting with image guided
interventional procedures would be
considered to be occupationally exposed, whereas a nurse working in
an in-patient ward where
occasional mobile radiography is performed by a medical radiation
technologist would not be
considered occupationally exposed but rather as subject to public
exposure. The term carer and
comforter has been introduced into the BSS to cover the persons who,
outside an occupational
capacity, willingly and voluntarily help in the care, support and
comfort of a patient undergoing a
radiological procedure. Carers and comforters are subject to medical
exposure, whereas a casual
acquaintance visiting a patient who has undergone radionuclide therapy
would be considered a
member of the public and hence subject to public exposure. More extensive
guidance is provided in
each of the specialty Sections 3 - 5.
2.6. Unintended and accidental medical exposures are covered in detail in
Sections 3-5. Such events
include the exposure of the wrong person¹.
APPLICATION OF THE RADIATION PROTECTION REQUIREMENTS
2.7. The three general principles of radiation protection, justification,
optimization of protection and
safety and the application of dose limits, are expressed in safety
principles 4, 5, 6 and 10 of the
Fundamental Safety Principles [2]. In terms of Requirement 1 of the
BSS, those responsible for
protection and safety must ensure that the relevant requirements applying these
principles are met.
2.8. Medical exposure differs from occupational and public exposure in
that persons (primarily
patients) are deliberately, directly, and knowingly exposed to radiation for
their benefit. In medical
exposures applying a ‘dose limit’ is inappropriate as it may limit
the benefit for the patient;
consequently, only two of the radiation protection principles apply –
justification and optimization.
Justification plays the role of gatekeeper, as it will determine whether the
exposure will take place or
not. If it is to take place, the radiological procedure
should be performed in such a way that the
radiation protection and safety is optimized.
TABLE 1. SUMMARY OF RADIATION PROTECTION PRINCIPLES AS APPLIED
TO
OCCUPATIONAL AND PUBLIC EXPOSURES IN COMPARISON WITH
MEDICAL
EXPOSURE
Application to occupational and public exposure Application to medical
exposure
Justification of practices: A practice that entails
exposure to radiation should only be adopted if it
yields sufficient benefit to the exposed individuals or
to society to outweigh the radiation detriment.
Justification: The diagnostic or therapeutic benefits
produced by exposures are weighed against the
radiation detriment they might cause, taking into
account the benefits and risks of available alternative
techniques that do not involve medical exposure.
¹ The definition of medical exposure in the BSS [3] was changed from the
previous edition to ensure that the
event of the ‘wrong person’ is kept within the radiation protection and safety
framework for medical exposure
so that it can be investigated by the appropriate people, with corrective
actions to minimize recurrence.
Optimization of protection and safety: Providing the
best available protection and safety measures under
the prevailing circumstances, so that the magnitudes
and likelihood of exposures and the numbers of
individuals exposed be as low as reasonably
achievable, economic and social factors being taken
into account.
Limitation of doses: Doses to individuals are limited
(for occupational and public exposure).
Optimization of protection and safety: In diagnostic
and interventional medical exposure, keeping the
exposure of patients to the minimum necessary to
achieve the required diagnostic or interventional
objective.
In therapeutic medical exposure, keeping the exposure
of normal tissue as low as reasonably achievable
consistent with delivering the required dose to the
planning target volume.
Limitation of doses: Not applicable to medical
exposure.
Justification
2.9. Justification in medical uses of ionizing radiation involves consideration
of all three categories of
exposure – medical, occupational and public exposure.
2.10. From an occupational and public exposure perspective, the
practice should be justified. This
aspect of justification is the process of determining whether the use
of the given radiological
procedure is expected to yield benefits to the individuals that undergo the
procedure and to society
that outweigh the harm (including radiation detriment) resulting from
the procedure. In almost all
cases the occupational and public considerations in justification are
overshadowed by the justification
of medical exposures (see para. 2.11). While medical radiological procedure is
expected to do more
good than harm to the patient, subsidiary account should be taken of the
radiation detriment from the
exposure of the radiological staff and of other individuals.
2.11. The application of the justification principle to medical exposures
requires a special approach,
using three levels . As an overarching justification of medical exposures, it
is accepted that the proper
use of radiation in medicine does more good than harm (level 1). At
the second level generic
justification of a given radiological procedure should be carried out
by the health authority in
conjunction with appropriate professional bodies,. The possibility of
accidental or unintended
exposures should also be considered at this level. This applies to the
justification of new technologies
and techniques as they evolve, but the decisions should be reviewed
from time to time, as more
information becomes available about the risks and effectiveness of the existing
procedure and about
new procedures. For the final level of justification (level 3), the
application of the radiological
procedure to a given individual should be considered. The specific
objectives of the exposure, the
clinical circumstances and the characteristics of the individual involved
should be taken into account.
National or international referral guidelines, developed by professional
bodies together with health
authorities, should be used (BSS paragraph 3.158). Those radiological
procedures that are not justified
should be eliminated. The approach to implementing of justification of a
procedure for an individual
patient (level 3) depends on whether it is a diagnostic procedure, an image
guided intervention, or a
treatment. Specific guidance on justification in each specialty is given in
Sections 3 to 5.
2.12. The level 3 justification of medical exposure for an individual
patient does not include
considerations of occupational exposure. If the proposed radiological
procedure is justified for that
patient, then the participation of particular staff in performing the
procedure is governed by the
requirements for optimization of occupational radiation protection and safety
and occupational dose
limitation.
Optimization of protection and safety
2.13. The optimization of protection and safety, when applied to the
exposure of workers and of
members of the public, and of ‘carers and comforters’ of patients undergoing
radiological procedures,
is a process for ensuring that the magnitude and likelihood of exposures and
the number of individuals
exposed are as low as reasonably achievable, with economic, societal and
environmental factors taken
into account. This means that the level of protection and safety would be the
best possible under the
prevailing circumstances.
2.14. As is the case with justification, the application of the
requirements for optimization to the
medical exposure of patients and to that of volunteers as part of a programme
of biomedical research
requires a special approach. Too low a radiation dose could be as bad as too
high a radiation dose, in
that the consequence could be that a cancer is not cured or the
images taken are not of suitable
diagnostic quality. The medical exposure should always lead to the required
clinical outcome.
2.15. Optimization is a prospective and iterative process that requires
judgements to be made using
both qualitative and quantitative information. Specialty specific guidance on
optimization of medical,
occupational and public radiation protection and safety is given in Sections 3
to 5.
2.16. Dose constraints are applicable to occupational exposure and to public
exposure in medical uses
of ionizing radiation. Dose constraints are also used in optimization
of protection and safety for
‘carers and comforters’ and for volunteers subject to exposure as part of a
programme of biomedical
research. Dose constraints are not applicable to the exposure of patients in
radiological procedures for
the purposes of medical diagnosis or treatment. See also paras. 2.46
to 2.50. Dose constraints are
used, in the planning stage, for optimization of protection and safety, the
intended outcome of which
is that all exposures are controlled to levels that are as low as
reasonably achievable (ALARA),
economic, societal and environmental factors being taken into account.
2.17. The dose constraint for each particular source of radiation exposure is
intended, among other
things, to ensure that the sum of doses from planned operations for all sources
under control remains
within the dose limits. Dose constraints are not dose limits;
exceeding a dose constraint does not
represent non-compliance with regulatory requirements, but it might result in
follow-up actions.
2.18. In X ray medical imaging, image guided interventional procedures
and diagnostic nuclear
medicine, diagnostic reference levels (DRLs) are a tool used in optimization of
protection and safety.
Periodic assessments are to be performed of typical patient doses or,
for radiopharmaceuticals,
activities administered in a medical radiation facility. Doses in this context
may be expressed in one
of the accepted dosimetric quantities as described in para 2.40 [8 - 10]. For
simplicity sake in this
Safety Guide in Sections 3 and 4, the term “dose” will be used when referring
generally to medical
exposure measurements in radiological imaging, with specific forms of dose or
activity used where
needed.
2.19. If comparison with established diagnostic reference levels shows that the
typical patient doses or
activities are either unusually high or unusually low, a local review
is to be initiated to ascertain
whether protection and safety has been optimized and whether any corrective
action is required. DRLs
are not dose limits. See also paras. 2.34 to 2.45.
2.20. Other tools used in optimization of protection and safety
include, inter alia, design and
operational considerations and programmes of quality assurance. These are
described in detail in the
specialty Sections 3 to 5.
Dose limits
2.21. Dose limits apply to occupational exposure and public exposure arising
from any use of ionizing
radiation, including medical applications. Schedule III of the BSS [3] sets out
these dose limits and
they are reproduced here for convenience in Table 2. Dose limits do not apply
to medical exposure –
i.e. exposure of patients, carers or comforters, and volunteers as part of a
programme of biomedical
research.
2.22. The occupational dose limit for the lens of the eye is lower
in the BSS [3] than previously
recommended. There are some areas of medical uses of ionizing
radiation, such as image guided
interventional procedures, where, if good radiation protection practice is not
being followed, there is a
possibility of exceeding this dose limit. Specific guidance is given in the
specialty Sections 3 to 5.
TABLE 2. DOSE LIMITS FOR PLANNED EXPOSURE SITUATIONS (ADAPTED FROM
SCHEDULE III OF THE BSS [3])
OCCUPATIONAL EXPOSURE
For occupational exposure of workers over the age of 18 years, the dose limits
are:
(a) An effective dose of 20 mSv per year averaged over five consecutive years
(100 mSv in 5 years), and
of 50 mSv in any single year;
(b) An equivalent dose to the lens of the eye of 20 mSv per year averaged
over 5 consecutive years (100
mSv in 5 years) and of 50 mSv in any single year;
(c) An equivalent dose to the extremities (hands and feet) or the skin of 500
mSv in a year.
Additional restrictions apply to occupational exposure for a female
worker who has notified pregnancy or is
breast-feeding (BSS para. 3.114).
For occupational exposure of apprentices of 16 to 18 years of age
who are being trained for employment
involving radiation and for exposure of students of age 16 to 18 who use
sources in the course of their studies,
the dose limits are:
(a) An effective dose of 6 mSv in a year;
(b) An equivalent dose to the lens of the eye of 20 mSv in a year;
(c) An equivalent dose to the extremities (hands and feet) or the skin of 150
mSv in a year.
PUBLIC EXPOSURE
For public exposure, the dose limits are:
(a) An effective dose of 1 mSv in a year;
(b) In special circumstances, a higher value of effective dose in a single
year could apply, provided that the
average effective dose over five consecutive years does not exceed 1 mSv per
year;
(c) An equivalent dose to the lens of the eye of 15 mSv in a year;
(d) An equivalent dose to the skin of 50 mSv in a year.
GRADED APPROACH
2.23. The so-called ‘graded approach’ is a concept that underpins the
application of the system for
protection and safety. The BSS in paragraph 2.12 states: “The application of
the requirements for the
system of protection and safety shall be commensurate with the radiation
risks associated with the
exposure situation.”
2.24. The risks associated with medical uses of ionizing radiation
vary significantly, depending
strongly on the particular radiological procedure. At the low risk end are
dental exposures (excluding
cone beam CT), and dedicated bone densitometry (DEXA) studies. At the high risk
end is radiation
therapy, where the doses involved could be lethal, and image guided
interventional procedures, where
radiation injuries can occur. Another aspect to consider when implementing a
graded approach is the
prevalence of a given application – an example is CT, which is increasingly
used in imaging.
2.25. The BSS places responsibilities for a graded approach on each of the
government, the regulatory
body, registrants and licensees, and employers. The government and the
regulatory body use the
graded approach in setting and enforcing regulatory requirements. For example,
it would be expected
that regulatory bodies devote fewer resources and less time to
regulating dental practices than to
regulating the use of radiation in radiation therapy or image guided
interventional procedures.
2.26. The registrants or licensees, and employers use the graded approach in
the measures they take
for protection and safety. For example, the registrant or licensee of a dental
practice would not need to
implement as comprehensive a quality assurance programme as would a radiation
therapy facility in
order to meet the requirements of the BSS.
2.27. Guidance incorporating the graded approach is given in the specific
guidance for each specialty
and the various modalities within those specialties – see Sections 3 to 5.
ROLES AND RESPONSIBILITIES
Government
General
2.28. The roles and responsibilities of the government² with regard to
protection and safety are set out
in requirement 2 and paragraphs 2.13 to 2.28 of the BSS, with further detailed
requirements given in
the IAEA Safety Requirements publication “Governmental, Legal and
Regulatory Framework for
Safety” [10]. These include:
• establishing an effective legal and regulatory framework for
protection and safety for all
exposure situations;
• establishing legislation that meets specified requirements;
• establishing an independent regulatory body with the necessary legal
authority, competence
and resources;
• establishing requirements for education and training in protection and
safety;
² States have different legal structures, and therefore the term
‘government’ as used in the IAEA safety
standards is to be understood in a broad sense, and is accordingly
interchangeable here with the term ‘State’.
• ensuring that arrangements are in place for:
o the provision of technical services (including radiation monitoring
services and
standards dosimetry laboratories), and
o education and training services.
All of these are relevant to the safe use of ionizing radiation in medicine.
2.29. As noted above in para. 1.7, this Safety Guide assumes that an effective
governmental, legal and
regulatory infrastructure for radiation protection and safety is in
place. However there are some
additional considerations that are important for ensuring radiation
protection and safety in medical
uses of ionizing radiation.
2.30. Government has a role to facilitate and ensure that the health authority,
the relevant professional
bodies, and the radiation protection regulatory body communicate and cooperate
in working towards
establishing the infrastructure necessary for radiation protection and safety
in medical uses of ionizing
radiation. The role of the health authority typically includes determining
policy, which in turn may
dictate the resources allocated to the various areas of healthcare, including
medical uses of ionizing
radiation. Up to date information on developments in medical uses of ionizing
radiation, and how that
might shape and influence medical practice, should be available so
that appropriate policy can be
developed and implemented. The professional bodies of the various health
professionals associated
with radiation in healthcare represent the collective expertise of the given
health profession and, as
such, can strongly influence the practice of radiation protection and safety.
The health authority and
the professional bodies should be active working partners with the
radiation protection regulatory
body to achieve effective regulation of medical uses of ionizing radiation. See
paras. 2.52 – 2.68 for
more guidance on the health authority and professional bodies.
2.31. Formal recognition of health professionals should be in place to ensure
that only persons with
the appropriate competencies are allowed to take on particular roles and
responsibilities. In medical
uses of ionizing radiation, this applies in particular to persons
undertaking the role of radiological
medical practitioner, medical radiation technologist, or medical physicist.
Detailed guidance is given
in the sub-section on education, training, qualifications and competence,
paras 2.117 to 2.135.
2.32. Other organizations can make a worthwhile contribution to radiation
protection and safety in
medical uses of ionizing radiation. These include technical standards
associations, medical devices
regulatory agencies, and health technology assessment agencies, which issue
standards or reports that
could have direct implications for radiation safety. Not all States have such
organizations but, where
they exist, the government should ensure that they interact cooperatively with
the radiation protection
regulatory body, the health authority and the relevant professional bodies. In
States that do not have
such organizations, the government should consider means to adopt or
adapt relevant standards or
reports from such organizations in other countries.
2.33. Other organizations can have an indirect, but not necessarily
insignificant, effect on radiation
protection and safety in medical uses of ionizing radiation. Such
organizations include health
insurance or re-imbursement companies and standards accreditation bodies. The
former, by deciding
on what radiological procedures (and other alternative techniques) are
covered. The latter, by
including radiation protection and safety in its scope, can positively
influence how well radiation
protection and safety is being implemented in medical facilities
seeking accreditation. Again,
government should be aware of these players in their country and utilize their
influence to improve
the practice of radiation protection and safety in medical uses of ionizing
radiation.
Diagnostic reference levels
2.34. Diagnostic reference level (DRL) should be used as an important
tool for optimization of
protection and safety for diagnostic medical exposures , (see para. 2.18).
Government has a particular
responsibility to ensure that DRLs are established for their country, or
regions within their country or,
in some cases, regions of several small countries. In establishing values for
the DRLs it is preferable
that, for common imaging procedures, typical (e.g. average or median) doses³
for patients are obtained
from a representative sample of rooms and facilities where these procedures are
being performed. In
this way a snap shot of current practice in the country or region is obtained,
reflecting both good and
poor practices, for that particular imaging procedure. The value of
the DRL for that particular
procedure is typically the rounded 75th percentile of the distribution of the
room/facility typical doses
[12, 13]. In diagnostic nuclear medicine, an ‘optimum’ value for a
DRL is used also instead of a
percentile: a reference level for administrations of activities of
radionuclides sufficient to obtain
information for standard groups of patients (adults and children),
based on the experience of the
professional groups (‘expert judgement’) [12]. In establishing DRLs, it is
fundamental to include only
radiological procedures whose image quality is adequate for the medical
purpose.
2.35. Once established, medical radiation facilities should compare
their typical doses (sometimes
called “facility reference levels”, or “local reference levels”) with the
relevant DRLs, as described in
Sections 3 and 4. Optimization of protection for a particular
radiological procedure should be
reviewed if the comparison shows that the facility’s typical dose
exceeds the DRL, or that the
facility’s typical dose is substantially below the DRL and it is
evident that the exposures are not
producing images of diagnostic usefulness or are not yielding the
expected medical benefit to the
patient. The resulting actions aimed to improve optimization of protection and
safety will usually, but
not necessarily, result in a lower facility typical doses for the procedure or
procedures. At some pre-
determined interval, typically 3 to 5 years, there should be a
review of the established national or
regional DRL values. A new national or regional survey will result in a new
distribution of facility
reference levels that will reflect the improvements made as a result of using
the existing DRLs. It is
likely that the new values of the DRLs will be lower than the
previous values. This cycle of
establishment of national or regional DRLs, use by imaging facilities,
corrective actions by imaging
facilities, and periodic review of national or regional DRLs brings
about a steady improvement in
optimization of protection and safety across the country or region.
2.36. There are several steps to the establishment of DRLs. At the national or
regional level decisions
should be made whether to use actual patients or phantoms to represent a
“standard patient” for each
modality. Phantoms avoid most of the issues with variations in patient size
(see paras. 2.38 and 2.41).
However their use does not truly represent clinical practice with patients and
clinical images and, as
such, would seem less appropriate for use in establishing DRLs.
Nevertheless, a phantom-based
approach, in the absence of adequate patient data, can be used to first
establish DRLs and then in their
utilization [14].
2.37. The imaging procedures, for which DRLs are to be established,
should be decided upon at
national or regional level. The criteria which may help in this decision are
the relative frequencies and
the magnitude of the doses of the imaging procedures – the more frequent and
higher dose procedures
should have a higher priority. Specific consideration should be given to
paediatric imaging. Based on
³ The term ‘doses’ in this subsection on DRLs includes activity for nuclear
medicine procedures, as described in
para 2.18.
national or regional resources, the actual number of procedures for which DRLs
are established will
vary [15, 16].
2.38. The next step is to perform preferably a wide-spread or at least a
representative survey for the
selected procedures. Most imaging radiological procedures are performed on
adults, and traditionally
national DRLs have been established first for adults. For each
room/facility performing the given
procedure, typically a sample size of about 20 patients is used. Not all
adults are the same size, so
many countries [12, 13] have established DRLs for an “standard” adult
patient, limiting patient
eligibility to the sample on the basis of weight, for example 70 kg ± 20 kg,
and aiming for a sample
average in a given weight range, for example 70 kg ± 5 kg. Other countries
adopt a more pragmatic
approach, accepting all adults in the initial sample but excluding extreme
outliers in terms of weight
[17].
2.39. Another consideration with DRLs is whether the procedure is simply
defined in terms of the
anatomical region being imaged, or whether there should be a further
refinement to include the
clinical purpose of the examination. For example, a CT of the abdomen may be
performed differently
depending on the medical question to be answered. For those embarking on
establishing DRLs for the
first time, it is advisable to take the simpler approach.
2.40. The dose metrics used to represent the dose to the patient should be
easily measurable and must
follow ICRU recommendations, as stated in the BSS (BSS para 1.46)
[8]. The following are
commonly used terms in this role: radiography – air kerma-area
product, incident air kerma, or
entrance surface air kerma (which includes backscatter) ; fluoroscopy – air
kerma-area product; CT –
CTDIvₒl, and DLP; mammography – incident air kerma or entrance
surface air kerma and mean
glandular dose; dentistry – incident air kerma or entrance surface air kerma
for intra-oral radiography
and air kerma-area product for panoramic radiography; and image guided
interventional procedures –
air kerma-area product and air kerma at the interventional reference
point. More guidance on dose
metrics is given in Section 3 paras 3.193 – 3.195. It is crucial that the dose
data collected for each
contributing room is for procedures where the image quality was
confirmed as adequate for the
clinical purpose. For nuclear medicine DRLs are set in activity administered to
patient.
2.41. Optimizing protection and safety for average adult patients does
not necessarily mean that
optimization is being achieved for other size or age groups. Past experience in
particular with children
undergoing CT examinations has clearly demonstrated that this is not
the case. This means that
consideration also should be given to establishing DRLs for children undergoing
imaging procedures.
The same problem of size and weight, as discussed in para. 2.38,
also pertains to children. Some
countries have adopted a simple age approach – for example, new born, 1, 5, 10,
and 15 years – while
others normalize patient dose data to particular representative sizes
or weights. More guidance on
grouping patient for establishing typical doses and DRL is given in
Section 3 para 3.204 for
diagnostic and interventional radiology and Section 4 para 4.203 for diagnostic
nuclear medicine.
2.42. The processes and steps towards establishing DRLs, as described
in paras. 2.37 to 2.41, are
likely to involve many players including the imaging facilities, the health
authority, the professional
bodies, and the regulatory body. In particular there should be
collective ownership of the DRLs –
deciding on: what procedures, what age groups, how to collect the data, who
will manage the data,
and when to review and update the DRLs. In some countries a
national governmental body
administers the national patient dose database that underpins the
establishing of DRLs. In other
countries, this role may be taken by the regulatory body or a professional
body. There is no preferred
custodian – what is important is that a patient dose database (for DRLs) is
established and maintained,
DRL values are set, these are promulgated through the regulatory
processes, and a process for
periodic review is established. It may be more appropriate to take a regional
rather than a national
approach to DRLs – either a region in a large country or a region of several
small countries.
2.43. The methodology used in performing the initial survey can range from a
paper-based approach
through to a web-based electronic submission approach. As the interconnectivity
of imaging systems,
with the availability of patient dose metrics, and radiology and
hospital information systems
improves, the process of gathering data for DRLs is likely to become easier.
Countries embarking on
establishing DRLs for the first time should consider the electronic approach.
2.44. The national or regional DRL values should be periodically
reviewed and updated, typically
with a cycle of 3 to 5 years (see para. 2.35). The review can be performed in
many ways, but in all
cases there is first a collection phase, followed by analysis of the data
collected. The collection of
facility reference levels can occur throughout the cycle, or it can be
restricted to a shorter time-frame
towards the end of the cycle. Pragmatically, the occasion of a medical
radiation facility comparing its
practice typical doses with the current DRLs would seem to be an appropriate
time for the facility to
submit their new facility typical doses to the national or regional database
being used for the DRLs.
At the end of the cycle the analysis of the submitted facility typical doses
would take place, and the
values of the DRLs updated accordingly. While increased digital
connectivity would technically
support collection and analysis to be continuous, a given set of DRL values
should be stable for a
period of time to allow the improvement cycle to take place.
2.45. Finally, if the government in a given country is not able to facilitate
the establishment of their
own national DRLs or to participate in a regional approach, there is
the option for government to
facilitate the adoption of another country’s or region’s DRLs. While
such DRLs do not reflect a
country’s own practice, with judicious choice, the adopted DRLs can still
perform the same role of
bringing about an improvement in the implementation of optimization of
protection and safety in the
adopting country. DRLs from countries with significantly different generations
of imaging systems
should be compared with a care
Dose constraints
2.46. Dose constraints are not dose limits; they are tools for
optimization of protection and safety,
including considerations of social and economic factors. The role of dose
constraints for occupational
exposure and for public exposure is introduced above in para. 2.16.
In particular the government,
typically through the radiation protection regulatory body, has
responsibilities with respect to public
exposure, where its primary role is to ensure that no member of public can
exceed the public dose
limit as a result of cumulative public exposure arising from multiple
authorized facilities, including
medical radiation facilities. A simple approach that can be taken is to set a
dose constraint, for public
exposure arising from a single facility, at some fraction of the dose limit.
Some countries use a dose
constraint of approximately one-third of the dose limit – namely an
effective dose of 0.3 mSv per
year. In establishing such a value the regulatory body should
consider the number and type of
radiation sources in use in a particular country or region that may expose the
public.
2.47. In addition to patients, there are two other groups of people that can
incur medical exposure.
These are the carers and comforters and the volunteers in biomedical research.
Because it is medical
exposure, neither of these groups of persons is subject to dose
limits for the exposures incurred.
Instead, reliance is placed on the use of dose constraints as a
means for ensuring optimization of
protection and safety takes place (see para. 2.16). For both of these groups of
people, the government,
through consultation between the health authority, the relevant professional
bodies and the radiation
protection regulatory body, has the responsibility to ensure that dose
constraints are established so that
they can be used.
2.48. For carers and comforters the usual approach is to apply the dose
constraints on an “episode by
episode” basis – i.e. the dose constraint applies to the cumulative exposure of
the carer and comforter
over the duration of that person giving care and comfort to a particular
patient. In the case of a mother
assisting with her child undergoing a diagnostic X ray procedure, the
episode is extremely short –
simply while the X rays are being produced. In the case of a person having
undergone treatment with
radiopharmaceuticals, the episode will last several days until such
time as the radionuclide has
decayed to negligible levels. Consideration should be given to the
cumulative dose of a carer or
comforter if he/she acts in this role for several distinct episodes. In such
cases, a dose constraint per
annum may be used in addition.
2.49. In setting dose constraints for carers and comforters consideration
should be given to the age of
the individual and for a woman the possibility of her being
pregnant. A particular issue is that of
children being in this role. The definition of a carer or comforter includes
that the person “willingly
and voluntarily” helps in this role. It could be argued that young children may
not understand such
concepts. None the less, it is reasonable and likely that the children of a
treated parent would want to
provide comfort, at least. The framework for radiation protection and
safety should accommodate
such human wishes. A pragmatic approach often taken is to effectively treat
children in this role as
members of the public and constrain their medical exposure to an effective dose
of 1 mSv per episode.
A pregnant carer or comforter presents a similar situation, and consideration
should be given to the
embryo or fetus. The same approach of constraining the effective dose to the
embryo/fetus to 1 mSv
per episode is often taken. For an adult carer or comforter, a value of dose
constraint commonly used
is 5 mSv effective dose per episode. For elderly persons more lenient dose
constraints may be used. In
any of these cases flexibility with respect to the dose constraint may need to
be used.
2.50. In setting dose constraints for diagnostic radiological
procedures that are performed on
volunteers participating in a programme of biomedical research, the
intention is that government,
through consultation between the health authority, the relevant professional
bodies and the radiation
protection regulatory body, provides broad guidance for the ethics
committees (see paras. 2.98 –
2.101) who, in turn, would adapt the dose constraints to suit the particular
programme of biomedical
research under consideration. Typical patient doses and national DRLs would be
two considerations
in setting such dose constraints.
Criteria and guidelines for release of patients after radionuclide therapy
2.51. Many factors can influence the exposure that members of the public and
carers and comforters
can incur following the release of a patient who has undergone a therapeutic
procedure with unsealed
sources or who retains implanted sealed sources. Detailed guidance on
these factors for unsealed
sources is given in the Safety Report Series No. 63 [18]. The role of
government, through consultation
between the health authority, the relevant professional bodies and the
radiation protection regulatory
body, is to ensure that criteria are established, with accompanying
guidance, to help simplify the
process when individual medical radiation facilities are considering
the release of their patients.
Guidance for these medical radiation facility actions are given in Sections 4
and 5.
Health authority
2.52. All medical facilities must be authorized by the health authority to
ensure that the facility meets
the applicable requirements for quality of medical services. When the medical
facility uses ionizing
radiation, authorization for medical practice and healthcare should be granted
by the health authority
only if radiation safety requirements are met (para 2.69 – 2.75). As noted
in para. 2.30, the health
authority should contribute to radiation protection and safety.
This includes participation in
establishing DRLs, dose constraints for carers and comforters and for
volunteers in biomedical
research, and criteria and guidance for the release of patients after
radionuclide therapy. See guidance
in paras. 2.34 to 2.51. Coordination and collaboration between the health
authority and the radiation
protection regulatory body should ensure radiation protection and
overall safety of the medical
facility.
2.53. Radiation protection and safety in medical uses of ionizing
radiation should be assured by a
proper specialization of health professionals – namely that only
health professionals with the
appropriate competencies can take on roles that include specific
responsibilities for radiation
protection and safety. The health authority has responsibilities in providing
policy and guidance with
respect to health profession specialties and their sub-specialties,
including scope of practice, and
requirements for competence. Guidance on recognition of competence in a
specialty is given paras.
2.117 to 2.130.
2.54. Adequate numbers of medical personnel and paramedical personnel should
be available for a
medical radiation facility to function correctly and safely. This includes
sufficient capacity to cover
absences of key personnel through sickness, leave or other official
reasons. The health authority,
through its policy making role, should set clear standards for acceptable
medical practice.
2.55. The health authority has particular roles in the application
of the radiation protection
requirements for justification – namely with respect to:
• generic justification of radiological procedures;
• justification of radiological procedures in health screening programmes;
and
• criteria for the justification of radiological procedures for health
assessment of asymptomatic
individuals intended for the early detection of disease, but not as part of a
health screening
programme.
2.56. Generic justification of radiological procedures is an on-going
process as new procedures
become available and as established procedures should be reviewed in the light
of new knowledge and
developments. It should be decided whether a new radiological
procedure should become a new
addition to the armamentarium of existing procedures. Conversely, an
existing radiological procedure
may need to be withdrawn from use because there is evidence that an alternative
technique has better
efficacy. The health authority, together with relevant professional
bodies, should make these
decisions.
2.57. The use of radiological procedures as part of a health screening
programme involves subjecting
asymptomatic populations to radiation exposure. The decision to embark
upon such a programme
should include consideration of, inter alia, the potential of the
screening procedure to detect the
disease, the likelihood of effective treatment of cases detected and,
for certain diseases, the
advantages to the community from the control of the disease. Sound
epidemiological evidence should
provide the basis for such health screening programmes. The health authority,
together with relevant
professional bodies, should consider all the factors before making a decision.
2.58. The use of radiological procedures on asymptomatic individuals, intended
for the early detection
of disease but not as part of an approved health screening programme, is now
increasingly common.
Such radiological procedures are not established medical practice, nor are
they being performed as
part of a programme of biomedical research. Therefore the health
authority, together with relevant
professional bodies, has a role in providing guidance on the applicability and
appropriateness of such
procedures. Such guidance would help the referring medical practitioner and the
radiological medical
practitioner carry out the justification for an individual patient (see Section
3, paras. 3.135 to 3.137).
2.59. National or international referral guidelines should be used as
an important tool in the
implementation of justification of medical exposure for an individual patient .
The health authority
should support the relevant professional bodies in developing and
implementing such referral
guidelines. See also para. 2.65.
2.60. The health authority should also encourage the development of and promote
the implementation
of practice guidelines and technical standards⁴ developed by professional
bodies.
Professional bodies
2.61. Professional bodies are the collective term used in the BSS and in this
Safety Guide to include
the various organizations and entities of health professionals within
a given State. These include
societies, colleges, and associations of health professionals often
within a particular specialty.
Examples, with direct involvement in the use of ionizing radiation,
include representation of
radiologists, radiation oncologists, nuclear medicine physicians, medical
physicists, medical radiation
technologists, and dentists. In large countries, such professional bodies may
be regional, within the
country. Conversely, there may be regional professional bodies covering several
countries. There are
also professional bodies in the wider medical arena that still influence some
aspects of radiation use.
Examples of these include societies, associations and colleges
representing specialties such as
cardiology, gastroenterology, urology, and neurology, who may use
radiation, and other
organizations, such as those that represent general practitioners and primary
care physicians.
2.62. Professional bodies, as stated in para. 2.30, represent the collective
expertise of the given health
profession and specialty and, as such, they also should play a role
in contributing to radiation
protection and safety in medical uses of ionizing radiation. This
includes setting standards for
education, training, qualifications and competence for a given
specialty, and setting technical
standards and giving guidance on practice. Further guidance on education,
training, qualifications and
competence is given in paras. 2.117 to 2.130.
2.63. Relevant professional bodies, in partnership with the health
authority and the radiation
protection regulatory body, have a role with respect to the establishment of
DRLs, dose constraints for
carers and comforters and for volunteers in biomedical research, and
criteria and guidance for the
release of patients after radionuclide therapy, as has been described
in paras. 2.42, 2.47, 2.50, and
2.51, respectively.
2.64. The role of the relevant professional bodies with respect to the
application of the requirements
for justification is described in paras. 2.56 - 2.60.
⁴ The term “practice guidelines and technical standards” is used to represent
the range of documents, statements
and other publications produced by professional bodies to help educate
and guide the respective health
professionals to carry out particular aspects of their specialty.
2.65. Professional bodies should take the lead in the development of
referral guidelines and
appropriateness criteria for use in justification of medical exposure
for an individual patient (para
2.59). It may not be possible for every State to develop its own referral
guidelines. The significant
work of a number of professional bodies around the world could be utilized in
many other countries
through adoption or adaption by the local professional bodies. See
also Section 3 para 3.137 and
Section 4 para 4.159.
2.66. With respect to medical imaging, the process of optimization of radiation
protection and safety
should aim at achieving adequate image quality – not the best possible image
quality, but certainly
sufficient to ensure that diagnosis or treatment is not compromised. From an
operational perspective,
there are many factors that influence the image quality versus patient dose
relationship for a given
procedure. Having standards or norms that specify acceptable image quality is
clearly advantageous,
and relevant professional bodies have a responsibility in this respect.
2.67. For the implementation of optimization of radiation protection
and safety a comprehensive
quality assurance programme for medical exposures should be established. As is
elaborated elsewhere
(para. 2.138), such quality assurance programmes are part of the wider quality
management system of
the medical radiation facility. Nonetheless, there is considerable
benefit in making use of resource
material and standards established by professional bodies for particular areas
of the programme. For
example, many medical physics professional bodies have developed
detailed guidance on
performance testing aspects of a quality assurance programme. Where such
material or standards are
lacking in a State, the relevant professional body could adopt or adapt such
resources from outside the
State.
2. 68. Professional bodies can play a proactive role by encouraging their
members to contribute to
relevant international or national anonymous and voluntary safety reporting and
learning systems, and
by contributing to developing of such systems. The large catchment of
such databases provides a
wealth of information that is educative in helping to minimize
unintended and accidental medical
exposures. Three such international safety reporting systems are the
IAEA’s Safety Reporting and
Learning System for Radiotherapy, SAFRON [19], and Safety in Radiological
Procedures, SAFRAD
[20], and the Radiation Oncology Safety Information System (ROSIS) [21].
Regulatory body
2.69. The radiation protection regulatory body must fulfil its regulatory
functions, such as establishing
requirements and guidelines, authorizing and inspecting facilities and
activities, and enforcing
legislative and regulatory provisions. Detailed requirements specifying these
roles and responsibilities
are given in the BSS [3] and in GSR Part 1 [11], and further general guidance
in Safety Guide GS-G-
1.5 [22]. Guidance on general regulatory body roles and responsibilities with
respect to occupational
radiation protection and radiation protection of the public are given in the
two general safety guides
[23, 24]. An pre-requisite for the regulatory body being able to
perform its regulatory functions
effectively is having staff with appropriate specialist expertise. This is
covered in detail in the IAEA
Safety Standards [3, 11, 22], but should be emphasized in the
context of medical uses of ionizing
radiation, where on the one hand persons are being deliberately exposed to
radiation and, on the other,
health outcome depends on the applicability, availability and use of radiation.
The regulatory controls
should be applied knowledgeably, and not just as an administrative exercise.
Authorization of medical radiation facilities
2.70. The graded approach to medical uses of ionizing radiation has
particular significance for
regulatory bodies because, as described in paras. 2.23 to 2.27, there
is a wide variation in the
complexity of medical radiation facilities. Regulatory bodies
should consider what form of
authorization is appropriate for a given type of medical radiation facility.
Coupled with the type of
authorization is the level of complexity in the documentation that
should be submitted to the
regulatory body prior to the authorization. This includes the degree of detail
in the safety assessment
(see paras. 2.148 - 2.152). The duration of an authorization is another
consideration for the regulatory
body – the more complex facilities would warrant a more frequent renewal
process.
2.71. Typical practices that are amenable to registration are those for which:
(a) safety can largely be
ensured by the design of the facilities and equipment; (b) the
operating procedures are simple to
follow; (c) the safety training requirements are minimal; and (d) there is a
history of few problems
with safety in operations. Registration is best suited to those practices for
which operations do not
vary significantly. These conditions are generally not met in medical uses of
ionizing radiation for the
following three reasons: patient exposure depends on human performance;
radiation protection and
safety is not largely ensured by design; and the training required
is significant. Medical radiation
facilities are, in principle, better candidates for individualized licensing
than for registration. It would
be expected that licensing would be used for radiation therapy facilities,
nuclear medicine facilities,
facilities performing image guided interventional procedures, and for
most diagnostic radiology
facilities. For some simple forms of diagnostic radiology, such as dental
radiography (without cone
beam CT) and DEXA, registration may be acceptable. For both forms of
authorization, the regulatory
body should have standardized forms or templates that help ensure
that the correct information is
submitted to the regulatory body. See also the paragraphs on safety
assessments, paras 2.148 - 2.152.
2.72. No matter what form of authorization is used for a medical radiation
facility, a crucial step prior
to the granting of the authorization is that the regulatory body
ascertains the credentials of key
personnel with responsibilities for radiation protection and safety – including
the radiological medical
practitioners, the medical radiation technologists, the medical physicists and
the radiation protection
officer. This step cannot be over-emphasized as all aspects of
radiation protection and safety in
medical uses of ionizing radiation depend ultimately on the competence of the
persons involved. See
also paras. 2.117 - 2.130.
2.73. Setting up a medical radiation facility may involve the
construction of facilities which are
difficult to modify at a later time. Regulatory bodies may choose a two-stage
process of authorization,
i.e. to require an initial application to build a facility before
construction begins. At this stage the
regulatory body should review the intended medical uses of ionizing radiation,
the facility’s design,
including structural shielding plans⁵, and the planned equipment. This is
followed at a later stage by
the full regulatory body review and assessment leading to the granting of the
authorization. For more
complex medical radiation facilities, such as a radiation therapy
facility, this latter process should
include an inspection by the regulatory body or authorized party.
2.74. Subsequent substantial modifications of a medical radiation
facility, including its medical
radiological equipment and its procedures, may have safety implications. The
regulatory body may
require an application for an amendment to the authorization.
⁵ Although not strictly a radiation safety issue, it is important to ensure
that the building will support the weight
of the structural shielding, for which it may have not been originally
designed.
2.75. The regulatory body should require the renewal of an authorization after
a set time interval. This
allows a review of the findings of inspections and of other information on the
safety performance of
the medical radiation facility. The frequency of renewal should be based on
radiation protection and
safety criteria, with consideration given to the frequency of inspections by
the regulatory body and the
safety record associated with a given type of practice, in general,
or with a particular medical
radiation facility. A renewal cycle longer than 5 years would seem
not appropriate for medical
radiation facilities.
2.76. The authorization of a medical radiation facility to use ionizing
radiation for medical purposes is
a separate exercise to that of the same facility, or the wider medical facility
of which it is part, being
authorized by the health authority to carry out medicine practice and
healthcare (para 2.52). Radiation
safety requirements are only a set of requirements before authorization is
granted to a medical facility
by the health authority. Meeting radiation safety requirements is a
condition necessary but not
sufficient to obtain an authorization to practice medicine. Coordination and
collaboration between the
radiation protection regulatory body and the health authority should be in
place to ensure radiation
protection and overall safety of the medical facility.
Inspection of medical radiation facilities
2.77. On-site inspection by the regulatory body is often the principal means
for face-to-face contact
with personnel in the medical radiation facility. The regulatory body should
have established a system
for prioritization and frequency of inspections, based on the risk and
complexity associated with the
particular medical uses of ionizing radiation. The regulatory body
inspection of medical radiation
facilities should be performed by staff with the specialist expertise to be
able to competently assess
the compliance of the facility with the radiation protection regulations and
authorization conditions.
For further detailed general guidance on inspections see GS-G-1.5 [22].
Particular considerations for the regulatory body with respect to medical,
occupational and public
exposure
2.78. The regulatory body should ensure that all the BSS requirements
with respect to medical,
occupational and public exposure are implemented in authorized medical
radiation facilities, as
described in detail in the relevant subsections of the specialty
Sections 3 to 5. To help medical
radiation facilities fulfil their obligations, there are some particular areas
where the regulatory body
should provide specific guidance.
2.79. Calibration of sources that give rise to medical exposure should be in
place to ensure radiation
safety in medical uses of ionizing radiation, as set out in the BSS
paragraph 3.167, and detailed
guidance is given in Sections 3 to 5. The regulatory body should
specify frequencies for re-
calibrations and, in doing so, should make use of applicable guidance
given by medical physics
professional bodies.
2.80. In the case of the calibration of radiation therapy units, independent
verification prior to clinical
use should be assured (BSS para. 3.167(c)). The regulatory body should be aware
of the limitations on
local resources in their State. The ideal independent verification, by
different independent medical
physicist using different dosimetry equipment, may not be realizable.
The regulatory body has the
responsibility of not allowing the radiation safety of the radiation therapy
unit to be compromised but
at the same time not to unnecessarily close down the facility. The
regulatory body should decide on
acceptable alternatives, such as verification by a second medical physicist
with the same equipment or
only verification using a second set of equipment, or using a form
of verification by postal
thermoluminescence, optically stimulated luminescence dosimetry or equivalent.
2.81. Unintended and accidental medical exposures do occur and the regulatory
body should require
that a system is in place and all practical measures are taken to prevent them,
and, if such exposure
happens, it is properly investigated and corrective actions are implemented.
Arrangements should be
in place to respond promptly in order to mitigate any consequences.
The regulatory body should
require written records to be kept of all unintended and accidental
medical exposures and should
provide guidelines on what information to be included in these reports. The
more significant events
should be reported to the regulatory body, (BSS para. 3.181(d)). The regulatory
body should provide
guidance on what events must be reported to them. One of the reasons for
reporting to the regulatory
body is to enable the regulatory body, in turn, to disseminate
information on the event to relevant
parties so that the recurrence of similar events is minimized. In addition to
mandatory reporting for
regulatory purposes, anonymous and voluntary safety reporting and learning
systems can significantly
contribute to enhance radiation safety and quality in health care.
The regulatory body can be pro-
active and encourage medical radiation facilities to participate in
relevant international or national
anonymous and voluntary safety reporting and learning systems, as discussed in
para 2.68. Further
guidance on this topic is given in the specialty Sections 3 to 5.
2.82. With respect to occupational exposure assessment, the regulatory body
should set requirements
and provide clear guidance on what form of monitoring should be in
place. The BSS (BSS [3],
paragraphs 3.99 to 3.102) requires the employers, registrants and licensees to
make arrangements for
occupational exposure assessment, and in the requirements gives broad criteria
for when individual
monitoring should be arranged and when workplace monitoring may be
sufficient. Occupational
exposures vary widely in medical uses of ionizing radiation, ranging from those
uses where it is quite
clear that individual monitoring should be undertaken, to those uses
where workplace monitoring
would suffice. It is where uses fall between these two situations
that specific direction should be
provided from the regulatory body. Further guidance on this topic is given in
the specialty Sections 3
to 5.
2.83. The regulatory body has a role as custodian of public radiation
protection. Because a member of
the public can be subject to exposure arising from any number of
authorized medical radiation
facilities (or indeed other facilities and activities using radiation), the
regulatory body has an oversight
role to ensure that the sum or cumulative effect of these multiple pathways
does not lead to public
exposure greater than the dose limits (see Table 2). Part of this
role includes the setting of dose
constraints and ensuring that safety assessments include considerations of
public and potential public
exposure.
2.84. The BSS has many requirements for registrants and licensees, and
employers with respect to
occupational radiation protection, to maintain and make available records on a
wide range of matters.
For records pertaining to occupational exposure, the BSS specifies how long
such records should be
maintained, namely to at least until the worker attains the age of 75 years and
for not less than 30
years after cessation of work in which the worker was subject to occupational
exposure (BSS, para.
3.104). For all other records, the period for which they should be maintained
is deferred to the State’s
regulatory body. The period of retention will depend on the type of
record and its usefulness or
relevance after the passage of time. Records relating to a person’s
health or healthcare arguably
should be kept for that persons’ lifetime, but there are significant variations
around the world. Some
States, for example, require medical records to be kept for the lifetime plus
10 years; others require a
much shorter period such as 7 to 10 years. Records for activities
such as calibrations, dosimetry,
quality assurance and investigations of accidents and unintended medical
exposures should be kept for
a significant period as there is always the possibility that the
records will be needed to perform
retrospective assessments of medical, occupational or public exposure. A
retention period of at least
10 years is recommended for these records. On the other hand,
records on personnel education,
training, qualification and competence may be of relevance only when that
person is working at the
medical radiation facility. Further guidance for the regulatory body and for
registrants, licensees and
employers is given in the IAEA Safety Guide on management systems [25].
Authorization for the installation, maintenance and servicing of medical
radiological equipment
2.85. The regulatory body should ensure that persons who install,
maintain or service medical
radiological equipment are appropriately authorized. See also the sub-sections
on responsibilities for
suppliers of sources, equipment and software, and for maintenance and servicing
organizations, and
the guidance on education, training, qualification and competence of
servicing engineers and
technicians given in para. 2.133.
Authorization of other practices related to medical uses of ionizing radiation
2.86. The regulatory body may also require authorization for other activities
related to medical uses of
ionizing radiation, including: import, distribution, assembly, sale, transfer
or transport of radioactive
sources or medical radiological equipment; decommissioning; disposal
of radioactive sources,
material or waste. The requirements to carry out these practices
should have been established by
regulations complemented by regulatory guidance documents.
Dissemination of information
2.87. The BSS (BSS [3], paragraph 2.38) requires that the regulatory body has
mechanisms in place
for the timely dissemination of information - in the context of this Safety
Guide - to medical radiation
facilities, manufacturers and suppliers, the health authority and professional
bodies, on lessons learned
for radiation protection and safety resulting from regulatory experience and
operating experience, and
from incidents and accidents and related findings. Information should
be exchanged through the
publication of newsletters (paper based or electronic) and the
periodic mailing of notices, by
presentations at scientific and professional association meetings, by
establishing a web site, or by co-
sponsoring educational seminars and workshops with professional and scientific
associations. More
rapid actions should be considered in response to actual or potential
problems that may result in
significant consequences.
Medical radiation facility
2.88. In medical uses of ionizing radiation, the prime responsibility for
radiation protection and safety
rests with the person or organization responsible for the medical radiation
facility – normally referred
to as the registrant or licensee. Almost all the requirements in the
BSS applicable to a medical
radiation facility for ensuring radiation safety in medical uses of
ionizing radiation place the
responsibility on the registrant or licensee (and on the employer, in the case
of occupational radiation
protection).
2.89. However medical uses of ionizing radiation involve a
multidisciplinary team led by a health
professional who often is not the registrant or licensee of the authorized
medical radiation facility.
Because of the medical setting in which such exposures occur, primary
responsibility for radiation
protection and safety for patients lies with the health professional
responsible for the radiological
procedure, who is referred to in the BSS and in this Safety Guide
as the ‘radiological medical
practitioner’. The radiological medical practitioner is the generic term that
the BSS uses to refer to a
health professional with specialist education and training in medical
uses of radiation, who is
competent to perform independently or to oversee procedures involving medical
exposure in a given
specialty. Health professionals that could take on the role of the
radiological medical practitioner,
depending on the particular use of radiation and on the laws and
regulations in a State, include
radiologists, nuclear medicine physicians, radiation oncologists,
cardiologists, orthopaedic surgeons,
other specialist physicians, dentists, chiropractors and podiatrists.
More guidance on the health
professionals who could be radiological medical practitioners is given in the
specialty Sections 3 to 5.
See also paras 2.122 - 2.123 on education and training.
2.90. The net effect of paras. 2.88 and 2.89 is that, for medical exposures,
the registrant or licensee
must ensure all requirements are implemented. This normally requires that the
radiological medical
practitioner ensures a given set of actions takes place, usually with the
involvement of further health
professionals – mainly medical radiation technologists and medical
physicists (see paras. 2.91 and
2.92, respectively). The medical exposure sub-sections of the specialty
Sections give guidance on the
many requirements that come under the responsibility of the radiological
medical practitioner.
2.91. The term medical radiation technologist is used in the BSS and this
Safety Guide as the generic
term for a second group of health professionals. A wide variety of
terms are used throughout the
world such as radiographer, radiological technologist, nuclear medicine
technologist, and radiation
therapist. In the BSS, a medical radiation technologist is a health
professional with specialist
education and training in medical radiation technology, competent
to carry out radiological
procedures, on delegation from the radiological medical practitioner, in one of
more of the specialties
of medical radiation technology (e.g. diagnostic radiology, radiation therapy,
nuclear medicine). The
medical radiation technologist is usually the interface between the
radiological medical practitioner
and the patient, and his/her skill and care in the choice of techniques and
parameters determines to a
large extent the practical realization of the optimization of radiation
protection and safety for a given
patient’s exposure in many modalities. More guidance on the roles and
responsibilities of medical
radiation technologists is given in the specialty Sections 3 to 5.
See also paras 2.124 - 2.125 on
education and training.
2.92. In the BSS, a medical physicist is a health professional with specialist
education and training in
the concepts and techniques of applying physics in medicine, and competent to
practise independently
in one or more of the subfields (specialties) of medical physics, (e.g.
diagnostic radiology, radiation
therapy, nuclear medicine). The medical physicist provides specialist
expertise with respect to
radiation protection of the patient. The medical physicist has responsibilities
in the implementation of
the optimization of radiation protection and safety in medical exposures,
including source calibration,
clinical dosimetry, image quality and patient dose assessment, and
physical aspects of the quality
assurance programme, including medical radiological equipment acceptance and
commissioning. The
medical physicist is also likely to have responsibilities in providing
radiation protection and safety
training for health professionals. In addition, he/she may also
perform the role of the radiation
protection officer (RPO), whose responsibilities are primarily in
occupational and public radiation
protection. More guidance on the roles and responsibilities of medical
physicists is given in the
specialty Sections 3 to 5, and in Ref [26]. See also paras 2.126 - 2.127 on
education and training.
2.93. There are other health professionals with responsibilities for radiation
protection of the patient.
These include, for example, radiopharmacists, radiochemists, dosimetrists and
biomedical or clinical
engineers. Detailed guidance is given in Sections 3 to 5.
2.94. For a medical radiation facility, the radiation protection and
safety responsibilities outlined
above for the radiological medical practitioner, the medical radiation
technologist, the medical
physicist and other health professionals with responsibilities for patient
radiation protection should be
assigned through an authorization (or other regulatory means) issued
by the radiation protection
regulatory body in that State.
2.95. The radiation protection officer (RPO) is a person technically competent
in radiation protection
and safety matters relevant for a given type of practice who is designated by
the registrant, licensee or
employer to oversee the application of relevant requirements [3]. For a medical
radiation facility, the
RPO oversee the application of requirements for occupational and
public radiation protection, and
may provide general radiation protection advice to the registrant or licensee.
The RPO have no direct
responsibilities or roles with respect to patient radiation protection.
In many medical radiation
facilities, especially smaller facilities with fewer personnel, the
role of the RPO and the medical
physicist are often performed by the same person who is a medical
physicist. However an RPO,
unless he or she has recognized competence in medical physics, cannot perform
the role of a medical
physicist with respect to medical exposure.
2.96. In addition to the above paragraphs, all health professionals
involved in the medical uses of
ionizing radiation have responsibilities with respect to occupational and
public radiation protection.
(See the occupational and public radiation protection sub-sections of the
specialty Sections 3 to 5.)
2.97. Medical radiation facilities, as they increasingly utilize
digital technologies, should assure
access to an information technology (IT) specialist⁶ who, through specialized
training and experience,
would have responsibilities for the maintenance and quality control of
information technology
software and hardware. The correct functioning of these systems is crucial for
radiation protection and
safety.
Ethics committees
2.98. Participants in a programme of biomedical research may be either
patients, with some disease or
ailment, or they may be healthy individuals. Regardless, they must be
volunteers. The ethics
committee ⁷ has a particular responsibility with respect to
justification of medical exposure of
volunteers exposed as part of a programme of biomedical research (BSS [3]
paragraph 3.161). The
first part of this responsibility is to decide whether to approve the programme
of biomedical research,
including the proposed use of radiation. The use of radiation in a programme of
biomedical research
can include: (a) the use of a diagnostic radiological procedure to assess the
efficacy of the treatment
under investigation (e.g. ranging from a DEXA scan to measure bone mineral
density before, during
and after a given treatment regime, to a CT or a PET-CT examination
to assess some clinical
indicators, again performed before, during and after the treatment);
(b) trials being performed to
assess a new radiopharmaceutical (i.e. the radiation itself is part of the
research, rather than a tool for
assessment); (c) trials being performed to assess a new radiotherapy protocol
alone or in combination
with other therapeutic modalities ; (d) trials being performed to compare
radiological procedures, for
example specificities and sensitivities of different imaging procedures
or efficacy of different
treatments. In making its decision, the ethics committee should be presented
with correct information
on the expected doses and estimates of the radiation risks based on the age,
gender and health status of
the participants. The ethics committee also should obtain information
on who will perform the
radiological procedures and how. The dose estimates and the associated
radiation risks should be
⁶ The IT specialist in this respect is an expert in imaging informatics, with
expertise to improve the efficiency,
accuracy, usability, reliability and interconnectivity of medical
imaging and radiotherapy services within the
medical radiation facility and, if relevant, its parent healthcare facility.
⁷ Ethics committee is the term used in the BSS to refer to a committee
dedicated to the rights and well-being of
research subjects. Other terms such as an Institutional Review Board are used
in some States.
assessed by a medical physicist. This information should be then considered by
the ethics committee
together with the information on the other risks and benefits of the programme.
2.99. The ethics committee has the responsibility to specify any dose
constraints that are to be applied
to the medical exposures incurred as part of the approved programme of
biomedical research. Such
dose constraints would be guided by nationally or regionally established dose
constraints (see para.
2.50). Dose constraints should be adjusted to the expected benefit of the
programme of biomedical
research – the lower the benefit to society, the more stringent the
dose constraint. The ICRP [27]
stratified doses incurred in biomedical research according to radiation risk
and in Publication 103 [4]
assigned numerical values of dose constraints ranging from less than 0.1 mSv to
greater than 10 mSv,
as the benefit to society ranged from “minor” through to “substantial”. Less
stringent dose constraints
may be applied to participants with short life expectancy, see for
example Ref [28]. Particular
attention should be given to setting dose constraints for healthy volunteers
who repeatedly take part in
biomedical research programs which expose them to increased risks.
2.100. Ethics committees may not be aware of these responsibilities. Therefore
it is desirable for the
radiation protection regulatory body to act as a facilitator in
promoting systems so that ethics
committees know about their responsibilities when a proposal for a
programme of biomedical
research which includes radiation exposure is submitted to the ethics
committee. Such a system may
include a “box” in the proposal form that asks the question “Will ionizing
radiation be used as part of
this programme of biomedical research?” And if the answer is yes,
the form should then ask for
information on radiation doses and risks to be provided, having been first
assessed and signed off by a
medical physicist.
2.101. In a parallel initiative, the regulatory body should inform
the registrants and licensees that
radiological procedures requested as part of a programme of biomedical research
are only justified if
that programme has been approved by an ethics committee, and that such an
approval is subject to
dose constraints which would then influence how the procedure would be
performed.
Suppliers of sources, equipment and software
2.102. Suppliers⁸ of medical radiological equipment and developers of software
that could influence
the delivery of the medical exposure have responsibilities with respect
to design and performance.
Generic requirements are given in the BSS paragraph 3.49 and specific
requirements in paragraph
3.161.
2.103. A particular issue with medical radiological equipment and
software in medical uses of
ionizing radiation is that of the language, terminology and icons used on
control panels, on software
screens and in instruction manuals. English and other major languages dominate.
However it is crucial
that the person using the equipment or software fully understands the options
being presented, and
translation into a local language is advisable. A passing knowledge of a major
language is not good
enough – there are documented instances of unintended or accidental medical
exposures arising from
incorrect understanding of the displayed language, see for example Ref [29].
2.104. Many items of medical radiological equipment can be configured and
supplied with different
options. For example, protective tools may be an optional extra, with
a higher price. Basic model
⁸ The definition of supplier (of a source) in the BSS [3] includes
designers, manufacturers, producers,
constructors, assemblers, installers, distributors, sellers, exporters or
importers of a source.
versions of a given piece of equipment should include as a default all the
relevant protective tools and
the features that provide the greatest control over patient radiation
protection. Paring the price back by
removing radiation protection and safety options in order to gain a
sale is not acceptable. Facility
management should not be placed in a position of saving money at
the expense of compromising
radiation safety.
2.105. When medical radiological equipment and software are to be
part of a digital network,
suppliers should facilitate interconnectivity with other relevant systems.
2.106. After installation of medical radiological equipment or
software, the supplier should go
through a formal hand-over to the medical radiation facility’s
registrant or licensee. This should
include the acceptance testing, described in more detail in the specialty
Sections 3 to 5. It should also
include ensuring that specific training in the use of the equipment or software
is given to the medical
radiation facility’s staff, including the radiological medical
practitioners, the medical radiation
technologists, the medical physicists and the local maintenance
engineers. The features of the
equipment or software should be fully understood, including their implications
for patient radiation
protection.
2.107. The radiation protection and safety responsibilities of
suppliers of refurbished medical
radiological equipment should be no different to the responsibilities for the
supply of new equipment.
Further guidance on refurbished equipment is given in Refs [30 – 32].
2.108. The radiation protection and safety responsibilities for donors
of medical radiological
equipment should be no different to those of commercial suppliers for
such equipment. See WHO
Refs for further guidance on donated equipment [33, 34].
2.109. Regulatory control of engineers and technicians who install
medical radiological equipment
varies around the world. In many countries they will be licensed to perform
installation and servicing
and a pre-requisite to obtaining such a licence would be that they
have had appropriate radiation
protection and safety training. Guidance on education, training,
qualification and competence of
installation and servicing personnel is given in para. 2.133.
Maintenance and servicing organizations
2.110. Maintenance and servicing of medical radiological equipment is
usually performed by an
engineer or technician employed either by a company offering such services (who
may also be the
manufacturer and/or the vendor) or by the medical facility itself (as
part of an engineering, bio-
medical / clinical engineering or service department, for example). In either
case, when the medical
radiological equipment is being serviced the equipment is not to be
used for medical exposures –
patients are not to be imaged or treated until service and hand back is
completed (see para 2.111). The
engineer or technician should follow both the radiation protection and
safety rules and procedures
established by his or her employer and the relevant rules and
procedures of the medical radiation
facility, including how to ensure a safe working environment for the
service and how to ensure
restricted access to the area where the service is taking place. Further
guidance on good maintenance
service practice is given in Refs [35].
2.111. Maintenance and servicing continues until the medical radiological
equipment is ready to be
handed back to the medical radiation facility’s registrant or
licensee. It is recommended that the
“hand-over” is formalized. Depending on the maintenance or servicing that has
taken place, there may
be a need for quality control tests to be performed by a medical
physicist before the hand-over is
complete (see Sections 3 – 5, paras. 3.45, 4.58, 5.87, respectively). The
engineering service should
collaborate with medical physicists and radiological practitioners in
setting optimal equipment
performance. The engineer or technician also should inform the registrant or
licensee of any changes
with respect to the medical radiological equipment that may have implications
for radiation safety. At
this stage the equipment is available for medical use. Pressures to
hand medical radiological
equipment back for medical use must not be allowed to compromise radiation
protection and safety –
for example, equipment being used clinically while still in “service mode”.
2.112. Regulatory control of servicing engineers and technicians varies
around the world. In many
countries they will be licensed to perform servicing and a pre-requisite to
obtaining such a licence
would be that they have had appropriate education and training in
radiation protection and safety.
Guidance on education, training, qualification and competence of servicing
engineers and technicians
is given in paragraph 2.133.
Referring medical practitioners
2.113. The health care of the patient is the responsibility of the
physician or health professional
managing the patient. This physician or health professional may decide
that the patient needs to
undergo a radiological procedure, at which point a referral to an appropriate
medical radiation facility
is initiated. “Referring medical practitioner” is the generic term
used in the BSS for the health
professional who initiates the process that may lead to the radiological
procedure being performed.
There may be different requirements in different States about who can act in
the role of a referring
medical practitioner. The referring medical practitioner has a joint
responsibility with the radiological
medical practitioner to decide on the justification of the proposed
radiological procedure. More
detailed guidance is given in the specialty Sections 3 to 5.
2.114. Usually the roles of the referring medical practitioner and the
radiological medical practitioner
are performed by two different persons. However there are some
instances where both roles are
performed by the same person – often called self-referral. A very common
example is the dentist, who
decides whether an X ray examination is needed and, if so, performs
the examination. Dental
professional bodies in many countries have established guidelines for when
dental X ray examinations
are appropriate or not, and using these guidelines should help the
dentist to fulfil both roles
acceptably. In other situations, typically involving medical imaging,
there may be very strong
financial incentives for self-referral because the performance of the
radiological procedure generates
significant income. Again there is a clear role for professional body
guidelines to help minimize
potential misuses of self-referral.
Patients
2.115. Patients are being increasingly involved in the decision making
processes concerning their own
health care, and this includes medical uses of ionizing radiation. The BSS
requires that the registrant
or licensee for the medical radiation facility ensures that the patient is
informed, as appropriate, of
both the potential benefit of the radiological procedure and the radiation
risks (BSS [3], paragraph
3.151(d). Information should always be provided in an understandable format
(e.g. verbally, leaflets,
posters, websites), and in a timely manner. The level of information should be
commensurate with the
complexity, dose and associated risks, and for some radiological procedures
informed consent may be
required, written or verbal. Female patients of childbearing potential
should be informed about the
risk to the embryo or fetus from radiological procedures for either diagnosis
or therapy.
2.116. “Self-presenting” patients are individuals demanding a particular
radiological procedure on the
basis that they believe that this procedure is needed to, for example, detect
cancer or heart disease in
its early stages before symptoms become manifest. These individuals should be
handled in the same
way as any other patient – namely through an appropriate referral and the
ensuing justification.
EDUCATION, TRAINING, QUALIFICATION AND COMPETENCE
2.117. Medical uses of ionizing radiation involve a number of health
professionals performing
radiological procedures – diagnostic examinations, interventional
procedures or treatment. In each
case the radiation protection and safety associated with the radiological
procedure depends strongly
on the skills and expertise of those health professionals involved as
the patient is necessarily and
deliberately exposed to radiation. In other words, education, training,
qualification and competence of
the respective health professionals underpin radiation safety in medical uses
of ionizing radiation.
2.118. The BSS places great emphasis on education and training for all persons
engaged in activities
relevant to protection and safety, with the responsibility placed on
government to ensure that
requirements for education, training, qualification and competence
are established and that
arrangements are in place for the provision of the necessary education and
training. The development
and implementation of a national strategy for education and training⁹
which is based on a national
needs assessment can be useful in this context. Further, the
regulatory body should ensure the
application of the requirements for education, training, qualification
and competence in radiation
protection. This should take place when an authorization application
has been submitted to the
regulatory body and during the periodic inspections of the medical
radiation facility. Finally, the
registrant or licensee of the medical radiation facility has the responsibility
to ensure that all the health
professionals in that facility with responsibilities for protection and safety
have appropriate education,
training, qualification and competence.
2.119. In medical uses of ionizing radiation medical exposure occurs,
and occupational and public
exposure may occur. For the health professionals involved, it is the education,
training, qualification
and competence in the medical exposure aspects that are the most
critical. To this end, the
requirements in the BSS for the health professionals involved in performing
radiological procedures
are quite stringent. For each of the key roles of the radiological
medical practitioner, the medical
radiation technologist, the medical physicist and the radiopharmacist, the
definition in the BSS takes
the same form. Namely, that the person is a health professional, that they have
specialist education
and training in the particular discipline (including radiation protection and
safety), and that they have
been assessed as being competent to carry out that particular role.
See the BSS for the complete
definitions (BSS [3], pp 114, 115, 122, 123]). The competence of a person is
normally assessed by the
State by having a formal mechanism for registration, accreditation or
certification of the particular
specialized health professional. States that have yet to develop such a
mechanism, should assess the
education, training and competence of an individual proposed by a
licensee to act as a specialized
⁹ IAEA Strategic Approach to Education and Training in Radiation, Transport and
Waste Safety, 2011–2020
(IAEA/2010/Note 44)
health professional and to decide, on the basis either of international
standards or standards of a State
where such a system exists, whether the individual can be considered competent.
2.120. A health professional intending to act in any of the roles of
radiological medical practitioner,
medical radiation technologist, medical physicist or radiopharmacist can do so
only if he or she has
the requisite education, training, qualification and competence. It is
the prime responsibility of the
registrant or licensee to ensure that their staff meets these requirements, and
it is the responsibility of
the regulatory body to use the authorization, inspection and
enforcement processes to ensure that
registrants and licensees are discharging their responsibilities in this
respect.
2.121. The institutes and organizations that provide education and training in
radiation protection to
health professionals should use the BSS and its companion Safety
Guides as resources on the
requirements for radiation protection and safety in medical uses of radiation.
Radiological medical practitioners
2.122. The term radiological medical practitioner is applied to a number of
health professionals who
independently perform or oversee radiological procedures within a given
specialty (see also para.
2.89). Some of these health professionals belong to a specialty with
a very long association with
medical uses of ionizing radiation, such as radiology, nuclear
medicine, radiation therapy and
dentistry. In countries where there are well established processes in
place for education, training,
qualification and competence in these specialties, this includes subjects not
only in the specialty itself
but also with respect to radiation protection (patient and
occupational). Typically these specialists
would become registered with the national medical or dental
registration board (or a body with a
similar function), and competence in the specialty should include competence in
radiation protection
and safety. It still falls on the regulatory body and the relevant
professional body to periodically
review the radiation protection and safety aspects of the education and
training to ensure that it is still
up to date and relevant. In States where there is a lack of infrastructure for
education and training in
these specialties, a prospective radiological medical practitioner should gain
the necessary education,
training, and qualification outside the State, both in the specialty itself and
in radiation protection and
safety. The competence of health professionals trained outside the State
should be assessed. In this
situation the regulatory body should seek advice from the health
authority and the relevant
professional body (if it exists), with respect to the adequacy of the
specialization of the individual.
Assessment of the radiation protection and safety aspects would be performed by
the regulatory body.
In time this approach should develop into a standardized process for
dealing with competence
assessments.
2.123. Other specialties, such as orthopaedic surgery and cardiology, have also
had a long association
with medical uses of ionizing radiations, but radiation protection and safety
may not traditionally have
been part of the processes for education, training, qualification and
competence in the specialty. Still
other specialties have a more recent association with medical uses of
ionizing radiation, especially
with respect to image guided interventional procedures. Radiation
protection (patient and
occupational) is often not included in the curriculum for education,
training, qualification and
competence in these specialties. For specialists from these two
groups, oftenadditional or separate
education and training and credentialing in radiation protection and safety
should be arranged, as it
applies to their specialty. The relevant professional bodies and the
regulatory body should work
together in establishing acceptable criteria on education and training
in radiation protection and
safety, and the means for recognition of competence in radiation protection.
Probably the preferred
approach would be for the relevant professional body to administer the process
and maintain a register
of specialists and their radiation protection and safety credentials.
Other possibilities include the
regulatory body taking on the role of overseeing the radiation
protection and safety training and
recognition processes. An individual medical radiation facility can
adopt a ‘credentialing and
privileging’ approach to cover radiation protection and safety education,
training, qualification and
competence [36]. In this approach the prospective radiological medical
practitioner would present all
their relevant data regarding training and experience (including in
radiation protection and safety),
and apply for permission to perform certain medical procedures
involving radiological procedures.
Detailed guidance on appropriate radiation protection and safety education and
training for various
specialties involved in medical use of ionizing radiation is given in
references [37, 38].
Medical radiation technologists
2.124. Medical radiation technologists should require a programme of
education and training in
medical radiation technology that usually includes significant
components of radiation protection
(patient and occupational). On completion of the programme the medical
radiation technologist
typically would become registered with the national registration board
(or a body with a similar
function), and this competence in medical radiation technology should
include competence in
radiation protection and safety.
2.125. Medical radiation technologists may be specialized in various
fields and sub-fields. The
approach to specialties and sub-specialties vary significantly among
countries. In many States, the
medical radiation technologist undergoes a programme specific to
diagnostic radiology, nuclear
medicine or radiation therapy and hence his/her competence would be in that
specialty only. Within
these specialties there may be specific sub-specialties, where the
programme does not necessarily
confer competence. For example, the diagnostic radiology programme in a country
may not cover CT
or image guided interventional procedures to the depth needed for competence.
Additional education
and training should be arranged to achieve competency in the sub-specialty. The
regulatory body, in
terms of reviewing the application for an authorization and during its periodic
inspections, needs to be
aware of issues of specialization and sub-specialization and ensure that only
persons with the correct
credentials can work in the particular roles. Likewise, the registrant or
licensee should ensure that they
employ only persons that have the requisite competence.
Medical physicists
2.126. Even though the International Labour Organization has stated that
medical physicists working
in clinical practice can be considered health professionals [39], the
recognition of medical physicists
as a specialist group of health professionals is not as widespread as needed.
In some countries there
are well established processes for education, training and qualification and
achieving competence in
medical physics, with academic training in medical physics at a university
(typically a post-graduate
programme), clinical training in a hospital or facility, and finally an
assessment of competence. In
some States, the professional body administers this whole process, with
approved universities for the
academic component, approved hospitals or facilities for the clinical
placement, and a professional
standards board for the competence assessment. More details on education,
training, qualification and
competence of medical physicists is given by the IAEA [26, 40 – 43]. There are
also national and
regional requirements and guidance on education, training and recognition of
medical physics expert
[44].The BSS requires specialization for the medical physicist, so, for
example, a medical physicist
with competence only in diagnostic radiology or image guided interventional
procedures cannot act in
the role of a medical physicist in radiation therapy, and vice versa.
2.127. More difficult is where either the State does not recognize medical
physics as a distinct health
profession or where there is no infrastructure in place for the
education and training of medical
physicists. In both cases there is likely to be little in the way of
infrastructure for medical physics in
the country. The problem is similar to that described above in the
second half of para. 2.123 for
radiological medical practitioners. The assessment of education,
training, qualification and
competence of a person seeking to act in the role of a medical
physicist still should take place.
Regardless of the educational process, the final competence assessment should
be specialty-specific as
required by the BSS (BSS [3], para. 3.150).
Radiopharmacists
2.128. A radiopharmacist is usually a pharmacist who has received additional
specialist education and
training and has competency in the preparation and dispensing of
radiopharmaceuticals. Post-graduate
courses in radiopharmacy are available in some countries. A few
countries have a radiopharmacy
professional body, or radiopharmacy can be a specialist sub-group
within the national nuclear
medicine professional body and/or a pharmacy professional body. More details on
education, training,
qualification and competence of persons working in a radiopharmacy is given by
the IAEA [45]. Even
in the absence of a formal infrastructure, the assessment of
education, training, qualification and
competence of a person seeking to act in the role of a radiopharmacist still
should take place.
Other health professionals in the medical radiation facility
2.129. Other health professionals are involved in medical uses of
ionizing radiation. However a
distinction should be made between those who have specific
responsibilities for patient radiation
protection and those whose responsibilities (in terms of radiation
protection) are for occupational
radiation protection only. Health professionals who fall into the
former group, and who are not a
radiological medical practitioner, a medical radiation
technologist, a medical physicist, or a
radiopharmacist, still should have appropriate specialization (as it applies to
the radiation use) and the
respective radiation protection and safety education, training,
qualification and competence. The
guidance given above for health professionals where infrastructure is
lacking would again be
applicable.
2.130. The latter group of health professionals and other professionals
involved in medical uses of
ionizing radiation include specialist nurses (working in a cardiac
investigation suite or theatre or
similar), specialist physicians (such as anaesthetists providing support
to a patient undergoing an
interventional procedure), biomedical engineers, clinical engineers,
radiochemists providing support
to the performance of the radiological procedure, either directly or
indirectly. All these persons should
have formal education and training on radiation protection. An example of such
training for radiation
oncology nurses is given in Ref [46].
Referring medical practitioners
2.131. The referring medical practitioner has a crucial role in the
justification of a given radiological
procedure for a given patient. The referring medical practitioner would be more
effective in this role if
he/she has a good understanding of radiation protection and safety as it
applies to medical uses of
ionizing radiation. Formal processes to require such education and
training under a radiation
protection and safety framework are difficult to put in place. Instead, a more
general approach may be
adopted of promoting education and training in radiation protection and safety
as part of the general
medicine degree curriculum, especially at the time when clinical rotations
begin, and/or as part of the
corresponding specialty education and training programme.
Radiation protection officers
2.132. As defined in the BSS and stated above in para. 2.95, the
RPO should be competent in
radiation protection and safety matters with respect to occupational and public
radiation protection,
relevant for given medical uses of ionizing radiation [3]. The RPO
could come from a range of
backgrounds, often in science, engineering, or health. The additional education
and training required
for the RPO role will depend on the complexity of the technology
and practice of the medical
radiation facility. In some facilities, the RPO may lead a team, all of whom
should have the requisite
education and training. As above for the health professionals, in the
absence of a third-party
recognition process, the regulatory body should liaise with the relevant
professional body (if it exists)
to set standards to allow assessment of persons seeking authorization to act in
the role of RPO. The
ILO has recognized the radiation protection expert as an “Environmental and
occupational health and
hygiene professional” [39].
Suppliers, installation, maintenance and servicing personnel
2.133. Persons who work as engineers or technicians for supply,
installation, maintenance and
servicing of radiological medical equipment and software should be qualified
and competent in such
work. Often they will have been trained by their employer specifically for this
role. Another aspect of
their training should be in the area of radiation protection and
safety – not only for their own
occupational radiation protection and radiation protection of the staff of the
medical radiation facility
where they are working, but also they should have a good working knowledge of
patient radiation
protection in the context of the types of medical radiological
equipment and software they are
servicing. For the last, this particularly includes understanding the
radiation protection and safety
implications of the various features of the equipment or software,
and how that changes when the
features undergo adjustments or revisions. Regulatory control of servicing
engineers and technicians
varies around the world. In some countries licence may be required to perform
servicing and a pre-
requisite to obtaining such a licence should be that they have had appropriate
radiation protection and
safety training.
Maintaining competence (CME and CPD)
2.134. The above paragraphs give guidance on the processes for the
initial education, training,
qualification and competence assessment of the health professional. The health
professionals should
maintain their core competencies, including radiation protection and safety,
and keep abreast of new
developments in medical uses of radiation. One way to demonstrate this is
through formal continuing
medical education (CME) or continuing professional development (CPD)
programmes. In many
countries, the professional bodies administer such programmes, and maintenance
of certification of
competence in a specialty is dependent on satisfactory participation in the
programme. The registrants
and licensees and regulatory bodies can use these programmes as evidence of
continuing competence.
Equipment and software specific training
2.135. Specific training should be assured on the actual medical
radiological equipment and the
software used in the medical radiation facility. This applies in
particular to radiological medical
practitioners and the medical radiation technologists who work directly
with the equipment and
software during radiological procedures, and the medical physicist. They should
understand how the
equipment and software function, including the available options and
how to customize these, and
their implications for patient radiation protection. Practical training should
take place in the medical
radiation facility when new equipment or software is installed and when
significant modifications are
made. From the vendors’ side, the servicing engineer, the applications
specialist and the IT specialist
have a role in providing this specific training for the medical radiation
facility.
MANAGEMENT SYSTEM FOR RADIATION PROTECTION AND SAFETY
2.136. The use of radiation in medicine is just one aspect of medical practice.
The application of the
radiation protection and safety requirements of the BSS should fit in
with the wider set of
requirements that ensure good medical practice. To this end, in
particular, the medical radiation
facility and its management should ensure complementarity between the
requirements for radiation
protection and safety and other healthcare delivery requirements within the
medical facility. This is to
be achieved through an appropriate management structure and management system.
2.137. The BSS has a specific requirement for radiation protection
and safety to be effectively
integrated into the overall management system of a given organization (BSS [3],
requirement 5). In
this Safety Guide, this applies to the medical radiation facility.
The BSS has additional detailed
requirements on the protection and safety elements of the management system,
for promoting a safety
culture, and taking into account human factors (BSS [3], paragraphs 2.47 to
2.52). Further detailed
requirements for facilities and activities, in general, are given in the IAEA
Safety Standards Series
No. GS-R-3 [47] and elaborated in the Safety Guide GS-G-3.1 [25].
The requirements for quality
management are established in these standards and will not be discussed further
in this Safety Guide
other than to emphasize that effective management for radiation
protection and safety requires
commitment from the highest level of management in the medical
radiation facility, including the
provision of all the required resources. The following guidance is
limited to a few particular
components of the management system related to radiation protection and safety.
2.138. The BSS has a requirement for a “protection and safety
programme” in general (BSS [3],
paragraphs 2.42 & 2.43) and a “radiation protection programme”
specifically for occupational
exposure (BSS [3], requirement 24). In addition the BSS has requirements for a
“quality assurance
programme for medical exposures” (BSS [3], paragraphs 3.170 – 3.172).
All three of these
programmes should be part of the overall management system of the
medical radiation facility¹⁰.
Detailed guidance on the radiation protection programme for occupational
exposure and the quality
assurance programme for medical exposures is given in the specialty Sections 3
to 5.
2.139. Depending on the size of the medical radiation facility, committees
might be formatedto help
the implementation of the radiation protection and safety programme
aspects of the management
system. One such committee might be a radiation safety committee, with the
function of advising on
safe operation and compliance with radiation protection and safety
regulatory requirements. The
members of the committee should be at the senior level and would typically
include an administrator
representing the management, a radiological medical practitioner, a medical
radiation technologist, a
medical physicist, and the radiation protection officer. For the day-to-day
oversight of the radiation
protection programme, a radiation protection officer should be appointed, who
should report to the
committee. The licensee should ensure that the RPO is provided with
the resources required to
oversee the programme, as well as the authority to communicate with the
committee on a periodic
basis. The RPO should be able to communicate directly with the licensee,
and with the regulatory
body as needed, such as in the case of breaches of compliance which may
compromise safety.
¹⁰ The medical radiation facility may be a “stand alone” entity, such as a
medical imaging centre, or it may be
part of a larger organization, such as a hospital. The focus of
this section on management systems is at the
medical radiation facility level, but it should be recognized that, where the
medical radiation facility is part of a
larger organization, the medical radiation facility management system should be
part of the larger organization’s
management system.
2.140. Another committee might be the quality assurance committee,
with oversight of the quality
assurance programme for medical exposures within the medical radiation
facility. The committee
would determine policy and give direction to the programme, ensure proper
documentation is being
maintained and review the effectiveness of the programme. The
radiation safety and the quality
assurance committees have some functions in common, especially with regard to
medical exposure,
and the health professional representation is likely to be the same.
Harmonization of the work of both
committees is required to avoid either the duplication of or the
inadvertent omission of some
functions.
2.141. Any management system should include continuous quality
improvement which implies a
commitment by staff to strive for continuous improvement in the medical uses of
ionizing radiation.
Feedback from operational experience and from lessons learned from
accidental exposures or near
misses should be used systematically, as part of the continuous quality
improvement.
2.142. The BSS requires that the medical radiation facility is able to
demonstrate effective fulfilment
of the requirements for protection and safety in its management system (BSS
[3], paragraph 2.50).
This will include monitoring, conducted to verify compliance with the
requirements for protection and
safety (BSS [3], requirement 14, and paragraphs 3.37 and 3.38).
2.143. There are requirements for records to be kept, and made available as
needed, in many sections
in the BSS. The management system of the medical radiation facility should
provide for such record
keeping and access. Details on what should be provided are described in
Sections 3 to 5.
2.144. Digital information systems are becoming increasingly available
to provide various support
functions to the management system of the medical radiation facility, including
handling requests for
radiological procedures, scheduling radiological procedures, tracking
patients, and the processing,
storage, and transmission of information pertaining to the patient.
Further, they may be used for
viewing imaging studies and providing reports of study interpretations.
Example of systems with
some or all of these functions include picture archiving and
communication systems (PACSs),
radiology information systems (RIS), hospital information systems (HIS),
and the electronic health
record (EHR). These systems should operate independently, but may also
interconnect with each
other. Imaging devices and other medical radiological equipment can be
interconnected by computer
networks and exchange information in accordance with standards such as
TCP/IP (Transmission
Control Protocol/Internet Protocol, or the Internet protocol suite), DICOM,
HL7, and IHE (Integrating
the Healthcare Enterprise) [48 – 50]. These information systems
are complex and expert
implementation and support should be assured. Digital information systems when
used appropriately
can have a positive effect on the practice of radiation protection and
safety in medical uses of ionizing
radiation. For example, use of these systems can help avoid performing
unnecessary or inappropriate
studies and repeat studies by making patient information available to
multiple users. Further,
connected digital systems should minimize the need for multiple manual
data entry, with its
associated risks, such as in radiation therapy. These systems can
also help in monitoring doses to
patients and image receptors, and monitor retakes; the information from this
monitoring can help in
the implementation of optimization of protection for imaging procedures.
2.145. These digital information systems and procedures for their use should be
designed to protect
against data loss, which in the context of the medical radiation
facility may compromise radiation
protection and safety by, for example, having to repeat examinations. It is
the responsibility of the
medical radiation facility to meet the records retention, security, privacy,
and retrieval requirements of
the relevant State authorities.
2.146. The management system should include a review cycle. The general
principles for audits and
reviews are well established [25, 47]. For a medical radiation facility, a
possible tool for this is the
clinical audit. Clinical audit may be considered as the systematic and critical
analysis of the quality of
clinical care, including the procedures used for diagnosis and
treatment, the associated use of
resources and the effect of care on the outcome and quality of life for the
patient [50]. A clinical audit
looks beyond a strict radiation protection and safety focus, and seeks to
assess the quality and efficacy
of the medical practice offered in the facility – ultimately the
patient health outcome. This should
include the radiation protection and safety aspects of medical uses
of ionizing radiation and,
importantly, keeps these aspects in the context of medical practice, ensuring a
common goal. Thus,
while the BSS does not require a clinical audit, its use may be seen as
fulfilling both the radiation
protection and safety and the medical aspects of the medical radiation
facility’s management system.
More detailed guidance on clinical audits is given in the three IAEA
publications [52 - 54].
2.147. The BSS in the context of medical exposures does require the performance
of a radiological
review and this should be incorporated into the medical radiation facility’s
management system ([3],
paragraph 3.182). At its simplest, the radiological review includes an
investigation and critical review
of the current practical application of the requirements for justification
and optimization of radiation
protection and safety for the radiological procedures that are being performed
in the medical radiation
facility. The radiological review involves at least the radiological medical
practitioners, the medical
radiation technologists and the medical physicists at the medical radiation
facility.
SAFETY ASSESSMENTS
2.148. In the context of medical uses of ionizing radiation, a safety
assessment means an assessment
of all relevant aspects of radiation protection and safety for a medical
radiation facility, including the
siting, design and operation of the facility. The safety assessment
can occur before a facility is
operational or when a major change in operation is contemplated. The safety
assessment deals with
finding ‘what can go wrong’ and how it can be prevented and, in
case it occurs, how it can be
mitigated. As noted above (paragraph 2.70), the regulatory body has the
responsibilities to establish
requirements for safety assessments and, once the safety assessment has been
submitted, to review
and evaluate it prior to granting an authorization (see BSS [3], requirement 13
and paragraph 3.29).
2.149. The BSS gives requirements on: what a safety assessment should include,
what the registrant
or licensee should take into account, its documentation and placement in the
management system, and
when additional reviews of the safety assessment should take place
(BSS [3], paragraphs 3.30 to
3.36). Further more detailed requirements on safety assessment for facilities
and activities, in general,
are given in the IAEA Safety Standards Series No. GSR Part 4 [55]. For medical
radiation facilities,
the safety assessment should include not only considerations of occupational
and public exposure, but
also medical exposure and the possibility of unintended or accidental medical
exposures.
2.150. The BSS specifies two types of safety assessments: generic,
and specific to the facility or
source. A generic safety assessment is usually sufficient for types of sources
with a high degree of
uniformity in design. A specific safety assessment is usually required
in other cases; however, the
specific safety assessment should not include those aspects covered by a
generic safety assessment, if
a generic safety assessment has been conducted for the source. The safety
assessments needed in the
context of medical uses of ionizing radiation will range in complexity, but
even if the source itself is
covered by a generic safety assessment, its placement in the medical
radiation facility will nearly
always require some form of specific safety assessment. It is very
useful if the regulatory body
develops a set of templates [11, 55] to be used by medical radiation facilities
for safety assessments
for the various modalities and specialties in medical uses of ionizing
radiation.
2.151. The BSS requires potential exposure to be considered in the safety
assessment of a new facility
being planned or a planned modification to an existing facility.
Potential exposure refers to
prospective exposure that may or may not occur, but could result from an
accident or from an event or
a sequence of events that may or may not occur. As discussed in paragraph 2.4,
in majority of cases in
medical use of radiation the potential exposure can be treated as a
planned exposure situation.
Facilities should have procedures in place to deal with the occurrence of such
events.
2.152. The BSS (BSS [3], paragraphs 3.43 and 3.44) requires, if the safety
assessment indicates that
there is a reasonable likelihood of an emergency affecting either workers or
member of the public, an
emergency plan to be prepared, including arrangements for the prompt
identification of an emergency,
and for effective response. Situations that can lead to emergency in
medical settings are loss of control
over a dangerous radiation therapy source in result of unauthorized or
malicious act, or conventional
emergencies such as fires and earthquakes. Further more detailed requirements
on emergency
preparedness and response are given in the GSR Part 7 [7].
3. SPECIFIC RECOMMENDATIONS FOR RADIATION PROTECTION AND
SAFETY IN DIAGNOSTIC RADIOLOGY AND IMAGE GUIDED
INTERVENTIONAL PROCEDURES
INTRODUCTION
3.1. This chapter covers radiographic and fluoroscopic diagnostic
procedures, image-guided
interventional procedures, and imaging studies using X-ray radiation
which are part of radiation
therapy or nuclear medicine processes. These radiological procedures
usually take place in permanent
facilities but they can also take place in mobile facilities.
3.2. The radiographic procedures aim to image a particular organ or tissue in 2
or 3 dimensions, and
include general radiography, computed tomography (CT), cone-beam CT
(CBCT), mammography,
tomosynthesis, dental radiography (intraoral, panoramic and CBCT) and
bone densitometry (dual
energy X ray absorptiometry, DEXA).
3.3. Fluoroscopic diagnostic procedures aim to provide real time
assessment of the anatomy and
pathology of a system or organ. Examples include cardiac,
gastrointestinal, urological, and
gynaecological examinations.
3.4. During image-guided interventional procedures, fluoroscopy
(primarily) or CT is used as an
imaging tool to facilitate the diagnosis and treatment of vascular and
non-vascular diseases. Examples
of vascular procedures include coronary angiography/angioplasty, uterine artery
embolizations, aortic
valve implantations and aortic endographs. Common non-vascular procedures
include, for example,
biliary drainage/stenting and liver cytostatic agent injections.
Fluoroscopically guided intra-operative
procedures include, for example, intramedullary nailing and vertebroplasty.
3.5. The generic term “medical radiation facility” is used widely in Section 2
to mean any medical
facility where radiological procedures are performed. In Section 3,
the narrower term “radiology
facility” is used to cover any medical radiation facility where
diagnostic radiology and/or image
guided interventional procedures are performed. A radiology facility includes
the traditional radiology
department in a hospital or medical centre; a stand-alone X ray imaging
facility; the interventional
cardiology (or other specialty) department, unit or facility, either
stand alone or as part of a larger
entity; or a dental practice.
3.6. Many different health professionals can take on the role of the
radiological medical practitioner
(see Section 2 para. 2.89) in diagnostic radiology or image guided
interventional procedures,
depending inter alia on national laws and regulations. They
typically include radiologists,
cardiologists, orthopaedic surgeons, neurosurgeons, plastic
surgeons, vascular surgeons,
gastroenterologists, urologists, respiratory and other specialist
physicians and surgeons, dentists,
chiropractors and podiatrists.
3.7. Section 2 of this Safety Guide provides general guidance on the
framework for radiation
protection and safety in medical uses of radiation, including roles
and responsibilities, education,
training, qualification and competence, and the management system for
protection and safety. This is
relevant to diagnostic radiology and image guided interventional procedures and
reference to Section
2 should be made as indicated.
SAFETY OF MEDICAL RADIATION FACILITIES AND MEDICAL RADIOLOGICAL
EQUIPMENT
Radiology facilities
Fixed facilities – X ray room design
3.8. The BSS, paragraph 3.51, set out the broad requirements that should
be met when choosing a
location and designing a radiology facility. Provisions for the
incorporation of radiation safety
features are best made at the facility design stage (X ray rooms and other
related rooms). The siting
and layout out should take into account the types of radiological
procedures, workload and patient
flow, both within the radiology facility and, in cases where the radiology
facility is part of a larger
hospital or medical centre, with other departments of the wider
facility. Guidance on setting up
diagnostic radiology and interventional radiology facilities is given in Refs
[56 - 58].
3.9. The three factors relevant to dose reduction (time, distance and
shielding) can be combined in the
design to optimize occupational and public radiation protection. Larger rooms
are preferable to allow
easy access for patients on a bed trolley. At the same time they allow for
easier patient positioning and
facilitate both equipment and patient movement during the procedure,
which in the case of
fluoroscopy and image-guided interventional procedures helps reduce time
and exposure. Larger
rooms should also reduce the levels of secondary radiation (scatter and
leakage) potentially reaching
areas occupied by staff and also public areas, typically reducing the level of
shielding required.
3.10. Shielding requirements should be individually tailored to suit the
practice requirements based on
the intended patient workload and the type of examinations to be
undertaken. Further assessments
should be undertaken when the intended use of a room changes; X ray
equipment is upgraded; or
surrounding room occupancy is altered.
3.11. Shielding should consider both structural and ancillary protective
barriers at the design stage
(see para. 2.73]. In rooms using fluoroscopy with staff working close to the
patients, such as rooms
for image guided interventional procedures, ceiling mounted protective
screens and table mounted
leaded curtains should be installed. Such ancillary protective barriers for
image guided interventional
procedures should be part of the initial facility plan, designed so as to not
interfere with the medical
procedure, e.g. sterility requirements. Shielding of walls should be at least
two metres high, and any
doors and viewing windows in walls or doors should have at least the same lead
equivalence as the
minimum shielding specifications for the shielded wall or barrier in
which they are located. Due
consideration should be given to the provision of floor and / or
ceiling shielding when rooms
immediately below and above the X ray installation respectively are occupied.
All penetrations and
joints in shielding should be arranged so that they are equally as
effective in shielding radiation. More
details with respect to structural shielding are given in paras 3.17 to 3.23.
3.12. General safety features of radiography, mammography, CT and fluoroscopy
rooms include:
(a) A protective barrier should be placed at the control console to
shield staff to the extent that
staff should not should wear protective clothing while at the
console. This is particularly
important in mammography where structural shielding in walls, ceiling and floor
may not be
necessary.
(b) For radiography, all possible intended directions of the X
ray beam should be taken into
consideration in the room design so that the X ray beam cannot be directed at
any area which
is not shielded and lead to potentially unacceptable doses being received in
this area.
(c) The doors should be calculated to act as a protective shield for
secondary radiation and be
shut when the X ray beam is on. For radiography, the X ray room should be
designed so as to
avoid the direct incidence of the X ray beam on the access doors.
(d) The medical radiation technologist should be able to clearly observe
the patient at all times
during an X ray diagnostic procedure and be able communicate with him/her.
3.13. Signs and warning lights, preferably positioned at eye level, should be
used at the entrances of
controlled and supervised areas to prevent inadvertent entry (see also para.
3.270 on control of public
access). For controlled areas, the BSS, paragraph 3.90, requires the
use of the symbol of the
International Organization for Standardization (ISO) [59]. The signs
should be clear and easily
understandable. Warning lights, such as illuminated and/or flashing signs as
appropriate, should be
activated when radiation is being produced inside the controlled or supervised
area. Door interlocks
are not appropriate in X ray diagnostic radiological procedures, because
should the X ray beam be
stopped, the medical procedure may have to be repeated. However, to
prevent unauthorized entry
once the procedure has started, some rooms have doors that can be open from the
inside only. This
particularly is used for rooms in which image guided interventional procedures
are performed.
3.14. A stable power supply should be available. An emergency diesel
generator may not be
sufficiently stable to power a CT or interventional radiology suites
and should not be relied upon.
Uninterruptible power supply (UPS) or battery backup systems should be
installed to capture the
active information at time of the outage and to power down all
software in a controlled manner.
Servers should be programmed to automatically shut down when the power supply
is interrupted.
3.15. The design of the facility may should include an air conditioning system
sufficient to maintain
the temperature in the examination room (and sometimes in areas with computer
equipment) within
the range defined by the equipment manufacturers.
Mobile facilities
3.16. Mammography and CT “vans” are commonly used in areas where
fixed facilities are not
available. Other modalities may also be offered via a mobile
facility. General safety features of
mobile facilities include:
(a) Mobile facilities should be built so that protection is optimized mainly
through shielding as
distance is often limited and time depends on the procedure.
(b) An appropriate power supply should be available with reliable connections.
(c) Entrance to the mobile facility should be under the control of the mobile
facility personnel.
(d) Waiting areas, if they exist, should be appropriately shielded to
afford levels of protection
consistent with public exposure limits. Waiting areas are common in mammography
mobile
facilities, but not in CT.
(e) To facilitate the imaging procedure, including patient flow, mobile CT
facilities are usually
operated adjacent to a hospital or clinic, from where they may draw water and
electricity, and
where patients can use the toilets, waiting and changing rooms and have access
to physician
offices. Similarly, mammography mobile facilities may also utilize hospital or
clinic facilities.
Shielding calculation considerations
3.17. Two widely used methodologies for shielding calculations are given in
Refs [60, 61], but other
methodologies are also available and used, e.g. Ref [62], as well as specific
shielding calculations for
the WHIS-RAD X-ray unit¹¹ [63]. The nominal design dose in occupied
areas is derived by the
process of constrained optimization, i.e. selecting a source related dose
constraint, with the condition
that the individual doses from all relevant sources is well below
the dose limits for the persons
occupying the area to be shielded. Nominal design doses are levels of air kerma
used in the design
calculations and evaluation of barriers for the protection of individuals, at a
reference point beyond
the barrier. Specifications for shielding are calculated on the basis
of the attenuation they should
provide to satisfy the nominal design doses.
3.18. The shielding thickness is obtained from the attenuation factor, which is
required to reduce the
dose that would be received by staff and the public if shielding were not
present (a) to a dose value
that can be considered as acceptable, as a result of an optimization process,
i.e. a nominal design dose
derived by a process of optimization (b):
(a) Doses that would be received without shielding are calculated by
using workload values, ‘use
factors’ for a given beam direction (fraction of the total amount of radiation
emitted in that
direction) and ‘occupancy factors’ (fraction of the total exposure
that will actually affect
individuals at a place, by virtue of the time permanence in that place). For
secondary barriers,
the ‘use factor’ is always unity, since scatter and leakage
radiation is propagated in all
directions all the time. If tabulated figures are used, care should be taken
that they reflect the
actual usage in the facility and not generic “national” scenarios.
Potential practice changes
and workload increases should be considered as part of the calculations.
(b) Once the dose that would be received without shielding is
known, attenuation should be
calculated to reduce this dose to a design level or to a level that can be
considered ‘optimized
protection’, i.e. a dose below which additional cost and effort in shielding is
not warranted by
the dose being averted. This may require successive calculations to determine
where this level
lies.
¹¹ World Health Imaging System is a general purpose X-ray equipment built in
accordance with specifications
developed by WHO for developing countries.
3.19. When using a shielding methodology to optimize occupational and public
radiation protection,
decisions should be made about many factors that can greatly
influence the final results for the
shielding specification. Those decisions may be based on conservative
assumptions, which together
may lead to an unduly over-conservative shielding specification. Realistic
assumptions should be used
as much as possible, with some allowance for future changes in use.
Adequateness of the shielding
specification should be ensuredas corrective actions after building has been
completed will invariably
be difficult and expensive. Further, it is likely that the building materials
used to provide the shielding
will come in specific discrete thicknesses or densities and this can
be used to provide a “safety
margin” over the calculated shielding values. If using a material
other than lead, tabulated values
should be used only for materials that exactly match those being
considered in terms of chemical
composition, density, and homogeneity. The following are some assumptions that
will each lead to
conservatism in the shielding specification:
(a) For primary barriers, the attenuation by the patient and image
receptor is not considered;
(b) Workload, use and occupancy factors are overestimated;
(c) Staff members are always in the most exposed place of the room;
(d) Distances are always the minimum possible;
(e) Leakage radiation is the maximum all the time;
(f) Field sizes used for the calculation of scatter radiation are
overestimated;
(g) Attenuation of the materials is usually considered for the maximum
beam quality used;
(h) The numerical value of calculated air kerma (in mGy) is directly
compared with dose limits or
constraints (mSv), which are given in terms of effective dose. However, the
actual effective
dose to personnel or members of the public is substantially lower than the air
kerma, given the
dose distribution within the body for the beam qualities used in diagnostic and
interventional
radiology.
3.20. Particular attention should be given to hybrid imaging systems where the
shielding should be
calculated for each modality and combined as appropriate [58, 64, 65]. See also
Section 4, paras. 4.32
– 4.35.
3.21. Considerations should be given during the design phase to make
sure that radiosensitive
equipment and consumables are appropriately shielded, for example
computed radiography (CR)
cassettes and X ray films. Where used, darkrooms for film processing may
require extra shielding to
prevent film fogging.
Adequacy of shielding
3.22. Specification of shielding, including calculations, should be
prepared by a RPO or medical
physicist. In some countries there may be a requirement for shielding
plans to be submitted to the
regulatory body for review or approval prior to any construction (see also
Section 2 para. 2.73).
3.23. The adequacy of the shielding should be verified, preferably during
construction, and certainly
before the room is placed in clinical use, and similarly after any
future structural modifications.
Clearly requirements of the regulatory body must be met (Section 2 para. 2.73).
Display and interpretation (reading) rooms design
3.24. To facilitate the interpretation by the radiological medical
practitioner, images should be
displayed in rooms specifically designed for these purposes. A proper level
of ambient light in the
viewing room should be ensured. See also paras. 3.40 to 3.41 on
image display devices and view
boxes.
3.25. Viewing rooms with workstations for viewing digital images should be
ergonomically designed
to facilitate image processing and manipulation so that reporting can be
performed accurately. The
viewing monitors of the workstations should meet applicable standards (see
para. 3.40).
Medical radiological equipment, software and ancillary equipment
3.26. This sub-section considers medical radiological equipment,
including its software, used in
diagnostic radiology or image guided interventional
procedures, including radiography,
fluoroscopy/angiography, CT, cone beam CT (CBCT), mammography, dental
radiology, bone mineral
densitometry (e.g. DEXA), tomography (including tomosynthesis). It is also
applicable to the X ray
based component of hybrid imaging modalities, including PET-CT,
SPECT-CT, and PET-
mammography, and the X ray based component of image-guided radiation therapy
systems. Some of
this equipment may be used in a nuclear medicine facility or in a radiation
therapy facility, rather than
a radiology facility.
3.27. The requirements for medical radiological equipment and its
software are given in the BSS
paragraphs 3.49 and 3.162. The International Electrotechnical Commission
(IEC), through its
Technical Committee 62 on Electrical equipment in medical practice and in
particular Sub-committee
62B on Diagnostic imaging equipment and Sub-committee 62C on Equipment
for radiotherapy,
nuclear medicine and radiation dosimetry, has published international standards
applicable to medical
radiological equipment. Current IEC standards relevant to X ray imaging include
the following Refs
[66 – 110]. For those relevant to the radiopharmaceutical based component of
hybrid imaging, see
Section 4 para. 4.40. It is recommended that the IEC website is visited to view
the most up-to-date list
of standards: http://www.iec.ch. The International Organization for
Standardization (ISO), through its
Technical Committee 85 on Nuclear energy, nuclear technologies, and
radiological protection and in
particular Sub-committee 2 on Radiological protection, may
publish international standards
applicable to medical radiological equipment. It is recommended that the
ISO website is visited to
view the most up-to-date list of standards: http://www.iso.org.
3.28. As licensees take responsibility for the radiation safety of medical
radiological equipment they
use, they should impose purchasing specifications that include
conditions to meet relevant
international standards of the IEC and ISO and/or equivalent national
standards. In some countries
there may be a medical devices agency or similar organization that gives type
approval to particular
makes and models of medical radiological equipment.
3.29. Displays, gauges and instructions on the operating consoles of medical
radiological equipment,
and accompanying instruction and safety manuals, may be used by staff who may
not understand, or
who may have a poor understanding of, the manufacturer’s original
language. In such cases, the
accompanying documents should comply with IEC and ISO standards and should be
translated into
the local language or into a language acceptable by the local staff.
The software should be designed so
that it can be easily converted into the local language resulting in displays,
symbols and instructions
that will be understood by the staff. The translations will require a quality
assurance process to ensure
proper understanding and avoid operating errors. The same applies to
maintenance and service
manuals and instructions for maintenance and service engineers and technicians,
where these persons
do not have an adequate understanding of the original language. See also
Section 2 para. 2.103.
3.30. All medical radiological equipment should be supplied with all
appropriate radiation protection
tools as a default, rather than as optional extras. This applies to both
patient radiation protection and
occupational radiation protection. See also Section 2 para. 2.104.
Design features of medical radiological equipment
3.31. The design of medical radiological equipment should be such
that its performance is always
reproducible, accurate and predictable, and that it has features that
facilitate the appropriate personnel
in carrying out the requirement in the BSS for operational
optimization of patient protection
(paragraph 3.163(b)) – namely that it provides “appropriate techniques and
parameters to deliver a
medical exposure of the patient that is the minimum necessary to
fulfil the clinical purpose of the
procedure, with account taken of the relevant norms of acceptable image quality
… .” Many design
features contribute to the performance of medical radiological equipment and
should be considered
when purchasing such equipment, as indicated briefly in the following
paragraphs. Further details on
design features and performance standards of medical radiological
equipment used in diagnostic
radiology or for image guided interventional procedures are given in Refs [72 –
81, 83, 85 - 91, 103,
105 – 115]. See also later paragraphs on quality assurance and acceptance
testing, and in particular
para. 3.231.
3.32. General design features for medical radiological equipment used
in diagnostic radiology and
image guided interventional procedures should consider the following:
(a) Means to immediately detect any malfunction of a single component
of the system that may
lead to an inadvertent under- or over-exposure of the patient or exposure of
staff so that the
risk of any unintended or accidental medical exposure is minimized;
(b) Means to minimize the frequency and impact of human error in the
delivery of unintended or
accidental medical exposure;
(c) Hardware and software controls are incorporated that minimize the
likelihood of unintended
or accidental medical exposures;
(d) All operational parameters for radiation generators, such as
generating tube potential,
filtration, focal spot position and size, source-image receptor
distance, field size indication
and either tube current and time or their product, should be clearly and
accurately shown;
(e) Radiation beam control mechanisms are provided, including
devices that indicate clearly
(visually and/or audibly) and in a fail-safe manner when the beam is ‘on’;
(f) X ray tubes should have inherent and added filtration
adequate to remove low energy
component of the X ray beam which do not provide diagnostic information;
(g) Collimating devices to define the radiation beam; in the case of a
light beam diaphragm the
light field should align with the radiation field;
(h) With the exception of mammography and CT equipment, diagnostic and
interventional X ray
equipment should be fitted with continuously adjustable beam
collimating devices. Such
devices allow the operator¹² to limit the area being imaged to the size of the
selected image
receptor or the region of interest, whichever is the smaller;
(i) When preset protocols are provided, the technique factors
being used should be readily
accessible and modifiable by appropriate personnel;
(j) Radiation leakage is kept as low as reasonably achievable and
should not exceed 1 mGy in an
hour measured at 1 metre from the focal spot, or is less than maximum levels
specified in
international standards or in local regulations.
3.33. Specific design features for medical radiological equipment used
for radiography should
include:
¹² The term “operator” is used generically in this section. The
operator is usually a medical radiation
technologist, but may sometimes be a radiological medical practitioner, such as
a radiologist.
(a) The provision of devices that automatically terminate the irradiation
after a preset time, tube
current–time product, or dose to the automatic exposure control detector, or
the dead man hand
switch is released;
b) Automatic exposure control (AEC) systems are incorporated in
radiographic units where
practicable. Such AEC systems should be able to compensate for energy
dependence, patient
thickness and exposure rate, for the expected range of clinical
imaging conditions, and be
suited to the type of image receptor being used – film/screen or digital.
3.34. Specific design features for medical radiological equipment used for CT
should include:
(a) Console display of all CT parameters that directly influence the
image acquisition (these may
be displayed over a number of screens);
(b) Console display of estimated volume CT air kerma index (CVOL or
CTDIvₒl) and CT air kerma
length product (PKL,CT or DLP) for the procedure/acquisition;
(c) Operator alert if exposure factors are set too high (usually
expressed in terms of volume CT air
kerma index and/or CT air kerma length product;
(d) Dose modulation (rotational and z-axis):
(e) A comprehensive range of beam widths and pitches and other ancillary
devices, e.g. dynamic
collimation, to ensure over ranging in CT is kept as low as
reasonably achievable by
facilitating the appropriate choice of beam width and pitch to limit
patient dose while
maintaining diagnostic image quality;
(f) Reconstruction algorithms that result in dose reduction without
compromising image quality,
such as iterative reconstruction algorithms.
3.35. Specific design features for medical radiological equipment used
for mammography (both
digital and film-screen systems) should include:
(a) Various anode/filter combinations;
(b) Compression and immobilization capabilities;
(c) Magnification views;
(d) Display on the console of a dose index, for example
incident air kerma or mean glandular
dose.
3.36. Specific design features for medical radiological equipment used
for fluoroscopy should
include:
(a) The provision of a device that energizes the X ray tube
only when continuously depressed
(such as an exposure footswitch or ‘dead man’s switch’);
(b) Indication or display of the elapsed time and air kerma area product,
and/or entrance surface
dose monitors;
(c) Automatic brightness control;
(d) Pulsed fluoroscopy and pulsed image acquisition modes;
(e) Last image hold – the capture and display of the last acquired frame;
(f) Road mapping;
(g) Interlocks which prevent energizing the X ray beam inadvertently when
the image detector is
removed from the imaging chain.
(h) The ability to disconnect the exposure footswitch between cases.
3.37. In addition to those listed in para 3.36, design features for medical
radiological equipment used
for image guided interventional procedures should include:
(a) X ray tubes that have high heat capacities to enable operation at
high tube currents and short
times;
(b) A generator with capability of at least 80 kilowatts (kW) of power;
(c) A generator with a large dynamic range of mA levels (to
minimize the range of kVp and
exposure time needed to compensate for differences in thickness);
(d) For paediatric work:
i. The generator supports an X ray tube with a minimum of three focal
spots;
ii. The anti-scatter grid is removable;
iii. Image acquisition frame rate capability extends up to at least 60
frames per second for
small children;
(e) Transmission chambers installed at the end of the
collimators to measure air kerma area
product;
(f) Imaging detectors that allow different fields of view
(magnification) to improve the spatial
resolution;
(g) Automatic collimation;
(h) Dual-shape collimators incorporating both circular and elliptical
shutters to be used to modify
the field for cardiac contour collimation;
(i) Additional filtration in the X ray beam (commonly copper filters)
that is selectable (often as
part of the automatic brightness control system)
(j) Extra filtration (0.2 mm - 0.9 mm) that may be automatically set
according to patient weight
and angulation of the C-arm;
(k) Dose per pulse and the number of pulses per second that are
selectable;
(l) Wedge filters that move automatically into the field of view to
block areas where there is no
tissue and thus no need for imaging;
(m) Possible manipulation of diaphragms while in ‘last image hold’;
(n) Display and recording in a digital format dose report of the
following parameters:
i. Reference air kerma rate;
ii. Cumulative reference air kerma;
iii. Cumulative air kerma area product;
iv. Cumulative time of fluoroscopy;
v. Cumulative number of image acquisitions, (acquisition runs and
frames per run);
vi. Integrated reference air kerma;
(o) System for Digital subtraction angiography (DSA).
3.38. All digital medical radiological equipment should have the following
additional features:
(a) Real time dose display and end-of-case dose report (radiation dose
structured report (RDSR),
DICOM object), including dose metrics export for the purpose of DRLs and
individual patient
dose calculation;
(b) Connectivity to RIS/PACS.
3.39. For medical radiological equipment used for performing diagnostic and
interventional radiology
procedures on children, there should be additional design features
that both facilitate successful
radiological procedures on patients that are typically uncooperative
and suit imaging very small
patients. Such features include the capability of very short
exposure times for radiography,
specifically designed automatic exposure controls, provision of “paediatric
modes” for the automatic
brightness control systems in fluoroscopy and image guided interventional
procedures, and paediatric
protocols for CT.
Other equipment
3.40. All equipment used for digital image display should meet appropriate
international and/or local
standards, for example meeting the performance specifications of the AAPM Task
Group 18 [116].
See paras. 3.24 to 3.25 for guidance on reporting rooms.
3.41. View boxes, for viewing films, should have sufficient uniform brightness
to facilitate diagnosis,
and the colour of view boxes should be matched through the complete
set of view boxes. Means
should be available (masks) to restrict the illuminated area of the radiograph
to avoid dazzling. View
boxes used for mammography should have higher luminance performance. Detailed
guidance is given
in Refs [117 – 122]. See paras. 3.24 to 3.25 for guidance on reporting rooms.
3.42. For radiology facilities where film is being used as an image receptor,
film processing plays a
crucial role in ensuring the medical exposure results in a diagnostic image.
Automatic film processors
should meet appropriate standards. Film-screen based mammography should
have dedicated film
processors with extended processing cycles. If manual processing is
being performed, specially
designed developer, fixer and washing tanks should be used, with
developer temperature-based
processing times. The darkroom for processing should meet relevant
international and/or local
standards for light-tightness and be equipped with an appropriately filtered
safe-light, compatible with
the film being used. Further details are given in Refs. [86, 117 – 122].
Maintenance
3.43. The BSS (BSS, paragraphs 3.15(i) and 3.41) gives requirements for
maintenance to ensure that
sources meet their design requirements for protection and safety
throughout their lifetime and to
prevent accidents as far as reasonably practicable. The registrant or
licensee should ensure that
adequate maintenance (preventive and corrective) is performed to ensure
that medical radiological
equipment retain, or improve through appropriate hardware and/or
software upgrades, their design
specifications for image quality and radiation protection and safety
for their useful lives. The
registrant or licensee should, therefore, establish the necessary
arrangements and coordination with
the manufacturer’s representative or installer before initial operation and on
an on-going basis.
3.44. All maintenance procedures should be included in the quality
assurance programme at the
frequency recommended by the manufacturer of the equipment and
relevant professional bodies.
Servicing should include a report describing the equipment fault, the work done
and the parts replaced
and adjustments made, which should be filed as part of the quality assurance
programme. A record of
maintenance carried out should be kept for each item of equipment: this should
include information
on any defects found by users (a fault log), remedial actions taken
(both interim and subsequent
repairs) and the results of testing before equipment is reintroduced to
clinical use.
3.45. In line with the guidance in Section 2 para. 2.111, after any
modifications or maintenance, the
person responsible for maintenance should immediately inform the licensee of
the medical radiation
facility before it is returned to clinical use. The person responsible for the
use of the equipment, in
conjunction with the medical physicist, the medical radiation
technologist and other appropriate
professionals, should decide whether quality control tests are needed
with regard to radiation
protection, including image quality, and whether changes to protocols are
needed.
3.46. The electrical and mechanical safety aspects of the medical radiological
equipment should be
part of the maintenance programme, and can have direct or indirect
effects on radiation safety.
Authorized persons who understand the specifications of the medical
radiological equipment should
perform this work. See also Section 2 paras. 2.110 – 2.112. Electrical and
mechanical maintenance
should be included in the QA programme at a frequency recommended and
preferentially performed
by the manufacturer of the medical radiological equipment or
authorized agent. Servicing should
include a written report describing the findings. These reports and follow up
corrective actions should
be archived as part of the QA programme.
OCCUPATIONAL RADIATION PROTECTION
Introduction
3.47. In diagnostic imaging procedures described in paras. 3.1 to
3.4, occupationally exposed
individuals are usually the medical radiation technologists and the
radiological medical practitioners
(including, for example, radiologists and, in dental practices, dentists
operating the X ray machine). In
a trauma centre, other health professionals such as nurses, emergency
department physicians and
anaesthetists who may have to be present when using portable or fixed X ray
machines, including C-
arm fluoroscopes, or who may have to be present in the CT room when the unit is
operating may also
be considered occupationally exposed.
3.48. In image guided interventional procedures and during surgery, as
described in para. 3.4, the
occupationally exposed individuals are the radiological medical
practitioners who perform the
interventions (including but not limited to radiologists, cardiologists,
vascular surgeons, orthopaedic
surgeons, neurosurgeons, urologists, anaesthetists¹³, respiratory physicians,
and gastroenterologists),
medical radiation technologists, and other health professionals who are
present and part of the
interventional team, including the anaesthetist, nurses, and
technicians who monitor patient
physiological parameters. Some complex and lengthy procedures may
require more than one
interventionist.
3.49 Additional occupationally exposed personnel may include medical
physicists, biomedical,
clinical or service engineers and some contractors, depending on their role.
3.50. Other radiology facility workers such as patient porters, orderlies,
assistants, cleaners and other
service support, for whom radiation sources are not directly related to their
work, require the same
level of protection as members of the public, as stated in the BSS paragraph
3.78.
3.51. This sub-section contains guidance very specific to diagnostic
radiology and image guided
interventional procedures. For more general and comprehensive guidance on
occupational radiation
protection, including guidance on radiation protection programmes,
assessment of occupational
exposure and providers of dosimetry services, applicable to all areas of
radiation use (including non-
medical uses), reference should be made to the IAEA Safety Guide
Occupational Radiation
Protection [23].
Arrangements under the radiation protection programme
Classification of areas
3.52. Various areas and rooms in a radiology facility should be classified as
controlled or supervised
areas, in line with the requirements given in BSS paragraphs 3.88 to 3.92. All
other rooms and areas,
not so-designated, are considered as “public domain” and levels of radiation in
these areas should be
low enough to ensure compliance with the dose limits for public exposure.
3.53. All X ray rooms should be designated as controlled areas; in addition,
areas where mobile X ray
units are used can also be categorized as controlled areas during
the time in which radiological
¹³ Called anaesthesiologists in some States.
procedures are being carried out. Open plan emergency departments (i.e. an area
without fixed walls
where curtains or similar are used to create cubicles), with either fixed or
mobile X ray units, can also
be categorized as controlled areas during the time in which radiological
procedures are being carried
out. In order to avoid uncertainties about the extent of controlled areas, the
boundaries should, when
possible, be walls and doors.
3.54. Supervised areas may involve areas surrounding X ray rooms. A typical
design of a radiology
department includes two basic areas: an area for staff circulation
and an area for circulation of
patients, which includes reception and waiting rooms, and corridors from which
the X ray rooms can
be accessed through the dressing cabinets. The staff area includes dark rooms,
film and workstation
reading rooms and internal corridors. Most of the staff area may be classified
as a supervised area, not
primarily because of the exposure level, which can be kept very low, but rather
as a “buffer zone”
owing to the potential for other individuals inadvertently entering the X ray
rooms and receiving an
exposure.
3.55. The control console may be inside the X ray room, separated by structural
shielding, or outside
the X ray room in the staff area, with visual control of
the X ray room and with patient
communication. Control console areas should have restricted access to
unauthorized individuals to
avoid the distraction of the operator, which might lead to
unnecessary exposure or repeated exposures.
For this reason, control panel areas should be either classified as controlled
or supervised, despite the
fact that the radiation levels may be very low.
Local rules and procedures
3.56. The BSS, in paragraph 3.93, establishes a hierarchy of preventive
measures for protection and
safety with engineered controls, including structured and ancillary
shielding, being supported by
administrative controls and personal protective equipment. To this end, and as
required in the BSS
paragraph 3.94, written local rules and procedures should be in place in any
radiology facility. Their
purpose is to ensure protection and safety for workers and other
persons. These local rules and
procedures should include measures to minimize occupational radiation exposure
during both normal
work and unusual events. The local rules and procedures also should cover the
wearing, handling and
storing of personal dosimeters, and specify investigation levels and
ensuing follow-up actions (see
paras. 3.98 – 3.123).
3.57. Since all personnel involved in using radiation in a radiology facility
should know and follow
the local rules and procedures, the development and review of these local rules
and procedures should
include representatives of all health professionals involved in diagnostic
radiology and image guided
interventional procedures.
3.58. Equipment (hardware and software) should be operated in a manner
that ensures satisfactory
performance at all times with respect to both the tasks to be accomplished and
radiation safety. The
manufacturer’s operating manual should be used as an important
resource in this respect, but
additional procedures are likely to be needed. The final documented
set of operational procedures
should be approved by the radiology facility’s licensee, and
incorporated into the facility’s quality
management system (see Section 2, paras. 2.136 – 2.147).
3.59. Radiology facility staff should understand the documented
procedures for their work with
radiation and for the operation of the equipment with which they are working,
including the safety
features, and should be trained, with periodic refresher training, in what to
do when things go wrong.
Additional training should occur when new medical radiological equipment is
brought into use in the
radiology facility.
3.60. Many local rules and procedures address aspects of some, or
all, of occupational radiation
protection, patient radiation protection and public radiation protection,
either directly or indirectly, as
well as ensuring a successful diagnostic examination or
intervention. This is the case with the
following paragraphs (3.61 to 3.82) – while placed in this section on
occupational radiation protection
because they are to be followed by workers, the local rules and
procedures often also have
significance for patient and/or public radiation protection. The
following paragraphs (3.61 to 3.82)
give recommendations that should be incorporated into the radiology
facility’s local rules and
procedures.
3.61. For those radiological procedures where there is no need for staff to be
in the room during an
exposure, all attending staff should position themselves in the appropriately
shielded areas.
3.62. Immobilization devices (e.g. CT head cradle) should be used
whenever possible and as
appropriate to minimize exposure to the patient, staff or carer or comforter.
Immobilization of patients
should not be performed by staff and, if possible, not by any person. If
immobilization requires the
use of a person, then this should be someone, such as a relative of the
patient, who has agreed to be a
carer and comforter, and is afforded radiation protection accordingly (see
paras. 3. 238 – 3.242).
3.63. For general radiography:
(a) At no time should the X ray tube be pointed at the control console
area;
(b) Given that the patient is the source of scatter, care should be
taken to ensure that the patient
position is as far from the control console as is feasible, taking
into account the room
configuration and accessories, and preferably more than one metre distant from
the console.
3.64. For mobile radiography:
(a) Operators¹⁴ should wear lead aprons and maintain as much
distance as possible between
themselves and the patient (to minimize exposure to scatter), whilst
still maintaining good
visual supervision of the patient and being able to communicate verbally with
him/her;
(b) Other staff (e.g., nursing, medical and ancillary staff) are
not considered as occupationally
exposed workers and hence should be afforded protection as a member of the
public. This is
achieved by ensuring such persons are as far away from the patient
as possible during the
exposure (at least 3 meters) or are behind appropriate barriers;
(c) In those situations where a member of staff needs to be close to the
patient, protective aprons
should be worn (e.g., an anaesthetist with a ventilated patient or a
nurse with an unstable
patient);
(d) Verbal warning of an imminent exposure is given;
(e) Consideration should be given to other patients who may be nearby.
(See also sub-section on
public radiation protection, para. 3.269.)
3.65. In many emergency departments, ceiling suspended X ray equipment
provides a versatile
environment for performing rapid trauma radiography. Appropriate occupational
radiation protection
can be afforded through the following:
(a) Lead aprons should be worn by staff members that need to be adjacent
to the patient being
exposed;
(b) The primary beam should be directed away from staff and other
patients whenever possible;
¹⁴ The term “operator” is used generically in this section. The
operator is usually a medical radiation
technologist, but may sometimes be a radiological medical practitioner, such as
a radiologist.
(c) Staff should keep as far away as possible from the patient
during exposure, whilst still
maintaining good visual supervision of the patient;
(d) Where available, mobile shields should be used;
(e) Any pregnant staff member (other than radiology staff)
should be asked by the medical
radiation technologist to leave the vicinity during exposure;
(f) Verbal warning of imminent exposure is given.
3.66. For CT, when staff need to be in the room during exposures,
additional measures should be
taken:
(a) In the case of CT interventions, the interventionist should use
appropriate personal protective
equipment (protective apron, thyroid shield, protective eye-wear). In addition,
care should be
exercised to avoid placing hands in the primary beam and immediate
notification to the
interventionist should be given if this does happen;
(b) In the case of persons providing medical support, e.g.,
anaesthetists, a protective apron should
be worn and the person should position themselves as far from the gantry as
possible, whilst
still maintaining good visual supervision of the patient.
3.67. For diagnostic fluoroscopic procedures, when staff need to be
in the room, the following
measures should be taken:
(a) The staff member performing the procedure should use
personal protective equipment
(protective apron, thyroid shield, protective eye-wear, gloves). In
addition, care should be
exercised to avoid placing hands in the primary beam;
(b) In the case of persons providing medical support, e.g.,
anaesthetists, a protective apron should
be worn and the person should position themselves as far from the patient as
possible during
screening.
3.68. For radiological procedures performed with mobile fluoroscopic units
(C-arm systems):
(a) The staff member performing the procedure should use
personal protective equipment
(protective apron, thyroid shield, protective eye-wear, gloves). In
addition, care should be
exercised to avoid placing hands in the primary beam and immediate
notification to the
fluoroscopist should be given if this does happen;
(b) Only essential staff should remain in the room. All such staff are
considered occupationally
exposed workers;
(c) In those situations where a member of staff needs to be close to the
patient, protective aprons
should be worn (e.g., an anaesthetist with a ventilated patient or a
nurse with an unstable
patient). At no time should a pregnant staff member take on this role;
(d) For other practical advice, including X ray tube orientation and
positioning, mobile shields,
technical parameter selection, see the section on image guided
interventional procedures
(paras 3.74 to 3.81).
3.69. For mammography, the medical radiation technologist should stand behind
the protective barrier
attached to the mammography unit when making the exposure.
3.70. For dental facilities with intraoral and panoramic (OPG)
equipment, personal protective
equipment is not usually needed. Radiation protection is afforded through the
use of distance from the
patient. Typically, a distance of at least two metres is recommended.
3.71. Cone beam CT (CBCT) is in use in some dental facilities, and should be
housed in a room that
has been designed and shielded accordingly (see para. 3.34). The staff should
be positioned behind the
protective barrier at the control console when exposures are being made.
3.72. For DEXA, the radiation levels around the unit are very low
and there are no specific
precautions that should be taken with respect to occupational
radiation protection. Typically the
operator can be in the room with the patient when the machine is
operating. The operators’ desk
should be positioned at least 1 m away from a pencil beam, and at least 2 m
from a fan-beam system.
In the case of fan-beam and cone-beam configurations or if the
distances above cannot be
accommodated, the use of protective screens should be considered.
3.73. Local rules for pregnant workers and persons under 18 should
reflect the guidance given in
paras. 3.127 – 3.129 and 3.130 respectively.
Local rules and procedures – image guided interventional procedures
3.74. Image guided interventional procedures, performed either in
fluoroscopy rooms or dedicated
interventional rooms, tend to be complex and are performed on patients who can
be quite ill and/or
have a life threatening condition. As a consequence, more staff will
be needed in the room to attend to
the patients’ individual medical needs (e.g., interventionists,
anaesthetists, medical radiation
technologists, nurses, and sometimes other specialists). Not only will more
staff be exposed during
interventional procedures, they may also be standing close to the
patient where dose rates from
radiation scattered by the patient are high.
3.75. Interventional procedures require specifically designed and dedicated
equipment. The exposure
rate in the vicinity of the patient is lower on the beam exit side
of the patient. For a vertical
orientation, an under-couch X ray tube with an over-couch image receptor has
lower levels of scatter
in the area of the operator’s trunk and head than an over-couch X ray
tube with an under-couch image
receptor. A similar situation exists with lateral projections, where the
maximum scatter radiation is on
the X ray tube side of the patient. Staff should, where practicable, always
stand on the image receptor
side of the patient during lateral or oblique projections.
3.76. There are simple methods of reducing exposure to staff as a
result of operational factors,
including choosing where to stand or to be positioned in the room. Since the
patient is the main source
of scatter radiation, the staff should remain as far away as practicable from
the patient when exposures
are taking place, to reduce exposure to staff. For the interventionist, taking
a step or even half a step
back during image acquisition results in a significant reduction in
occupational dose. As discussed in
para. 3.75, the X ray tube orientation and positioning will determine where it
is best to stand in order
to be in an area subject to relatively low scatter.
3.77. Staff should never be subject to direct beam exposure. This includes
avoiding placing the hands
in the beam whenever possible. When hands of the operator are close to the
direct beam, an under-
couch X ray tube with an over-couch image receptor should be used,
because the exposure rate is
lower on the beam exit side of the patient and the exposure to the operator’s
hands is significantly
reduced.
3.78. There are many operational factors that affect the
patient dose during image guided
interventional procedures, and these factors in turn affect staff dose because
the dose to the patient
determines the amount of scatter being produced. Methods to reduce
patient dose are described in
paras 3.182 to 3.187, and should always be used to reduce both patient and
staff doses.
3.79. Medical radiological equipment specifically designed for
image guided interventional
procedures often incorporates protective devices, such as ceiling
suspended lead acrylic viewing
screens, and under-table and lateral shielding attachments to the X ray
couch, and personal mobile
shields. Alternatively, such devices can be purchased separately. These devices
can afford individuals
significant radiation protection, but they can sometimes be cumbersome
to use. However, the
appropriate use of these devices will result in a significant reduction in
staff doses.
3.80. Higher incidence of radiation injuries to the lens of the eye has been
reported for interventionists
and nurses performing image guided interventional procedures [123]. For
this reason it is strongly
recommended that interventionists, and other staff who routinely work close to
the patient, always use
ceiling mounted screens and/or protective eye-wear. This is further reinforced
by the requirement to
comply with the relatively low dose limit (20 mSv per year) for the lens of the
eye. It is quite likely
that the dose limit would be exceeded for an interventionist performing several
hundred image guided
interventional procedures in a year if that person used no protection for the
eyes.
3.81. Further specific guidance on interventional radiology and interventional
cardiology, endorsed by
several regional professional societies, can be found in references [124, 125].
3.82. Some image guided interventional procedures are performed using CT and
the guidance given in
para 3.66 applies.
Personal protective equipment
3.83. As required in the BSS paragraph 3.93 and 3.95, personal
protective equipment and in-room
protective equipment should be available and used when structural
shielding and administrative
controls alone cannot afford the required level of occupational
radiation protection. This typically
arises when staff are required to be in the room where and when the
radiological procedures are taking
place, such as with image guided interventional procedures and
fluoroscopy, and with mobile
radiography. The need for these protective devices should be established by the
radiology facility’s
RPO or medical physicist.
3.84. Personal protective equipment is worn on the person and
includes protective aprons, thyroid
protectors, protective eye-wear, and protective gloves. Protective aprons are
available in many shapes,
configurations, materials and lead equivalence, and should be chosen to
best suit the intended use.
Some aprons require using fully overlapping panels to provide complete
coverage. Expert advice on
personal protective equipment should be sought from the RPO or medical
physicist.
3.85. For image guided interventional procedures, wrap around aprons,
preferably consisting of vests
and skirts to spread the weight, should be used. They should cover:
(a) From the neck down to at least 10 cm below the knees;
(b) The entire breast bone (sternum) and shoulders;
(c) The sides of the body from not more than 10 cm below the armpits to
at least halfway down
the thighs;
(d) The back from the shoulders down to and including the buttocks.
3.86. Protective gloves are useful to protect the hands near the beam but may
produce the opposite
effect during fluoroscopy with automatic brightness control (ABC) when
the hands enter the area
covered by the sensor of the ABC, because this would drive the exposure to
higher levels for both the
staff and the patient and would be ineffective in protecting the hands. Even if
the fluoroscopy system
operates without ABC, leaded gloves may prolong the procedure because
they do no afford the
necessary tactile sensitivity and thus their value is questionable.
3.87. Protective eye-wear, especially for use in image guided interventional
procedures, should cover
the entire orbit. This requires lateral protection provided by shielded sides
and that the glasses are a
close fit.
3.88. The lead-equivalence of personal protective equipment should be
specified at the maximum
operating X Ray tube potential (kVp) applicable for its intended use.
3.89. Non-lead based personal protective equipment, incorporating
shielding materials, such as tin,
tungsten, bismuth, and antimony, may be preferable if they are lighter and
easier to use. Care should
be taken in interpreting claimed lead equivalences for non-lead based
protective equipment, and
expert advice from the RPO or medical physicist should be sought.
3.90. Protective equipment for pregnant workers should be carefully
considered, as wrap around
aprons may no longer provide adequate protection for the embryo/fetus
to meet BSS requirements
(BSS paragraph 3.114). The RPO or medical physicist should be consulted as
needed.
3.91. Personal protective equipment, in particular protective aprons,
can lose their protective
effectiveness if mistreated or not appropriately used or cared for.
All personnel that use personal
protective equipment have the responsibility for its appropriate use and
care, for example ensuring
aprons are correctly hung and stored to minimize damage.
3.92. Personal protective equipment should be examined under
fluoroscopy at least annually to
confirm its shielding integrity.
In-room protective devices
3.93. Additional protective devices for use in fluoroscopy and image guided
interventional procedures
include:
(a) Ceiling suspended protective screens for protecting eyes and the
thyroid while keeping visual
contact with the patient. Technical advances with such screens include
systems that move
with the operator;
(b) Protective lead curtains or drapes mounted on the patient table;
(c) Mobile shields either attached to the table (lateral shields) or
mounted on coasters (full body);
(d) Disposable patient protective drapes.
Monitoring of the workplace
3.94. The BSS, in paragraphs 3.96-3.98, sets out the requirements and
responsibilities for workplace
monitoring. Workplace monitoring comprises measurements made in the working
environment and
the interpretation of such results. Workplace monitoring serves several
purposes, including routine
monitoring, special monitoring for specific occasions, activities or tasks, and
confirmatory monitoring
to check assumptions made about exposure conditions. Workplace monitoring can
be used to verify
the occupational doses of personnel whose work involves exposure to
predictable low levels of
radiation. It should be particularly used for staff members who are not
individually monitored. Further
general guidance on workplace monitoring is given in Ref [23].
3.95. Workplace monitoring in areas around each of the medical
radiological equipment in the
radiology facility, when it is being operated, should be carried out when:
(a) The room and shielding construction has been completed,
regardless whether it is a new
construction or a renovation, and before the room is first used clinically;
(b) New or substantially refurbished equipment is commissioned
(both direct and indirect
radiation such as leakage and scatter should be measured);
(c) New software for the medical radiological equipment is
installed or there is a significant
upgrade;
(d) New techniques are introduced;
(e) Servicing on the medical radiological equipment has been
performed, which may have an
impact on the radiation delivered.
3.96. Workplace monitoring should be performed and documented as part of the
radiology facility’s
radiation protection programme. The radiology facility’s RPO or medical
physicist should provide
specific advice on the workplace monitoring programme, including any
investigations that arise
through investigation levels being exceeded (see para. 3.115).
3.97. The survey meters used for radiation monitoring should be calibrated in
terms of ambient dose
equivalent. For diagnostic radiology and image guided interventional
procedures, the quantity is
H*(10) and the unit the Sv and its sub-multiples.
Assessment of occupational exposure and workers’ health surveillance
Occupational exposure assessment
3.98. The purpose of monitoring and dose assessment is, inter alia, to provide
information about the
actual exposure of workers and confirmation of good working practices and
regulatory compliance. It
contributes to reassurance and motivation. The BSS require individual
monitoring for any worker who
normally works in a controlled area or any worker who occasionally works in a
controlled area and is
likely to receive significant occupational exposure (see BSS paragraphs 3.99 to
3.102). Workers who
may require individual monitoring include radiologists,
cardiologists, gastroenterologists,
endoscopists, urologists, orthopaedic surgeons, neurosurgeons, respiratory
physicians, anaesthetists,
medical physicists, biomedical/clinical engineers, medical radiation
technologists, nurses, and the
RPO.
3.99. Monitoring involves more than just measurement. It includes
interpretation, assessment,
investigation and reporting, which may lead to corrective measures, if
necessary. Individual external
doses are assessed using individual monitoring devices that
include thermoluminescent dosimeters
(TLD), optical stimulated luminescence dosimeters (OSLD), film badges, and
electronic dosimeters.
Care should be taken when using electronic dosimeters in pulsed X ray
fields that they function
correctly. Individual monitoring devices must be calibrated, traceable
to a standards dosimetry
laboratory. For more detailed guidance see Ref [23].
3.100. Each dosimeter is to be used for monitoring only the person to whom it
is issued, for work
performed at that radiology facility and should not be taken to other
facilities where that person may
work. For example, if a person is issued with a dosimeter at
hospital A it should be worn only at
hospital A and not at any other hospitals or medical centres where he/she may
also work. Monitoring
results can then be interpreted for the person working in a specific
radiology facility, and allow
appropriate review of the effectiveness of the optimization of
protection for that individual in that
facility. See also paras. 3.117 to 3.118.
3.101. The monitoring period (period of dosimeter deployment) specified
by regulatory bodies in
most countries is typically in the range of one to three months. A one
month monitoring period is
usually used for persons performing procedures associated with higher
occupational exposures, such
as image guided interventional procedures. A longer monitoring period (two or
three months) is more
typical for personnel exposed to lower doses, as a one month cycle would
usually mean that the actual
occupational dose was less than the minimum detection level of the
dosimeter resulting in no
detectable doses. With a longer cycle it is more likely to obtain a
reading. The radiology facility
should send the dosimeters to the dosimetry service provider who should then
process the dosimeters
and return the dose reports, all in a timely manner. Some regulatory
bodies may specify a
performance criterion for timely reporting.
3.102. The operational dosimetric quantity used is the personal dose
equivalent Hp(d). For weakly
penetrating and strongly penetrating radiation, the recommended depths, d, are
0.07 mm and 10 mm,
respectively. Radiation used in diagnostic radiology and image guided
interventional procedures is
usually relatively strongly penetrating, and therefore d = 10 mm for dosimeters
being used to assess
effective dose. Hp(10) is used to provide an estimate of
effective dose that avoids both
underestimation and excessive overestimation [23]. In diagnostic
radiology and image guided
interventional procedures, the overestimation is somewhat larger because
of the lower photon
penetration from X ray beams in the kV range [126, 127]. If a protective apron
or thyroid shield is
being worn, the relationship between Hp(10) effective dose becomes
more complex and additional
guidance is given below in para. 3.109.
3.103. For monitoring the skin and extremities, a depth of 0.07 mm (d = 0.07)
is recommended, and
Hp(0.07) is used to provide an estimate of equivalent dose to the skin and
extremities.
3.104. For monitoring the lens of the eye, a depth of 3 mm (d = 3) is
recommended, and Hp(3) is used
to provide an estimate of equivalent dose to the lens of the eye. In practice,
however, the use of Hp(3)
has not been widely implemented for routine individual monitoring. In cases
where eye doses are a
concern, such as in image guided interventional procedures, Hp(0.07), and to a
lesser extent Hp(10),
can be considered as an acceptable surrogate operational quantity. More
guidelines are provided in the
IAEA publication [128].
3.105. There are three dose limits applicable to workers in
diagnostic radiology and image guided
interventional procedures – effective dose; and equivalent dose to the lens of
the eye, and to skin and
extremities. The dosimeter being worn will be used to estimate one or more of
the quantities used for
the dose limits. Depending on the work being performed by the person being
individually monitored,
there may be a preferred position for wearing the dosimeter, and more than one
dosimeter may be
used.
3.106. For individual monitoring with only one dosimeter in diagnostic
radiology and image guided
interventional procedures:
(a) If the monitored worker never wears a protective apron, the
dosimeter should be worn on the
front of the torso between the shoulders and the waist;
(b) If the monitored worker sometimes wears a protective apron, the
dosimeter should be worn on
the front of the torso between the shoulders and the waist, and
under the apron when it is
being worn;
(c) If the monitored worker always wears a protective apron, the
dosimeter should be worn on the
front of the torso at shoulder or collar level outside the apron (see also
para. 3.107);
(d) If the working situation is such that the radiation is always or
predominantly coming from one
side of the person, such as in image guided interventional procedures, the
dosimeter should be
placed, in addition to the guidance in (a) to (c), on the front of the torso on
the side closest to
the source of radiation.
3.107. For individual monitoring with two dosimeters, such as in
image guided interventional
procedures where the monitored worker always wears a protective apron,
one dosimeter should be
worn on the front of the torso at shoulder or collar level outside the apron
on the side closest to the
source of radiation. The other dosimeter should be worn on the front
of the torso between the
shoulders and the waist and under the apron, preferably on the side closest to
the source of radiation.
3.108. Specialized dosimeters, such as ring dosimeters for monitoring
finger doses, will have their
own specific wearing instructions which should be followed.
3.109. When a protective apron is being used, the assessment of
effective dose may not be
straightforward:
(a) A single dosimeter placed under the apron, reported in Hp(10),
provides a good estimate of the
contribution to the effective dose by the parts of the body
protected by the apron, but
underestimates the contribution of the unprotected parts of the body (thyroid,
head and neck,
and extremities).
(b) A single dosimeter worn outside the apron, reported in
Hp(10), provides a significant
overestimate of effective dose and should be corrected for the
protection afforded by the
apron by using an appropriate algorithm [ 123, 127 129].
(c) Where two dosimeters are worn, one under the apron and the
other outside the apron, an
algorithm should be applied to estimate effective dose from the two reported
values of Hp(10)
[127, 129].
3.110. As noted in para. 3.104, dosimeters for reporting Hp(3) are not widely
available. A dosimeter
worn outside the apron at collar or neck level, reported in either Hp(0.07) or
Hp(10), can provide a
surrogate estimate for the equivalent dose to the lens of the eye. Whether
protective eye-wear was
being worn or not should be taken into account to correctly interpret the dose
estimate.
3.111. When not in use, individual dosimeters should be kept in an established
place and protected
from damage or from irradiation. If an individual’s dosimeter is lost, the RPO
should perform a dose
assessment, record this evaluation of the dose and add it to the worker’s dose
record. Where there is a
national dose registry, it should be informed of the dose estimate
in a timely manner. The most
reliable method for estimating an individual’s dose is to use his or her recent
dose history. In those
cases where the individual performs non-routine types of work, it may be better
to use the doses of co-
workers having similar exposure conditions as the basis for the dose estimate.
3.112. In some radiology facilities and for some individuals with a low level
of occupational exposure
(e.g. general dental practitioners), area dosimetry to estimate the level of
dose per procedure can be an
acceptable alternative to individual monitoring. Knowing the typical level of
dose per procedure for
positions where personnel are placed during exposures and the number of
procedures per year, the
RPO can estimate personnel doses.
3.113. Similarly, occupational doses can be estimated from the results of
workplace monitoring. The
effective dose for personnel can be inferred from the measured ambient dose
equivalent H*(10). ICRP
Publication 116 provides conversion coefficients from ambient dose equivalent
to effective dose for
different types of radiation and energies [126]. The conversion coefficients
for photons are close to
unity except for very low energy, such as the energy of scattered photons from
a mammography X ray
beam.
3.114. An additional direct reading operational dosimeters, such as
appropriately calibrated electronic
dosimeters, may be also used in image guided interventional procedures, as
these devices can give the
worker an instant indication of both the cumulative dose and the current dose
rate and are a useful
educative tool for the optimization of occupational radiation protection [23].
Investigation levels for staff exposure
3.115. Investigation levels are separate from dose constraints and dose limits;
they are a tool used to
provide a ‘warning’ of the need to review procedures and performance,
investigate what is not
working as expected and take timely corrective action. Exceeding an
investigation level should
prompt such actions. The following are examples for radiology facilities of
levels and their related
tasks that should not normally be exceeded and, therefore, could be suitable as
investigation levels.
For example, monthly values higher than 0.5 mSv (for the dosimeter worn under
the protective apron)
should be investigated. Values higher than 2 mSv per month [125]
from the over-apron dosimeter
may indicate that eye doses may be of concern. Values higher than 15 mSv per
month for hand or
finger dosimeters should also be investigated. Abnormal conditions and/or
events should also trigger
an investigation. In all cases the investigation is carried out with a view to
improve implementation of
optimization of occupational protection and results should be recorded.
Investigation levels should
also be set for workplace monitoring.
3.116. The investigation is to be initiated as soon as possible
following the trigger or event and a
written report is to be prepared concerning the cause, including determination
or verification of the
dose, corrective or mitigating actions, and instructions or recommendations to
avoid recurrence. Such
reports should be reviewed by quality assurance and radiation safety
committees, as appropriate, and
the holder of the licence should be informed. In some cases, the regulatory
body may also need to be
informed.
Persons who work in more than one place
3.117. Personnel may work regularly in more than one radiology facility. The
facilities may be quite
separate entities in terms of ownership and management, or they may have common
ownership but
separate management, or they may have common ownership and management, but be
physically quite
separate. No matter which, the occupational radiation protection
requirements for the particular
radiology facility apply when the person is working in that facility. As
described above in para. 3.100,
a dosimeter issued for individual monitoring should be worn only in the
facility for which it is issued
as this facilitates effective implementation of optimization of protection in
that facility. This approach
is logistically more easily implemented as each physical site has its own
dosimeters – there is no need
to transport dosimeters between facilities, with the risk of either
loss or forgetting them. In cases
where the facilities are under common ownership it may be seen as an
unnecessary financial burden to
provide more than one set of dosimeters for staff that work in more than one of
its facilities. However
the radiation protection advantages of having the dosimeter results linked to a
person’s work in only
one radiology facility remain. See also para. 3.119.
3.118. There is however an additional consideration, namely
ensuring compliance with the
occupational dose limits. Any person who works in multiple radiology
facilities should notify the
licensee for each of those facilities. Each licensee, through their RPO,
establishes formal contact with
the licensees of the other radiology facilities, and their RPOs, so that each
facility has an arrangement
to ensure that a personal dosimeter is available and that there is
an on-going record of the
occupational doses for that person in all the facilities where they work.
3.119. Some personnel, such as consultant medical physicists or service
engineers, may perform work
in many radiology facilities and, maybe as well, other medical
radiation facilities. They may be
employed by a company or be self-employed, providing contracted services to the
radiology and other
facilities. In such cases it is simpler for the company or the
self-employed person to provide the
dosimeters for individual monitoring. In other words, in these cases
for each person the same
dosimeter is used for work performed in all radiology facilities (and other
medical radiation facilities)
in the monitoring period.
Records of occupational exposure
3.120. Paragraphs 3.103 to 3.107 of the BSS state the requirements
for records of occupational
exposure, placing obligations on the employer, registrant and licensee.
Apart from demonstrating
compliance with legal requirements, records of occupational exposure
should be used within the
radiology facility for additional purposes, including assessing the
effectiveness of the facility’s
implementation of optimization of protection, and evaluating trends in
exposure. National or local
regulatory bodies might specify additional requirements for records of
occupational exposure and for
access to the information contained in those records. Employer shall provide
workers with access to
records of their own occupational exposure (BSS, para 3.106 (a). Further
general guidance on records
of occupational exposure is given in [23].
Workers’ health surveillance
3.121. The primary purpose of health surveillance is to assess the
initial and continuing fitness of
employees for their intended tasks, and requirements are given in BSS
paragraphs 3.108 to 3.109.
3.122. No specific health surveillance related to exposure to ionizing
radiation is necessary for staff
involved in diagnostic radiology or image guided interventional procedures,
with perhaps the possible
exception of initial and periodic eye assessments for visual acuity
and contrast resolution for
personnel performing significant numbers of image guided interventional
procedures. Only in cases of
overexposed workers, at doses much higher than the dose limits (e.g. a few
hundred millisieverts or
higher), would special investigations involving biological dosimetry and
further extended diagnosis
and medical treatment be necessary [23]. Under normal working conditions, the
occupational doses
incurred in diagnostic radiology and image guided interventional procedures are
low and no specific
radiation related examinations are required for persons who are occupationally
exposed to ionizing
radiation, as there are no diagnostic tests that yield information
relevant to normal exposure. It is,
therefore, rare for considerations of occupational exposure arising from the
working environment of a
radiology facility to significantly influence the decision about the
fitness of a worker to undertake
work with radiation or to influence the general conditions of service [23].
3.123. Counselling should be available to workers who have or may have been
exposed in excess of
dose limits, and information, advice and, if indicated, counselling should be
available to workers who
are concerned about their radiation exposure. In diagnostic radiology and image
guided procedures,
the latter group may include women who are or may be pregnant such as, for
example, female medical
radiation technologists and nurses working in therapy wards.
Counselling should be given by
appropriately experienced and qualified practitioners. Further guidance is
given in [23].
Information, instruction and training
3.124. All staff involved in diagnostic radiology and image guided
interventional procedures should
fulfil the respective training and competence criteria described in
Section 2, paras. 2.117 to 2.135.
This will include general education, training, qualification and competence for
occupational radiation
protection. Radiological medical practitioners, medical radiation
technologists and nurses working
with hybrid units (such as PET-CT, SPECT-CT) may have trained
exclusively in their original
specialty. They should undertake radiation protection training relevant
to the additional imaging
modality.
3.125. The BSS, in paragraph 3.110, places responsibilities on
employers to provide, inter alia,
specific instruction and training for protection and safety as it
pertains to their radiology facilities.
This is not only for new staff but also for all staff as part of
their continuing professional
development. Specific instruction and training should be provided when
new medical radiological
procedures, equipment, software and technologies are introduced.
Conditions of service and special arrangements
3.126. As required in BSS paragraph 3.111, no special benefits are to be
offered to staff because they
are occupationally exposed. It is simply not acceptable to offer benefits as
substitutes for measures for
protection and safety.
Female workers (pregnant)
3.127. A female worker should notify the licensee that she is pregnant as soon
as she knows of her
condition, so that radiation protection requirements for the embryo/fetus as a
member of the public
can be met.
3.128. The limitation of the dose to the embryo/fetus does not mean
that pregnant women should
avoid work with radiation, but it does imply that the employer should carefully
review the exposure
conditions with regard to both normal exposure and potential exposure. A
possible solution includes
reassigning a pregnant worker to a location that may have lower
ambient dose equivalent; for
example, from fluoroscopy to radiography or to CT. Adequate training
should accompany such
reassignments.
3.129. When applying the dose limit of 1 mSv to the fetus, the
reading of the dosimeter may
overestimate fetal dose by a factor of 10. If the reading corresponds to a
dosimeter worn outside a lead
apron, the overestimation of fetal dose may rise to a factor of 100 [130].
Counselling for pregnant
workers should be available (see also para. 3.123).
Persons under 18
3.130. In many countries there is the possibility of students aged
16 or more, but under 18,
commencing their studies and training to become a medical radiation
technologist or other health
professional that may involve occupational exposure to ionizing radiation. The
BSS paragraph 3.116
states the requirements for access to controlled areas and the dose limits for
such persons are more
restrictive – see Table 2 in this Safety Guide and Schedule III of the BSS.
RADIATION PROTECTION OF PATIENTS, CARERS AND COMFORTERS, AND
VOLUNTEERS IN BIOMEDICAL RESEARCH
Introduction
3.131. The section covers what is more formally called ‘medical exposure’
in radiation protection.
Medical exposure is defined in the BSS and described in Section 2, para. 2.5.
It concerns radiation
protection of the patient, carers and comforters, and volunteers in
biomedical research. The term
‘patient’, when used in the context of medical exposure, means the person
undergoing the radiological
procedure. Other patients in the radiology facility, including those who may be
waiting for their own
radiological procedure, are considered as members of the public and
their radiation protection is
covered in paras. 3.264 to 3.273.
3.132. As described in Section 2 para. 2.8, there are no dose limits for
medical exposure, making it
very important that there is effective application of the requirements for
justification and optimization.
Justification of medical exposures
3.133. The requirements of the BSS for justification of medical exposure,
paragraphs 3.156 to 3.163,
incorporate the ‘3 level approach’ to justification [4, 131, 132].
3.134. The roles of the health authority and professional bodies with respect
to ‘level 2’ or generic
justification of radiological procedures, justification of health screening
programmes, and justification
of screening intended for the early detection of disease, but not as
part of a health screening
programme, are described in Section 2 paras. 2.55 to 2.58, and 2.64.
Justification for the individual patient
3.135. The BSS invokes a joint approach to justification at the individual
patient level, with a shared
decision involving both the referring medical practitioner (who initiates the
request for a radiological
procedure) and the radiological medical practitioner. A referral should be
regarded as a ‘request for a
professional consultation or opinion’ rather than an ‘instruction or order to
perform’. The referring
medical practitioner brings the knowledge of the medical context and
the patient’s history to the
decision process, while the radiological medical practitioner has the
specialist expertise on the
radiological procedures. The efficacy, benefits and risks of alternative
methods (both involving and
not involving ionizing radiation) should be considered. In all cases the
justification is informed by
national or international referral guidelines, for example Refs [133 – 141].
Ultimate responsibility will
be identified by individual member states’ regulations.
3.136. The patient also should be informed about the expected benefits, risks
and limitations of the
proposed radiological procedure(s), as well as the consequences of not
undergoing the procedure.
3.137. Justification, which is a radiation protection principle, is implemented
more effectively as part
of the medical process of determining ‘appropriateness’. Appropriateness
uses an evidence-based
approach to choose the best test for a given clinical scenario,
taking into account the diagnostic
efficacy of the radiological procedure as well as alternative
procedures that do not use ionizing
radiation, for example, ultrasound, MRI or endoscopy. Useful tools to support
this decision making
process include national or international imaging referral guidelines
developed by professional
societies [133 – 141]. Imaging referral guidelines can be disseminated through
electronic requesting
systems¹⁵ and clinical decision support tools or systems.
3.138. In determining appropriateness of the radiological procedure for
an individual patient, the
following questions should be asked by the referring medical practitioner
[138];
(a) Has it been done already? A radiological procedure that has already
been performed within a
reasonable time period (depending on the procedure and clinical
question) should not be
repeated. The results (images and reports) of previous examinations should be
available, not
only within a given radiology facility but also between different
facilities. Digital imaging
¹⁵ Such electronic requesting systems include the so-called CPOE – computerized
physician order entry, noting
that for imaging such a system should generate a request rather than an order.
modalities and electronic networks should facilitate this process. Individual
patient exposure
records should facilitate decision making process if available.
(b) Is it needed? The results of the proposed radiological procedure
(positive or negative) should
influence the patient’s management.
(c) Is it needed now? The timing of the proposed radiological
procedure in relation to the
progression of the suspected disease and the possibilities for
treatment, all should be
considered as a whole.
(d) Is this the best investigation to answer the clinical question?
Advances in imaging techniques
are taking place continually, and the referring medical practitioner may need
to discuss with
the radiological medical practitioner what is currently available for a given
problem.
(e) Has the clinical problem been explained to the radiological
medical practitioner? The
medical context for the requested radiological procedure is crucial to
ensure the correct
technique is performed with the correct focus.
3.139. For a large percentage of radiological procedures, primarily ‘well
established’ and low dose
procedures, the practical implementation of justification may be
achieved through the medical
radiation technologist, who is effectively representing the radiological
medical practitioner with the
formal understanding that, if there is uncertainty, the radiological
medical practitioner is contacted
and the final decision is taken by the radiological medical
practitioner in consultation with the
referring medical practitioner. Such justification is guided by
national or international referral
guidelines.
3.140. For the smaller percentage of radiological procedures, primarily because
of a combination of
complexity, difficult medical context, and higher dose, the
justification is likely to be led by the
radiological medical practitioner with the referring medical practitioner
providing any needed further
clarification on the medical context. Again the justification is informed by
national or international
referral guidelines.
3.141. Two particular groups of patients are identified in the BSS,
paragraph 3.157, for special
consideration with respect to justification – patients who may be pregnant and
paediatrics.
(a) Due to the higher radiosensitivity of the embryo/fetus, it
should be ascertained whether a
female patient is pregnant before performing an X ray examination for diagnosis
or an image
guided interventional procedure. Determining pregnancy status is a requirement
in the BSS,
paragraph 3.176, for those radiological procedures that could result in
significant dose to the
embryo or fetus. Pregnancy would then be a factor in the
justification process and might
influence the timing of the proposed radiological procedure or whether another
examination is
more appropriate. Confirmation of pregnancy may occur after the
initial justification and
before the radiological procedure is performed. Repeat justification is
required taking into
account the additional sensitivity of the pregnant female patient and embryo or
fetus.
(b) As children are at greater risk of incurring
radiation-induced stochastic effects, paediatric
examinations require special consideration in the justification process.
3.142. Review of the justification may need to take place if circumstances
change. For example, the
performance of a low dose procedure that has been justified but, at
the time of performing the
examination, a high dose protocol is needed. Such a case might be the
justification for low dose CT
for renal colic would have to be reviewed if high-dose enhanced CT urography is
actually needed to
answer the clinical question.
3.143. ‘Self-referral’ occurs when a health professional undertakes a
radiological procedure for
patients as a result of justification based on their own previous
clinical assessment. Examples of
acceptable self-referral practice include: dentistry, cardiology,
orthopaedics, vascular surgery,
urology, and gastroenterology. Relevant professional bodies in many
countries develop appropriate
guidance for their specialty, for example dental associations [142].
3.144. ‘Self-presentation’ occurs when a member of the public asks
for a radiological procedure
without a referral from a health professional. This may have been
prompted by media reports or
advertising. Examples include ‘individual health assessment’ which often
involves CT procedures in
asymptomatic individuals for early detection of cancer (e.g. whole body CT,
lung CT or colon CT)
and/or quantification of coronary artery calcification (coronary artery CT).
Justification is required, as
for all radiological procedures. Relevant professional bodies have an
important role in considering
evidence for developing guidance when new practices are proposed, as for
example in the case of CT
[143]. Member states may choose to incorporate this into legislation [144].
3.145. Approaches to support the implementation of the requirements
for justification for medical
exposures should consider the development of means to improve
‘awareness, appropriateness and
audit’. Awareness of the need for justification underpins the whole
process. Means for promoting
awareness include traditional education and training, such as at
medical school or during specialty
training, web-based learning or learning through work flow at the right
time,(e.g. junior doctors in the
emergency department), and the use of feedback in the reporting process.
Appropriateness has been
described in paras. 3.137 – 3.138, and the process of audit is used for
monitoring and feedback to
improve both awareness and appropriateness.
Justification – biomedical research volunteers
3.146. The role of the ethics committee in the justification of medical
exposure of volunteers exposed
as part of a programme of biomedical research is described in Section 2 para
2.98.
Justification – carers and comforters
3.147. The justification of medical exposure incurred by a carer or comforter
is effectively carried out
by the radiological medical practitioner or medical radiation technologist
involved in the radiological
procedure, prior to the performance of the procedure. It depends on
the carer or comforter being
correctly informed about the radiation doses and risks involved, and his or her
understanding of this
information and consequent agreement to take on the role of carer or comforter.
Optimization of protection and safety
3.148. In medical exposures, optimization of protection and safety has
several components, some
applied directly to the radiological procedure about to be performed and others
providing the support
or framework for the other components. These components of optimization of
protection and safety
are described in the following paras. 3.149 to 3.243.
Design considerations
3.149. The use of appropriate and well-designed medical radiological
equipment underpins any
radiological procedure in diagnostic radiology or any image guided
interventional procedure. X ray
generators and their accessories should be designed and manufactured so as to
facilitate the keeping of
medical exposures as low as reasonably achievable consistent with
obtaining adequate diagnostic
information or guidance for the intervention. Extensive guidance on design
considerations is given in
the sub-section on medical radiological equipment, paras. 3.26 to 3.39. This
guidance is applicable to
both stand-alone and hybrid systems. Ultimately, as stated in the BSS
paragraph 3.162, it is the
responsibility of the radiology facility licensee to ensure that
his/her facility uses only medical
radiological equipment and software that meets applicable international or
national standards.
Operational considerations - general
3.150. Following justification, the diagnostic radiological procedure or image
guided interventional
procedure is performed in such a way as to optimize patient
protection, as required in the BSS
paragraph 3.163. The level of image quality sufficient for diagnosis is
determined by the radiological
medical practitioner and is based on the clinical question posed and the
anatomical structures imaged
(e.g., the diagnosis of the pattern of sinusitis on CT requires only
a low dose procedure as high
contrast structures viz. air and bone are being imaged). With image guided
interventional procedures,
additionally, the level of image quality should be sufficient to guide the
intervention.
3.151. The following points apply to all diagnostic radiological
procedures or image guided
interventional procedures:
(a) There should be an effective system for correct identification of
patients, with at least two
forms of verification, for example name, birthday, address, medical record
number.
(b) Patient details should be correctly recorded, such as age, gender, weight,
height, pregnancy
status.
(c) Patient clinical history should be reviewed.
3.152. The first step in operational considerations of optimization is to
select the appropriate medical
radiological equipment. For example, a chest X ray should be performed using
dedicated equipment
with a generator producing high output enabling the use of a long source
to image receptor distance
(typically 1.8 m) and a short exposure time to ensure a reproducible
diagnostic quality image by
minimizing patient respiratory and cardiac motion.
3.153. The volume (area) of the patient that is exposed should
bestrictly limited to that of clinical
interest. This is achieved through collimation for radiography, mammography,
fluoroscopy and image
guided interventional procedures, and through choice of scan
parameters in CT. For digital
radiography image cropping, performed after the exposure, does not achieve the
same reduction in
exposed volume as collimation. See also the paragraphs below for more
specific guidance for the
different modalities.
3.154. Patient co-operation should be achieved to produce a diagnostic
quality image. This is
particularly relevant when imaging children. Good communication helps
to achieve this. Verbal
interaction between the medical radiological technologist or the medical
radiological practitioner and
the patient should be in place before, during and after the procedure.
3.155. Optimization of protection for women undergoing radiological
procedures during pregnancy
should take into account the woman and the embryo/fetus. Routine diagnostic CT
examinations of the
pelvic region with and without contrast injection can reach lead to a dose of
50 mSv to the uterus
which is assumed to be equivalent to the fetal dose in early
pregnancy. When CT scanning is
indicated in a pregnant patient, Low-dose CT protocols should be used and the
scanning area should
be reduced to a minimum possible.
3.156. Shielding of radiosensitive organs such as the gonads, lens of
the eye, breast and thyroid
should be used when appropriate.
3.157. For each modality there are a number of factors which can be adjusted to
influence the image
quality and patient dose relationship. Written protocols that specify the
operational parameters to be
used for common diagnostic radiological procedures should be developed, adopted
and implemented
in each radiology facility. The protocol ‘technique charts’ should be posted
adjacent to each X ray
generator and be specific for each piece of equipment. The protocols should
take into account patient
habitus, especially mass. The protocols are best developed using
guidelines from national or
international professional bodies, and hence reflect current best practices, as
for example in Refs [145
– 155]. For modern digital equipment many of the factors are
automated through menu-driven
examination selections on the console. Nevertheless, in setting up
these options, significant scope
exists for optimization of protection through the appropriate
selection of values for the various
technical parameters, thereby producing effectively an electronic ‘technique
chart’.
3.158. Size specific written protocols should be developed for children, from
neonates to teenagers,
and include additional operational considerations, such as the use of
additional filtration or the
removal of grids when appropriate [151, 154, 155].
3.159. The BSS paragraph 3. 166 (b) set special requirement to the
optimization of protection for
individuals subject to medical exposure as part of an approved health
screening programme. All
aspects should be considered before the approval of the program and
during its implementation:
selection of X-ray equipment suitable for the particular screening, parameters
settings, etc. Dedicated
quality assurance program should be implemented to meet screening objectives,
as described in more
details in paras 3.225 – 3.237.
Operational considerations - radiography
3.160. In developing protocols for radiography, many technique factors should
be considered which
can influence the image quality and the patient dose for the radiographic
projection. Detailed guidance
on appropriate choices for those factors is widely available [145, 150, 151,
156 – 162]. Such factors
include:
(a) The tube potential (kV); current (mA); exposure time; focal
spot size; filtration; source to
image receptor distance (SID, FID or FFD); choice of anti-scatter
grids or Bucky device;
collimation; image receptor size; patient positioning, immobilisation and
compression;
(b) The number of projections needed (e.g. PA chest X ray rather than PA
and lateral X rays);
(c) Organ shielding where appropriate (e.g. testicular shielding for
pelvic radiographs in boys);
3.161. Suitably calibrated and maintained automatic exposure controlled
systems (AEC) should be
used when available and appropriate. Particular attention should be given to
paediatric radiography to
ensure that the AEC sensor(s) is(are) within the radiation field. AEC systems
are calibrated based on
the radiation exposure at the detector required to produce the desired level of
optical density (OD) for
film-screen systems or a pre-determined acceptable level of signal to noise
ratio (SNR), or surrogate,
for digital systems. The value for the SNR should be established as part of
setting up the protocols for
radiographic projections for each particular X ray unit. When AEC is
not available, consideration
should be made of the patient size and thickness of the body part imaged in
determining technique
factors.
3.162. For digital systems, users should understand how the selection of
“exposure index” (or similar
term for exposure indicator) affects patient dose. For some manufacturers,
increasing the index lowers
the dose; for others, it increases it [163].
3.163. For film-based image acquisition systems, additional factors
include: the type (speed and
spectral response) of film–screen combination and film processing conditions
(e.g. the chemicals used
and developing time and temperature).
3.164. Mobile and portable radiographic equipment usually produce
images of lower quality
compared with fixed units, and should only be used for examinations where it is
impractical or not
medically acceptable to transfer patients to a fixed unit.
3.165. Patient should be properly positioned and immobilized. In addition,
instructions should be clear
and in the language understood by the patient (e.g. “Please hold your breath”).
Operational considerations - mammography
3.166. In developing protocols for mammography, consideration of
radiographic technique factors
should be made as for radiography, in para 3.160. Additional factors
that should be considered
include: adequate compression; tissue composition (e.g. dense glandular breasts
identified on previous
mammograms); and correct choice of anode and filters. Detailed guidance on
appropriate choices for
technique factors and additional factors is available [119 - 122, 147, 164 –
166].
3.167. For film-based mammographic systems, additional factors include:
the type of film–screen
combination and film processing conditions (e.g. the chemicals used,
and developing time and
temperature), as discussed in Refs [119 – 121].
3.168. Breast tomosynthesis is an evolving technique where guidance
for optimization is likely to
become available as the modality matures. A review of features that influence
image acquisition has
been made in Ref. [167].
3.169. Viewing conditions are of paramount importance for both
digital and film-based
mammography systems, and the operational performance should be meet the
conditions described in
paras. 3.24, 3.25 and 3.41. Poor viewing conditions not only
compromise the reporting of a good
quality image, but they may also lead to changes in technique
factors, in a mistaken attempt to
compensate for the poor viewing conditions, that actually result in
sub-optimal image quality. For
example, use of low luminance viewing boxes may lead to radiographs being
produced that have a
low density with insufficient diagnostic content. Although the dose may be
reduced, there might be an
unacceptable loss of diagnostic information.
Operational considerations - CT
3.170. In developing protocols for CT, many technique factors and
features should be considered
which can influence the image quality and the patient dose for the
examination, including: tube
potential; tube current; tube current modulation with noise index; pitch; beam
width; total scan length,
over-ranging and over-beaming. These and other factors may be
optimized through automatic
exposure control systems where available. The choice of protocol will be
determined by the clinical
question to be answered (e.g. for cardiac CT, a low dose protocol is sufficient
for stratifying risk in
patients with intermediate probability for coronary artery disease,
whereas a higher dose contrast
enhanced protocol is needed for patients with suspected coronary artery
disease). Detailed guidance
on appropriate choices for these factors and features is available [65, 146,
152, 153, 155, 157, 160,
161, 168 – 172].
3.171. Careful consideration should be made as to the need for multiple phase
studies to answer the
clinical question (e.g. CT abdomen portal venous phase only for routine
detection of liver metastases
rather than triple phase - arterial, portal venous and delayed phase
acquisitions). Protocols for
optimized CT procedures for common clinical conditions should be
agreed, put in place and used.
This applies particularly to children where protocols based on size should be
used [153, 160, 161].
3.172. Consideration of a spiral or axial technique will depend on the
indication and will have an
implication on image quality and dose (e.g. for diffuse lung disease a
non-contiguous single slice
protocol is preferred for high resolution lung CT, and delivers lower patient
dose).
3.173. Improved image presentation, reconstruction algorithms and
post-processing features to reduce
image noise can have the potential to enable the choice of a protocol with
reduced patient dose. An
example is the use of iterative reconstruction algorithms.
3.174. Proper positioning of the patient and proper setting of the scanned
anatomical area of interest
should be achieved, e.g. CT thorax with both arms raised or CT of the wrist in
the ‘superman position’
(i.e. with the patient lying prone with the affected arm stretched out
above his or her head) are of
considerable advantage to avoid artefact and to reduce dose. Immobilizing
devices may be used where
appropriate. Special attention should be made for proper immobilisation of
paediatric patients by use
of straps, swaddling clothes, plastic holders for the head or
body, foam pads, sponges, sand bags,
pillows or other objects.
3.175. Including the lens of the eye in the primary beam should be avoided.
This may be achieved in
brain scans by using a head cradle or, in some cases, tilting the gantry.
3.176. For CT angiography, use of software to detect the arrival of contrast
medium in the relevant
vessel to trigger the volume acquisition has image quality advantages and
avoids repeat acquisitions
(e.g. detecting contrast medium in the pulmonary artery in CT pulmonary
angiography).
3.177. For cardiac CT and CT angiography, use of software to control
acquisition with respect to the
patient electrocardiograph (ECG-gated or ECG triggered studies) should
be considered, when
appropriate, to reduce radiation dose.
3.178. For hybrid imaging with CT (e.g. PET-CT and SPECT-CT), consideration
should be given to
the use of a low dose CT protocol to correct for PET or SPECT attenuation,
which may necessitate a
second diagnostic procedure of the primary area of interest or a
higher dose CT protocol (often
contrast-enhanced) as part of the hybrid procedure.
3.179. Cone beam CT (CBCT) is used for imaging the head and neck
(e.g. sinuses where there is
significant artefact from dental amalgam) as well as for dental indications.
Operational aspects with
respect to optimization are still evolving. Guidance is available [173, 174]
and factors that should be
considered include: tube potential, tube current-exposure time product, field
of view, voxel size, and
the number of projections.
Operational considerations - dentistry
3.180. In developing protocols for conventional intra-oral radiography, factors
that can influence the
image quality and the patient dose include: tube potential (kV);
current (mA); exposure time;
collimation; focus to skin distance; and for analogue systems, film speed and
processing development
time and temperature. Detailed guidance on appropriate choices for those
factors is available [175,
176].
3.181. In developing protocols for panoramic imaging, additional factors that
can influence the image
quality and the patient dose include: patient positioning (e.g. jaw open /
closed); collimation (e.g. for
temporomandibular joint examinations only these areas should be
included); and for analogue
systems, film/screen speed and processing development time and temperature.
Detailed guidance on
appropriate choices for those factors is available [175, 176].
Operational considerations - image guided interventional procedures
3.182. The choice of imaging modality for guidance will depend on
the clinical scenario (e.g.
fluoroscopic guidance for percutaneous coronary intervention, CT guidance for
biopsy). Occasionally
more than one modality may be used in a single interventional procedure to
improve effectiveness and
safety. This may result in a lower dose when the second modality is
non-ionizing (e.g. ultrasound is
used to locate the renal pelvis in percutaneous nephrostomy before fluoroscopic
catheter placement).
Furthermore the correct selection of equipment with appropriate size
(and shape) of flat panel or
image intensifier will improve diagnostic image quality.
3.183. Patient co-operation should be ensured forthe successful
interventions (e.g. movement may
compromise accuracy of roadmaps when performing aneurysm embolization in
neuro-intervention). It
is imperative that patients are briefed about the intervention prior
to the commencement of the
procedure so that they know what to expect and how to cooperate.
3.184. In developing protocols for fluoroscopically guided interventional
procedures, many technique
factors and features should be considered which can influence the image quality
and the patient dose
for the intervention, including: tube potential; tube current; use of pulsed
fluoroscopy (hence pulse
width and rate); dose rate mode (effectively the image intensifier or flat
panel detector input air kerma
rate); collimation, and collimation tracking with focus to detector
distance; filtration (fixed and
variable); use of magnification; total fluoroscopy time for the
intervention; image acquisition dose
mode (effectively the image intensifier or flat panel detector input
air kerma per frame); image
acquisition frame rate; number of frames per run and the total
number of acquisitions. Detailed
guidance on appropriate choices for these factors and features is available
[123, 157, 158, 160, 161,
177 – 180].
3.185. Many of the factors in para. 3.182 are automated through an
algorithm-driven ‘automatic
brightness control system’ (ABC). Nevertheless, in setting up the
algorithm, scope exists for
optimization of protection through the selection of values for these
parameters. For example, image
intensifier or flat panel detector input air kerma rates (for fluoroscopy) and
input air kerma per frame
(for image acquisition) are set during installation and adjusted thereafter
during periodic maintenance
and servicing. The values actually used for these settings can vary
considerably. High rate dose modes
in fluoroscopy should be used only during the minimum indispensable
time necessary to the
procedure. The use of magnification modes should be kept to a minimum
consistent with a successful
intervention.
3.186. In the course of the intervention the tube orientation and position may
need to be changed – for
long procedures the area of skin where the X ray beam is incident upon the
patient should be changed
during the procedure to avoid deterministic skin effects. As a
default from a radiation protection
perspective, it is preferable to have the X ray tube under the
patient (i.e. “under couch”). Steep
oblique projections should be avoided. The distance between the X ray tube and
patient should always
be maximized to reduce patient dose. Typically for a vertical beam this
is achieved by having the
table as high as possible for the primary operator. In conjunction with this,
the image intensifier or flat
panel detector is positioned as close to the patient as possible.
3.187. Particular paediatric considerations include: use of special filtration,
removal of the grid; gonad
protection.
3.188. In developing protocols for CT guided interventional procedures,
technique factors that should
be considered, which can influence the image quality and the patient
dose for the intervention,
include: tube potential; tube current; and beam width. The number of image
acquisitions should be
kept to a minimum consistent with a successful intervention.
Operational considerations - fluoroscopy
3.189. See fluoroscopic considerations in the image guided
interventional procedures sub-section,
paras. 3.184 – 3.187.
Operational considerations - bone densitometry
3.190. Selection of the appropriate site for densitometry will take into
account both the anatomical
area of clinical concern as well as the likelihood of non-representative images
and measurements due
artefacts (e.g. massive vertebral osteophytes may obviate the value of lumbar
densitometry).Guidance
is given in Ref [181].
Operational considerations - emergency radiology
3.191. Special considerations for the emergency department include:
judicious patient positioning
taking into account injury or disease (e.g. lateral shoot through
projection of the hip); and CT
protocols with the minimum number of acquisitions (e.g. contrast enhanced CT
for polytrauma when
one acquisition only is needed for diagnosis and expedience).
Calibration - general
3.192. Following the BSS paragraph 1.46, for diagnostic radiology and image
guided interventional
procedures the dosimetric quantities and units of ICRU are to be used [8, 182].
Further guidance on
dosimetry in diagnostic radiology is given in Refs [9, 183].
3.193. Calibration requirements for medical radiological equipment and
dosimetry equipment are
given in the BSS paragraph 3.167. Responsibility is assigned to the
radiology facility’s medical
physicist. After the initial calibration, the intervals for periodic
calibrations may differ, depending on
the complexity of the medical radiological equipment. Related to calibrations
are the constancy tests
on equipment performance performed as quality control tests. These are
described in paras 3.228 and
3.230 – 3.231.
Calibration – medical radiological equipment
3.194. In diagnostic radiology, including medical radiological equipment used
for radiation therapy
simulation and treatment verification and hybrid imaging systems,
and for image guided
interventional procedures, ‘source calibration’ is to be interpreted as
the measurement of certain
dosimetric quantities which are modality-dependent and should be carried out in
reference conditions.
3.195. For diagnostic radiographic and fluoroscopic medical radiological
equipment, including
conventional radiation therapy simulators, the dosimetric quantities are:
incident air kerma, Kₐ,i, in Gy,
incident air kerma rate, Ķₐ,i, in Gy.s⁻¹ and air kerma-area product,
PKA, in Gy.m² (noting that some
manufacturers use μGy.m² or mGy.cm² or Gy.cm²).
3.196. In CT, the dosimetric quantities are (see also [8, 9, 182 - 186]):
(a) CT air kerma index, CK, usually in mGy. In many countries
the more colloquial term
computed tomography dose index, CTDI, is used, and has been accepted by ICRU
[182];
(b) Weighted CT air kerma index, CW, usually in mGy, which is the CT air
kerma calculated from
measurements at the centre and periphery of a standard PMMA CT head or body
phantom. As
in (a), this quantity is often simply called weighted CTDI or CTDIW;
(c) Volume CT air kerma index, CVOL, usually in mGy which takes into
account the helical pitch
or axial scan spacing. As in (a), this quantity is often simply called volume
CTDI or CTDIVOL;
(d) CT air kerma-length-product, PKL,CT, usually in mGy.cm. In
many countries the more
colloquial term dose-length-product, DLP, is used, and has been accepted by
ICRU [182].
3.197. In mammography, three dosimetric quantities are used: incident air
kerma, entrance surface air
kerma and mean glandular dose, DG, usually in mGy [8, 9].
3.198. Measurements of these dosimetric quantities, when being used to
calibrate or characterize a
given X ray, CT or mammography unit output or performance, should be
made for a range of
representative technique factors used clinically, and following recognized
protocols such as in Ref [9].
Calibration – dosimetry instrumentation
3.199. Dosimetry instrumentation used at a radiology facility should
be calibrated at appropriate
intervals. A period of two to three years is recommended. See also para. 3.235
on associated quality
assurance guidance.
3.200. The BSS in paragraph 3.167(d) requires the calibration of
dosimetry instrumentation to be
traceable to a standards dosimetry laboratory (SDL). Ideally this
would be to the national SDL
(primary or secondary) in the State concerned, with access either directly or
through a duly accredited
calibration facility. However, it may be necessary for dosimetry
instruments to be sent to another
country or state if there is no national SDL in the country or state where the
instruments are used. At
present only some of the secondary SDLs of the IAEA/WHO network provide
calibration services
using diagnostic radiology spectra and dose rates representative of clinical
practice. However, since
dosimetry accuracy is not as critical in diagnostic radiology as in radiation
therapy, calibrations with
comparable radiation qualities should be sufficient. Alternatively, the
regulatory body may accept
instrument manufacturers’ “calibrations” as spelled out in the “certificate of
calibration” issued by the
instrument manufacturer, provided that the manufacturer operates or uses a
calibration facility that is
itself traceable to a SDL and appropriate calibration conditions have been
used. This certificate should
state the overall uncertainty of the calibration factors.
3.201. Records of calibration measurements and associated calculations,
including uncertainty
determinations (budgets), should be maintained as described in
para. 3.263. Uncertainty
determinations for several radiological examinations have been calculated by
the IAEA [9, 161].
3.202. There is a role for cross-calibration of dosimeters, where the
radiology facility’s dosimeters
that have been officially calibrated are used to check or compare
with other dosimeters. This is
particularly important for field KAP meters which should be calibrated (or
cross-calibrated) against a
reference KAP meter or air kerma dosimeter in situ in the clinical rather than
in a SDL environment
[9]. It might also occur when a radiology facility has many dosimeters, and
to calibrate all dosimeters
could be too costly. Cross-calibration can also be utilized as a constancy
check, as part of periodic
quality control tests.
Dosimetry of patients
3.203. The BSS paragraph 3.168 requires facilities to ensure that
patient dosimetry in diagnostic
radiology and image guided interventional procedures is performed and
typical doses for their
radiological procedures are determined. Knowledge of a facility’s typical
doses forms the basis for
applying methods of dose reduction as part optimization of protection. It also
enables the radiology
facility to use diagnostic reference levels (see paras. 3.217 to 3.224) as
another tool for optimization
of protection.
3.204. Clearly, the more radiological procedures at the radiology facility for
which typical doses are
known, the better the basis for optimization of protection. However,
pragmatically, the BSS requires
typical dose determination for only common radiological procedures in radiology
facilities. What is
common will vary from facility to facility, and country to country, but in
general there are some core
common examinations, including:
(a) Radiography – head, chest, abdomen and pelvis;
(b) CT – head, chest, abdomen and pelvis, for specified clinical
indications;
(c) Fluoroscopy – barium swallow, barium enema;
(d) Mammography – cranio-caudal, medio-lateral-oblique;
(e) Dentistry – intraoral, panoramic (orthopantomography), cone-beam CT
(f) Bone densitometry (DEXA) – spine and hip.
3.205. For image guided interventional procedures, the facility should
ascertain typical doses for the
broad types of procedures they perform. For example, an interventional
cardiology facility would
characterize typical doses for percutaneous coronary interventions,
including PTCAs. A facility
performing neurological procedures might characterize typical doses for
their diagnostic cerebral
angiograms and for their embolization interventions. Other image guided
interventional procedures
might include ERCP (endoscopic retrograde cholangiopancreatography) and
TIPS (transjugular
intrahepatic portosystemic shunt).
3.206. The term ‘typical dose’, as used in the BSS paragraph 3.168, means the
average or median of
the doses for a sample of relatively standard-sized patients, at
clinically acceptable image quality.
Patient size has a large influence on dose, so some selection or grouping of
patients is required. Such
groupings include ‘average adult’, often based around an average weight of 70
kg with a range of ± 20
kg. Groupings for children have sometimes been based on age, such as new born
(0 years), infant (1
year), small child (5 years), child (10 years) and teenager (15 years), but
more recently size-specific
groupings are being used. The radiology facility should adopt patient size
groupings that correspond
with the groupings used in their country or state for DRLs. The
sample size used for each patient
grouping and radiological procedure should be sufficient to assure confidence
in the determination of
the mean dose. Such sample sizes are typically in the range of 20 patients, but
clearly the larger the
sample the better the statistics.
3.207. Dose in the term ‘typical dose’, as used in the BSS
paragraph 3.168, means for the given
radiological procedure an accepted dosimetric quantity as described in
paras. 3.195 - 3.196. For
particular reasons, e.g. for risk estimation or for collective dose estimation,
organ doses or effective
dose can be estimated from typical dose.
3.208. Patient dosimetry to determine typical doses should be carried
out in conjunction with an
assessment of the diagnostic image quality. Exposure alone is not meaningful if
it does not correspond
to images that are adequate for an accurate diagnosis. Therefore patients
included in the sample used
for determining typical doses should be only those whose radiological
procedure resulted in
acceptable image quality.
3.209. The results of the surveys used to determine typical doses at the
radiology facility should be
used as part of the facility’s on-going review of the implementation of
optimization of protection, and
that additionally will be used for comparison with established DRLs (see paras.
3.217 to 3.224). The
results should also be submitted to the organization in their country or
state that is responsible for
establishing and reviewing the national or regional DRLs. Patient dosimetry
surveys, required by the
BSS, should take place at intervals of no more than 5 years and
preferably no more than 3 years.
Another trigger for a survey would be the introduction of new
equipment or technology into the
radiology facility or when significant changes have been made to the protocols
or the equipment.
3.210. Sometimes patient dosimetry in diagnostic radiology or
image guided interventional
procedures may be required for specific individual patients,
either through measurements or
calculations. Reasons might include an unintended or accidental medical
exposure, where estimation
of patient doses is required as part of the investigation and report (see
para. 3.256) or because there is
a need to estimate embryo or fetal doses (see para. 3.155).
3.211. There are several indirect and direct methods to estimate patient dose
in diagnostic radiology
and image guided interventional procedures. Methodologies for these
determinations are explained in
detail in Refs [8, 9, 180, 182 – 187] and summarized below:
(a) Estimations based on incident air kerma (Kₐ,i ) or
entrance surface air kerma (Kₐ,ₐ)
measurements corrected for the techniques used (X ray tube potential,
current and time,
source-skin-distance, etc.). This approach can be used for radiography,
fluoroscopy and
mammography.
(b) Estimations based on measured air kerma-area product (PKA). This
approach can be used for
radiography, fluoroscopy and CBCT.
(b) Estimations based on measurements of CT air kerma index (CVOL) and
CT air kerma-length
product (PKL,CT). This approach can be used for CT.
(c) Reported values of dose quantities from DICOM headers or
the DICOM radiation dose
structured reports. The accuracy of the reported dose quantities should have
been validated at
acceptance testing and commissioning and during QA procedures as explained in
para. 3.235.
This approach is applicable to all digital modalities.
(d) Direct measurements in selected organs such as the skin in
interventional procedures. For this
purpose thermoluminescent and optically-stimulated dosimeters (TLDs and OSLs)
as well as
radiochromic or silver halide film can be used.
(e) In the case of CT, size-specific dose estimates, SSDE, can be made,
where CVOL values are
corrected taking into consideration the size of the patient using linear
dimensions measured on
the patient or patient images [182, 187].
3.212. When needed, organ doses can be derived from the quantities
mentioned in para. 3.211 by
using conversion coefficients derived from Monte Carlo codes applied to
anatomical models. Methods
for doing this are discussed in Ref [9].
Dosimetry of patients - specific considerations for image guided interventional
procedures
3.213. For interventional procedures using X rays, in addition to the
quantities that are related to
stochastic effects, such as air kerma-area product, it is critical to monitor
cumulative doses to the most
exposed areas of skin because of the potential for reaching the
threshold for tissue effects in
complicated cases [188, 189].
3.214. The determination of the dose to the most exposed area of skin is not
straightforward since
exposure parameters and projection angles change during the procedure and the
most exposed area
cannot always be anticipated. This makes the knowledge of the skin
dose distribution (sometimes
called ‘dose mapping’ over the skin) necessary. A comprehensive review of
approaches to obtain dose
mapping and to determine the most exposed area of the skin is given in Ref
[180].
3.215. An established method for dose mapping uses low sensitivity X ray films,
such as films used in
radiation therapy and radiochromic films. However, determination of the dose is
only possible after
the procedure.
3.216. The cumulative dose at the interventional reference point Kₐ,r, defined
as the kerma in air at
15 cm from the isocentre in the direction of the X ray tube [74], either
displayed during the procedure
and/or obtained from the DICOM header, may be used as a conservative estimate
for peak skin dose.
The degree of over-estimation depends on several factors, including how often
the beam projection
was changed. The cumulative dose at the interventional reference point gives
the least over-estimation
when most of the radiation is delivered in just one beam projection.
The accuracy of the reported
cumulative dose at the interventional reference point should have been
validated at acceptance testing
and commissioning and during QA procedures, as explained in para. 3.233.
Diagnostic reference levels
3.217. The BSS, paragraphs 3.168 - 3.169, requires that radiology facilities
perform patient dosimetry
surveys, as described in paras 3.203 to 3.209, and that these results are
compared with the established
DRLs for their country or region. The purpose is to ascertain whether the
typical dose in the facility
for a given radiological procedure compares favourably, or not, with the value
of the DRL for that
radiological procedure. Guidance on establishing national or regional
DRLs in given in Section 2,
paras. 2.34 to 2.45.
3.218. A review of optimization of protection for that particular radiological
procedure is triggered if
the comparison shows that the facility’s typical dose exceeds the DRL, or that
the facility’s typical
dose is substantially below the DRL and it is evident that the exposures are
not producing images of
diagnostic usefulness or are not yielding the expected medical benefit to the
patient.
3.219. Given the uncertainties in determining a facility’s typical dose
(see paras 3.206 and 3.207),
questions can arise over whether a DRL has really been exceeded or not. Some
countries adopt an
algorithmic approach, for example where the facility’s typical dose,
minus two times its standard
error, should be greater than the value of the DRL [13]. A simpler
approach, based purely on the
facility’s typical value, may be sufficient as the purpose is to
identify the need for a review, and
perhaps it is better to perform more reviews than fewer.
3.220. No individual patient’s dose should be compared with a DRL. It is the
facility’s typical dose, as
determined by the representative patient sample, which is compared.
3.221. Further, the comparison is not a case of “Does the radiology facility
comply with the DRL?”
DRLs are not dose limits. It is mandatory to use them for the comparison
exercise to identify practices
that warrant further investigation.
3.222. The review of how the given radiological procedure is being performed
and its optimization of
protection, triggered by the DRL comparison, may conclude that there are valid
reasons supported by
sound clinical judgement why the radiology facility has a typical dose that
exceeds the DRL. These
reasons should be documented as part of the facility’s QA programme.
Adequateness of image quality
should be always taken into account. On the other hand, the review
may identify areas for
improvement resulting in revised protocols for that radiological procedure.
The results of the DRL
comparison and any ensuing review and actions should be documented as part of
the facility’s QA
programme.
3.223. The fact that a radiology facility’s typical dose for a
radiological procedure is less than the
DRL for that procedure does not mean necessarily that optimization of
protection for that radiological
procedure has been fully achieved. DRLs are only one of the tools for
optimization, aimed specifically
at identifying the outliers in performance.
3.224. The regulatory body in a given country may specify frequencies
for performing DRL
comparisons. Otherwise, the general guidance on patient dosimetry, described
above in para. 3.207,
would be applicable.
Quality assurance for medical exposures
3.225. The BSS paragraph 3.170 requires radiology facilities to have a
comprehensive programme of
quality assurance for medical exposures. General guidance on management
systems is given in
Section 2, paras 2.135 to 2.146, and it is simply re-iterated here
that the programme of quality
assurance for medical exposures should fit in with, and be part of, the
facility’s wider management
system. The paragraphs in this subsection give guidance on the technical
aspects of the programme of
quality assurance for medical exposures.
3.226. The purpose of the programme of quality assurance for medical exposures
is to help ensure
successful implementation of optimization of patient protection in the
radiology facility and to
minimize the occurrence of unintended and accidental medical exposures.
3.227. The complexity of the programme of quality assurance for medical
exposures will depend on
the type of facility. A dental practice with intra-oral only radiography will
have a simpler programme
compared with a facility that offers all modalities of diagnostic
radiology as well as image guided
interventional procedures. Nonetheless, most of the elements of the programme
are common, and it is
more in the degree of implementation that there are differences. The BSS
paragraph 3.171 establishes
the common elements of the programme.
3.228. Measurements on medical radiological equipment are one
of the components of the
programme. Acceptance tests are required for new or significantly refurbished
or repaired equipment,
or after the installation of new software or modification of existing
software that could affect
protection and safety of patients. The acceptance test is followed immediately
by commissioning, and
then on-going periodic quality control tests, including constancy tests. The
purpose is to ensure that,
at all times, all medical radiological equipment is performing correctly,
accurately, reproducibly and
predictably. Acceptance and commissioning tests should be performed in the same
way for equipment
and/or software that has been donated.
3.229. Depending on the equipment purchase agreement, acceptance tests may be
performed by the
manufacturer’s representative in the presence of the local medical
physicist and the radiological
medical practitioner representing the user, or, if acceptable to the
manufacturer and/or the purchaser,
by a medical physicist jointly with the manufacturer’s
representative. The process involves
verification of all specifications and features of the equipment.
3.230. After acceptance and before clinical use on patients, commissioning is
carried out by or under
the supervision of the medical physicist. Commissioning should include
measurements of all
parameters and conditions of use that are expected in clinical use, including
setting up or validating
image acquisition protocols. For most modalities (CT, image guided
interventional procedures,
tomosynthesis, mammography, DR/CR, fluoroscopy) the medical
physicist should be directly
involved in the measurements, calculations and interpretation of data to
characterize the equipment’s
performance. For the least complex modalities (dental and DEXA) the
medical physicist should
provide documented advice on how the commissioning should be performed. At
commissioning, the
baseline for subsequent constancy tests is established.
3.231. There are many published reports from international and national
organizations and national or
regional professional bodies giving detailed guidance on the
performance tests and quality control
tests that should be performed on the various modalities, including recommended
frequencies [111,
112, 117 – 122, 166, 170, 175, 176, 179 – 181, 183, 190 – 211].
In addition, many of these
organizations have extensive websites where any new or updated
publications on the topic can be
found. The regulatory body may have its own specific requirements on
the tests that should be
performed and their frequencies. It would be expected that such
specific requirements would have
been established with consultation between the regulatory body and the relevant
professional bodies.
3.232. Quality control tests also should be performed on other
equipment or devices that have an
impact on the successful outcome of the radiological procedure. Such equipment
and devices include,
but are not limited to: film processors, darkrooms, and cassettes for
facilities using film-based
imaging; CR plates and CR readers for facilities with CR systems; and view
boxes, workstations, and
reporting rooms. Many of the references given in para. 3.231 are applicable
here.
3.233. The results of the quality control tests should be compared with
established tolerance limits.
These limits may have been established to ensure compliance with a regulatory
requirement for the
performance of particular physical parameters or they may be set on the basis
of recommended values
given in published reports, such as referenced in para. 3.231.
As required in the BSS paragraph
3.171(b), if the measured values fall outside the tolerance limits, corrective
actions should take place.
Such corrective actions are likely to include maintenance or servicing of the
equipment, and hence the
radiology facility should have a preventive maintenance programme in
place. In some cases, the
equipment may be outside the tolerance limits by a significant amount and the
equipment should be
immediately taken out of clinical use and not returned until the servicing has
taken place and it has
been ascertained that the equipment meets the performance requirements.
3.234. The programme of quality assurance for medical exposures in
the radiology facility should
include use of “checks and balances” to ensure that the facility’s protocols
and procedures for imaging
and interventional procedures, including radiation protection and
safety, are being followed. The
periodic review of the protocols and procedures themselves is part of the
facility’s radiological review
(see 3.260 – 3.262). In addition a review of imaging procedures may
have been triggered by a
comparison with DRLs (see paras. 3.217 to 3.224).
3.235. The BSS paragraph 3.171(e) specifically requires that dosimetry and
monitoring equipment are
part of the QA programme. This is to ensure that such
instrumentation has a current calibration,
typically within 2 or 3 years (see para. 3.199), and that it is functioning
correctly. The programme of
quality assurance for medical exposures should establish a calibration cycle
for each instrument and a
set of quality control checks on the operation of each instrument to be
performed at set intervals. This
applies to stand alone dosimetry equipment and to dosimeters integrated into
the medical radiological
equipment, such as KAP meters in fluoroscopic systems, and to software of the
medical radiological
equipment itself that calculates, displays and reports dose metrics such as
CTDI and DLP in CT and
Kₐr in image guided interventional procedures.
3.236. Maintaining records is a crucial aspect of the programme of
quality assurance for medical
exposures. This includes the procedures used in the programme, and the results
of the quality control
tests, the dosimetry surveys, the DRL comparisons, corrective actions
and the investigations of
unintended and accidental medical exposures. When planning and
developing an effective QA
programme, licensees should recognize that it demands strong managerial
commitment and support in
the form of training and time, personnel and equipment resources.
The regulatory body, during its
inspections of a radiology facility, should review the records of the programme
of quality assurance
for medical exposures.
3.237. In line with standard practices for quality management, the BSS in
paragraph 3.172 requires
regular and independent audits of the programme of quality assurance for
medical exposures, adding
that the frequency of such audits will depend on the complexity of the
radiological procedures being
performed in the facility. Such audits may be internal or external.
Internal audits are usually
logistically simpler to implement, while the external audit generally has the
advantage of bringing in
an outside perspective. The audit of the radiology facility’s
programme of quality assurance for
medical exposures can be incorporated into more comprehensive audits that the
facility is performing
for its management system. Further, the results of the audit of the programme
of quality assurance for
medical exposures will be a major input into the radiology facility’s
radiological review (see paras.
3.260 to 3.262).
Dose constraints – carers and comforters
3.238. Some diagnostic radiological procedures, particularly of children, can
be better performed with
the assistance of a helper – the carer or comforter, for example a relative in
the case of a paediatric
patient, or a relative or friend for a disabled or very elderly or very ill
patient. In these circumstances,
the carer or comforter will be exposed, usually to a low dose.
3.239. The BSS paragraph 3.153 states that a carer or comforter must be
informed about the radiation
risks involved in helping with the radiological procedure and about the means
to be taken to afford
appropriate radiation protection to the carer or comforter. The carer or
comforter should indicate that
he/she has understood the information and is still willing to help with the
radiological procedure.
3.240. The radiation protection afforded the carer or comforter should be
optimized, and as part of this
process dose constraints should be applied, as required in the BSS
paragraph 3.173. These are the
dose constraints established by government, as a result of
consultation with the health authority,
relevant professional bodies and the regulatory body, as required by the BSS
paragraph 3.149(a)(ii)
(see also Section 2, paras, 2.48 to 2.49).
3.241. The radiology facility should have written protocols for the
optimization of protection
measures for carers and comforters who hold patients during radiological
procedures. The measures
should utilize the basic methods for radiation protection, i.e., time and
distance. The protocol should
include the following:
(a) Methods to avoid the need for holding patients, for example the
administration of sedatives
(especially for long procedures such as CT examinations) and the use of infant
restraints;
(b) Criteria specifying which carers and comforters are allowed
to hold patients, for example
friends and relatives, provided that they are not pregnant, but not employees
such as porters
and nurses (see also Section 2 para. 2.49);
(c) Methods for positioning and protecting the carer or comforter so
that his or her exposure is as
low as reasonably achievable, for example by ensuring that the carer or
comforter is not in the
direct beam of the radiation device and that appropriate personal protective
equipment is used,
for example a protective apron or ancillary shields of a specified lead
equivalence;
(d) The values of the dose constraints to be applied (see
Section 2 para. 2.49) depend on the
radiological exam or intervention; a common value may be 5 mSv per event as
stated in para
2.49. Although it is unlikely that a child, closely related to the patient,
would be a carer or
comforter for a diagnostic radiological procedure, in cases where this is
unavoidable, his or
her dose should be constrained to less than 1 mSv.
3.242. Registrants and licensees should be able to demonstrate that the
effective dose to the carer or
comforter, by applying the protocol, is unlikely to exceed the dose
constraint. It is relatively
straightforward to estimate effective doses to carers and comforters
from measurements of the
ambient dose equivalent rates at the positions where they will be
situated. These determinations
should be made in advance to verify that dose constraint will not be exceeded.
Therefore, individual
dose monitoring is normally not necessary.
Dose constraints – volunteers in biomedical research
3.243. Some individuals will undergo diagnostic radiological procedures
as part of their voluntary
participation in an approved programme of biomedical research (see Section 2
para. 2.98). Part of the
approval process for the biomedical research will have been the
setting of dose constraints for the
radiological procedures (see Section 2 paras 2.98 - 2.99). When the volunteer
presents him/herself at
the radiology facility, he/she is to be afforded the same radiation protection
as if he/she were a patient
ready to undergo a radiological procedure within a normal healthcare pathway,
but with the additional
restriction that his/her exposure will be subject to a dose
constraint, either a nationally established
dose constraint or as specified by the ethics committee that approved
the biomedical research
programme (see Section 2 paras. 2.50, 2.98 and 2.99).
Pregnant female patients
3.244. Female patients who are pregnant form a special sub-group of
patients that should be
particularly considered with respect to radiation protection. These
considerations have been described
in para. 3.141(a) for justification and para. 3.155 for optimization. None of
these considerations can
take place if it is not known that the patient is pregnant. Therefore it is
crucial, as is required in the
BSS paragraphs 3.175 -3.176, for the radiology facility to have in
place means for ensuring that
pregnancy status is known for female patients.
3.245. The first approach is through posting of clear signs, in
languages easily understood by the
range of people using the radiology facility, asking the question
“Are you pregnant or possibly
pregnant?” and, if so, “Please tell the staff”. Such signs should be
posted widely in the facility,
including waiting rooms and cubicles. The second approach is to ask female
patients directly whether
they are or might be pregnant. This may not always be so easy given social and
cultural sensitivities,
but it should be done when needed.
3.246. Neither of the approaches described in para. 3.245 work if the woman
does not know if she is
or may be pregnant. For this reason, the BSS (paragraph 3.176) has
an additional requirement on
facilities to have “procedures in place for ascertaining the pregnancy
status of a female patient of
reproductive capacity before the performance of any radiological
procedure that could result in a
significant dose to the embryo or fetus”. Such radiological procedures
would include those that
involve primary beam irradiation of the abdomen or pelvis area
delivering relatively high patient
doses directly to the embryo or fetus, or to volumes near the uterus
such that significant scattered
radiation reaches the embryo or fetus. Cooperation with the referring
medical practitioner, through
standard requests for pregnancy status for specified radiological procedures,
is one approach. In case
of doubt, a pregnancy test or hormonal level determination to assess menopausal
status may be carried
out.
Unintended and accidental medical exposures
Prevention of unintended and accidental medical exposures
3.247. The BSS, paragraphs 3.179 to 3.181, sets out requirements both for
minimizing the likelihood
of unintended and accidental medical exposures and for the ensuing
investigation if such exposures
occur. The pathways identified in the BSS for potentially leading to unintended
or accidental medical
exposures include flaws in the design of medical radiological
equipment, failures of medical
radiological equipment while in operation, failures and errors in software that
control or influence the
delivery of the radiation, and human error. General strategies for addressing
those pathways include
regular maintenance of medical radiological equipment and software, a
comprehensive quality
assurance programme, and continuing education and training of staff. The
lessons learned from events
that have occurred should be used for preventing or minimizing unintended and
accidental medical
exposures, as described in para. 3.257.
3.248. A reduction in the probability of unintended or accidental
medical exposures in diagnostic
radiology and image guided interventional procedures can be brought about by:
(a) The introduction of safety barriers at identified critical points
in the radiology pathway, with
specific quality control checks at these points. Quality control is not
confined to physical tests
or checks but can include such as actions as correct identification of the
patient.
(b) Actively encouraging a culture of always working with awareness and
alertness.
(c) Providing detailed protocols and procedures for each process in the
radiology pathway.
(d) Education and training, including continuous professional
development, of all staff involved
in providing radiology services.
(e) Clear definitions of the roles, responsibilities and functions of
staff in the radiology facility
that are understood by all staff.
3.249. Preventive measures should include incident and near incident reporting,
analysis and feedback
including lessons learnt from international experience [123]. Preventive
measures should also include
each radiology facility checking the robustness of their safety system against
reported incidents. ICRP
85 [123] contains a review of case histories from a collection of unintended
and accidental medical
exposures in image guided interventional procedures.
3.250. Building on the guidance from paras 3.248 to 3.249, the following
three-step strategy can help
to prevent unintended and accidental medical exposures in the radiology
facility:
(a) allocate responsibilities to appropriately qualified health
professionals only and ensure that a
quality management system is in place that includes radiation safety;
(b) use the lessons learned from unintended and accidental medical
exposures to test whether the
quality management system, including for radiation safety, is robust
enough against these
types of events;
(c) identify other latent risks by posing the question “what else could
go wrong” or “what other
potential hazards might be present” in a systematic, anticipative
manner to all steps in the
diagnostic and interventional radiology process.
Investigation of unintended and accidental medical exposures
3.251. The events that constitute unintended or accidental medical exposures
are detailed in the BSS
paragraph 3.180. Unintended and accidental medical exposures may occur in all
imaging procedures;
however, the consequences in CT may be more severe and in image guided
interventional procedures
may be even more severe [123, 169].
3.252. Exposure of the wrong patient or wrong body part is always a possibility
in a radiology facility.
Many patients have similar names, for example, or patients may not have a
clear understanding of
what procedures are meant to take place. Facilities should have procedures
in place that consist of
several independent methods of patient identification, and verification of
examination requisition and
patient orientation.
3.253. One of the events requiring investigation is when the exposure was
“substantially greater than
was intended”. This situation may occur when the radiological procedure did not
go according to plan
– for example, the AEC in radiography may not have terminated the exposure when
expected because
the wrong sensors were selected or there was a hardware malfunction; or one or
more of the technique
factors in the examination protocol, for example for a CT examination, were
incorrectly set giving a
much higher dose than intended.
3.254. Another event that should be investigated is the inadvertent exposure of
the embryo or fetus in
the course of a radiological procedure, where at the time of the procedure it
was not known that the
woman was pregnant.
3.255. Radiation injuries have and do occur in image guided
interventional procedures. A given
procedure that may be performed in accordance with the facility’s protocol
still has the potential to
result in tissue effects because there were difficulties with the particular
patient. However, most cases
of reported severe radiation injuries involving ulceration and
necrosis were associated with
unnecessary and extreme exposure conditions, such as: (a) very short
distance between the X ray
source and the patient; (b) use of high dose-rate mode for a time much longer
than necessary; (c) fixed
projection exposing the same area of skin; and (d) malfunction of automatic
exposure control systems.
These situations cannot be considered to be normal, their occurrence can be
avoided and their severity
can be substantially reduced by optimization; they should be considered
accidental medical exposure
and investigated with a view to improve. Facilities performing
image guided interventional
procedures should have systems in place for identifying patients who
may be at risk from late
radiation injuries, typically based on estimates of peak skin
dose, cumulative dose at the
interventional reference point or kerma area product. For so-identified
patients, information should be
added to the patient’s medical record so that appropriate observation and
follow up is implemented.
For example, it is recommended that patients with estimated skin doses of 3 Gy
should be followed up
10-14 days after exposure [123]. Further information on trigger levels
for patient follow up are
available in the IAEA SAFRAD website [20]. Any resulting radiation
injury should receive
appropriate medical care.
3.256. The BSS, paragraph 3.181, states what should be done in the course of
the investigation. This
includes calculation or estimation of patient doses, which should be performed
by a medical physicist,
and notification of the event to the patient’s referring medical practitioner.
A record of the calculation
method and results should also be placed in the patient file. When required,
counselling of the patient
should be undertaken by someone with appropriate experience and clinical
knowledge.
3.257. The investigation of unintended and accidental medical
exposures, as required by the BSS
paragraphs 3.180 – 3.181, has three main purposes. The first is to
assess the consequences for the
patient (s) affected and provide remedial and health care actions if
necessary. The second is to
establish what went wrong and how to prevent or minimize the
likelihood of a recurrence in the
radiology facility – i.e. the investigation is for the facility’s and
their patients’ benefit. The third
purpose is to provide information to other persons or other radiology
facilities. Dissemination of
information about unintended and accidental medical exposures and
radiation injuries has greatly
contributed to increasing awareness worldwide of methods for avoiding
radiation injuries, for
example, by the FDA and ICRP Publication 85 [123, 188, 212, 213].
This might be through the
regulatory body and/or the health authorities for more significant events or as
required by a country’s
regulations, where the regulatory body disseminates information on the
event and the corrective
actions to other facilities that might learn from the event. Independently from
any legal requirement
for reporting to the regulatory body, the implementation of voluntary and
anonymous safety reporting
and learning systems can significantly contribute to improving safety and
safety culture in health care.
This includes participation in voluntary international or national
databases designed as educative
tools. One such database for image guided interventional procedures is
the so-called SAFRAD
reporting system (Safety in Radiological Procedures) [20]. Facilities
performing image guided
interventional procedures should participate in SAFRAD or similar databases.
3.258. As noted in para. 3.257, the BSS requires reporting to the regulatory
body and to the health
authority if appropriate for “significant” events. Further, the regulatory body
in a given country may
also specify their own requirements for reporting events to them. It
is difficult to quantify
“significant” – specifying a numerical trigger value immediately
creates an artificial distinction
between values immediately below that value (and hence should not be reported)
and those just above
the value (which should be reported). However, the attributes of
“significant” events can be
elaborated, and events with one or more of these attributes should be reported
to the regulatory body
and the health authority. Such attributes would include the occurrence of,
or the potential for, serious
unintended or unexpected health effects due to radiation, the likelihood of a
similar event occurring in
other radiology facilities, a large number of patients were affected,
and gross misconduct or
negligence by the responsible health professionals. As stated in para. 3.257
above, one of the roles of
the regulatory body for such a reported event is to disseminate
information on the event and the
lessons learned to all potentially affected parties – typically other
radiology facilities and relevant
professional bodies, but also in some cases manufacturers, suppliers, and
maintenance companies.
3.259. Irrespective of whether the event is also reported to the
regulatory body, feedback to staff
should be provided in a timely fashion and, where changes are
recommended, all staff should be
involved in bringing about their implementation.
Records and review
Radiological review
3.260. The BSS in paragraph 3.182 requires the performance of a periodic
radiological review at the
radiology facility. This involves considering both justification and
optimization aspects of radiation
protection. For the latter, the results of the programme of quality assurance
for medical exposures,
including the periodic independent audit, would be a significant input into the
process. As described
in Section 2, paras. 2.146 and 2.147, the wider clinical audit [52]
could include the radiological
review with its assessment of the effective application of the
requirements for justification and
optimization in the facility for the radiological procedures being performed.
3.261. To facilitate compliance with the BSS paragraph 3.182 and to learn from
periodic radiological
reviews, the methodology used, the original physical, technical and
clinical parameters considered,
and the conclusions reached should be documented and taken into account prior
to any new review
that may result in an update of the radiology facility’s policies and
procedures.
3.262. Radiological reviews should consider changes in patient
management that result from the
diagnostic or interventional procedure, the effect of introducing new
technologies on efficiency and
cost, and comparisons of different imaging modalities and of protocols for the
same pathologies.
Records
3.263. Records should be in place to demonstrate on-going compliance
with radiation protection
requirements. The BSS, paragraphs 3.183 to 3.185, specify requirements
for keeping personnel
records, records of calibration, dosimetry and quality assurance, and
records for medical exposure.
These records should be kept for the period specified by the country’s
regulatory body. In the absence
of such a requirement, a suggested period for keeping records is 10
years. In the case of children,
records should be kept for a longer time.
RADIATION PROTECTION OF THE PUBLIC
Introduction
3.264. Public exposure may arise from the performance of diagnostic
radiology and image guided
interventional procedures to persons in and around the radiology facility.
3.265. The requirements of the BSS for public protection, paragraphs 3.117 to
3.123, 3.125 to 3.127,
and 3.135 to 3.137, apply to radiology facilities. This sub-section contains
guidance very specific to
radiology facilities. For more general and comprehensive guidance on
radiation protection of the
public, reference should be made to the IAEA Safety Guide Radiation Protection
of the Public [24].
Non-occupationally exposed workers and visitors
3.266. Non-occupationally exposed workers are those persons who work at the
radiology facility but
not in a role that is directly involved in the use of X rays, for example ward
nurses, imaging staff who
work exclusively with non-radiation imaging modalities (ultrasound or
MRI), clerical staff, and
cleaning personnel. It also includes those persons who work at the wider
medical facility where the
radiology facility is located. These persons are to be afforded the same level
of radiation protection as
any member of the public, as required by the BSS paragraph 3.78.
3.267. Visitors to the radiology facility include those persons who will be
undergoing a radiological
procedure, for the time during the visit when the radiological
procedure is not taking place – for
example, while they are sitting in the waiting room. Similarly for
carers and comforters – any
exposure other than during the radiological procedure in which they
are helping will be public
exposure. Other visitors, including persons delivering goods or
supplies, sales personnel,
accompanying persons and escorts, and other patients in the facility, are also
considered members of
the public.
External exposure
3.268. The primary means for protecting the public (non-occupationally exposed
workers and visitors)
is to ensure that the shielding of the radiology facility (see paras. 3.17 –
3.23) is sufficient so that
public exposure resulting from being in any immediate adjacent area,
including rooms above and
below, accessible by either non-occupationally exposed workers or visitors
would be in compliance
with the public dose limits, and preferably less than any dose constraint that
the regulatory body may
have applied (see Section 2 paras. 2.16 and 2.46.)
3.269. Particular consideration should be given to patients in the
radiology facility, who are not
undergoing a radiological procedure, but are in the vicinity when
mobile radiography is being
performed in their ward or area, or when fixed radiography is being performed
in an open area, such
as in an emergency department. In these cases a combination of
distance, placement of mobile
shielding and careful control of the X ray beam direction should
ensure that appropriate public
radiation protection is being afforded.
Control of access
3.270. Access to areas where radiation is being used should be controlled to
provide for controlling
doses to visitors and non-occupationally exposed workers. According to the BSS,
paragraph 3.128,
access of visitors to radiology or interventional rooms or other
controlled areas while in use is
restricted. Exceptionally, a visitor, for example a health professional from
another medical facility,
may be accompanied by a staff member who knows the protection and
safety measures for the
controlled area. The radiology facility should have written procedures
specifying when such
exceptions can take place and who may do the accompanying. Similarly,
the facility should have
established the rules regarding non-occupationally exposed workers,
especially regarding access to
controlled and supervised areas. Particular attention, in all cases,
should be made with respect to
potentially pregnant women.
3.271. Controlled and supervised areas should be clearly identified to help
prevent inadvertent entry
to areas where diagnostic radiology or image guided interventional procedures
are being performed
[59]. See also para. 3.13. Further control can be afforded by the use of keys
(or passwords) to restrict
access to the control panels of medical radiological equipment to only
authorized persons.
Monitoring and reporting
3.272. The BSS, requirement 32 and paragraph 3.137, sets out the requirements
that should be met by
the radiology facility with respect to monitoring and reporting. In the
radiology facility, procedures
should be in place to ensure that:
(a) The requirements regarding public exposure are satisfied and to assess
such exposure;
(b) Appropriate records of the results of the monitoring programmes are kept.
3.273. The programme for monitoring public exposure arising from diagnostic
radiology and image
guided interventional procedures should include dose assessment in the areas in
and surrounding the
radiology facility which are accessible to the public. This can be
achieved from the shielding
calculations in the planning stage, combined by area monitoring at the initial
operation of the facility
and periodically thereafter. Records of these assessments should be kept for
typically 7-10 years, but
in any case for periods that meet any relevant regulatory requirements.
PREVENTION AND MITIGATION OF ACCIDENTS
Safety assessments
3.274. To comply with the BSS requirements for safety assessment
(paragraphs 3.29 to 3.36), the
registrant or licensee should conduct a safety assessment applied to
all stages of the design and
operation of the radiology facility. The safety assessment report should be
submitted to the regulatory
body if required. Basically, the safety assessment deals with determining
‘what can go wrong’ and
how it can be prevented and, in case it occurs, how it can be mitigated.
Section 2, paras 2.148 – 2.152,
describes general considerations for facilities using ionizing radiation for
medical purposes.
3.275. The safety assessment should be systematic and contain
information on identification of
possible events leading to accidental exposure (see Appendix I for a summary of
typical causes and
contributing factors to accidental exposures in diagnostic radiology and image
guided interventional
procedures). The safety assessment should not only cover these events, but also
aim at anticipating
other events that have not previously been reported. Clearly the
safety assessment should be
documented.
3.276. The safety assessment should be revised when:
(a) New or modified medical radiological equipment or their accessories are
introduced;
(b) Operational changes occur, including workload;
(c) Operational experience or information on accidents or errors
indicates that the safety
assessment should be reviewed.
Accident prevention
3.277. Accident prevention is clearly the best means for avoiding
potential exposure and the BSS,
paragraphs 3.39 to 3.41, set out requirements based on good
engineering practice and defence in
depth, as well as facility-based arrangements, to achieve this. Design
considerations for medical
radiological equipment and the radiology facility are described in paras 3.8 to
3.46.
3.278. The licensee should incorporate:
(a) Defence in depth measures to cope with identified events, and evaluation of
the reliability of the
safety systems (including administrative and operational procedures, equipment
and facility design).
(b) Operational experience and lessons learned from accidents and errors. This
information should be
incorporated into the training, maintenance and quality assurance programmes.
3.279. For diagnostic radiology and image guided interventional
procedures, possible scenarios for
potential exposure include flaws in the design of medical radiological
equipment, failures of medical
radiological equipment while in operation, failures and errors in software that
control or influence the
delivery of the radiation, and human error. In addition in the
radiology facility, another possible
scenario is the rupture of radioactive sources, such as those used in dosimetry
systems calibration.
3.280. Potential public exposure from a radiation generator may occur
if a person (for example a
cleaner) enters an interventional or conventional fluoroscopy room in between
cases and depresses the
exposure footswitch (usually a foot pedal placed on the floor). To prevent
such potential exposure
equipment should be provided with a special X ray interlock in the control
panel to disconnect the
exposure footswitch in between cases, as described in para. 3.36(h).
3.281. Inadvertent entry to the room when a patient is undergoing a
radiological procedure is another
way for potential public exposure. Means for control of entry has
been addressed in paras. 3.268-
3.269.
3.282. For medical exposure, potential exposure, when it does occur, is
manifest as an unintended or
accidental medical exposure. Means for preventing or minimizing unintended and
accidental medical
exposures are described in para. 3.245, and the ensuing investigation
and corrective actions are
described in paras, 3.251 – 3.257.
Mitigation and contingency plans
3.283. Because the radiation source in almost all cases is an X ray
generator and tube, turning the
primary electrical source off immediately stops any radiation being
produced. All relevant staff
should be adequately trained to be able to recognize when medical
radiological equipment is not
functioning correctly or, for example, when a programming error in the software
is suspected. If there
are implications for occupational and/or patient protection and if medical
considerations allow it, the
radiological procedure should be discontinued and the X ray unit turned off.
3.284. Some radiology facilities may have sealed radioactive sources
for dosimetry instrument
calibration purposes or even possibly I-125 sources for
low-intensity X ray imaging scopes
(lixiscopes). Source loss or encapsulation rupture may lead to
contamination. See Section 4 para.
4.291 for guidance on source loss.
4. SPECIFIC RECOMMENDATIONS FOR RADIATION PROTECTION AND
SAFETY IN NUCLEAR MEDICINE
INTRODUCTION
4.1. This chapter covers nuclear medicine, the branch of clinical
medicine in which unsealed
radioactive materials are administered to patients for diagnosis or treatment
of disease, or for clinical
or pre-clinical research. Treatment using sealed sources is covered in Section
5. X ray imaging such
as CT, which may occur in conjunction with a nuclear medicine procedure such as
in hybrid imaging,
is mainly covered in Section 3 with appropriate cross-references.
4.2. All nuclear medicine procedures involve the administration of a
radiopharmaceutical to the
patient. For diagnostic nuclear medicine procedures, trace amounts of
compounds are labelled with
photon or positron emitters, forming what is called a radiopharmaceutical. For
photon emitters, the
distribution of the radiopharmaceutical in the human body can be imaged in
several ways, such as
planar imaging -including whole body imaging- or single photon emission
computerized tomography
(SPECT). In the case of positron emitters, the detection of annihilation
photons allow registering the
three-dimensional spatial distribution of the radiopharmaceutical using
positron emission tomography
(PET). In hybrid imaging, SPECT and PET are combined with an X ray based
modality, such as in
PET-CT and SPECT-CT, and more recently also with magnetic resonance imaging,
such as in PET-
MRI. In addition, probes may be used for the intra-operative localization of
tumours and lymph nodes
or leaks, and for uptake measurements in specific organs, such as the thyroid
or lungs. In therapeutic
nuclear medicine, therapeutic activities of radiopharmaceuticals are
administered that are usually
labelled with beta or beta-gamma emitting radionuclides, more recently
also with alpha emitters;
therapy with Auger electrons is mostly experimental. The nuclear medicine
facility may also perform
in vitro studies, although these are not a primary focus of this Safety Guide.
Some nuclear medicine
facilities may also have an associated cyclotron facility for on-site
radionuclide production. Detailed
guidance for such cyclotron facilities is beyond the scope of this Safety
Guide.
4.3. The generic term “medical radiation facility” is used widely in Section 2
to mean any medical
facility where radiological procedures are performed. In Section 4,
the narrower term “nuclear
medicine facility” is used to cover any medical radiation facility where
nuclear medicine procedures
are performed. A nuclear medicine facility may be a nuclear medicine
department inside a wider
hospital or medical centre, or may be a stand-alone facility providing nuclear
medicine services. In
some cases, the nuclear medicine facility may be a mobile facility.
4.4. The defined term “radiological procedure” is used in the BSS to cover all
imaging and therapeutic
procedures using ionizing radiation. In a nuclear medicine facility
both imaging and therapeutic
radiological procedures may occur, and this needs to be borne in mind when
reading the following
guidance in Section 4. In cases where the guidance is specific to one or other
of imaging or treatment,
additional qualifiers, such as “imaging”, “diagnostic”, “therapy” or
“treatment”, are used.
4.5. Different health professionals can take on the role of the
radiological medical practitioner in
nuclear medicine procedures, depending inter alia on national laws and
regulations. They primarily
include nuclear medicine physicians, but may include other
specialists such as radiologists,
cardiologists or radiation oncologists.
4.6. Section 2 of this Safety Guide provides general guidance on the
framework for radiation
protection and safety in medical uses of radiation, including roles
and responsibilities, education,
training, qualification and competence, and the management system for
protection and safety. This is
relevant to nuclear medicine and reference to Section 2 should be made as
indicated or needed.
SAFETY OF MEDICAL RADIATION FACILITIES AND MEDICAL RADIOLOGICAL
EQUIPMENT
Nuclear medicine facilities
4.7. Provisions for the incorporation of radiation safety features are best
made at the facility design
stage. The siting and layout should take into account workload and
patient flow, both within the
nuclear medicine facility and, in cases where the nuclear medicine facility is
part of a larger hospital
or medical centre, with other departments of the wider facility. The nuclear
medicine facility is likely
to provide services to both inpatients and outpatients, so the location of the
facility should give easy
access to both groups. Consideration should also be given to provide
easy exit routes for patients,
after the examination or treatment has been performed, that minimize movement
through the facility.
4.8. A typical nuclear medicine facility using unsealed sources¹⁶
requires areas for the following:
source storage and preparation (radiopharmacy or radioisotope
laboratory or “hot lab”),
radiopharmaceutical administration to patient, uptake rooms, imaging (in vivo),
sample measurement
(in vitro), radioactive waste storage and predisposal processing
. For those nuclear medicine facilities
performing therapy with radiopharmaceuticals, a dedicated ward for
patients undergoing such
treatments should be considered. In addition, a nuclear medicine
facility requires separate waiting
areas for patients before and after radiopharmaceutical administration,
changing areas and toilets. The
facility will also have areas where radioactive materials are not
expected to be found, such as in
offices, reporting areas and staff rooms, including cloakrooms, showers
and toilets for staff. For
detailed guidance on setting up nuclear medicine facilities, including
PET-CT facilities, reference
should be made to Refs [65, 214 - 219]. The following paragraphs give a general
overview.
4.9. For security purposes, nuclear medicine facilities should be
located in areas where access by
members of the public to the rooms where sources, including
generators, and radiopharmaceutical
dispensing equipment are used and stored can be restricted. Further, the
proximity of source storage
facilities to personnel that may respond in the event of a security breach
should also be considered.
4.10. As a general rule, the design of the nuclear medicine facility should
make provisions for safety
systems or devices associated with the equipment and rooms. This includes
electrical wiring related to
emergency ‘off’ switches, as well as safety interlocks and warning signals.
4.11. A stable power supply should be available for the facility.
Uninterruptible power supply (UPS)
or battery backup systems should be installed to capture the active information
at time of any outage
¹⁶ In a nuclear medicine facility the only sealed sources present are those
used as check sources for calibration of
the activity meters and the flood sources to check the uniformity of the gamma
cameras.
and to power down all software in a controlled manner. Servers
should be programmed to
automatically shut down when the power supply is interrupted.
4.12. The design of the facility may include an air conditioning
system sufficient to maintain the
temperature in the examination room within the parameters defined by the
equipment manufacturers.
For example, temperature control is needed for uptake rooms in a PET facility
to prevent artefacts
(brown fat uptake) occurring if room temperatures are too low.
4.13. Issues to be considered for the design of the nuclear medicine
facility include: optimizing
protection against external radiation and contamination, maintaining low
radiation background levels
to avoid interference with imaging equipment, meeting radiopharmaceutical
requirements (see para.
4.38), and ensuring safety and security of sources (locking and control of
access).
4.14. For external exposure, the three factors relevant to dose reduction
(time, distance and shielding)
should be combined in the design to optimize occupational and public
radiation protection. Larger
rooms are preferable to allow easy access for patients on a bed trolley and to
reduce exposure of the
staff as well as the public, and at the same time allow for
patient positioning and easy movement
during the procedures.
4.15. The design of the nuclear medicine facility should include
provision for secure and shielded
storage for the radioactive sources. Shielding should be appropriate
to the type and energy of the
emitted radiation. Storage may be provided in a room or a separate space
outside the work area or in a
locked cupboard, safe, refrigerator or freezer situated in the work area.
Separate radiopharmaceutical
storage compartments and an area for temporary storage of radioactive waste
should be provided with
appropriate protection.
4.16. Special consideration should be given to avoiding interference with work
in adjoining areas,
such as imaging or counting procedures, or where fogging of films stored nearby
can occur.
4.17. Signs and warning lights should be available at the entrances of
controlled and supervised areas
to prevent inadvertent entry (see also paras. 4.266 - 4.267 on control of
public access). For controlled
areas, the BSS, paragraph 3.90, requires the use of the symbol of the
International Organization for
Standardization (ISO) [59]. Signs and warning lights should be particularly
available at the entrances
of source preparation and storage areas, hybrid imaging rooms, and for
rooms for hospitalized patients
undergoing radiopharmaceutical therapy (see, also below on treatment wards,
paras. 4.28 – 4.30). The
signs should be clear and easily understandable. Warning lights, such
as illuminated and flashing
signs, should be activated when CT is being used in hybrid imaging.
4.18. Bathrooms designated for use by nuclear medicine patients should be
finished in materials that
are easily decontaminated. Nuclear medicine facility staff should not use the
patient bathrooms, as it
is likely that the floors, toilet seats and sink tap handles will be
contaminated.
Mobile facilities
4.19. In some countries, PET-CT scanners are mounted on a truck and this
mobile unit provides a
service to specific regions of that country. These mobile units should meet the
same requirements of
the BSS as fixed facilities and the relevant guidance in this Safety Guide is
applicable.
Areas where unsealed radioactive materials are handled
4.20. Radiopharmacies or laboratories where unsealed radioactive materials are
handled, such as the
source preparation area, should have:
(a) Means to prevent access by unauthorized persons;
(b) Adequate storage space for equipment used in the given room or area to be
available at all
times, to minimize the potential for spreading contamination to other areas;
(c) A contained workstation for easy decontamination;
(d) Shielded storage for radioactive sources;
(e) Shielded temporary storage for both solid and liquid radioactive waste,
and places designated
for the authorized discharge of liquid radioactive waste;
(f) Shielding to protect workers where significant external exposure may
occur;
(g) A wash-up area for contaminated articles, such as glassware;
(h) An entry area where protective clothing can be stored, put on
and taken off, and which is
provided with a hand wash-up sink and a contamination monitor;
(i) Taps and soap dispenser that are operable without direct hand contact and
disposable towels
or a hot air dryer;
(j) An emergency eyewash, installed near the hand washing sink; and
(k) An emergency shower for decontamination of persons.
Detailed guidance is given in Refs [66, 214 – 219].
4.21. Radiopharmacies, laboratories and other work areas for
manipulation of unsealed radioactive
materials should be provided with equipment kept specifically for this purpose,
which should include:
(a) Tools for maximizing the distance from the source, for example tongs and
forceps;
(b) Syringe shields;
(c) Containers for radioactive materials, with shielding as close as possible
to the source;
(d) Double walled containers (the outer being unbreakable) for liquid samples;
(e) Drip trays for minimizing the spread of contamination in the case of
spillage;
(f) Disposable tip automatic pipettes (alternatively, hypodermic syringes to
replace pipettes);
(g) Lead walls or bricks for shielding;
(h) Lead barriers with lead glass windows;
(i) Barriers incorporating a low atomic number material (i.e. acrylic) for
work with beta emitters;
(j) Radiation and contamination monitoring equipment (surface and air);
(k) Carrying containers, wheeled if necessary, for moving radioactive
materials from place to
place;
(l) Equipment to deal with spills (decontamination kits).
4.22. Drainpipes from sinks in the radiopharmacy or laboratory should go as
directly as possible to the
main building sewer and should not connect with other drains within the
building, unless those other
drains also carry radioactive material. This is to minimize the possibility of
a ‘backup’ contaminating
other non-controlled, areas. The final plans of the drainage system, which are
supplied to maintenance
personnel, should clearly identify the drains from radiopharmacies and
laboratories. Pipelines through
which radioactive materials flow should be marked to ensure
that monitoring precedes any
maintenance.
4.23. Some countries require that drainpipes from a nuclear medicine
facility and especially from
radionuclide therapy wards terminate in a delay tank. Requirements on this
issue differ very much
among countries but each nuclear medicine facility should comply with their
country’s regulations.
4.24. The floors of areas with the potential for contamination should be
finished in an impermeable
material which is washable and resistant to chemical change, curved to the
walls, with all joints sealed
and glued to the floor. The walls should be finished in a smooth and washable
surface, for example
painted with washable, non-porous paint. The surfaces of the room
where unsealed radioactive
materials are used or stored, such as benches, tables, seats, and door and
drawer handles, should be
smooth and non-absorbent, so that they can be cleaned and
decontaminated easily. Supplies (for
example, gas, electricity and vacuum equipment) should not be mounted on bench
tops, but on walls
or stands.
4.25. The floor and benches, including worktops, should be strong enough to
support the weight of
any necessary shielding materials or of radionuclide generators. The need for
lifting equipment for
radionuclide generators should be assessed.
4.26. Radiopharmacies or laboratories in which radioactive aerosols or
gases may be produced or
handled should have an appropriate ventilation system that includes a fume
hood, laminar air flow
cabinet or glove box. The fume hood should be constructed of material that is
smooth, impervious,
washable and resistant to chemicals. The working surface should have a slightly
raised lip to contain
any spills. The ventilation system should be designed such that the
radiopharmacy or laboratory is at
negative pressure relative to surrounding areas.
4.27. The airflow should be from areas of minimal likelihood of
airborne contamination to areas
where such contamination is likely. All air from the radiopharmacy or
laboratory should be vented
through a fume hood and should not be recirculated either directly,
in combination with incoming
fresh air in a mixing system, or indirectly, as a result of proximity of the
exhaust to a fresh air intake.
For reasons of asepsis, some radiopharmacies may require a positive rather than
a negative pressure.
In this case, the pressure gradient can be obtained by placing other
workstations requiring negative
pressure next to the radiopharmacy workstation.
Treatment rooms and wards
4.28. Floors and other surfaces of rooms designated for patients
undergoing radiopharmaceutical
therapy should be covered with smooth, continuous and non-absorbent materials
that can be easily
cleaned and decontaminated. Shielding should be designed using appropriate dose
constraints for staff
and public. Secure areas should be provided with bins for the temporary storage
of linen and waste
contaminated with radioactive materials. Storage areas should be clearly
marked, using the radiation
sign.
4.29. Rooms designated for patients undergoing radiopharmaceutical
therapy should have separate
toilet and washing facilities. A sign requesting patients to flush the toilet
at least twice and wash their
hands should be displayed to ensure adequate dilution of excreted radioactive
materials and minimize
contamination. The facilities should include a wash-up sink as a
normal hygiene measure. For
guidance on bathrooms and their use see para. 4.18.
4.30. The design of safe and comfortable accommodation for carers and
comforters (see also paras.
4.232 – 4.236) should beconsidered for nuclear medicine facilities with
radiopharmaceutical therapy
patients.
Shielding calculation considerations
4.31. The shielding should be designed to meet the requirements for
optimization of protection and
taking into consideration the classification of the areas within the facility,
the type of work to be done
and the radionuclides (and their activity) intended to be used.
Shielding should consider both
structural and ancillary protective barriers at the design stage (see para
2.73). It is convenient to shield
the source, where possible, rather than the room or the person. The need for
wall shielding should be
assessed, for example in the design of therapy and of PET-CT facilities to
reduce occupational and
public exposure to acceptable levels. Wall shielding may be needed in the
design of rooms housing
sensitive instruments (to keep a low background), such as well
counters, probes and imaging
equipment (gamma cameras and PET scanners). In designing such wall
shielding, consideration
should be given to the height of the wall to ensure scattered radiation, such
as from a CT scanner,
does not pass over the wall into the area being protected.
4.32. Methodologies and data for shielding calculations for nuclear medicine
facilities are given in
Refs [58, 64, 215], and reference should be made to Section 3, paras 3.17 –
3.21) for shielding with
respect to X ray imaging systems (e.g. CT) used as part of hybrid imaging
equipment. The nominal
design dose in occupied areas is derived by the process of constrained
optimization, i.e. selecting a
source related dose constraint, with the condition that the individual doses
from all relevant sources
are well below the dose limits for the persons occupying the area
to be shielded. Nominal design
doses are levels of air kerma used in the design calculations and
evaluation of barriers for the
protection of individuals, at a reference point beyond the barrier.
Specifications for shielding are
calculated on the basis of the attenuation they should provide to
satisfy the nominal design doses.
Potential practice changes and workload increases should be considered.
4.33. Care should be taken to avoid multiplication of conservative
assumptions which can lead to
unrealistic overestimates of required shielding. Typical conservative
assumptions are: attenuation by
the patient is usually not considered; decay of short-lived radionuclides, such
as ¹⁸F, is not considered;
workload, use and occupancy factors are overestimated; and the persons
to be protected are
considered as remaining permanently in the most exposed place of the
adjacent roomTherefore, a
balanced decision should be achieved and accumulation of overly conservative
measures that may go
beyond optimization should be avoided.
4.34. Specification of shielding, including calculations, should be
performed by a RPO or medical
physicist. In some countries there may be a requirement for shielding
plans to be submitted to the
regulatory body for review or approval prior to any construction (see also
Section 2 para 2.66).
4.35. The adequacy of the shielding should be verified, preferably during
construction, and certainly
before the facility, room or area comes into clinical use. Clearly requirements
of the regulatory body
must be met (Section 2 para. 2.73).
Display and interpretation (reading) rooms design
4.36. To facilitate the interpretation by the radiological medical
practitioner, images should be
displayed in rooms specifically designed for these purposes. A low
level of ambient light in the
viewing room should be achieved. See also Section 3, paras. 3.40 - 3.41 on
image display devices and
view boxes.
4.37. Viewing rooms with workstations for viewing digital images should be
ergonomically designed
to facilitate image processing and manipulation so that reporting can be
performed accurately. The
viewing monitors of the workstations should meet applicable standards (see
Section 3, para. 3.40).
Radiopharmaceuticals
4.38. Radiopharmaceuticals should be manufactured according to good
manufacturing practice
following relevant international standards [217, 218, 222 - 224] for:
(a) Radionuclide purity;
(b) Specific activity;
(c) Radiochemical purity;
(d) Chemical purity;
(e) Pharmaceutical aspects: toxicity, sterility and pyrogenicity.
Medical radiological equipment, software and ancillary equipment
4.39. This sub-section considers medical radiological equipment,
including its software, used in a
nuclear medicine facility. Such equipment falls into two categories –
those that detect ionizing
radiation from the unsealed or sealed sources and those that generate ionizing
radiation. The former
includes probes, gamma cameras, SPECT scanners, and PET scanners, since these
have an influence
on the activity to be administered to the patient in order to obtain the
desired diagnosis. The latter
includes CT, typically as part of a hybrid imaging system such as a PET-CT or
SPECT-CT scanner.
Some hybrid equipment utilizes MRI, and although these are not generating
ionizing radiation and are
outside the scope of this Safety Guide, their performance can influence the
efficacy of the nuclear
medicine procedure and hence such equipment should meet relevant IEC standards
or equivalent.
4.40. The requirements for medical radiological equipment and its
software are given in the BSS
paragraphs 3.49 and 3.161. The International Electrotechnical Commission
(IEC), through its
Technical Committee 62 on Electrical equipment in medical practice and in
particular Sub-committee
62B on Diagnostic imaging equipment and Sub-committee 62C on Equipment
for radiotherapy,
nuclear medicine and radiation dosimetry, has published international standards
applicable to medical
radiological equipment. Current IEC standards relevant to nuclear
medicine include the following
Refs [225 – 234]. For those relevant to the X ray based component of hybrid
imaging, see Section 3
para. 3.27. It is recommended that the IEC website is visited to
view the most up-to-date list of
standards: http://www.iec.ch. The International Organization for
Standardization (ISO), through its
Technical Committee 85 on Nuclear energy, nuclear technologies, and
radiological protection and in
particular Sub-committee 2 on Radiological protection, may
publish international standards
applicable to medical radiological equipment. It is recommended that the
ISO website is visited to
view the most up-to-date list of standards: http://www.iso.org .
4.41. As the licensees take responsibility for the radiation safety of medical
radiological equipment
they use, they should impose purchasing specifications that include
conditions to meet relevant
international standards of the IEC and ISO and/or equivalent national
standards. In some countries
there may be a medical devices agency or similar organization that gives type
approval to particular
makes and models of medical radiological equipment.
4.42. Some nuclear medicine facilities may operate a cyclotron for on-site
radionuclide production.
As the cyclotrons are not directly involved in the exposure of the patient,
they should not comply with
the BSS requirements for medical radiological equipment. Nevertheless, they
should comply with the
more general requirements of the BSS for radiation generators (requirement 17
and paragraphs 3.49 -
3.60), as well as additional regulatory requirements, in a given Member State,
for the preparation and
control of radiopharmaceuticals.
4.43. Displays, gauges and instructions on operating consoles of medical
radiological equipment, and
accompanying instruction and safety manuals, may be used by staff who may not
understand, or who
may have a poor understanding of, the manufacturer’s original
language. In such cases, the
accompanying documents should comply with IEC and ISO standards and should be
translated into
the local language. The software should be designed so that it can be easily
converted into the local
language resulting in displays, symbols and instructions that will be
understood by the staff. The
translations will require a quality assurance process to ensure proper
understanding and avoid
operating errors. The same applies to maintenance and service manuals
and instructions for
maintenance and service engineers and technicians, where these persons
do not have an adequate
understanding of the original language. See also Section 2 para. 2.103.
Design features for medical radiological equipment
4.44. The performance of probes, gamma cameras, SPECT systems and PET scanners
determine the
efficacy of the diagnostic radiological procedures and hence can influence the
amount of radioactivity
needed to be administered to the patient, even whether the procedure
is diagnostically successful.
Many design features contribute to their performance and should be considered
when purchasing such
equipment, as indicated briefly in the next paragraphs and detailed in Refs
[192, 210, 211, 219, 225 –
241].
4.45. Design features for probes used for uptake measurements that
should be considered include
energy response, energy resolution, sensitivity, counting precision,
linearity of count rate response
and geometrical dependence.
4.46. Design features for probes used intra-operatively that should be
considered include energy
resolution, background count rate, sensitivity in scatter, sensitivity
to scatter, shielding (side and
back), counting precision, linearity of count rate response (with scatter), and
count rate recorded by
visual display and by an audible sound, the intensity of which is proportional
to the count rate.
4.47. Design features for gamma and SPECT cameras as well as their
accessories that should be
considered include:
(a) Detector performance;
(b) Detector head and gantry design;
(c) Detector head motion;
(d) Automatic patient–detector distance sensing;
(e) Collimators;
(f) Pulse height analysis;
(g) Imaging table and attachments;
(h) Data acquisition, including:
a. General acquisition features;
b. Static acquisition;
c. Dynamic acquisition;
d. List mode acquisition;
e. Gated cardiac acquisition;
f. Whole body imaging;
g. Tomography.
(i) Data processing system, including data display, image manipulation,
region of interest (ROI)
generation and display, curve generation, display and arithmetic, processing of
SPECT data,
quality control software and test data;
(j) Accessories that include physiological triggering, anatomical
landmarking, and phantoms.
4.48. Design features for PET scanners that should be considered include:
(a) Detector design and performance;
(b) Spatial resolution;
(c) Scatter fraction, count losses and random events;
(d) Sensitivity;
(e) Accuracy;
(f) Time of flight capability;
(g) Data acquisition, including 2D and 3D whole body imaging, and
cardiac and respiratory
gating;
(h) Data processing system, including image reconstruction algorithms and
image manipulation.
4.49. Guidance on medical radiological equipment using X rays, used for imaging
as part of nuclear
medicine, is given in Section 3, paras. 3.26 – 3.39.
4.50. All digital medical radiological equipment should have connectivity to
RIS/PACS.
Ancillary equipment
4.51. All equipment used for digital image display should meet appropriate
international and/or local
standards, for example meeting the performance specifications of the AAPM Task
Group 18 [115].
Workstations and image processing and display software should be specifically
designed for nuclear
medicine, ensuring DICOM conformance, and network interconnectivity. Guidance
on DICOM image
and data management for nuclear medicine is given in Ref [242]. See paras. 4.36
- 4.37 for guidance
on reporting rooms.
4.52. The nuclear medicine facility should have equipment, instruments
and test objects for
measurements, dosimetry and quality control which may include liquid
scintillation counters, well
counters, activity meters, dose calibrators, check sources, flood
sources, phantoms, geometry and
mechanical test tools. Where applicable, such instrumentation should adhere to
relevant IEC standards
or national equivalents. Further guidance on appropriate equipment,
instruments and test objects is
given in Refs [227, 240, 243].
4.53. The nuclear medicine facility should be equipped with radiation
monitoring, including survey
meters and portable contamination monitors.
4.54. Radiopharmaceutical dispensing equipment should adhere to relevant IEC
standards or national
equivalents.
Security of sources
4.55. The objective of source security is to ensure continuity in the control
and accountability of each
source at all times in order to meet the requirement in BBS paragraph 3.53. In
the nuclear medicine
facility the sources include unsealed radiopharmaceuticals as well as
radionuclide generators,
radiopharmaceutical dispensing equipment, and sealed sources used for
calibration or QC tests.
Requirements for the identification and documentation of unsealed
radioactive uses are given by ISO
[228]. Situations that are particularly critical with respect to source
security in the nuclear medicine
facility include receipt of radiopharmaceuticals, storage of sources, movement
of sources within the
facility, and storage of radioactive waste. The nuclear medicine
facility licensee should develop
procedures to ensure the safe receipt and movement of radioactive sources
within the institution and
establish controls to prevent theft, loss, unauthorized withdrawal of
radioactive materials or entrance
of unauthorized personnel to the controlled areas. An inventory of sources
should be maintained, with
procedures in place to check and confirm the sources are in their assigned
locations and are secure.
Procedures to stimulate proactive behaviour should be in writing, for
example, to trigger a search
when a delivery of radiopharmaceuticals is not received at the expected
time.
Maintenance
4.56. The BSS (BSS, paragraphs 3.15(i) and 3.41) gives requirements for
maintenance to ensure that
sources meet their design requirements for protection and safety
throughout their lifetime and to
prevent accidents as far as reasonably practicable. The licensee
should ensure that adequate
maintenance (preventive and corrective) is performed as necessary to ensure
that medical radiological
equipment used in the nuclear medicine facility retains, or improves
through appropriate hardware
and/or software upgrades, its design specification for image quality, radiation
protection and safety for
its useful life. The licensee should, therefore, establish the necessary
arrangements and coordination
with the manufacturer’s representative or installer before initial operation
and on an on-going basis.
4.57. All maintenance procedures should be included in the quality
assurance programme at the
frequency recommended by the manufacturer of the equipment and
relevant professional bodies.
Servicing should include a report describing the equipment fault, the work done
and the parts replaced
and adjustments made, which should be filed as part of the quality assurance
programme. A record of
maintenance carried out should be kept for each item of equipment: this should
include information
on any defects found by users (a fault log), remedial actions taken
(both interim and subsequent
repairs) and the results of testing before equipment is reintroduced to
clinical use.
4.58. In line with the guidance in Section 2, para. 2.111, after any
modifications or maintenance, the
person responsible for maintenance should immediately inform the licensee of
the nuclear medicine
facility before it is returned to clinical use. The person responsible for the
use of the equipment, in
conjunction with the medical physicist, the medical radiation
technologist and other appropriate
professionals, should decide whether quality control tests are needed
with regard to radiation
protection, including image quality, and whether changes to protocols are
needed, especially in the
amount of administered activity.
4.59. The electrical and mechanical safety aspects of the medical
radiological equipment are an
important part of the maintenance programme, and can have direct or
indirect effects on radiation
safety. Authorized persons who understand the specifications of the medical
radiological equipment
should perform this work. See also Section 2 paras. 2.110 – 2.112.
Electrical and mechanical
maintenance should be included in the QA programme at a
frequency recommended and
preferentially performed by the manufacturer of the medical radiological
equipment. Servicing should
include a written report describing the findings. These reports should be
archived as part of the QA
programme.
OCCUPATIONAL RADIATION PROTECTION
Introduction
4.60. In nuclear medicine, described in paras. 4.1 to 4.5,
occupationally exposed individuals are
usually the medical radiation technologists, the radiological medical
practitioners (including, for
example, nuclear medicine physicians), the radiopharmacists and the medical
physicists. Other health
professionals such as nurses, particularly in nuclear medicine
facilities providing therapy services,
may also be considered occupationally exposed.
4.61. Additional occupationally exposed personnel may include
biomedical, clinical or service
engineers and some contractors, depending on their role.
4.62. Other nuclear medicine facility workers such as administrative
personnel, patient porters,
orderlies, assistants, cleaners and other service support personnel, for whom
radiation sources are not
directly related to their work, require the same level of protection as members
of the public, as stated
in the BSS paragraph 3.78.
4.63. This sub-section contains guidance very specific to nuclear
medicine. For more general and
comprehensive guidance on occupational radiation protection, including
guidance on radiation
protection programmes, assessment of occupational exposure and providers
of dosimetry services,
applicable to all areas of radiation use (including non-medical uses),
reference should be made to the
IAEA Safety Guide Occupational Radiation Protection [23].
Arrangements under the radiation protection programme
Classification of areas
4.64. Various areas and rooms in a nuclear medicine facility should
be classified as controlled or
supervised areas, in line with the requirements given in BSS paragraphs 3.88
and 3.91, respectively.
Once designated, these areas should meet the requirements detailed in
the BSS paragraphs 3.89 to
3.90 (for controlled areas) and 3.92 (for supervised areas),
including requirements for area
delineation, signage, protection and safety measures, control of
access, provision of personal
protective equipment, provision of individual and area monitoring,
provision of equipment for
monitoring for contamination, and provision of personal decontamination
facilities. All other rooms
and areas, not so-designated, are considered as “public domain” and levels of
radiation in these areas
should be low enough to ensure compliance with the dose limits for public
exposure.
4.65. In a nuclear medicine facility, rooms for radiopharmaceutical preparation
(i.e. radiopharmacies
or hot labs), injection of the radiopharmaceuticals and for storage and decay
of radiopharmaceuticals
meet the criteria for controlled areas and should be so designated. Imaging
rooms, particularly those
housing radiopharmaceutical dispensing equipment (i.e. PET radiopharmaceutical
and radioactive gas
and aerosol dispenser devices), should also be considered controlled
areas. Rooms with patients
undergoing radiopharmaceutical therapy should be controlled areas. Rooms
housing hybrid machines
that have an X ray component (SPECT-CT, PET-CT) should be considered
controlled areas when the
X ray unit is energized. A warning light at the entry to the room should
indicate the machine is on to
prevent unintended passage.
4.66. Supervised areas may include examination rooms (with probes, gamma
cameras and SPECT and
PET systems) and waiting rooms dedicated to patients who
have been injected with
radiopharmaceuticals (e.g. uptake rooms in a PET facility).
4.67. The area around the control panel of hybrid imaging equipment (e.g.
PET-CT and SPECT-CT)
should be classified as either a controlled or a supervised area, even though
the radiation levels may
be very low due to the shielding design. In either case, this should ensure
restricted access and hence,
among other things, avoid distraction of the operator which may lead
to accidental or unintended
medical exposure of patients. See also Section 3, para. 3.55.
4.68. In order to avoid uncertainties about the extent of controlled
and supervised areas, the
boundaries should, when possible, be walls and doors or other physical
barriers, clearly marked or
identified with ‘radiation area’ signs.
Local rules and procedures
4.69. The BSS, in paragraph 3.93, establishes a hierarchy of preventive
measures for protection and
safety with engineered controls, including structured and ancillary
shielding, specific physical
barriers, signs and interlocks, being supported by administrative
controls and personal protective
equipment. To this end, and as required in the BSS paragraph 3.94, written
local rules and procedures
should be established in a nuclear medicine facility. Their purpose is to
ensure protection and safety
for workers and other persons. These local rules and procedures should include
measures to minimize
occupational radiation exposure during both normal work and unusual
events. The local rules and
procedures should also cover the wearing, handling and storing of personal
dosimeters, and specify
investigation levels and ensuing follow-up actions (see paras. 4.117 – 4.131).
4.70. Since all personnel involved in using radiation in nuclear medicine
should know and follow the
local rules and procedures, the development and review of these local rules
and procedures should
include representatives of all health professionals involved in nuclear
medicine.
4.71. Equipment (hardware and software) should be operated in a manner
that ensures satisfactory
performance at all times with respect to both the tasks to be accomplished and
radiation safety. The
manufacturer’s operating manual is an important resource in this respect, but
additional procedures
should be also considered. The final documented set of operational
procedures should be approved by
the nuclear medicine facility’s licensee, and incorporated into the
facility’s quality management
system (see Section 2, paras. 2.136 – 2.147).
4.72. Nuclear medicine staff should understand the documented
procedures for their work with
radiopharmaceuticals and for the operation of the equipment with which they are
working, including
the safety features, and should be trained, with periodic refresher training,
in what to do when things
go wrong. Additional training should occur when new radiopharmaceuticals or
devices are brought
into nuclear medicine practice.
4.73. Many local rules and procedures address aspects of some, or
all, of occupational radiation
protection, patient radiation protection and public radiation protection,
either directly or indirectly, as
well as ensuring a successful diagnostic examination or application of the
treatment. This is the case
with the following paragraphs (4.74 to 4.108) – while placed in this section on
occupational radiation
protection because they are to be followed by workers, the local rules and
procedures often also have
significance for patient and/or public radiation protection. These following
paragraphs (4.74 to 4.108)
give recommendations that should be incorporated into the nuclear medicine
facility’s local rules and
procedures. In addition, reference should also be made to para. 4.55 on the
security of sources.
4.74. Work procedures should be formulated so as to minimize exposure from
external radiation and
contamination, to prevent spillage from occurring and, in the event of
spillage, to minimize the spread
of contamination (surface and airborne). For instance, all manipulation
for dispensing radioactive
materials should be carried out over a drip tray. Work with unsealed sources
should be restricted to a
minimum number of specifically designated areas.
4.75. No food or drink, cosmetic or smoking materials, crockery or cutlery
should be brought into an
area where unsealed radioactive materials are used. An exception to this is
when food is radiolabelled
for patient studies. Food or drink should not be stored in a refrigerator used
for unsealed radioactive
materials. Personal cell phones and handkerchiefs should never be used in these
areas; an adequate
supply of paper tissues should be provided. Before a person enters an area
where radioactive material
is handled, any cut or break in the skin should be covered by a waterproof
dressing.
4.76. In areas classified as controlled, protective clothing should be worn as
determined by prior risk
assessment of potential contamination. Protective clothing is unlikely
to be necessary for persons
accompanying patients into gamma camera rooms. On leaving the controlled area,
protective clothing
that is contaminated should be placed in an appropriate container. The method
of removing gloves
should be based on the surgical technique, in order to avoid transferring
activity to the hands.
4.77. Staff leaving a controlled area, classified as such on account of the
potential for contamination,
should, after removal of their protective clothing, wash their hands
and then monitor their hands,
clothing and body. Liquid soap should be provided unless aseptic considerations
require an alternative
cleaner. Non-abrasive nail brushes should only be used if contamination
persists after simple washing.
See also paras. 4.104 – 4.108 below on decontamination of persons.
4.78. Pipettes should never be operated by mouth. Syringes used for
handling radioactive liquids
should be appropriately shielded wherever practicable. The distance
between the fingers and the
radioactive liquid should be as large as can be achieved.
4.79. The work area should be kept tidy and free from articles not required for
work. A monitoring
and cleaning programme should be established to ensure minimal spread of
contamination. Cleaning
and decontamination can be simplified by covering benches and drip trays with
disposable material
such as plastic backed absorbent paper.
4.80. All containers used for radioactive material should be
clearly labelled, indicating the
radionuclide, chemical form and activity at a given date and time. Batch number
and expiry date and
time should be added as appropriate. All such containers are to be adequately
sealed and shielded at
all times. Except for very small activities, containers are not to be handled
directly and, if possible,
tongs or forceps for vials and syringe shields should be used. Records of
stocks, administrations and
pre-disposal waste management should be kept.
4.81. The amount of shielding material required can be minimized by
positioning it close to the
source. A variety of materials can be used for this purpose, such as lead,
tungsten, lead glass and lead
composite. Shielding incorporating acrylic is more suitable for beta emitters,
as it lowers the amount
of bremsstrahlung produced. Lead should be coated to provide a cleanable
surface.
4.82. The attenuation by lead aprons at the typical gamma energies
used in nuclear medicine is
modest, and even less for non-lead based protective aprons. Automatic
dispensers and injectors, and
mobile shields are preferred alternatives.
4.83. The following protective approaches can reduce occupational exposure
significantly:
(a) For preparation and dispensing of radiopharmaceuticals, working behind a
lead glass bench
shield, using shielded vials and syringes, and using disposable gloves.
(b) During examinations, when the distance to the patient is short, using a
movable transparent
shield.
4.84. All radioactive sources should be returned to safe storage immediately
when no longer required.
4.85. All operations involving radioactive gases or aerosols should be carried
out in a fume hood or
similar ventilated device to prevent airborne contamination. Exhaust
vents should be situated well
away from air intakes. The administration of aerosols to patients,
such as for ventilation studies,
should be performed using a mouthpiece or mask for the patient.
4.86. Glassware and implements for use in the radiopharmacy should be
appropriately marked and
under no circumstances removed from that area.
4.87. Packaging and containers for radioactive material should be
checked for contamination on
opening.
4.88. Containers, lead pots, etc., that no longer contain radioactive material
and require to be managed
as non-radioactive waste should have any radiation warning labels
removed or obliterated before
removing them from regulatory control.4.89. For X ray based imaging
(e.g. CT) in the nuclear
medicine facility, reference should be made to the guidance, where appropriate,
in Section 3, paras.
3.61 - 3.73.
4.90. Local rules for pregnant workers and persons under 18 should
reflect the guidance given in
paras. 4.144 – 4.148 and 4.149 respectively.
Specific local rules and procedures for radiopharmaceutical therapy
4.91. Administration is normally by the oral route, intravenous
injection (systemic), intra-arterial
injection (locoregional) or instillation into closed joints
(intra-articular/radiosynoviorthesis) or body
cavities (intracavitary).
(a) Shielded syringes should be utilized during the intravenous or
intra-arterial administration of
radiopharmaceuticals as necessary to ensure extremity doses are
maintained below
occupational dose constraints. Absorbent materials or pads should be placed
underneath an
injection or infusion site. The facility RPO should be consulted to determine
the necessity of
other protective equipment (e.g., shoe covers, step-of-pads,
etc.) for particular
radiopharmaceutical therapies.
(b) For intravenous or intra-arterial administrations by bolus injections,
when dose-rates warrant,
the syringe should be placed within a syringe shield (plastic for beta-emitting
radionuclides to
minimize bremsstrahlung, high atomic number materials for photon-emitting
radionuclides)
with a transparent window to allow for visualization of the
material in the syringe. For
intravenous administrations by slower drip or infusions, the activity
container should be
placed within a suitable shield. For high-energy photons, a
significant thickness of lead or
other high atomic number material may need to be used. In addition,
consideration should be
given for shielding pumps and lines.
(c) For oral administrations of therapeutic radiopharmaceuticals, the
radioactive material should
be placed in a shielded, spill-proof container. Care should be taken to
minimize the chance for
splashing liquid, or for dropping capsules. Appropriate long-handled tools
should be utilized
when handling unshielded radioactive materials.
4.92. Patients hospitalized for therapy with radiopharmaceuticals should
be attended by staff
(physicians, nurses, aids, cleaning staff) trained in radiation protection.
This also includes night staff.
The training should cover radiation protection and specific local
rules, in particular, for situations
where there is a risk of significant contamination from urine, faeces or
vomiting. Ward nurses should
be informed when a patient may pose a radioactive hazard.
4.93. Local rules should be established concerning the type of
nursing that can be performed
according to the level of ambient dose equivalent. In general,
non-essential nursing should be
postponed to take advantage of the reduction of activity by decay
and excretion. Blood and urine
analyses should be performed prior to therapy Procedures should be established
for the handling of
any potentially contaminated item (bed linen, clothing, towels, crockery,
etc.).
4.94. As described above in para. 4.65, rooms occupied by patients treated with
radiopharmaceuticals
should be controlled areas, and both a radiation sign and a warning sign
should be posted. Access
should be restricted and a list of relevant contacts (such as nuclear medicine
and on-call physicians,
medical radiation technologists, radiation protection officer) should be
provided. Protective clothing,
such as laboratory coats, gloves and shoe covers, should be available at the
entrance to the room. The
nursing staff should be familiar with the implications of the procedures for
controlled areas, the time
and date of administration, and any relevant instructions to carers and
comforters.
4.95. Values of ambient dose equivalent at suitable distances should be
determined by the RPO or
medical physicist. This information will assist in deriving appropriate
arrangements for entry by staff
and by carers and comforters. These arrangements should be made in writing in
the local rules.
4.96. On leaving the work area, staff should remove any protective clothing and
wash their hands.
4.97. Patients treated with radiopharmaceuticals should use designated toilets.
Measures to minimize
contamination should be implemented (such as laying plastic backed
absorbent paper on the floor
around the toilet bowl, sitting down when using the toilet and instructions to
flush the toilet at least
twice in the absence of delay tanks).
4.98. Particular attention and measures to limit spread of contamination are
required in the case of
incontinent patients and in the risk of vomiting after oral administration of
the radiopharmaceutical.
Plastic backed absorbent paper on the bed and floor can help reduce
spread of contamination.
Contaminated bedding and clothing should be changed promptly and retained for
monitoring.
4.99. Crockery and cutlery may become contaminated. Local rules should
specify washing up and
segregation procedures, and the management of single use dishes, cutlery and
food waste.
4.100. Nursing care items should be covered when possible to prevent
contamination. For example,
the stethoscope can be covered with a glove. The blood pressure cuff
and the thermometer should
remain in the room until the release of patient, and then checked
for contamination before being
returning to regular use again.
4.101. The staff should be informed about the treatment procedure and any
relevant medical history.
If the medical condition of a patient deteriorates such that
intensive care becomes necessary, the
advice of the RPO should be sought immediately. While urgent medical care is a
priority and should
not be delayed, it may be necessary to restrict the maximum time that
individual health professionals
spend with a patient.
Specific local rules and procedures in PET facilities
4.102. PET imaging personnel can receive relatively large annual
occupational radiation doses
compared to their counterparts in general nuclear medicine. The main
contribution to the occupational
dose for personnel comes from patient handling. PET radiopharmacists,
at facilities performing
radiopharmaceutical synthesis and unit dose preparations, can receive
significant hand and body
doses, even where heavily shielded ‘hot cells’ are available to
moderate dose. For these reasons, local
rules and procedures for PET facilities should emphasize the means described
above for minimizing
the dose to personnel when handling radiopharmaceuticals and the injected
patient.
4.103. Radiopharmaceuticals should be stored and transported in lead
or tungsten containers
specifically designed to limit external radiation levels from radionuclides
used for PET. An additional
plastic shield inside a lead or tungsten syringe shield will absorb
positrons before striking the
tungsten, minimizing unwanted production of bremsstrahlung radiation. The use
of tongs to handle
unshielded radiopharmaceutical vials markedly reduces hand doses. Automatic
systems are available
which allow safe and quick radiopharmaceuticals dispensing into syringes, thus
minimizing operators’
actions.
Decontamination of persons
4.104. Hands should be washed on completing work with unsealed
radioactive materials and on
leaving an area that is controlled, because of potential
contamination. If detectable contamination
remains on the hands after simple washing, use of a surfactant or
chelating agent specific to the
chemical form of the contaminant agent may be more successful. A
decontamination kit and
procedures for its use shoul be available on site.
4.105. The RPO should be consulted when contamination of parts of the body
other than the hands is
suspected, or when the procedures for decontamination of the hands
are ineffective. Special care
should be taken in the decontamination of the face to restrict entry of
radioactive material into the
eyes, nose or mouth.
4.106. If the skin is broken or a wound is sustained under
conditions where there is a risk of
radioactive contamination, the injury should be irrigated with water
as soon as appropriate, taking
care not to wash contamination into the wound. As soon as the first aid
measures have been taken, the
person should seek further treatment, including decontamination if
necessary The RPO should be
consulted as needed.
4.107. Contaminated clothing should be removed as soon as practicable,
taking care not to spread
contamination.
4.108. All staff working with unsealed sources should be trained in the
procedures for dealing with
accidents, spills or contaminated persons, with refresher training at
appropriate intervals. This
includes instructions on appropriate use of showering and eye washing.
Personal protective equipment and in-room protective devices
4.109. The BSS, paragraphs 3.93 and 3.95, requires that personal protective
equipment and in-room
protective equipment are available and used when structural shielding
and administrative controls
alone cannot afford the necessary level of occupational radiation
protection. The need for this
protective equipment should be established by the nuclear medicine
facility’s RPO or medical
physicist.
4.110. In a nuclear medicine facility, protective equipment includes the
following:
(a) Shields for bench tops, vials, syringes, activity meters and for
the preparation of the
radiopharmaceuticals (i.e. L-blocks and side blocks) of a material and
thickness appropriate to
the type and energy of the radiation. Particular considerations for the choice
of shield include:
a. Alpha emitters such as ²¹¹At may need to be shielded by high
atomic number
materials because of their characteristic X rays and high-energy gamma
components;
b. 223Ra does not need a high atomic number shield because the gamma
component does
not contribute significantly to the dose;
c. Solutions containing pure low-energy beta emitters such as ¹⁴C
and ⁹⁰Sr require a
plastic shield to attenuate the beta particles;
d. Solutions containing high-energy beta emitters such as ³²P and ⁹⁰Y
require a plastic
shield to attenuate the beta particles followed by a high atomic
number material
shield for the bremsstrahlung radiation;
e. Solutions containing radionuclides which have both beta and gamma
radiations (such
as ⁸⁹Sr, ¹⁵³Sm, ¹⁶⁹Er, ¹⁷⁷Lu, and ¹⁸⁶Re) may need in addition a lead shield to
attenuate
the high-energy gamma components;
f. Gamma emitters always require high-Z material shielding.
(b) Protective clothing should be used in work areas where there is a
likelihood of contamination,
such as radiopharmaceutical preparation areas. The protective
clothing may include
laboratory gowns, waterproof gloves (made of latex or non-latex such as
neoprene, polyvinyl
chloride or nitrile), overshoes, and caps and masks for aseptic work. The
clothing serves both
to protect the body of the wearer and to help to prevent the transfer of
contamination to other
areas. The clothing should be monitored and removed before leaving
designated areas;
however, when moving between supervised areas such as the camera room and the
injection
area, it may not be necessary to change the protective clothing
unless a spill is suspected.
Protective clothing should be removed prior to going to other areas such as
staff rooms.
(c) When lower energy beta emitters are handled, the gloves should be thick
enough to protect
against external beta radiation.
(d) Lead aprons should be worn when entering a room with hybrid imaging (e.g.
PET-CT) if the
X rays are about to be used and the staff member needs to be in the room with
the patient.
Lead aprons may also be worn when preparing and administering high
activities of ⁹⁹ᵐTc,
although their use is not recommended, as other protective measures are more
effective.
(e) Tools for remote handling of radioactive material, including tongues and
forceps.
(f) Containers for radioactive waste and radioactive source transportation.
(g) Fume hoods, fitted with appropriate filters, should be used with volatile
radiopharmaceuticals
such as ¹³¹I, ¹³³Xe and ²¹¹At. The sterility of the intravenous
radiopharmaceuticals should be
preserved.
Monitoring of the workplace
4.111. The BSS, in paragraphs 3.96-3.98, sets out the requirements and
responsibilities for workplace
monitoring. Workplace monitoring comprises measurements made in the working
environment and
the interpretation of such results. Workplace monitoring serves several
purposes, including routine
monitoring, special monitoring for specific occasions, activities or tasks, and
confirmatory monitoring
to check assumptions made about exposure conditions. Workplace monitoring
can be used to verify
the occupational doses of personnel whose work involves exposure to
predictable low levels of
radiation. It is particularly important for staff members who are not
individually monitored. In the
nuclear medicine facility, workplace monitoring should address
both external exposure and
contamination. Further general guidance on workplace monitoring is given in Ref
[23].
4.112. Laboratories and other areas in which work with unsealed
sources is undertaken should be
monitored, both for external radiation and for surface
contamination, on a systematic basis.
Contamination monitoring is required for:
(a) All working surfaces (including the interior of enclosures),
tools, equipment and devices
(including dosimetry systems, computers and peripherals, and stress testing
units), the floor
and any items removed from these areas;
(b) Contained workstations, ventilation systems and drains during maintenance;
(c) Protective and personal clothing, and shoes, particularly when
leaving an area that is
controlled due to the risk of contamination (monitors should be available near
the exit);
(d) Clothing, bedding and utensils used by radiopharmaceutical therapy
patients.
4.113. Periodic monitoring with a survey meter and contamination monitor or by
wipe tests should be
conducted for controlled and supervised areas. Continuous monitoring with an
area monitor should be
considered for source storage and handling areas. If a package
containing radioactive sources is
damaged upon arrival, a survey of removable contamination and the external
radiation field should be
carried out.
4.114. Workplace monitoring with respect to X ray based imaging systems used in
nuclear medicine
should follow the guidance given in Section 3, para. 3.95.
4.115. Workplace monitoring should be performed and documented as part of the
nuclear medicine
facility’s radiation protection programme. The nuclear medicine facility’s RPO
or medical physicist
should provide specific advice on the workplace monitoring programme, including
any investigations
that arise when investigation levels are being exceeded (see paras. 4.130 –
4.131).
4.116. The survey meters used for external radiation monitoring should
be calibrated in terms of
ambient dose equivalent. For nuclear medicine, the quantity is H*(10) and the
unit the Sv and its sub-
multiples. Contamination monitors should be calibrated in appropriate
operational quantities. See also
further guidance on calibration in paras. 4.195 – 4.200.
Assessment of occupational exposure and workers’ health surveillance
Occupational exposure assessment
4.117. The purpose of monitoring and dose assessment is, inter alia, to provide
information about the
actual exposure of workers and confirmation of good working practices and
regulatory compliance. It
contributes to reassurance and motivation. The BSS require individual
monitoring for any worker who
normally works in a controlled area or who occasionally works in a controlled
area and is likely to
receive significant occupational exposure (see BSS paragraphs 3.99 to
3.102). Workers who may
require individual monitoring include nuclear medicine physicians, other
specialist doctors, medical
radiation technologists, medical physicists, the RPO, radiopharmacists and any
other persons involved
in the preparation, dispensing and administering of radiopharmaceuticals to
patients for diagnosis and
therapy, staff dealing with radioactive waste, biomedical engineers,
maintenance and servicing
personnel, and any nursing or other staff who need to spend time with nuclear
medicine patients or
work in controlled areas.
4.118. Monitoring involves more than just measurement. It includes
interpretation, assessment,
investigation and reporting, which may lead to corrective measures, if
needed. Individual external
doses are assessed by using individual monitoring devices that include
thermoluminescent dosimeters
(TLD), optical stimulated luminescence dosimeters (OSLD), film badges, and
electronic dosimeters.
Individual monitoring devices must be calibrated, traceable to a standards
dosimetry laboratory. For
more detailed guidance see Ref [23].
4.119. Each dosimeter is to be used for monitoring only the person to whom it
is issued, for work
performed at that nuclear medicine facility and should not be taken
to other facilities where that
person may work. For example, if a person is issued with a dosimeter at
hospital A it should be worn
only at hospital A and not at any other hospitals or medical
centres where he/she may also work.
Monitoring results can then be interpreted for the person working in
a specific nuclear medicine
facility and allow appropriate review of the effectiveness of the
optimization of protection for that
individual in that facility. See also paras. 4.132 to 4.134.
4.120. The monitoring period (period of dosimeter deployment) specified
by regulatory bodies in
most countries is typically in the range of one to three months. It is
determined by such factors as
service availability, work load and type of work. A one month monitoring period
is usually used for
persons performing procedures associated with higher occupational exposures. A
longer monitoring
period (two or three months) is more typical for personnel exposed to lower
doses, as a one month
cycle would usually mean that the actual occupational dose was less than the
minimum detection level
of the dosimeter resulting in no detectable doses. With a longer cycle
it is more likely to obtain a
reading. In certain circumstances (e.g. introduction of new procedures, high
dose rate work) shorter
monitoring periods may be needed. In these situations, the supplementary use of
electronic dosimeters
may be appropriate. Unnecessary delays in the return, reading and reporting of
dosimeters should be
avoided – the nuclear medicine facility should send the dosimeters to the
dosimetry service provider
who should then process the dosimeters and return the dose reports, all
in a timely manner. Some
regulatory bodies may specify a performance criterion for timely reporting.
4.121. The operational dosimetric quantity used for external radiation is the
personal dose equivalent
Hp(d). For weakly penetrating and strongly penetrating radiation, the
recommended depths, d, are 0.07
mm and 10 mm, respectively. Both weakly penetrating and strongly penetrating
radiation are used in
nuclear medicine. Hp(10) is used to provide an estimate of
effective dose that avoids both
underestimation and excessive overestimation [23].
4.122. For monitoring the skin and extremities, a depth of 0.07 mm (d = 0.07)
is recommended, and
Hp(0.07) is used to provide an estimate of equivalent dose to the skin and
extremities. When there is a
possibility of high exposure to the hands, such as in the
preparation and administration of
radiopharmaceuticals, extremity dosimeters should be worn (if
compatible with good clinical
practice).
4.123. For monitoring the lens of the eye, a depth of 3 mm (d = 3) is
recommended, and Hp(3) is used
to provide an estimate of equivalent dose to the lens of the eye. In practice,
however, the use of Hp(3)
has not been widely implemented for routine individual monitoring. In
nuclear medicine it would
generally be expected that the dose to the lens of the eye is not higher than
for the rest of the body. A
possible exception is during the handling of sources for preparation
and administration, but with
accepted practices (as described in 4.69 to 4.90) the lens of the eyes should
be adequately protected.
Nonetheless, monitoring of eyes doses may need to be considered.
4.124. There are three dose limits applicable to workers in nuclear
medicine – effective dose, and
equivalent dose to the lens of the eye, and to skin and extremities.
However, in nuclear medicine,
exposures from both external radiation and internal contamination are relevant.
The dosimeter being
worn will measure external radiation only and will be used to estimate one or
more of the quantities
used for the dose limits. Depending on the work being performed by the person
being individually
monitored, there may be a preferred position for wearing the dosimeter, and
more than one dosimeter
may be used. In nuclear medicine, dosimeters are usually worn on the front of
the upper torso (and
under any protective clothing), as occupational exposure arising from
most nuclear medicine
procedures results in the whole body being fairly uniformly exposed.
See para. 4.122 for when
extremity dosimeters should be worn.
4.125. When a protective apron is being used, the assessment of
effective dose may not be
straightforward:
(a) A single dosimeter placed under the apron, reported in Hp(10),
provides a good estimate of the
contribution to the effective dose by the parts of the body
protected by the apron, but
underestimates the contribution of the unprotected parts of the body (thyroid,
head and neck,
and extremities).
(b) A single dosimeter worn outside the apron, reported in
Hp(10), provides a significant
overestimate of effective dose and should be corrected for the
protection afforded by the
apron by using an appropriate algorithm [ 123, 127 129].
(c) In nuclear medicine, a single dosimeter under the apron provides an
estimate of the effective
dose that is sufficient for radiation protection purposes.
4.126. In nuclear medicine, certain workers may be at risk of both surface
(skin) contamination and
internal contamination by ingestion, inhalation or adsorption of radioactive
material. Employers are
responsible (BSS paragraph 3.102) to identify those persons and arrange
appropriate monitoring. This
requirement is typically met by monitoring the thyroid with an external
detector assessing the iodine
uptake for individuals handling radioiodine and by monitoring the hands after
the protective gloves
have been removed. In some special cases, it may be required to
measure the activity of urine
samples. The committed effective dose should be calculated as part of
the worker’s total effective
dose [23].
4.127. When not in use, individual dosimeters should be kept in an established
place and protected
from damage or from irradiation. If an individual’s dosimeter is lost, the RPO
should perform a dose
assessment, record this evaluation of the dose and add it to the worker’s dose
record. Where there is a
national dose registry, it should be informed of the dose estimate
in a timely manner. The most
reliable method for estimating an individual’s dose is to use his or her recent
dose history. In those
cases where the individual performs non-routine types of work, it may be better
to use the doses of co-
workers having similar exposure conditions as the basis for the dose estimate.
4.128. In some cases occupational doses may be estimated from the results of
workplace monitoring.
The effective dose for personnel can be inferred from the measured ambient dose
equivalent H*(10),
provided the dose gradient in the workplace is relatively low. ICRP
[125] provides conversion
coefficients from ambient dose equivalent to effective dose for
different types of radiation and
energies.
4.129. The use of additional direct reading operational dosimeters,
such as electronic dosimeters,
should be considered for use in a nuclear medicine facility, e.g. in
a new department or with the
introduction of new procedures, as these devices can give the worker an instant
indication of both the
cumulative and the current dose rate and also allow pre-setting of an alarm to
alert when a given level
has been reached [23]. These dosimeters are also useful for staff
involved in radiopharmaceutical
therapies and for pregnant workers where a ‘real-time’ radiation exposure
reading is recommended.
Investigation levels for staff exposure
4.130. Investigation levels are separate from dose constraints and dose limits;
they are a tool used to
provide a ‘warning’ of the need to review procedures and performance,
investigate what is not
working as expected and take timely corrective action. Exceeding an
investigation level should
prompt such actions. In nuclear medicine, one could use predetermined values
such as 0.5 mSv per
month for effective dose or 15 mSv per month for finger dose. Suitable
alternatives may be doses that
exceed an appropriate fraction (e.g. 25%), pro rata per monitoring period, of
the annual dose limits or
a pre-set value above a historical average. Abnormal conditions and/or events
should also trigger an
investigation. In all cases the investigation is with a view to improve
implementation of optimization
of occupational protection and results should be recorded. Investigation levels
should also be set for
workplace monitoring.
4.131. The investigation is to be initiated as soon as possible
following the trigger or event and a
written report is to be prepared concerning the cause, including determination
or verification of the
dose, corrective or mitigating actions, and instructions or recommendations to
avoid recurrence. Such
reports should be reviewed by quality assurance and radiation safety
committees, as appropriate, and
the holder of the licence should be informed. In some cases, the
regulatory body should also be
informed.
Persons who work in more than one place
4.132. Personnel may work in more than one nuclear medicine facility. The
facilities may be quite
separate entities in terms of ownership and management, or they may have common
ownership but
separate management, or they may even have common ownership and management, but
be physically
quite separate. No matter which, the occupational radiation protection
requirements for the particular
nuclear medicine facility apply when the person is working in that
facility. As described above in
para. 4.119, a dosimeter issued for individual monitoring should be worn only
in the facility for which
it is issued as this facilitates effective implementation of optimization of
protection in that facility.
This approach is logistically more easily implemented as each physical site has
its own dosimeters –
there is no need to transport dosimeters between facilities, with the risk of
losing them or forgetting
them. In cases where the facilities are under common ownership it may be seen
as an unnecessary
financial burden to provide more than one set of dosimeters for staff that work
in more than one of its
facilities. However the radiation protection advantages of having the
dosimeter results linked to a
person’s work in only one nuclear medicine facility remain. See also para.
4.134.
4.133. There is however an additional consideration, namely
ensuring compliance with the
occupational dose limits. Any person who works in more than one nuclear
medicine facility should
notify the licensee for each of those facilities. Each licensee, through their
RPO, establishes formal
contact with the licensees of the other nuclear medicine facilities, and their
RPOs, so that each facility
has an arrangement to ensure that a personal dosimeter is available
and that there is an on-going
record of the occupational doses for that person in all the facilities where
they work.
4.134. Some personnel, such as consultant medical physicists or service
engineers, may perform work
in many nuclear medicine facilities and, maybe as well, other medical radiation
facilities. They may
be employed by a company or be self-employed, providing contracted
services to the nuclear
medicine and other facilities. In such cases it is simpler for the company or
the self-employed person
to provide the dosimeters for individual monitoring. In other words, in these
cases for each person the
same dosimeter is used for his/her work performed in all nuclear
medicine facilities (and other
medical radiation facilities) in the monitoring period.
Records of occupational exposure
4.135. Paragraphs 3.103 to 3.107 of the BSS state the detailed
requirements for records of
occupational exposure, placing obligations on the employer, registrant
and licensee. Apart from
demonstrating compliance with legal requirements, records of occupational
exposure should be used
within the nuclear medicine facility for additional purposes, including
assessing the effectiveness of
the facility’s implementation of optimization of protection, and
evaluating trends in exposure.
National or local regulatory bodies might specify additional requirements for
records of occupational
exposure and for access to the information contained in those records. Further
general guidance on
records of occupational exposure is given in [23].
Workers’ health surveillance
4.136. The primary purpose of health surveillance is to assess the
initial and continuing fitness of
employees for their intended tasks, and requirements are given in BSS
paragraphs 3.108 to 3.109.
4.137. No specific health surveillance related to exposure to ionizing
radiation is necessary for staff
involved in nuclear medicine. Under normal working conditions, the occupational
doses incurred in
nuclear medicine are low and no specific radiation related medical surveillance
is required as there are
no diagnostic tests that yield information relevant to low dose
exposure. It is, therefore, rare for
considerations of occupational exposure arising from the working environment of
a nuclear medicine
facility to significantly influence the decision about the fitness of a worker
to undertake work with
radiation or to influence the general conditions of service [23].
4.138. Only in cases of overexposed workers, at doses much higher than the dose
limits (e.g. a few
hundred millisieverts or higher), would special investigations involving
biological dosimetry and
further extended diagnosis and medical treatment be necessary [23]. In case of
internal contamination,
additional investigations to determine uptake and retention may be required.
Interventions to facilitate
excretion or limit uptake of the radioactive agent should be considered, as
appropriate.
4.139. Counselling should be available to workers who have or may have been
exposed in excess of
dose limits, and information, advice and, if indicated, counselling should be
available to workers who
are concerned about their radiation exposure. In nuclear medicine, the
latter group may include
women who are or may be pregnant. Counselling should be given by appropriately
experienced and
qualified practitioners. Further guidance is given in [23].
Information, instruction and training
4.140. All staff involved in nuclear medicine should fulfil the
respective training and competence
criteria described in Section 2, paras. 2.117 to 2.135. This will include
general education, training,
qualification and competence for occupational radiation protection in
nuclear medicine. Nuclear
medicine physicians, medical radiation technologists, medical physicists
and nurses may not have
been trained with respect to the X ray based component of hybrid imaging
systems, such as PET-CT,
and as such should undertake radiation protection training relevant to
the additional imaging
modalities in their nuclear medicine facility.
4.141. The BSS, in paragraph 3.110, places responsibilities on
employers to provide, inter alia,
specific instruction and training for protection and safety as it
pertains to their nuclear medicine
facilities. This is not only for new staff but also for all staff as part of
their continuing professional
development. Specific instruction and training should be provided when new
radiopharmaceuticals,
medical radiological equipment, software and technologies are introduced.
4.142. Information on potential contamination risks should be given to
ancillary staff, including
information technology specialists, and contractors doing occasional
work in a nuclear medicine
facility and/or radiopharmaceutical laboratory.
Conditions of service and special arrangements
4.143. As required in BSS paragraph 3.111, no special benefits are to be
offered to staff because they
are occupationally exposed. It is simply not acceptable to offer benefits as
substitutes for measures for
protection and safety.
Special arrangements – female workers (pregnant or breast-feeding)
4.144. A female worker should notify the licensee that she is pregnant as soon
as she knows of her
status, or if she is breast feeding, so that radiation protection requirements
for the embryo/fetus and
baby as a member of the public can be met respectively.
4.145. Limitation of the dose to the embryo/fetus does not mean that pregnant
women should avoid
working with radiation, but it does imply that the employer should
carefully review the exposure
conditions with regard to both normal exposure and potential exposure.
For example, a pregnant
worker may be restricted from spending a lot of time in the radiopharmacy or
working with solutions
of radioiodine [130]. The main risk with radioiodine is that it
crosses the placental barrier and
concentrates in the fetal thyroid.
4.146. Other possible solutions include reassigning a pregnant worker to duties
where the likelihood
of an accident or incident is unlikely or to a location that may have lower
ambient dose equivalent.
Adequate training should accompany such reassignments. A further consideration
is to avoid having
pregnant workers respond to an incident or accident such as a radioactive spill
(see also paras. 4.291 –
4.296).
4.147. The use of personal electronic dosimeters are valuable in assessing
radiation doses to pregnant
workers and subsequently the embryo/fetus. See also para. 4. 129.
4.148. When applying the dose limit of 1 mSv to the embryo/fetus, embryo/fetal
doses are not likely
to exceed 25% of the personal dosimeter measurement of external exposure. This
value depends on
the penetration of the radiation, i.e. on the photon energy of the
radionuclides in use. Information,
advice and, if indicated, counselling for pregnant workers should be available
(see also para. 4.139).
Special arrangements – persons under 18
4.149. In many countries there is the possibility of students aged
16 or more, but under 18,
commencing their studies and training to become a medical radiation
technologist or other health
professional that may involve occupational exposure to ionizing radiation. The
BSS paragraph 3.116
states the requirements for access to controlled areas, and the dose limits for
such persons are more
restrictive – see Table 2 in this Safety Guide and Schedule III of the BSS.
Protection of workers responding to incidents in the nuclear medicine facility
4.150. The practice of nuclear medicine is a planned exposure
situation, and when circumstances
result in incidents that lead to, or could lead to, unintended or accidental
exposures of patients or staff,
they are still within the framework of a planned exposure situation. The
potential occurrence of such
should be considered in advance in the facility’s safety assessment and
contingency plans developed
accordingly – see guidance in paras. 4.280 – 4.297 in the section
on prevention and mitigation of
accidents.
4.151. Occupational exposure of staff responding to these incidents is still
subject to the occupational
dose limits and the contingency plans for incidents should include
considerations for the optimization
of protection for the responding workers. The contingency plans should
also include allocation of
responsibilities and provide for the training of the relevant staff in
executing the mitigation measures,
which should be periodically rehearsed. Most of these situations, for example
spillage of radioactive
materials on work surfaces, can be executed in a planned manner so that doses
can be kept low. There
may be cases with high doses, for example in medical emergencies
involving immediate care of
patients in the case of a stroke or cardiac arrest, when large amounts of
radioactive material have been
incorporated (e.g. 2 GBq of ¹³¹I), but in these events the dose is
justified because the procedure is
lifesaving. However, even in the case of urgent surgery, rotation of personnel
may be utilized if the
surgical procedure is lengthy, to help maintain optimized occupational
radiation protection for this
situation. The advice of the facility’s RPO is needed for these situations. See
the guidance in paras.
4.295 – 4.296 for more details.
RADIATION PROTECTION OF PATIENTS, CARERS AND COMFORTERS, AND
VOLUNTEERS IN BIOMEDICAL RESEARCH
Introduction
4.152. This section covers what is more formally called ‘medical exposure’ in
radiation protection.
Medical exposure is defined in the BSS and described in Section 2, para. 2.5.
It concerns radiation
protection of the patient, carers and comforters, and volunteers in
biomedical research. The term
‘patient’, when used in the context of medical exposure, means the person
undergoing the radiological
procedure. Other patients in the nuclear medicine facility, including
those who may be waiting for
their own radiological procedure, are considered as members of the
public and their radiation
protection is covered in paras. 4.260 – 4.269.
4.153. As described in Section 2 para. 2.8, there are no dose limits for
medical exposure, making it
very important that there is effective application of the requirements for
justification and optimization.
Justification of medical exposures
4.154. The requirements of the BSS for justification of medical exposure, BSS
paragraphs 3.155 to
3.161, incorporate the ‘3 level approach’ to justification [4, 131, 140].
4.155. The roles of the health authority and professional bodies with respect
to ‘level 2’ or generic
justification of radiological procedures, justification of health screening
programmes, and justification
of screening intended for the early detection of disease, but not as
part of a health screening
programme, are described in Section 2 paras. 2.55 – 2.59 and 2.64 – 2.65.
Justification for the individual patient
4.156. The BSS invokes a joint approach to justification at the individual
patient level, with a shared
decision involving both the referring medical practitioner (who initiates the
request for a radiological
procedure) and the radiological medical practitioner. A referral for a
nuclear medicine procedure
should be regarded as a ‘request for a professional consultation or opinion’
rather than an ‘instruction
or order to perform’. The referring medical practitioner brings the knowledge
of the medical context
and the patient’s history to the decision process, while the radiological
medical practitioner has the
specialist expertise on the radiological procedures. The efficacy,
benefits and risks of alternative
methods (both involving and not involving ionizing radiation, such as for
example CT or ultrasound)
should be considered. Ultimate responsibility for justification will be
identified by individual member
states’ regulations.
4.157. In the case of radiopharmaceutical therapy, the requirements for
justification are implemented
more effectively as part of the medical process of determining the best
approach to treatment. When a
patient is referred by a referring medical practitioner for treatment, careful
consideration should be
made by a multidisciplinary team, including such specialists as
radiation oncologists or
endocrinologists, regarding whether to treat the patient with
radiopharmaceutical therapy or some
other form of radiation therapy, another modality, a combined
treatment approach (sequential or
concomitant) or not to be treated at all. Ideally every treatment
decision should be discussed and
documented in a tumour board or equivalent multidisciplinary meeting.
4.158. The patient also should be informed about the expected benefits, risks
and limitations of the
proposed radiological procedure(s), as well as the consequences of not
undergoing the procedure.
4.159. In nuclear medicine imaging, requirements for justification are applied
more effectively as part
of the medical process of determining ‘appropriateness’. Appropriateness
uses an evidence-based
approach to choose the best test for a given clinical scenario, taking into
account diagnostic efficacy
and justification as well as alternative procedures that do not use
ionizing radiation, for example,
ultrasound or MRI. Useful tools to support this decision making
process include national or
international imaging referral guidelines developed by professional
societies [133 – 140, 245], in
some cases disseminated through electronic requesting systems¹⁷ and clinical
decision support tools or
systems.
4.160. In determining appropriateness of the nuclear medicine imaging
procedure for an individual
patient, the following questions should be asked by the referring medical
practitioner [138];
(a) Has it been done already? A radiological procedure that has already
been performed within a
reasonable time period (depending on the procedure and clinical
question) should not be
repeated. In some cases an alternative procedure may have already been
performed in another
facility, making the proposed radiological procedure unnecessary – for
example a CTPA
performed recently in one facility for a patient who presents as a
potential V/Q patient at
another facility. The results (images and reports) of previous
examinations should be
available, not only within a given nuclear medicine facility but also
between different
facilities. Digital imaging modalities and electronic networks should
facilitate this process.
(b) Is it needed? The results of the proposed radiological procedure
(positive or negative) should
influence the patient’s management.
(c) Is it needed now? The timing of the proposed radiological
procedure in relation to the
progression of the suspected disease and the possibilities for
treatment, all should be
considered as a whole.
(d) Is this the best investigation to answer the clinical question?
Advances in imaging techniques
are taking place continually, and the referring medical practitioner may need
to discuss with
the radiological medical practitioner what is currently available for a given
problem.
(e) Has the clinical problem been explained to the radiological
medical practitioner? The
medical context for the requested radiological procedure is crucial to
ensure the correct
technique is performed with the correct focus.
4.161. Three particular groups of patients are identified in the BSS,
paragraph 3.157, for special
consideration with respect to justification in nuclear medicine –
patients who may be pregnant or
breast-feeding, and paediatric patients.
(a) Due to the higher radiosensitivity of the embryo/fetus, it should
be ascertained whether a
female patient is pregnant before performing a nuclear medicine
procedure. Determining
pregnancy status is a requirement in the BSS, paragraph 3.176, for
those radiological
procedures that could result in significant dose to the embryo or fetus.
Pregnancy would then
be a factor in the justification process and might influence the
timing of the proposed
radiological procedure or whether another examination or treatment is more
appropriate. Care
should be taken to ascertain that the examination is indeed indicated for a
medical condition
that requires prompt medical treatment. Confirmation of pregnancy may occur
after the initial
justification and before the radiological procedure is performed.
Repeat justification is
required taking into account the additional sensitivity of the pregnant woman
and embryo or
fetus.
a. Most diagnostic procedures with ⁹⁹ᵐTc do not cause high
fetal doses. For
radionuclides that do not cross the placenta, fetal dose is
derived from the
radioactivity in maternal tissues. Some radiopharmaceuticals, or their
breakdown
¹⁷ Such electronic requesting systems include the so-called CPOE – computerized
physician order entry, noting
that for imaging such a system should generate a request rather than an order.
components, that do cross the placenta and concentrate in a specific organ or
tissue
can pose a significant risk to the fetus. Particular attention should
be drawn to
radiopharmaceuticals labelled with iodine isotopes. Radiopharmaceuticals
labelled
with other radionuclides, in particular positron emitters, need special
consideration. In
all these instances the medical physicist should estimate the fetal
dose. Detailed
information on doses to the embryo and fetus from intakes of radionuclides
by the
mother is given by the ICRP [246].
b. As a rule, a pregnant woman should not be subject to radioiodine therapy
unless the
application is lifesaving. Otherwise, the therapeutic application should
be deferred
until after the pregnancy and after any period of breast feeding
[130, 247]. In
particular, radioiodine will easily cross the placenta, and the fetal thyroid
begins to
accumulate iodine at about ten weeks of gestation.
(b) In breast feeding patients excretion through the milk and
potentially enhanced breast dose
should be considered in the justification process. Detailed information
on doses to infants
from the ingestion of radionuclides in mother’s milk is given by the ICRP [248]
and can be
calculated using available software, for example [249].
(c) As children are at greater risk of incurring radiation-induced
stochastic effects, paediatric
examinations require special consideration in the justification process [245].
4.162. ‘Self-referral’ occurs when a health professional undertakes a
radiological procedure for
patients as a result of justification based on their own previous clinical
assessment. Most examples of
acceptable self-referral practice occur with X ray imaging, such as dentistry,
and relevant professional
bodies in many countries develop appropriate guidance for their specialty (see
Section 3 para. 3.143).
Self-referral in nuclear medicine, if it occurs, would need to be
guided by such professional
guidelines.
4.163. ‘Self-presentation’, including ‘individual health assessment’,
occurs when a member of the
public asks for a radiological procedure without a referral from a health
professional. This may have
been prompted by media reports or advertising. Self-presentation for nuclear
medicine procedures is
not widely prevalent, but for any such case justification is required, as for
all radiological procedures.
Relevant professional bodies have an important role in considering evidence for
developing guidance
when new practices are proposed. Member states may choose to
incorporate this approach into
legislation [144].
4.164. Approaches to support applications of the requirements for justification
for medical exposures
should consider the development of means to improve ‘awareness,
appropriateness and audit’.
Awareness of the need for justification underpins the whole process. Means for
promoting awareness
include traditional education and training, such as at medical school or during
specialty training, web-
based learning or learning through work flow at the right time, (e.g. junior
doctors in the emergency
department), and the use of feedback in the reporting process. Appropriateness
has been described in
paras. 4.159 to 4.161, and the process of audit is used for monitoring and
feedback to improve both
awareness and appropriateness.
Justification – biomedical research volunteers
4.165. The role of the ethics committee in the justification of medical
exposure of volunteers exposed
as part of a programme of biomedical research is described in Section 2 para
2.98.
Justification – carers and comforters
4.166. The justification of medical exposure incurred by a carer or comforter
is effectively carried out
by the radiological medical practitioner or delegated to the medical radiation
technologist involved in
the radiological procedure, prior to the performance of the procedure.
It depends on the carer or
comforter being correctly informed about the radiation doses and risks
involved, and his or her
understanding of this information and consequent agreement to take on the role
of carer or comforter.
Optimization of protection and safety
4.167. In medical exposures, optimization of protection and safety has
several components, some
applied directly to the radiological procedure(s) about to be
performed and others providing the
support or framework for the other components. These components of optimization
of protection and
safety are described in the following paras. 4.168 to 4.236.
Design considerations
4.168. The use of appropriate and well-designed medical radiological
equipment and associated
software underpins any nuclear medicine procedure. Gamma cameras,
SPECT-CT and PET-CT
scanners and their accessories should be designed and manufactured so as to
facilitate the keeping of
medical exposures as low as reasonably achievable consistent with
obtaining adequate diagnostic
information. Guidance on design considerations is given in the sub-section on
medical radiological
equipment, paras. 4.44 to 4.50. Guidance on design considerations
applicable for X ray imaging
systems as part of hybrid systems is given in Section 3, paras. 3.31 to 3.39.
Ultimately, as stated in the
BSS paragraph 3.162, it is the responsibility of the nuclear medicine facility
licensee to ensure that
his/her facility uses only medical radiological equipment and
software that meets applicable
international or national standards.
Operational considerations - general
4.169. Following justification, the nuclear medicine procedure is
performed in such a way as to
optimize patient protection, as required in the BSS paragraph 3.163 for
diagnostic procedures and in
paragraph 3.166 for radiopharmaceutical therapy procedures. The level of image
quality sufficient for
diagnosis is determined by the radiological medical practitioner and is based
on the clinical question
posed.
4.170. The following points apply to all nuclear medicine patients, whether
undergoing diagnostic or
therapeutic procedures:
(d) There should be an effective system for correct identification of
patients, with at least two
forms of verification, for example name, birthday, address, medical record
number.
(e) Patient details should be correctly recorded, such as age, gender, weight,
height, pregnancy
and breast-feeding status, current medications, allergies.
(f) Patient clinical history should be reviewed.
Operational considerations – diagnostic imaging
4.171. The nuclear medicine facility should have a written protocol
for each diagnostic procedure
performed in the facility, designed to maximize the clinical information to be
obtained from the study,
taking into consideration the appropriate diagnostic reference level for the
procedure (see Section 2
paras. 2.34 – 2.45). These protocols are best developed using guidelines from
national or international
professional bodies, and hence reflect current best practices, as for example
in Refs [65, 214, 215, 250
– 255]. For modern digital equipment many of the factors are
automated through menu-driven
selections on the equipment console. Nevertheless, in setting up these options,
significant scope exists
for optimization of protection through the appropriate selection of
values for the various technical
parameters, thereby producing effectively an electronic protocol.
Protocols should be periodically
reviewed in line with the requirements for quality assurance and
radiological reviews (see paras.
4.231 and 4.256 - 4.258).
4.172. Deviations from such protocols may be necessary owing to the special
needs of a particular
patient or because of the local unavailability of components for a test. In
these cases the radiological
medical practitioner should record a valid reason for his or her decision.
4.173. Equipment should be operated within the conditions established in the
technical specifications,
and in any licence conditions, ensuring that it will operate satisfactorily at
all times, in terms of both
the tasks to be accomplished and radiation safety, so that optimal image
acquisition and processing
can be achieved with the minimum of patient exposure.
4.174. Many factors influence the relationship between image quality and
patient dose in diagnostic
nuclear medicine procedures. Detailed guidance on appropriate choices
for those factors is widely
available and should be followed [65, 214, 215, 219, 250 – 255]. Such factors
include:
(a) Appropriate selection of the best available radiopharmaceutical and
its activity, noting the
special requirements for children and for patients with impairment of organ
function;
(b) Adherence to patient preparation requirements specific to the
study to be performed.
Examples include:
a. Use of methods for blocking the uptake in organs not under study and for
accelerated
excretion when applicable;
b. Correct hydration.
(c) The storage or retention of radiopharmaceuticals within specific organs
can be influenced by
drugs such as diuretics or gall bladder stimulants, whenever they do not
interfere adversely
with the procedure. This method is sometimes used to increase the
specificity of the
examination, but has also a positive influence on radiation protection – e.g.
the use of diuretic
challenge in renography.
(d) For children undergoing diagnostic procedures, the amount of activity
administered should be
chosen utilizing methodologies described in international or national
guidelines [66, 214, 215,
219, 251, 252, 256 – 258].
(e) Using appropriate image acquisition parameters:
a. For nuclear medicine with a gamma camera, this may include selection of
collimator,
acquisition matrix, energy windows, acquisition zoom, time per frame,
imaging
distance;
b. For PET systems, this may include 2D and 3D acquisitions, matrix size,
field of view,
time of flight, attenuation correction, slice overlap, scatter
correction, coincidence
timing;
(f) Using appropriate reconstruction parameters such as
algorithm, matrix, filters, scatter
correction, zoom factor,
(g) Utilizing quantitative and qualitative capabilities, such as the
generation of region-of-interest
(ROI) analysis, time-activity curve generation, image reformatting, or
tissue uptake ratios,
specific for the clinical need.
4.175. Many radionuclides are excreted by the kidneys. Bladder doses can be
minimized by drinking
plenty of fluid and frequent bladder emptying. Patients, particularly in the
case of children, should be
encouraged to void frequently, especially in the immediate interval following
the examination.
4.176. While most adults can maintain the required position without
restraint or sedation during
nuclear medicine examinations, it may be necessary to immobilize or
sedate children so that the
examination can be completed successfully. Increasing the administered
activity to reduce the
examination time is an alternative that can be used in elderly patients with
pain.
4.177. In some cases, if the patient is healthy and cooperative, activity
might be reduced and scan
times increased – e.g. lung scans for pregnant women. In all cases, however,
diagnostic information
produced should not be compromised by reduction in activity.
4.178. Care should be taken to ensure that there is no contamination
on the collimator surface or
elsewhere as this might impair the quality of the images.
Operational considerations – radiopharmaceutical therapy
4.179. The nuclear medicine facility should have written
protocols for each type of
radiopharmaceutical therapy performed in the facility, designed to meet the
requirements of the BSS
paragraph 3.165. These protocols are best developed using guidelines from
national or international
professional bodies, and hence reflect current best practices, as for example
in Refs [214, 215, 250 –
254, 259, 260]. For modern digital equipment many of the factors are automated
through menu-driven
selections on the equipment console. Nevertheless, in setting up these options,
significant scope exists
for optimization of protection through the appropriate selection of
values for the various technical
parameters, thereby producing effectively an electronic protocol.
Protocols should be periodically
reviewed in line with the requirements for quality assurance and
radiological reviews (see paras.
4.231 and 4.256 - 4.258).
4.180. In addition to the guidance in paras. 4.169 – 4.171 (for both
diagnostic and therapy nuclear
medicine procedures), the following provisions should be in place:
(a) Verbal and written information and instructions to patients about
their radiopharmaceutical
therapy and about how to minimize exposure to family members and the public,
and advice
on pregnancy and contraception after therapy. For detailed guidance,
including sample
information sheets, see Refs [214, 18, 261, 262];
(b) Special attention to preventing spread of contamination due to patient
vomit and excreta;
(c) A protocol for the release of patients after
administration of therapeutic doses of
radiopharmaceuticals (see guidance in paras. 4.243 – 4.245).
4.181. The BSS, paragraph 3.165, gives requirements concerning the
activity of the therapeutic
radiopharmaceutical to be administered. Algorithms for determining appropriate
activities for a given
patient based on radiation doses to critical organs do exist, but there
is no standardized algorithm.
Methodologies are discussed in Refs [214, 251, 263-268]. Typically, therapeutic
radiopharmaceuticals
are administered at standard fixed activities (in gigabecquerel or millicurie),
standard fixed activities
per unit body mass (MBq kg–¹ or mCi kg–¹) or standard fixed activities per
unit body surface area
(MBq m–² or mCi m–²), based on the results of toxicity and side effects in
clinical trials.
4.182. For female patients, their pregnancy and breast-feeding status
should be evaluated before
administration of a therapy dose (see also paras. 4.238 – 4.242). Immediately
prior to administration
of a therapeutic radiopharmaceutical, the following information, as
applicable, should be verified
preferably by two individuals:
(a) Dose on the radiopharmaceutical label matches the prescription;
(b) Identification of the patient by two independent means;
(c) Identity of the radionuclide;
(d) Identity of the radiopharmaceutical;
(e) Total activity;
(f) Date and time of calibration.
4.183. The administered activity should be verified in an activity
meter (dose calibrator) or other
suitable device to ensure that the total activity does not deviate
significantly from the prescribed
administered activity (e.g. < 10 % deviation), and the measured value recorded.
Corrections should be
calculated for residual activity in the syringe, cups, tubing, inline filter,
or other materials used in the
administration.
4.184. Radiopharmaceutical therapy patients should be informed in advance that
it will be necessary
for medical personnel to minimize close or direct contact, so that
this precaution will not be
interpreted as a lack of concern.
4.185. Both females and male patients should be advised about avoiding
conception after therapeutic
administrations. Data on the periods during which conception should be avoided
after administration
of a radiopharmaceutical to females for therapeutic purposes are
given in Appendix II, with further
guidance in Ref. [251]. These times have been derived with a view to the need
for further therapy,
which may be compromised if the patient becomes pregnant.
4.186. The administration of therapeutic doses of relatively long lived
radionuclides in ionic chemical
forms to males is a possible source of concern because of the appearance of
larger quantities of these
radionuclides in ejaculate and in sperm. It may be prudent to advise sexually
active males who have
been treated with for example ³²P (phosphate), ⁸⁹Sr (strontium chloride), ¹³¹I
(iodide), ²²³Ra (chloride),
to avoid fathering children for a period of four months after treatment. The
period of four months is
suggested as this is longer than the life of a sperm cell [251, 269].
Operational considerations – pregnant female patients
4.187. Administration of radiopharmaceuticals for therapy to female
patients who are pregnant or
might be pregnant should be generally avoided. There may be
exceptions when the treatment is
lifesaving. See also paras. 4.161 on justification and 4.238 – 4.240 on the
need to ascertain pregnancy
status.
4.188. Diagnostic nuclear medicine procedures with ⁹⁹ᵐTc and
radiopharmaceuticals that do not cross
the placenta do not cause high fetal doses. Protection of the fetus can be
optimized by using smaller
administered activities and longer imaging times. This is feasible if the
patient is able to remain still.
4.189. Specific assessment of individual fetal doses is not usually necessary
after diagnostic nuclear
medicine studies involving ⁹⁹ᵐTc radiopharmaceuticals. In the case of
other radiopharmaceuticals
(such as iodine or gallium), calculation of dose to the individual
fetus and risk estimation may be
necessary.
4.190. In the case of radiopharmaceuticals that are rapidly eliminated by
the maternal kidneys, the
urinary bladder is the major source of fetal irradiation.
After the administration of such
radiopharmaceuticals, maternal hydration and frequent voiding
should be encouraged. Some
radiopharmaceuticals, for example radioactive iodides, including those
administered for diagnostic
purposes, cross the placenta freely and are taken up by the fetal
tissues, for example the thyroid.
Failure to ascertain whether a patient is pregnant when administering ¹³¹I for
a scan, for example, may
lead to a severe accidental exposure of the fetus.
4.191. Of special concern is also the use of CT in PET-CT or
SPECT-CT examinations. Routine
diagnostic CT examinations of the pelvic region with and without contrast
injection can lead to a dose
of 50 mSv to the uterus, which is assumed to be equivalent to the fetal
dose in early pregnancy. When
PET-CT or SPECT-CT scanning is indicated in a pregnant patient, low-dose CT
protocols should be
used and the scanning area should be reduced to a minimum. See also Section 3,
paras. 3.170 – 3.178.
4.192. If FDG is being used for the PET component of hybrid
imaging, a lower activity of FDG
should be used. Further guidance is given in Refs [65, 270].
Operational considerations – breast feeding
4.193. Female patients should be advised that breast feeding is
generally contraindicated after
therapeutic administration of some radiopharmaceuticals, because of both the
external irradiation of
the suckling baby and the potential excretion of radioactivity through the
breast milk. See also paras.
4.161 on justification and 4.241 – 4.242 on the need to ascertain
breast-feeding status.
4.194. Depending on the radiopharmaceutical, breast feeding may need to be
interrupted for a period
or even stopped following its administration for a diagnostic procedure. More
specific advice is given
in Appendix III and Refs. [214, 247, 251].
Calibration
4.195. Requirements for calibration of sources and instruments used
for dosimetry of patients are
given in the BSS paragraph 3.167. For nuclear medicine, responsibility
is assigned to the nuclear
medicine facility’s medical physicist. Unsealed sources for nuclear
medicine procedures should be
calibrated in terms of activity of the radiopharmaceutical to be administered,
with the activity being
determined and recorded at the time of administration. Detailed guidance on
acceptable protocols for
making activity measurements can be found in the following references [243,
271].
4.196. Radionuclides should be checked for radioactive impurities when these
are liable to be present.
This applies particularly to short lived radionuclides, as longer- lived
impurities may be present and
could deliver a significant fraction of the absorbed dose.
4.197. The calibration of X ray based imaging devices that are part of hybrid
imaging systems, such
as CT in PET-CT or SPECT-CT, should follow the guidance for such modalities in
Section 3. See
also Section 5 for hybrid imaging systems used in radiation therapy
simulation and treatment
planning.
4.198. In the nuclear medicine facility, instruments used for dosimetry of
patients, such as activity
meters (dose calibrators), also should be calibrated at appropriate intervals
using calibrated reference
sources that cover the energy range used in clinical practice. After the
initial calibration, the intervals
for periodic calibrations may differ, depending on the availability of
calibration radioactive sources in
the department. A period of one to two years is recommended.
4.199. The BSS in paragraph 3.167(d) requires the calibration of
dosimetry instrumentation to be
traceable to a standards dosimetry laboratory (SDL). Ideally this
would be to the national SDL
(primary or secondary) in the State concerned, with access either directly or
through a duly accredited
calibration facility. However, it may be necessary for instruments used for
dosimetry of patients to be
sent to another country or state if there is no national SDL in
the country or state where the
instruments are used.
4.200. Records of calibration measurements and associated calculations,
including uncertainty
determinations (budgets), should be maintained as described in para. 4.230.
Dosimetry of patients – diagnostic procedures
4.201. The BSS paragraph 3.168 requires nuclear medicine facilities to ensure
that patient dosimetry
is performed and typical doses for their diagnostic
radiological procedures are determined.
Knowledge of a facility’s typical doses forms the basis for applying
methods of optimization of
protection. It also enables the nuclear medicine facility to use diagnostic
reference levels (see paras.
4.210 to 4.217) as another tool for optimization of protection. Administered
activity (in MBq) is the
most widely used surrogate for dose in diagnostic nuclear medicine,
however organ doses and
effective doses can be calculated from activity using established methodologies
– see para 4.208.
4.202. Clearly, the more radiological procedures at the nuclear
medicine facility for which typical
doses are known, the better the basis for optimization of protection. However,
pragmatically, the BSS
requires only common radiological procedures for diagnostic medical
exposures. What is common
will vary from facility to facility, and country to country, but in general
there are some core common
examinations, including: thyroid scans, bone scans, myocardial perfusion
imaging, FDG-PET/CT in
oncology, renal scans, and lung scans.
4.203. The term ‘typical dose’, as used in the BSS paragraph 3.168, refers to
the average or median
dose or activity for a particular size of patients. Patient size has a large
influence on doses, so some
selection or grouping of patients is required. Such groupings include
‘average adult’, often based
around an average weight of 70 kg with a range of ± 20 kg. Groupings for
children have sometimes
been based on age, such as new born (0 years), infant (1 year), small child (5
years), child (10 years)
and teenager (15 years), but more recently size-specific groupings are
being used. The nuclear
medicine facility should adopt patient size groupings that correspond with the
groupings used in their
country or state for DRLs. The sample size used for each patient grouping and
radiological procedure
should be of sufficient size to assure confidence in the determination of the
mean dose. Such sample
sizes are typically in the range 10 to 20 patients, but clearly the
larger the sample the better the
statistics.
4.204. Dose in the term ‘typical dose’ means for the given procedure the
effective dose, organ doses,
or, in the case of X ray imaging, an accepted dosimetric quantity as
described in Section 3 paras.
3.195 - 3.196. For combining doses from radiopharmaceuticals and X rays, organ
doses will need to
be used.
4.205. Patient dosimetry to determine typical doses in diagnostic nuclear
medicine should be carried
out in conjunction with an assessment of the diagnostic image
quality. Exposure alone is not
meaningful if it does not correspond to images that are sufficient for an
accurate diagnosis. Therefore
patients included in the sample used for determining typical doses
should be only those whose
radiological procedure resulted in acceptable image quality.
4.206. The results of the surveys used to determine typical doses at
the nuclear medicine facility
should be used as part of the facility’s on-going review of the
implementation of optimization of
protection, and additionally will be used for comparison with established DRLs
(see paras. 4.210 to
4.217). The results should also be submitted to the organization in
their country or state that is
responsible for reviewing the national or regional DRLs. With these
considerations in mind, the
patient surveys of administered activities, from which patient doses can be
calculated as required by
the BSS, should take place at intervals of no more than 5 years and preferably
no more than 3 years.
Another trigger for a survey would be the introduction of new
radiopharmaceuticals, equipment or
technology into the nuclear medicine facility or when significant
changes have been made to the
protocols or the equipment.
4.207. Sometimes patient dosimetry in diagnostic nuclear medicine procedures
may be required for
specific individual patients. Reasons might include an unintended or
accidental medical exposure
where an estimation of patient doses is required as part of the
investigation and report (see para.
4.252), or there may be the need to estimate embryo or fetal doses (see para.
4.189).
4.208. There are several indirect and direct methods to estimate
patient dose in diagnostic nuclear
medicine procedures. In the case of hybrid systems, the contribution
from each of X rays and
radionuclides should be calculated and combined. Methodologies and data for the
determination of
doses from radiopharmaceuticals are given in Refs [232, 234, 240,
241, 255 – 259] and for
methodologies for X ray imaging see Section 3, paras. 3.211.
Dosimetry of patients – radiopharmaceutical therapy procedures
4.209. The BSS paragraph 3.168 requires nuclear medicine facilities to
determine typical absorbed
doses to patients for their therapeutic radiological procedures. As in para.
4.208, methodologies for
the determination of doses from therapy radiopharmaceuticals are explained in
detail in Refs [249,
251, 257, 258, 272 – 278].
Diagnostic reference levels
4.210. The BSS, paragraphs 3.168 - 3.169, requires that nuclear medicine
facilities perform patient
dosimetry surveys for their diagnostic procedures, as described in paras 4.201
to 4.208, and that these
results are compared with the established DRLs for their country or region. The
purpose is to ascertain
whether the typical dose or activity in the facility for a given diagnostic
nuclear medicine procedure
compares favourably, or not, with the value of the DRL for that
nuclear medicine procedure.
Guidance on establishing national or regional DRLs in given in Section 2,
paras. 2.34 to 2.45.
4.211. A review of optimization of protection for that particular
nuclear medicine procedure is
triggered if the comparison shows that the facility’s typical dose or activity
exceeds the DRL, or that
the facility’s typical dose or activity is substantially below the
DRL and it is evident that the
exposures are not producing images of diagnostic usefulness or are not yielding
the expected medical
benefit to the patient.
4.212. Given the uncertainties in determining a facility’s typical dose or
activity, questions can arise
over whether a DRL has really been exceeded or not. Some countries adopt an
algorithmic approach,
for example where the facility’s typical dose or activity, minus two times its
standard error, should be
greater than the value of the DRL [13]. A simpler approach, based
purely on the facility’s typical
value, may be sufficient as the purpose is to identify the need for a review,
and perhaps it is better to
perform more reviews than fewer [12, 13].
4.213. No individual patient’s dose or activity should be compared with
a DRL. It is the facility’s
typical dose or activity, as determined by the representative patient sample,
which is compared.
4.214. Further, the comparison is not a case of “Does the nuclear medicine
facility comply with the
DRL?” DRLs are not dose limits. It is mandatory to use them for the comparison
exercise to identify
practices that warrant further investigation.
4.215. The review of how the given nuclear medicine procedure is
being performed and its
optimization of protection, triggered by the DRL comparison, may
conclude that there are valid
reasons supported by sound clinical judgement why the nuclear medicine facility
has a typical dose or
activity that exceeds the DRL. These reasons should be documented as
part of the facility’s QA
programme. On the other hand, the review may identify areas for improvement
resulting in revised
protocols for that nuclear medicine procedure. The results of the DRL
comparison and any ensuing
review and actions should be documented as part of the facility’s QA programme.
4.216. The fact that a nuclear medicine facility’s typical dose or
activity for a nuclear medicine
procedure is less than the DRL for that procedure does not mean
necessarily that optimization of
protection for that nuclear medicine procedure has been fully
achieved. DRLs are only one of the
tools for optimization, aimed specifically at identifying the outliers in
performance.
4.217. The regulatory body in a given country may specify frequencies
for performing DRL
comparisons. Otherwise, the general guidance on patient dosimetry, described
above in para. 4.206,
would be applicable.
Quality assurance for medical exposures
4.218. The BSS paragraph 3.170 requires nuclear medicine facilities to
have a comprehensive
programme of quality assurance for medical exposures. General guidance on
management systems is
given in Section 2, paras 2.136 to 2.147, and it is simply
re-iterated here that the programme of
quality assurance for medical exposures should fit in with, and be
part of, the facility’s wider
management system. The paragraphs in this subsection give guidance on the
technical aspects of the
programme of quality assurance for medical exposures.
4.219. The purpose of the programme of quality assurance for medical exposures
is to help ensure
successful implementation of optimization of patient protection in the nuclear
medicine facility and to
minimize the occurrence of unintended and accidental medical exposures.
4.220. The complexity of the programme of quality assurance for medical
exposures will depend on
the type of nuclear medicine facility. A facility with only limited diagnostic
procedures will have a
simpler programme compared with a facility that offers a comprehensive
diagnostic service, including
PET-CT imaging, radiopharmaceutical therapy, and that has a radiopharmacy.
Nonetheless, most of
the elements of the programme are common, and it is more in degree of
implementation that there are
differences. The BSS paragraph 3.171 establishes the common elements of the
programme.
4.221. Measurements on medical radiological equipment are one
of the components of the
programme. Acceptance tests are required for new or significantly refurbished
or repaired equipment,
or after the installation of new software or modification of existing
software that could affect
protection and safety of patients. The acceptance test is followed immediately
by commissioning, and
then on-going periodic quality control tests, including constancy tests. The
purpose is to ensure that,
at all times, all medical radiological equipment is performing correctly,
accurately, reproducibly and
predictably. Acceptance and commissioning tests should be performed in the same
way for equipment
and/or software that has been donated.
4.222. Depending on the equipment purchase agreement, acceptance tests may be
performed by the
manufacturer’s representative in the presence of the local medical
physicist and the radiological
medical practitioner representing the user, or, if acceptable to the
manufacturer and/or the purchaser,
by a medical physicist jointly with the manufacturer’s
representative. The process involves
verification of all specifications and features of the equipment; in
particular, protection and safety
features including displayed and reported dose metrics.
4.223. After acceptance and before clinical use on patients, commissioning is
carried out by or under
the supervision of the medical physicist. Commissioning should include
measurements of all
parameters and conditions of use that are expected in clinical use. For
most situations the medical
physicist should be directly involved in the measurements, calculations and
interpretation of data to
characterize the equipment’s performance. In some simple situations it
may be sufficient for the
medical physicist to provide documented advice on how the commissioning should
be performed. At
commissioning, the baseline for subsequent constancy tests is established.
4.224. There are many published reports from international and national
organizations and national or
regional professional bodies giving detailed guidance on the quality
control tests that should be
performed in nuclear medicine, including recommended frequencies [192, 193,
197, 210, 211, 214,
215, 217, 225 – 241, 243, 271, 277, 279 – 281]. In addition, many
of these organizations have
extensive websites where any new or updated publications on the topic will be
able to be found. The
regulatory body may have its own specific requirements on the tests that
should be performed and
their frequencies.
4.225. For guidance on the quality control tests for X ray imaging devices used
in nuclear medicine,
see Section 3, para. 3.231.
4.226. In nuclear medicine there is the additional factor of the
radiopharmaceuticals themselves. The
quality assurance programme for medical exposures should ensure that
radiopharmaceuticals intended
for administration to patients are prepared in a manner that meets
clinical needs and satisfies both
radiation safety and pharmaceutical quality requirements [214, 217, 218].
Therefore, radiopharmacist
should be involvemed.
4.227. The BSS paragraph 3.171(e) specifically requires that dosimetry and
monitoring equipment are
part of the QA programme. This is to ensure that such
instrumentation has a current calibration,
typically within 2 years (see para. 4.198), and that it is
functioning correctly. The programme of
quality assurance for medical exposures should establish a calibration cycle
for each instrument and a
set of checks on the operation of each instrument to be performed at set
intervals. This applies to stand
alone dosimetry equipment and to equipment software related to dosimetry –
e.g. software used to
calculate “specific uptake values (SUV)” from which doses can be estimated.
Detailed guidance on
quality control tests for such nuclear medicine instrumentation is given in
[192, 193, 197, 210, 211,
214, 215, 217, 225 – 241, 243, 271, 277, 279 – 281].
4.228. The results of the quality control tests should be compared with
established tolerance limits.
These limits may have been established to ensure compliance with a regulatory
requirement for the
performance of particular physical parameters or they may be set on the basis
of recommended values
given in published reports, such as referenced in para. 4.224.
As required in the BSS paragraph
3.171(b), if the measured values fall outside the tolerance limits, corrective
actions should take place.
Such corrective actions are likely to include maintenance or servicing of the
equipment, and hence the
nuclear medicine facility should have a maintenance programme in
place. In some cases, the
equipment may be outside the tolerance limits by a significant amount and the
equipment should be
immediately taken out of clinical use and not returned until the servicing has
taken place and it has
been ascertained that the equipment now meets the performance requirements.
4.229. The programme of quality assurance for medical exposures in nuclear
medicine should include
use of “checks and balances” to ensure that the facility’s protocols and
procedures for imaging and
therapy, including radiation protection and safety, are being followed.
The periodic review of the
protocols and procedures themselves is part of the facility’s radiological
review (see 4.256 – 4.258).
In addition a review of imaging procedures may have been triggered by a
comparison with DRLs (see
paras. 4.210 to 4.217).
4.230. Maintaining records is a crucial aspect of the programme of
quality assurance for medical
exposures. This includes the procedures used in the programme, and the results
of the quality control
tests including trend analysis, the dosimetry surveys, the DRL comparisons,
corrective actions and the
investigations of unintended and accidental medical exposures. When
planning and developing an
effective QA programme, licensees should recognize that it demands strong
managerial commitment
and support in the form of training and time, personnel and
equipment resources. The regulatory
body, during its inspections of a nuclear medicine facility, should
review the records of the
programme of quality assurance for medical exposures.
4.231. In line with standard practices for quality management, the BSS in
paragraph 3.172 requires
regular and independent audits of the programme of quality assurance for
medical exposures, adding
that the frequency of such audits will depend on the complexity of the
radiological procedures being
performed in the facility. Such audits may be external or internal.
Internal audits are usually
logistically simpler to implement, while the external audit generally has the
advantage of bringing in
an outside perspective. The audit of the nuclear medicine facility’s programme
of quality assurance
for medical exposures can be incorporated into more comprehensive
audits that the facility is
performing for its management system. Further, the results of the audit of the
programme of quality
assurance for medical exposures will have a major input into the
nuclear medicine facility’s
radiological review (see paras. 4.256 to 4.258).
Dose constraints – carers and comforters
4.232. Some diagnostic radiological procedures in nuclear medicine,
particularly of children, can be
better performed with the assistance of a carer or comforter, for example a
relative in the case of a
paediatric patient, or a relative or friend for a disabled patient. In these
circumstances, the carer or
comforter will be exposed. This is usually to a low dose, such as when caring
for a child undergoing a
renal examination, but in some cases the dose is not insignificant e.g. staying
with a child during a
PET examination. Further, in nuclear medicine there is also the additional
consideration of exposure
of carers and comforters after the diagnostic procedure, or in the case of
radiopharmaceutical therapy
with radioiodine, their exposure during the course of the treatment.
This exposure is defined as
medical exposure (see the BSS) and as such is not subject to dose
limits. However the BSS,
paragraphs 3.153 and 3.173, requires that such carers and comforters are
afforded radiation protection
through the application of the requirements for optimization of
protection and safety and, in
particular, the use of dose constraints in this process. These are the dose
constraints established by
government, as a result of consultation with the health authority, relevant
professional bodies and the
regulatory body, as required by the BSS paragraph 3.149(a)(ii). Guidance on
setting dose constraints,
including considerations for children and pregnant women, is given in Section
2, paras, 2.48 to 2.49.
4.233. The nuclear medicine facility should have written protocols for
implementing measures for the
optimization of protection for carers and comforters of patients
during or after nuclear medicine
procedures. The measures should utilize the basic methods for radiation
protection, i.e. shielding, time
and distance (proximity), and measures to minimize spread of
contamination. The protocol should
include the following:
(a) Criteria specifying who would be acceptable for acting as a carer or
comforter;
(b) Methods for ensuring that the carer or comforter receives a dose
that is as low as reasonably
acceptable;
(c) The values of the dose constraints to be applied (see Section 2
para. 2.49).
4.234. Licensees should be able to demonstrate that the effective dose to the
carer or comforter, by
applying the protocol, is unlikely to exceed the dose constraint. In
some cases it is relatively
straightforward to estimate effective doses to carers and comforters
from measurements of the
ambient dose equivalent rates at the positions where they will be
situated. These determinations
should be made in advance to verify that dose constraint will not be exceeded.
Therefore, individual
dose monitoring is normally not necessary. For carers and comforters in a
therapy ward, consideration
may be given to the use of electronic dosimeters.
4.235. The BSS paragraph 3.153 states that a carer or comforter must be
informed about the radiation
risks involved in helping with the radiological procedure and about the means
to be taken to afford
appropriate radiation protection to the carer or comforter. The carer or
comforter should indicate that
he/she has understood the information and is still willing to provide support,
care and comfort to the
patient that is or has undergone a nuclear medicine procedure. In
the case of radiopharmaceutical
therapy with iodine, for both patients still in the hospital and for those that
have been released (see
also para. 4.245), appropriate written instructions should be provided to the
carer or comforter of the
patient (including for example, time and proximity to the patient, minimizing
physical contact and not
sharing food or drinks). Further guidance is given in Refs [18, 261].
4.236. Guidance applicable to carers and comforters supporting patients
undergoing X ray imaging
radiological procedures as part of the nuclear medicine procedure in the
nuclear medicine facility is
given in Section 3 paras. 3.236 to 3.240.
Dose constraints – volunteers in biomedical research
4.237. Some individuals will undergo diagnostic nuclear medicine
procedures as part of their
voluntary participation in an approved programme of biomedical research (see
Section 2 para. 2.98).
Part of the approval process for the biomedical research will have been the
setting of dose constraints
for the nuclear medicine procedures (see Section 2 para. 2.99). When the
volunteer presents at the
nuclear medicine facility, they are to be afforded the same radiation
protection as if they were a
patient presenting for a nuclear medicine procedure within a normal health care
pathway, but with the
additional measure that their exposure will be subject to a dose
constraint, either a nationally
established dose constraint or as specified by the ethics committee
that approved the biomedical
research programme (see Section 2 paras. 2.50 and 2.99).
Pregnant female patients
4.238. Female patients who are pregnant form a special sub-group of
patients that should be
particularly considered with respect to radiation protection. These
considerations have been described
in para. 4.161 for justification and paras. 4.187 - 4.192 for optimization.
None of these considerations
can take place if it is not known that the patient is pregnant. Therefore it is
crucial, as is required in
the BSS paragraphs 3.175 -3.176, for the nuclear medicine facility to have in
place means for ensuring
that pregnancy status is known for female patients.
4.239. The first approach is through posting of clear signs, in
languages easily understood by the
range of people using the nuclear medicine facility, asking the question “Are
you pregnant or possibly
pregnant?” and, if so, “Please tell the staff”. Such signs should be
posted widely in the facility,
including waiting rooms and cubicles. The second approach is to ask female
patients directly whether
they are or might be pregnant. This may not always be so easy given social and
cultural sensitivities,
but it may should be done when needed.
4.240. Neither of the approaches described in para. 4.239 will work if the
woman does not know if
she is or may be pregnant. For this reason, the BSS (paragraph 3.176) has an
additional requirement
on facilities to have “procedures in place for ascertaining the pregnancy
status of a female patient of
reproductive capacity before the performance of any radiological
procedure that could result in a
significant dose to the embryo or fetus”. In nuclear medicine, pregnancy status
should be ascertained
for all radiopharmaceutical therapy and it is advisable for all diagnostic
procedures, in particular for
those radiopharmaceuticals that are known to cross the placental
barrier. Cooperation with the
referring medical practitioner, through standard requests for
pregnancy status for specified
procedures, is one approach. In case of doubt, a pregnancy test or hormonal
level determination to
assess menopausal status may be carried out.
Breast-feeding patients
4.241. As above for pregnant women, breast-feeding patients form a special
sub-group of patients that
should be particularly considered with respect to radiation protection
in nuclear medicine. These
considerations have been described in para. 4.161 for justification
and paras. 4.193 – 4.194 for
optimization. None of these considerations can take place if it is not known
that the patient is breast-
feeding. Therefore it is crucial, as is required in the BSS
paragraphs 3.175 -3.176, for the nuclear
medicine facility to have in place means for ensuring that breast-feeding
status is known for female
patients.
4.242. The first approach is through posting of clear signs, in languages able
to be understood by the
range of people using the nuclear medicine facility, simply asking
the question “Are you breast
feeding?” and, if so, informing staff. Such signs should be posted
widely in the facility, including
waiting rooms and cubicles. The second approach is to directly ask female
patients whether they are
breast feeding. This may not always be so easy given social and
cultural sensitivities, but it may
should be done when needed.
Release of patients after radiopharmaceutical therapy
4.243. As required in the BSS paragraph 3.178, a nuclear medicine facility
should have arrangements
in place to manage the release of patients who have undergone
radiopharmaceutical therapy. Once the
patient is released, two groups of persons should be afforded appropriate
radiation protection – the
general public with whom the patient may encounter or interact, and members of
the patient’s family
and close friends who may be viewed simply as also being members of the public
or as carers and
comforters. Exposure of members of the public is subject to the
public dose limits (see Section 2
Table 2), while exposure of carers and comforters is not subject to dose
limits but is instead controlled
through dose constraints (see paras. 4.232 – 4.236). Further, as
discussed in para 2.46, public
exposure arising from a single “source”, such as a patient who has undergone
radiopharmaceutical
therapy, should be subject to dose constraints set at some fraction of the dose
limits.
4.244. The nuclear medicine facility’s medical physicist or radiation
protection officer should
establish prior to the release of a patient that the retained radioactivity in
the patient is such that the
doses that could be received by members of the public would not
exceed public dose limits, and
would be unlikely to exceed the relevant dose constraints for both members of
the public and carers
and comforters. An acceptable method to estimate the acceptable retained
activity for patients being
discharged from hospitals is to calculate the time integral of the
ambient dose equivalent rate,
considering the activity, energy and the effective half-life of the
radionuclides. When deciding on the
discharge for a particular patient, the living conditions of the patient, such
as the extent to which he or
she can be isolated from other family members, in particular children and
pregnant women, should
also be considered. Safe management of the patient’s contaminated
excreta should be addressed.
Special consideration should be given to the case of incontinent patients.
In the case of carers and
comforters, the assumptions made for the calculations should be
consistent with the written
instructions that will be given at the time the patient is discharged from the
facility. Published data
would suggest that systematic dose monitoring is not necessary. For detailed
guidance on all aspects
pertaining to the release of patients see the Refs [18, 261].
4.245. As indicated in 4.244, the patient or legal guardian of the
patient should be provided with
written instructions on how to keep doses to members of the public and carers
and comforters as low
as reasonably achievable. Areas of particular concern are children and pregnant
partners of patients.
Detailed guidance, including sample information sheets, is given in the Refs
[18, 261].
Unintended and accidental medical exposures
Prevention of unintended and accidental medical exposures
4.246. The BSS, paragraphs 3.179 to 3.181, sets out requirements both for
minimizing the likelihood
of unintended and accidental medical exposures and for the ensuing
investigation if such exposures
occur. The pathways identified in the BSS for potentially leading to unintended
or accidental medical
exposures include flaws in the design of medical radiological
equipment, failures of medical
radiological equipment while in operation, failures and errors in software that
control or influence the
delivery of the radiation, and human error. General strategies for addressing
those pathways include
regular maintenance of medical radiological equipment and software, a
comprehensive quality
assurance programme, and continuing education and training of staff. The
lessons learned from events
that have occurred should be usedfor preventing or minimizing
unintended and accidental medical
exposures, as described below in para. 4.248.
4.247. A reduction in the probability of unintended or accidental
medical exposures in nuclear
medicine can be brought about by:
(f) The introduction of safety barriers at identified critical
points in the nuclear medicine
pathway, with specific quality control checks at these points. Quality control
is not confined
to physical tests or checks but can include such as actions as
double checks of the
radiopharmaceutical and activity to be administered, and correct identification
of the patient.
(g) Actively encouraging a culture of always working with awareness and
alertness.
(h) Providing detailed protocols and procedures for each process
in the nuclear medicine
pathway.
(i) Education and training, including continuous professional
development, of all staff involved
in providing nuclear medicine services.
(j) Clear definitions of the roles, responsibilities and functions of
staff in the nuclear medicine
facility that are understood by all staff.
4.248. Preventive measures should include incident and near incident reporting,
analysis and feedback
including lessons learnt from international experience [282]. Preventive
measures should also include
each nuclear medicine facility checking the robustness of their safety
system against reported
incidents. IAEA Safety Reports Series No. 17 [282] contains reviews
of case histories from an
extensive collection of accidental medical exposures, including
examples relevant to nuclear
medicine.
4.249. Building on the guidance from paras 4.247 to 4.248, the following
three-step strategy can help
to prevent unintended and accidental medical exposures in nuclear medicine:
(d) allocate responsibilities to appropriately qualified health
professionals only and ensure that a
quality management system is in place that includes radiation safety;
(e) use the lessons learned from unintended and accidental medical
exposures to test whether the
quality management system, including for radiation safety, is robust
enough against these
types of events;
(f) identify other latent risks by posing the question “what else could
go wrong” or “what other
potential hazards might be present” in a systematic, anticipative
manner to all steps in the
nuclear medicine process.
Investigation of unintended and accidental medical exposures
4.250. The events that constitute unintended or accidental medical exposures
are detailed in the BSS
paragraph 3.180, and for a nuclear medicine facility include those
associated with diagnostic
procedures and with radiopharmaceutical therapy. For diagnostic procedures,
reference should also be
made to Section 3 paras. 3.249 - 3.252, for aspects relating to X
ray imaging. Unintended and
accidental medical exposures may occur at any stage in the
nuclear medicine process. For
radiopharmaceutical therapy, unintended or accidental medical
exposures may be either
underexposures or overexposures. The events in the BSS paragraph 3.180 also
include ‘near misses’,
and these should be considered in the same way as actual events.
4.251. One of the events identified in the BSS paragraph 3.180 is when the
activity administered in
radiopharmaceutical therapy was “substantially different from (over or under)”
the prescribed dose.
Consensus recommendations regarding the level of activity difference
that would be considered as
substantially different appear to be lacking, but a pragmatic approach
for use within the nuclear
medicine facility might be deviations greater than 10%. The nuclear medicine
facility should have in
place a system with clear procedures for identifying when this type of event
occurs.
4.252. The BSS, paragraph 3.181, states what should be done in the course of
the investigation. This
includes calculation or estimation of patient doses, which should be performed
by a medical physicist.
A record of the calculation method and results should also be placed
in the patient file. When
required, counselling of the patient should be undertaken by someone with
appropriate experience and
clinical knowledge.
4.253. The investigation of unintended and accidental medical
exposures, as required by the BSS
paragraphs 3.180 – 3.181, has three main purposes. The first is to
assess the consequences for the
patient (s) affected and provide remedial and health care actions if
necessary. The second is to
establish what went wrong and how to prevent or minimize a
recurrence in the nuclear medicine
facility – i.e. the investigation is for the facility’s and their patients’
benefit. The third purpose is to
provide information to other persons or other nuclear medicine
facilities. Dissemination of
information about unintended and accidental medical exposures and
radiation injuries has greatly
contributed to improving methods for minimizing their occurrence. This
might be through the
regulatory body and/or the health authorities for more significant events or as
required by a country’s
regulations, where the regulatory body disseminates information on the
event and the corrective
actions to other facilities that might learn from the event. Independently from
any legal requirement
for reporting to the regulatory body, the implementation of voluntary and
anonymous safety reporting
and learning systems can significantly contribute to improve safety and safety
culture in health care.
This includes participation in voluntary international or national
databases designed as educative
tools, as is the case for image guided interventional procedures and radiation
therapy – see Sections 3
and 5, paras. 3.257 and 5.272, respectively.
4.254. As noted in 4.253, the BSS requires reporting to the regulatory body and
to the health authority
if appropriate for “significant” events. Further, the regulatory body
in a given country may also
specify their own requirements for reporting events to them. It is difficult to
quantify “significant” –
specifying a numerical trigger value immediately creates an artificial
distinction between values
immediately below that value (and hence should not be reported) and
those just above the value
(which should be reported). However, the attributes of “significant”
events can be elaborated, and
events with one or more of these attributes should be reported to the
regulatory body. Such attributes
would include the occurrence of, or the potential for, serious unintended or
unexpected health effects
due to radiation (in this case the health authority should be also informed),
the likelihood of a similar
event occurring in other nuclear medicine facilities, a large number
of patients were affected, and
gross misconduct or negligence by the responsible health professionals. As
stated in 4.253 above, one
of the roles of the regulatory body for such a reported event is to disseminate
information on the event
and the lessons learned to all potentially affected parties – typically other
nuclear medicine facilities
and relevant professional bodies, but also in some cases manufacturers,
suppliers, and maintenance
companies.
4.255. Irrespective of whether the event also reported to the regulatory body,
feedback to staff should
be provided in a timely fashion and, where changes are recommended, all staff
should be involved in
bringing about their implementation.
Records and review
Radiological review
4.256. The BSS in paragraph 3.182 requires the performance of a periodic
radiological review at the
nuclear medicine facility. This involves considering both justification
and optimization aspects of
radiation protection. For the latter, the results of the programme of
quality assurance for medical
exposures, including the periodic independent audit, would be a significant
input into the process. As
described in Section 2, paras. 2.146 – 2.147, the wider clinical audit could
include the radiological
review with its assessment of the application of the requirements for
justification and optimization in
the facility for the nuclear medicine procedures being performed [53].
4.257. To facilitate compliance with the BSS paragraph 3.182 and to learn from
periodic reviews, the
methodology used, the original physical, technical and clinical
parameters considered, and the
conclusions reached should be documented and taken into account prior to any
new review that may
result in an update of institutional policies.
4.258. Radiological reviews should consider changes in patient
management that result from the
diagnostic nuclear medicine procedures, and the effect of
introducing new technologies or
radiopharmaceuticals on efficiency and cost. In radiopharmaceutical therapy,
reviews should consider
patient outcome (survival, acute and late side effects, etc.), and
the effect of introducing new
radiopharmaceuticals on efficiency and cost. The nuclear medicine facility
should have a system for
the on-going collection of relevant data to support such reviews.
Records
4.259. Records should be in place to demonstrate on-going compliance
with radiation protection
requirements. The BSS, paragraphs 3.183 to 3.185, specify requirements
for keeping personnel
records, records of calibration, dosimetry and quality assurance, and
records of medical exposure.
These records should be kept for the period specified by the country’s
regulatory body. In the absence
of such a requirement, a suggested period for keeping records is 10
years. In the case of children,
records should be kept for a longer time.
RADIATION PROTECTION OF THE PUBLIC
Introduction
4.260. Public exposure may arise from the performance of nuclear medicine, to
persons in and around
the nuclear medicine facility, but also in the wider public domain. The latter
can occur as a result of
the release from the nuclear medicine facility of patients with some remaining
radioactivity. Persons
who may be so exposed fall into three categories – those who work at the
nuclear medicine facility
but not in a role that is directly involved in the use of radiation, those who
are visitors to the facility,
and the wider public. Radiation exposure of carers and comforters
while performing that role is
considered occupational exposure and not public exposure and is not
covered by this section. See
paras. 4.232 – 4.236 for guidance on carers and comforters. In addition there
is the possibility, albeit
low, of public exposure from pathways associated with the discharge of
radioactive waste.
4.261. The requirements of the BSS for public protection, paragraphs 3.117 to
3.123, 3.125 to 3.127,
and 3.135 to 3.137, apply to nuclear medicine facilities. This
sub-section contains guidance very
specific to nuclear medicine facilities. For more general and
comprehensive guidance on radiation
protection of the public, reference should be made to the IAEA Safety Guide
Radiation Protection of
the Public [24].
Non-occupationally exposed workers and visitors
4.262. Non-occupationally exposed workers are those persons who work
at the nuclear medicine
facility but not in a role that is directly involved in the use of
radiation, for example non-nuclear
medicine ward nurses, clerical staff, and cleaning personnel. It also includes
those persons who work
at the wider medical facility where the nuclear medicine facility is
located. These persons are to be
afforded the same level of radiation protection as any member of the public, as
required by the BSS
paragraph 3.78.
4.263. Visitors to the nuclear medicine facility include those persons who will
be undergoing nuclear
medicine procedures, for the time during the visit when their treatment or
diagnostic procedure is not
taking place – for example, while they are sitting in the “cold” waiting room.
Similarly for carers and
comforters – any exposure other than that arising from the nuclear medicine
procedure with which
they are helping will be public exposure. Other visitors, including
persons delivering goods or
supplies, sales personnel, accompanying persons and escorts, and other patients
in the facility, are also
considered members of the public.
External exposure and contamination
4.264. The primary means for protecting the public (non-occupationally exposed
workers and visitors)
from external exposure is to ensure that the shielding of the nuclear medicine
facility (see paras. 4.31
– 4.35) is sufficient so that public exposure resulting from being in
any immediate adjacent areas,
including rooms above and below, accessible by either
non-occupationally exposed workers or
visitors would be in compliance with the public dose limits, and
preferably less than any dose
constraint that the regulatory body may have applied (see Section 2 paras. 2.16
and 2.46.)
4.265. Patients that have been administered radiopharmaceuticals may expose
members of the public
in the nuclear medicine facility and upon release. For the latter situation
(release) see paras. 4.243 to
4.245. In the nuclear medicine facility, the RPO should establish rules to
ensure that the exposure of
any member of the public will be less than the public dose limit
and, preferably, lower than any
applicable dose constraint. At the design stage of the nuclear medicine
facility, consideration should
given to the respective flow of patients and visitors in the facility so that
their contact or proximity is
minimized, thereby reducing the potential for both external exposure and spread
of contamination.
Control of access
4.266. Following adequate shielding, access to areas where radiation is being
used should be control
to provide forcontrolling doses to visitors and non-occupationally exposed
workers. This is effective
against both external exposure and contamination. Visitors should not be
allowed to enter controlled
areas, in particular radiopharmaceutical therapy wards. Exceptionally, a
visitor may enter, but he/she
should be accompanied to some controlled areas by a staff member who knows the
protection and
safety measures for the controlled area. The nuclear medicine facility should
have written procedures
specifying when such exceptions can take place and who may do the
accompanying. Similarly, the
facility should have established the rules regarding non-occupationally exposed
workers, especially
regarding access to controlled and supervised areas. Particular attention, in
all cases, should be made
with respect to potentially pregnant women.
4.267. Controlled and supervised areas should be clearly identified to help
prevent inadvertent entry.
This includes areas such as toilets designated for nuclear medicine patients.
Further control can be
afforded by the use of keys (or passwords) to restrict access to
the control panels of medical
radiological equipment to only authorized persons.
Members of the public in the wider public domain
4.268. Usually there are no restrictions with respect to public exposure for
the release of patients that
have undergone diagnostic nuclear medicine procedures. Patients should be
advised on measures to
enhance elimination of the residual radioactivity (such as hydration
and frequent voiding) and to
avoid prolonged contact with sensitive members of the public (young children,
pregnant women), if
appropriate.
4.269. The exposure of other persons, in the wider public domain, by
patients who have received
radiopharmaceutical therapy can occur through external irradiation of
persons close to the patient,
such as on public transport, and through internal contamination of persons as a
result of excreted or
exhaled radionuclides. The RPO of the nuclear medicine facility should
establish rules to ensure that
the exposure of any member of the public, following release of a
radiopharmaceutical therapy patient,
will be less than the public dose limit and, preferably, lower than any
applicable dose constraint. As
stated in para. 4.245, the patient should have been given written instructions
that include means for
avoiding external and internal exposure of the public. An acceptable
method to estimate the
acceptable retained activity for patients being discharged is described in
para. 4.244. Results of the
calculations should be recorded. When deciding on the appropriate discharge
activity for a particular
patient, the licensee and RPO should take into account the transport and the
living conditions of the
patient, such as the extent to which the patient can be isolated from other
family members and the safe
management of the patient’s excreta and body fluids. Detailed
guidance on release of
radiopharmaceutical therapy patients and public radiation protection is given
in Refs [18, 261].
Death of a patient who has undergone a nuclear medicine procedure
4.270. Precautions may be required after the death of a patient to whom
radiopharmaceuticals have
been administered, particularly in the case of radiopharmaceutical
therapy. This applies to the
immediate handling of the dead patient, both in the hospital and in home or
other place, but also with
respect to autopsy, embalming, burial or cremation. The radiation
protection precautions should be
determined by the RPO, based on a generic safety assessment of the need for
monitoring personnel
who carry out these procedures, the need for monitoring the premises and the
need for minimizing
external radiation exposure and the potential for contamination. In addition to
whole body monitoring,
finger monitoring may be required for autopsy and embalming personnel,
as contamination and
radioactive waste are likely to be generated. A particular problem is the
cremation of patients injected
with bone-seeking radiopharmaceuticals such as ⁸⁹Sr for pain management
of skeletal metastases.
Because of the relatively long half-life of this radionuclide (50 days), the
crematorium should store
the ashes until adequate decay is achieved before releasing them to the family.
Detailed guidance is
given in Refs [18, 261]. Other considerations, e.g. cultural o legal,
may prevail over the radiation
protection considerations.
Radioactive waste
4.271. Another potential pathway for public exposure is radioactive waste and
hence there should be
systems and procedures in place to manage radioactive waste and its discharge,
as stated in the BSS,
requirement 31 and paragraphs 3.131 to 3.134. Detailed guidance on the
management of radioactive
waste, applicable to nuclear medicine facilities, is given in Ref [283].
4.272. Most radioactive waste from nuclear medicine is short lived
radionuclides, and it is feasible to
consider them as non-radioactive waste either directly or after some
time for decay. A formal
mechanism should be in place, including rigorous control measures, to
demonstrate compliance with
regulatory requirements in respect of release of radioactive waste
from regulatory control. Further
guidance is given in Ref [283].4.273. Since waiting for decay until the
waste meets the regulatory
criteria for clearance or authorized discharge is an essential method in
nuclear medicine, a room for
interim storage of radioactive waste should be available. The room should be
locked, properly marked
and ventilated. Records should be kept from which the origin of the
waste can be identified. The
process requires grouping (segregation) radionuclides according to the expected
time for their decay
(initial activity and physical half-life) and their physical form. Examples of
different physical forms
include: vials that may contain residual radioactivity,
biological waste which may undergo
decomposition, infectious waste requiring sterilization, broken
glassware, syringes, and needles
requiring collection in separate containers to prevent personnel being injured,
radionuclide generators,
bed linen and clothing from hospital wards (therapeutic applications),
and liquid scintillation
solutions. Containers to allow segregation of different types of radioactive
waste should be available
in areas where the waste is generated. The containers should be
suitable for their purpose (for
example, in volume, shielding and leak tightness).
4.274. In practice, it is mainly ¹³¹I and the waste from
radiopharmaceutical therapy patients that
require special precautions. Appropriate storage of radioactive
materials to allow for decay will
minimize the environmental impact of the release. The majority of diagnostic
studies are performed
using ⁹⁹ᵐTc, which has a physical half-life of 6 h. Following storage of 2.5
days (10 half-lives, i.e. a
decay of a factor of more than 1000) most of this waste can be treated
conventionally. The technetium
generators contain ⁹⁹Mo with a half-life of 2.75 days; depending on their
initial activity, the decay
time at the nuclear medicine facility is 1.5–2 months.
4.275. 18F is the most commonly used radionuclide in PET. The short
physical half-life of 110
minutes generally allows discharge within 24 hours.
4.276. Management of radioactive waste containing longer-lived radionuclides
should consider initial
activity and half-life. The nuclear medicine facility’s RPO should give advice
in these situations.
4.277. Following the above considerations, a summary of practical advice for
specific situations in
nuclear medicine can be given:
(a) Technetium generators. There are two options: (1) returning to the
supplier after use, ensuring
compliance with regulations for transport of radioactive materials (see paras.
4.298 – 4.300),
and (2) waiting for decay. After 1.5–2 months, the generator can
be dismantled and the
elution column removed, as the material is considered non-radioactive. Labels
should then be
removed.
(b) Used syringes and needles. These can be collected in a shielded container
in the rooms used
for preparation and injection of radiopharmaceuticals. When the container is
full, it should be
sealed and the expected date of release from regulatory control be marked on
it. After this
time, the external dose rate can be monitored. The container can be released
from regulatory
controlwhen the external ambient dose equivalent rate is the same as the
background or in line
with national/local regulations.
(c) Vials containing residues of ⁹⁹ᵐTc, ⁶⁷Ga, ¹¹¹In and ²⁰¹Tl. A
similar procedure should be
established as for the syringes, but segregation based on the
physical half-life is needed.
Caution should be exercised in storing waste containing very low
levels of longer lived
residues such as ⁶⁸Ge (half-life 271 days) as these could over time accumulate
to activities
where they should be considered as radioactive waste and could
require prolonged storage
before release from regulatory control.
(d) Gloves and cover paper. These should be collected in plastic
bags in the rooms used for
preparation and injection of radiopharmaceuticals. When a bag is filled, it
should be sealed.
After waiting for decay or with appropriate monitoring, they can be released
from control and
treated as ordinary waste.
(e) Sealed sources for calibration of activity meters, quality control
of gamma cameras and
counters, and anatomical marking of images are released from
regulatory controlas
determined by the RPO in accordance with national regulations and
authorization by the
regulatory body (clearance).
(f) Small activities of ³H and ¹⁴C in organic solutions can usually be treated
as non-radioactive
waste. In certain instances, because of their potential toxicity, special
precautions may apply,
and appropriate bio-hazard precautions taken.
(g) Patients’ excreta, such as urine with ¹³¹I. For diagnostic
patients there is no need for
collection of excreta and ordinary toilets can be used. For therapy patients,
policies vary for
different countries, but in principle follow the dilution and decay
methodologies to meet
national or local requirements. Some precautions may be required where sewerage
systems
allow rapid processing of effluent with subsequent usage for
irrigation of land used for
growing vegetables.
Monitoring and reporting
4.278. The BSS, requirement 32 and paragraph 3.137, sets out the requirements
that should be met by
the nuclear medicine facility with respect to monitoring and
reporting. In the nuclear medicine
facility, procedures should be in place to ensure that:
(c) The requirements regarding public exposure are satisfied and such exposure
is assessed;
(d) The requirements regarding discharge of radioactive materials to
the environment are
satisfied;
(e) Appropriate records of the results of the monitoring programmes are kept.
4.279. The programme for monitoring public exposure arising from nuclear
medicine should include
dose assessment in the areas in and surrounding the nuclear medicine facility,
which are accessible to
the public. This can be achieved from the shielding calculations in the
planning stage, combined with
area monitoring and contamination monitoring at the initial operation of the
facility and periodically
thereafter. Records of these assessments should be kept for typically 7-10
years, but in any case for a
period that meets any relevant regulatory requirements.
PREVENTION AND MITIGATION OF ACCIDENTS
Safety assessments
4.280.To comply with the BSS requirements for safety assessment (BSS,
paragraphs 3.29 to 3.36), ,
the the registrant or licensee should conduct a safety assessment applied to
all stages of the design and
operation of the nuclear medicine facility. The safety assessment report
should be submitted to the
regulatory body if required. Basically, the safety assessment deals
with determining ‘what can go
wrong’ and how it can be prevented and, in case it occurs, how it can be
mitigated. Section 2, paras
2.148 – 2.152, describes general considerations for facilities using
ionizing radiation for medical
purposes.
4.281. The safety assessment should be systematic and contain
information on identification of
possible events leading to accidental exposure (see Appendix I for a summary of
typical causes and
contributing factors to accidental exposures in nuclear medicine). The safety
assessment should not
only cover these events, but also aim at anticipating other events
that have not previously been
reported. Clearly the safety assessment should be documented.
4.282. The safety assessment should be revised when:
(d) New or modified radiopharmaceuticals, equipment, or their accessories are
introduced;
(e) Operational changes occur, including workload;
(f) Operational experience or information on accidents or errors
indicates that the safety
assessment is to be reviewed.
4.283. Safety assessments in nuclear medicine include consideration of
all the steps in the use of
radiopharmaceuticals for diagnosis and treatment in the nuclear medicine
facility. The steps include
ordering, transport and receipt of radiopharmaceuticals, unpacking,
storage, preparation and
administration of the radiopharmaceuticals to the patient, examination or
treatment, care of therapy
patients with high amounts of radioactivity, and storage and handling of
radioactive waste.
Accident prevention
4.284. Accident prevention is clearly the best means for avoiding
potential exposure and the BSS,
paragraphs 3.39 to 3.41, set out requirements based on good engineering
practice, defence in depth,
and facility-based arrangements, to achieve this. Design considerations
for the nuclear medicine
facility, medical radiological equipment and ancillary equipment are described
in paras 4.7 to 4.58.
4.285. The licensee should incorporate:
(a) Defence in depth measures to cope with identified events, and evaluation of
the reliability of the
safety systems (including administrative and operational procedures, equipment
and facility design).
For example, theft of sources can be minimized through multiple layers of
security including having
sources locked up in a safe within a locked room, in an area that has a
restricted access with camera
surveillance and is routinely patrolled.
(b) Operational experience and lessons learned from accidents and errors. This
information should be
incorporated into the training, maintenance and quality assurance programmes.
Mitigation and contingency plans
4.286. On the basis of events identified by the safety assessment for
the nuclear medicine facility,
contingency plans should be prepared for events associated with
potential exposure, including
allocation of responsibilities and resources, the development and
implementation of procedures, and
the provision of training and periodic retraining of the relevant
staff in executing the mitigation
measures. Contingency plans in a nuclear medicine facility should, as
a minimum, cover the
following:
(a) Predictable incidents and accidents, and measures to deal with them;
(b) The persons responsible for taking actions, with full contact details;
(c) The responsibilities of individual personnel in an accident or
emergency procedures (for
example, nuclear medicine physicians, medical physicists, nuclear medicine
technologists, the
RPO);
(d) Equipment and tools necessary to carry out the emergency procedures;
(e) Training and periodic rehearsals;
(f) Recording and reporting systems;
(g) Immediate measures to avoid unnecessary radiation doses to patients, staff
and the public;
(h) Measures to prevent access of persons to the affected area;
(i) Measures to prevent spread of contamination, including leakage from fume
hoods and room
ventilation systems.
4.287. Kits should be kept readily available for use in a nuclear medicine
incident. These may include
the following:
(a) Protective clothing, for example overshoes and gloves;
(b) Decontamination materials for the affected areas, including absorbent
materials for wiping up
spills;
(c) Decontamination materials for persons;
(d) Warning notices and barrier tape;
(e) Portable monitoring equipment;
(f) Bags for waste, tape, labels and pencils.
4.288. The exposure of workers involved in such nuclear medicine events cannot
be considered an
unexpected exposure and whether deliberate or not should be
controlled, and the dose limits for
workers in planned exposure situations should apply.
4.289. For medical exposure, potential exposure when it does occur is manifest
as an unintended or
accidental medical exposure. Means for preventing or minimizing unintended and
accidental medical
exposures in nuclear medicine are described in para. 4.246 – 4.249, and the
ensuing investigation and
corrective actions are described in paras, 4.250 – 4.255.
4.290. The BSS, paragraphs 3.43 to 3.44, sets out the requirements for
emergency preparedness and
response. It is unlikely that emergency exposure situation can arise from
nuclear medicine practice,
but if the safety assessment identified the need, an emergency plan should be
prepared according to
the requirements in the GSR Part 7 and guidelines by the regulatory authority
[7].
Lost sources
4.291. It is critical for this type of event that an up-to-date inventory
exists (see para. 4.55) so that it
can be determined immediately when a source is missing, what its type and
activity are, when and
where it was last known to be, and who last took possession of it.
A proactive attitude is
recommended for the case that sources are ordered and not received at the
expected time. Making a
check for the arrival of a source at the expected receipt time should be part
of the procedures. The
actions to be part of the contingency plans in this case include:
(a) Obtain assistance from the RPO when needed.
(b) Conduct a local search.
(c) Check and ensure security and control of the other sources if a
theft in the facility is
suspected.
(d) If the source is not found, call the supply company and inform them of the
loss so that they
can trace the shipment and find out where the radioactive material is.
(e) If not found, report the loss of the material according to the
rules given by the regulatory
body.
Damage to radionuclide generators
4.292. Radionuclide generators, such as for ⁸²Rb, ⁹⁹ᵐTc and ⁶⁸Ga, contain a
relatively large amount of
activity. In the event of a generator being damaged, the measures to be taken
should include:
(a) Evacuate the area immediately and institute measures to prevent entry to
the area.
(b) Inform the RPO, who should confirm the spillage, define the safety
boundaries and supervise
the decontamination and monitoring procedures, including when restrictions to
enter the area
can be lifted.
(c) Record the event and make a report according to the rules given by the
regulatory body.
Spillage of small amounts of radioactivity
4.293. After a spillage of a small amount of radioactivity, for
example low volumes of non-toxic
radiopharmaceuticals which are easily removed, such as 10 MBq of ⁹⁹ᵐTc or ¹⁸F,
the following actions
should be taken:
(a) Use protective clothing and disposable gloves.
(b) Quickly blot the spill with an absorbent pad to keep it from spreading.
(c) Remove the pad from the spill.
(d) Wipe with a tissue from the edge of the contaminated area towards the
centre.
(e) Monitor the tissue for residual activity, for example using a
contamination monitor or
performing a wipe test.
(f) Continue the cycle of cleaning and monitoring until the measurements
indicate that the spill
has been removed, trying to keep the volume of contaminated waste as small as
possible. In
some cases, such as with short-lived radionuclides, it may be simpler to
“quarantine” the area
for sufficient time to allow decay – e.g. cover the spill site, such as with a
laboratory coat, and
prevent access to the area.
(g) Use a plastic bag to hold contaminated items. Suitable bags should be
always available, as
well as damp paper towels.
(h) If the decontamination process is not successful, the RPO should be
contacted.
(i) Monitor all people involved in the spill for contamination when leaving
the room, particularly
the monitoring of shoes if the spill was on the floor.
Spillage of large amounts of radioactivity
4.294. After a spillage of a large amount of radioactivity, for
example if a patient undergoing ¹³¹I
therapy vomits shortly after administration, the following actions should be
taken:
(a) Throw absorbent pads over the spill to prevent further spread of
contamination.
(b) All people not involved in the spill should leave the area immediately.
(c) The RPO should immediately be informed and directly supervise the
clean-up.
(d) Monitor all people involved in the spill for contamination when leaving
the room.
(e) If clothing is contaminated, remove and place it in a plastic bag labelled
‘RADIOACTIVE’.
(f) If contamination of skin occurs, wash the area immediately.
(g) If contamination of an eye occurs, flush with large quantities of water.
(h) When the contamination is contained, the procedures outlined for cleaning
small spills may
be followed, taking particular care that the contaminated waste bags are
appropriately labelled
and stored.
Medical emergencies involving patients who have received therapeutic
radiopharmaceuticals
4.295. There may be medical emergencies, such as in the case of a stroke or
cardiac arrest, involving
immediate care of patients who have been administered large amounts of
radioactive material (e.g. of
the order of several GBq of ¹³¹I) for radiopharmaceutical therapy. In these
cases dose rates near the
patient are high, and attendant medical personnel may receive significant
doses. However, the dose
will be acceptable because the procedure is lifesaving. Measures should
be used to minimize such
doses. All members of the emergency team should wear impermeable protective
gloves. Medical staff
should be informed and trained on how to deal with such patients.
Rehearsals of the procedures
should be held periodically.
Need for urgent patient attention, including surgery
4.296. Radiation protection considerations should not prevent or delay
lifesaving operations in the
event that surgery is required on a patient who has been
administered radiopharmaceuticals. The
following precautions should be observed:
(a) Notify the operating room staff.
(b) Modify operating procedures under the supervision of the RPO to
minimize exposure and
spread of contamination.
(c) Use protective equipment as long as efficiency and speed are not affected.
(d) Rotate personnel as needed if the surgical procedure is lengthy.
(e) Measure personnel doses.
Fires, earthquakes and other disasters affecting the nuclear medicine facility
4.297. The normal facility drill should be observed, providing for safe
evacuation of patients, visitors
and staff. When the first responders (for example the fire brigade) attend,
they should be informed of
the presence of radioactive material. No one, other than emergency responders,
should re-enter the
building until it has been checked for contamination by the RPO or by the
radiation safety staff of the
agency in charge of emergency response.
SAFETY IN THE TRANSPORT OF RADIOACTIVE MATERIALS
4.298. The BSS paragraph 2.25 sets out the requirements for the transport of
radioactive materials,
invoking in particular the IAEA Regulations for the Safe Transport of
Radioactive material [283]. The
IAEA Regulations for the Safe Transport of Radioactive material use the defined
terms “consignor” to
mean any person, organization or government that prepares a
consignment for transport, and
“consignee” to mean any person, organization or government that is
entitled to take delivery of a
consignment. “Consignment” is also a defined term, meaning any package or
packages, or load of
radioactive material, presented by a consignor for transport.
4.299. The licensee of a nuclear medicine facility may be both a consignee and
a consignor, and hence
may have responsibilities for both receipt and shipment of
radioactive materials. Receipt of
radioactive materials will be a regular occurrence for all nuclear medicine
facilities. Shipments may
take place when the facility has a cyclotron or laboratory that sends
radiopharmaceuticals to other
sites, or when expired generators, old sealed calibration sources or
radioactive liquids (for example
¹⁴C solutions) should be returned to the supplier or disposed off-site, as
applicable.
4.300. The detailed requirements for the safe transport of radioactive
material, including general
provisions, activity limits and classification, requirements and controls for
transport, requirements for
radioactive material and for packagings and packages, test
procedures, and approval and
administrative requirements, are given in the IAEA Regulations for the Safe
Transport of Radioactive
material [284]. Emergency arrangements during the transport of
radioactive material should be in
place, in line with the requirements in the GSR Part 7 and guidelines by the
regulatory authority [7].
The licensee and the RPO of the nuclear medicine facility should be familiar
with these regulations to
ensure that their transport of radioactive materials will be in compliance.
5. SPECIFIC RECOMMENDATIONS FOR RADIATION PROTECTION AND
SAFETY IN RADIATION THERAPY
INTRODUCTION
5.1. This chapter covers radiation therapy, the branch of clinical medicine
that uses ionizing radiation
(teletherapy and brachytherapy), either alone or in combination with
other modalities, for the
treatment of patients with malignancies or other diseases. It includes
responsibility for the treatment
decision, treatment preparation and planning, treatment delivery, follow-up and
supportive care of the
patient as an integral part of the multidisciplinary management of patients.
Treatment using unsealed
sources is covered in Section 4. Imaging studies used in treatment preparation,
planning, verification
and delivery are covered in Section 3, with appropriate cross-references.
5.2. External beam radiation therapy (teletherapy) is mainly performed
with linear accelerators
(Linacs), superficial and orthovoltage units, tomotherapy units
incorporating a CT, or radioactive
source based equipment (primarily using cobalt-60). In the case of Linacs, both
photons and electrons
are widely used and, for each, a range of energies can be utilized. External
beam radiation therapy can
be delivered using a wide range of techniques including: 2D, 3D conformal
radiotherapy (3DCRT),
intensity modulated radiotherapy (IMRT), stereotactic radiosurgery (SRS),
stereotactic radiotherapy
(SRT), volumetric-modulated arc therapy (VMAT), Rapid Arc, and
intraoperative radiotherapy
(IORT). More recently, proton and heavy ion therapy has become available in
some facilities.
5.3. Brachytherapy can be performed by implanting radioactive sources
directly into the patient or
using after-loading devices that allow for the sources – either manually or
remotely – to be placed into
catheters that then are inserted in the body. Techniques can be
interstitial, intracavitory, surface or
intraoperative and a range of sources are used. Low dose rate (LDR), medium
dose rate (MDR), high
dose rate (HDR) and pulsed dose rate (PDR) brachytherapy techniques are used.
5.4. The generic term “medical radiation facility” is used widely in Section 2
to mean any medical
facility where radiological procedures are performed. In Section 5,
the narrower term “radiation
therapy facility” is used to cover any medical radiation facility
where radiation therapy is being
performed. A radiation therapy facility may be a radiation therapy department
inside a wider hospital
or medical centre, or it may be a stand-alone facility.
5.5. The defined term “radiological procedure” is used in the BSS to cover all
imaging and therapeutic
procedures using ionizing radiation. In a radiation therapy facility
both imaging and therapeutic
radiological procedures occur, and this should be borne in mind when reading
the following guidance
in Section 5. In cases where the guidance is specific to one of either imaging
or treatment, additional
qualifiers, such as “imaging” or “treatment”, are used.
5.6. The health professionals that can take on the role of the
radiological medical practitioner in
radiation therapy depend inter alia on national laws and regulations.
Most typically this will be a
radiation oncologist, but may also include other specialists – for example,
neurosurgeons in the case
of stereotactic radiosurgery.
5.7. Section 2 of this Safety Guide provides general guidance on the
framework for radiation
protection and safety in medical uses of radiation, including roles
and responsibilities, education,
training, qualification and competence, and the management system for
protection and safety. This is
relevant to radiation therapy and reference to Section 2 should be made as
indicated or needed.
SAFETY OF MEDICAL RADIATION FACILITIES AND MEDICAL RADIOLOGICAL
EQUIPMENT
Radiation therapy facilities
Location and site
5.8. A radiation therapy facility should be located on a site that gives ready
access for in-patients and
outpatients, and that at the same time makes fulfilling radiation protection
requirements as simple as
possible. Operational efficiency, initial cost, as well as provision
for future expansion and/or
increased workload, should be considered when locating a new radiation therapy
facility. Radiation
therapy facilities are often located on the periphery of the hospital
complex to minimize radiation
protection problems arising from treatment rooms being adjacent to high
occupancy areas. The option
of being able to construct rooms below ground level, with the
potential for a reduced need for
substantial shielding, may also influence the choice of site. Further guidance
on location and site of
radiation therapy facilities is given in Refs [285 – 288].
5.9. In addition to on-site considerations, surrounding environment should be
also considered. This
includes presence of, and implications for, adjacent residential or
industrial areas, and the level of
general public access to, and use of, the area. This relates to ensuring that
public exposure outside –
and above and below if there are occupied areas – the radiation
therapy facility is consistent with
public exposure requirements.
5.10. When considering expansion of an existing radiation therapy facility
consideration should be
given to the areas beside, above and below the proposed expansion site.
5.11. For physical security purposes, radiation therapy facilities
using sealed radioactive sources
should be located in areas where access by members of the public to the rooms
where sources are
used and stored can be restricted.
Design of rooms within the radiation therapy facility – general considerations
5.12. A typical radiation therapy facility consists of 5 main
functional areas: reception, clinical
consulting areas, external beam radiation therapy, brachytherapy,
and imaging and treatment
planning. Within these areas there are several types of rooms and,
depending on the treatment
modalities being provided, may include rooms or areas for patient
imaging, treatment simulation,
treatment planning, treatment control, treatment delivery, mould preparation,
and patient examination,
as well as patient changing cubicles, public waiting rooms, operating
theatres and source storage and
preparation rooms. Provisions for the incorporation of radiation safety
features into these areas and
rooms are best made at the facility design stage. (See para. 2.73). Because
the structural shielding of
radiotherapy facilities is very heavy, care should be taken that the shielding
weight can be supported
by the building structure, especially in cases when machines are
replaced by higher energy ones, such
as is the case of a cobalt-60 unit being replaced by a linear accelerator. The
layout should take into
account workload and staff and patient flow, both within the radiation therapy
facility and, in cases
where the radiation therapy facility is part of a larger hospital or
medical centre, with other
departments and wards of the wider facility. Wherever possible,
treatment rooms should be
surrounded with rooms that have low or controlled occupancy. Physical signage
giving information
on where different areas are located and designating hazardous areas is
beneficial here and should be
preferably in both word and picture format. Colour coding of
different areas is also very helpful.
General guidance on radiation therapy facility design is given in Refs. [285 –
289].
5.13. The three factors relevant to dose reduction for workers and the public
(namely, time, distance
and shielding) should be combined in the design to optimize
occupational and public radiation
protection.
5.14. Access to the radiation therapy facility and its treatment,
imaging, consultation and patient
preparation rooms should be considered. This includes provision for the
delivery of equipment and for
ease of access for patients undergoing clinical assessment and daily
treatment. Patients may arrive in
wheelchairs or on trolleys or beds.
5.15. As a general rule, the design of the radiation therapy facility should
make provisions for safety
systems or devices associated with the equipment and room. This
includes ventilation, electrical
wiring related to emergency ‘off’ switches, as well as standby lighting, safety
interlocks and warning
signals.
5.16. A reliable and stable power supply should be available for all
modern equipment and
information technology systems. An emergency diesel generator alone is
generally not sufficiently
stable to power a linear accelerator or orthovoltage unit and should
not be used in this way.
Uninterruptible power supply (UPS) or battery backup systems should be
installed to capture the
active information at the time of an outage and to power down all software in a
controlled manner.
Servers should be programmed to automatically shut down when the
power supply is interrupted.
Diesel generators could be used to run systems that depend only on
timers, such as in the case of
cobalt-60 teletherapy units.
5.17. The design of the facility should include an air conditioning system
sufficient to maintain the
temperature in the treatment room within the parameters defined by the
equipment manufacturers.
5.18. Room lights should be dimmable so that the alignment lasers and the field
defining lights can be
seen easily to facilitate patient set up. It is useful to be able to control
the treatment and imaging room
lights and lasers from the control pendant in the respective room. When the
field light is switched on,
the room lights should dim to a pre-set (but adjustable) level, and the
alignment lasers should also be
switched on. Since fluorescent lights do not dim very satisfactorily,
it is recommended that
incandescent lights are used for the dim level. Four alignment lasers are
recommended. Three lasers
projecting across: two aligned with the gantry positions of 90° and
270°, and one mounted in the
ceiling directly above the isocentre. A fourth laser should project a sagittal
line along the gantry axis.
This laser is usually mounted on an angled bracket on the wall
opposite the gantry. The laser
switching should be controlled from the hand pendant, but it is also useful to
be able to switch them
off independently for quality control tests.
5.19. In addition to interlocks, as described in para 5.29, signs and warning
lights should be placed at
the entrances of controlled areas to prevent inadvertent entry (see also para.
5.287 on control of public
access). The BSS, paragraph 3.90, requires the use of the symbol of the
International Organization for
Standardization (ISO) [59]. An illuminated warning sign should be displayed at
the entrance to the
maze or treatment room as well as several inside the treatment room. It should
be possible to see a
warning sign from any position within the treatment room. These signs should be
interlocked with the
treatment unit control. The illuminated signs may have two or three stages. For
a two stage sign, the
first stage will be illuminated when there is power to the treatment unit,
and the second stage will
illuminate when the beam or the source is on. For a three stage sign, stage one
will be illuminated
when there is power to the treatment unit, stage two will light when the
treatment unit is programmed
to deliver a radiation beam and stage three will illuminate when the beam or
the source is on. Other
possibility is that the warning lights flash when the beam is on. Other rooms
which are also controlled
areas, such as imaging, simulator and source storage rooms, also should have
appropriate signs and
warning lights.
5.20. Radiation therapy facilities that use radioactive sources should
implement technical measures so
that unauthorized access to sources can be detected in a timely fashion,
including afterhours. These
technical measures should be independent of any interlocks that terminate the
radiation beam during
normal operation. Such measures could include a video camera that
provides continuous remote
surveillance of the device, a photoelectric beam or motion detector
system installed in the maze
and/or treatment room, or a door interlock. If these devices indicate
the potential presence of an
unauthorized person, an alarm should indicate this locally and remotely so that
personnel can respond
in a timely fashion. Further guidance on security provisions for teletherapy
sources, and HDR, PDR,
MDR and LDR brachytherapy sources is given in Ref. [286]. See also para. 5.84.
5.21. Firefighting equipment should be available in all areas. For example, in
brachytherapy this is in
order to preserve the integrity of radioactive sources in the event
of a fire. Further guidance is
available in Ref [287].
Design of rooms within the radiation therapy facility – treatment rooms for
external beam radiation
therapy and afterloading brachytherapy
5.22. External beam radiation therapy and HDR/PDR brachytherapy should be
carried out within the
radiation therapy facility in treatment rooms designed for that purpose.
5.23. A shielded treatment room should not be shared between HDR brachytherapy
and external beam
radiation therapy, as this can negatively influence procedure flow and
efficiency. Further guidance is
given in Ref [285].
5.24. The size of the treatment room will depend on many factors, including the
treatment equipment
and the intended techniques of the treatments. The room should be
large enough to allow full
extension of the couch in any direction, with sufficient space for staff to
walk around it. The design
should also take account of the need for larger treatment rooms to allow for
specific procedures. For
example, total body irradiation (TBI) will require a larger treatment
distance to one wall; IORT
procedures require additional support staff and equipment, and the
room may need to be larger.
Imaging systems for image guided radiation therapy (IGRT), especially
CT-on-rails, also need extra
space. Easy access for patients on a bed or trolley, correct storage of
accessory equipment such as
electron applicators or breast positioning boards, and ease of patient
positioning and staff movement
during the setup procedures may be better facilitated with a larger
room. Careful placement of
accessory equipment within the room can help minimize the walking distance for
each patient set up.
Further guidance is given in Refs [285 – 287].
5.25. Care should be taken when a new machine or unit is to be introduced into
an existing treatment
room or bunker. The room size and shielding specification should be
consistent with the new
equipment and practices. This can be particularly relevant in the case of IMRT
or the installation of a
non-isocentric unit, for instance.
5.26. Some current or future equipment integrations, such as MRI/Cobalt/ MRI or
MRI/Linac/ MRI,
may have particular requirements that should be considered in the
room design to ensure both
efficient and effective operation and radiation safety.
5.27. The treatment and imaging room designs should include an open access
conduit for the control
panel, and monitoring and dosimetry equipment cables. This duct should
not be aligned with the
primary beam.
5.28. Entrance to the treatment room may be through a shielded door or via a
maze or a combination
of both. A maze reduces the need for a heavy shielded door and provides a route
for ventilation ducts
and electrical conduits without compromising the shielding. However a
maze requires more room.
More guidance on mazes and entrances is given in Refs [286, 287, 289].
5.29. Access to the treatment room should be furnished with a visible signal
indicating whether the
radiation source is ‘on’ or ‘off’. An interlock barrier to prevent
unauthorized access should be
provided. This could include a light beam or a physical barrier such as a gate
or door. Preferably two
such interlock barriers should be in place. The interruption of irradiation
should be maintained until
the interlock is reset after it has been verified that no person but the
patient is inside the room and that
the patient setup has not changed. After an interruption, provided no
operating parameters are
changed or reselected, it should be possible to resume irradiation,
but only from the equipment’s
control panel. See also para. 5.68.
5.30. The design should be such that access to the treatment (and imaging)
rooms should be visible to
the operators at all times. Further, the controls should be installed in such a
way that access to the
treatment room can be monitored at all times.
5.31. A safety system, such as the ‘last man out button’, should be in place to
ensure that all staff have
left the room prior to treatment commencement.
5.32. One or more emergency off switches should be conveniently placed inside
the treatment room,
in addition to those on the control panel and the equipment itself,
to allow interruption of the
irradiation from inside the treatment room.
5.33. Adequate systems, audio-visual devices or other means should be provided
to allow staff to have
communication with and a clear and full view of the patient. Oral communication
from the control
panel should be possible with the patient in the treatment (and imaging) room
using an intercom or
other communication system.
5.34. When using sealed sources a power fail-safe radiation area
monitor (audio-visual) should be
visible upon entering the room.
5.35. Provision should be made in each treatment room to enable the safe
removal of the patient in the
event of a power outage (e.g. availability of flashlights or torches). This
also means manual operation
of heavy doors should be possible.
5.36. Enclosed patient changing cubicles should not be located within the
treatment room.
Design of rooms within the radiation therapy facility – rooms for manual
brachytherapy
5.37. Typical radiation safety features for rooms used for the
storage and preparation of sealed
radioactive sources for manual brachytherapy include:
(a) The room should be provided with a lockable door to
control access and maintain source
security (see also paras. 5.11 and 5.84).
(b) There should be shielded storage (e.g. a safe) for all
sources, the outer surface of which
should be made of fireproof materials. The safe should be located
near the preparation
workbench to reduce the exposure of personnel during handling and
transfer of sources if
applicable.
(c) The safe should have compartments for different source activities.
Each compartment should
be marked so as to permit immediate and easy identification of its contents
from the outside
with a minimum of exposure.
(d) Sources should be readily identifiable by sight. When
radioactive sources of the same
appearance but of different activities or activity distribution are
used, they should be
distinguishable, e.g. by different coloured threads or beads.
(e) The workbench should be provided with L block shielding,
and with a lead glass viewing
window and a magnifying glass.
(f) The working surface for source preparation should be smooth and
seamless to avoid losing
small sources such as ¹⁹²Ir wire fragments or small ¹²⁵I seeds.
(f) The source handling area should be well illuminated and a
magnifying glass in a fixed
mounting should be available for viewing, in order to handle sources
efficiently and with a
minimum of radiation exposure.
(g) Devices for handling sources, typically forceps, should be
available. They should be as long
as practicable, compatible with efficient source handling. A device
should be provided for
threading sources expeditiously with the fingers protected by distance.
(h) The source storage and preparation laboratory should have a
sink with a filter or trap to
prevent sources being lost into the sewerage system.
(j) There should be a clear indication of the radiation level in terms
of ambient dose equivalent.
This should be achieved either by an area radiation monitor that should be
visible on entering
the room and during any handling of the unshielded sources, or by a survey
meter that should
be available and in use during source handling.
(k) Hand carried transport containers should be provided with
long handles. The lid of the
container should be securely fastened to prevent tipping and dropping
of sources during
transport. Containers should bear the radiation symbol as well as a warning
sign.
(l) Space should be available for source transport trolleys.
5.38. It is preferable that patients’ rooms be single and adjacent
to one another. Where this is not
possible, appropriate shielding between patients is necessary to
minimize to the external exposure
from other patients in the room. Within patients’ rooms, whenever possible
movable shielding for the
nurses and potential visitors should be provided (see also para. 5.146).
5.39. The treatment room should contain a shielded storage container
(large enough to accept the
applicators if necessary) and a remote handling tool (forceps) in the event of
a dislodged source.
5.40. An area monitor should be placed at the entrance so as to detect when a
source or a patient with
a source is leaving the room or the controlled area. In order to
ensure that after the treatment no
source remains within the patient, clothes or bed linen, or anywhere in the
area, a portable monitor
should be available for monitoring these items.
Design of rooms within the radiation therapy facility – imaging and other
non-treatment rooms
5.41. Patient preparation and imaging areas where radiation is used,
such as simulator rooms (CT,
PET-CT, conventional), together with their console areas and patient
changing areas should be
designed to ensure that requirements for occupational and public protection are
met. Details are given
in the appropriate sub-sections of Sections 3 and 4, paras. 3.8 –
3.15 and paras. 4.7 – 4.27,
respectively, and further guidance is given in Ref [285, 287].
Design of rooms within the radiation therapy facility – shielding
considerations
5.42. Radiation therapy facilities typically require significant shielding,
especially for the treatment
rooms, to ensure that the requirements for occupational and public radiation
protection are met. The
nominal design dose in occupied areas is derived by the process of
constrained optimization, i.e.
selecting a source related dose constraint, with the condition that
the individual doses from all
relevant sources be well below the dose limits for the persons occupying the
area to be shielded. The
following paragraphs (5.43 to 5.50) highlight some considerations with respect
to shielding design,
but a full discussion and details on methodologies and data for shielding
calculations for treatment
rooms are presented in Refs. [286, 290, 291].
5.43. Care should be taken to avoid multiplication of conservative
assumptions which can lead to
unrealistic overestimates of required shielding. Typical conservative
assumptions are: workload, use
and occupancy factors are overestimated; and the persons to be protected are
permanently in the most
exposed place of the adjacent room. A balanced decision should be
achieved and accumulation
avoided of overly conservative measures that may go beyond optimization.
5.44. However, from the other perspective, since corrections or
additions after radiation therapy
facilities are completed can be difficult and expensive, it is also
advisable that the design includes
consideration of possible future needs for new equipment and changes in
practice or use, increased
workloads, and changes in the occupancy of adjacent, above and below spaces.
5.45. The design and specification for the radiation shielding should
be performed by a medical
physicist or RPO to ensure that the required level of occupational and public
radiation protection is
achieved. The medical physicist or RPO should be involved from the
very beginning because
shielding requirements may influence decisions on where to site treatment and
imaging rooms, and
the type of building construction. The medical physicist or RPO should be
provided with all relevant
information regarding the proposed medical radiological equipment and
its use, type of building
construction, and occupancy of nearby areas. The shielding assumptions and
specifications should be
documented and signed off by the medical physicist or RPO and all
documentation, including
calculations, should be archived for the lifetime of the facility.
Depending on a Member State’s
regulatory requirements, it may also be necessary to submit the final
shielding specifications to the
radiation protection regulatory body for review prior to construction.
5.46. The shielding of the radiation treatment room should be so constructed
that the integrity of the
radiation protection is not compromised by joints, by openings for
ducts, pipes or other objects
passing through the barriers, or by conduits, service boxes, or other
structural elements embedded in
the barriers.
5.47. Treatment room door and/or maze design for high-energy
machines requires special
consideration to ensure adequate radiation protection without sacrificing
operational efficiency.
5.48. Whenever and wherever possible, there should be site visits by the
medical physicist or RPO
during construction to ensure that there has been, from the radiation safety
perspective, the correct
positioning of the joins in the structure and to ensure that the concrete has
been poured to avoid gaps
or cracks in the shielding and that the ducting does not go through the primary
shielding or is aligned
with the primary beam.
5.49. A final assessment of the adequacy of the shielding should be
performed by the medical
physicist or RPO after construction or installation of the equipment has been
completed.
5.50. Shielding considerations for imaging and simulator rooms, depending on
the modalities used,
are given in Sections 3 and 4, paras. 3.17 – 3.23 and 4.31 – 4.35,
respectively.
Medical radiological equipment, software and ancillary equipment
5.51. This sub-section considers medical radiological equipment,
software and ancillary equipment
used in a radiation therapy facility, including for diagnosis, simulation,
treatment planning, treatment
delivery, verification and follow up. For treatment pre-planning and
simulation, the equipment used
may include C-arms, conventional simulators, CT scanners, PET-CT,
SPECT-CT, MRI, and
ultrasound units. Medical radiological equipment used for external beam therapy
includes superficial
units (including units using Grenz or Bucky rays), orthovoltage units,
gamma-ray teletherapy units,
linear accelerators, and proton or heavy ion accelerators. While the
radiological equipment used for
external beam therapy falls into two main camps – linac-based and cobalt-based
– the techniques used
and hence how the equipment is constructed, its features and
configurations vary enormously
depending on whether treatment is via conventional external beam radiation
therapy, SRS, stereotactic
body radiation therapy (SRBT), 3-dimensional conformal radiation therapy, IMRT,
VMAT, intensity
modulated arc therapy (IMAT), Rapid Arc, or some other technique.
Some hybrid external beam
radiation therapy units incorporate imaging systems, such as
radiography, fluoroscopy, CT, cone
beam CT or MRI, and can perform IGRT. Brachytherapy may be manual
or remote and is
characterized by contact, intracavitary, or interstitial applications,
which may be temporary or
permanent. Almost all brachytherapy is performed with sealed
radioactive sources but electronic
brachytherapy systems with miniature X ray tubes are available [292, 293].
Radiation therapy with
unsealed sources is covered in Section 4.
5.52. The requirements for medical radiological equipment and its
software are given in the BSS
paragraphs 3.49 and 3.162. The International Electrotechnical Commission
(IEC), through its
Technical Committee 62 on Electrical equipment in medical practice and in
particular Sub-committee
62C on Equipment for radiotherapy, nuclear medicine and radiation
dosimetry, has published
international standards applicable to medical radiological equipment
used in radiation therapy.
Current IEC standards relevant to radiation therapy include the following Refs
[289, 294 – 310]. For
those relevant to the X ray based imaging systems used in radiation therapy see
Section 3 para. 3.27,
and for those relevant to radiopharmaceutical based imaging used in radiation
therapy, see Section 4
para. 4.40. It is recommended that the IEC website is visited to
view the most up-to-date list of
standards: http://www.iec.ch . The International Organization for
Standardization (ISO), through its
Technical Committee 85 on Nuclear energy, nuclear technologies, and
radiological protection and in
particular Sub-committee 2 on Radiological protection, publishes international
standards applicable to
medical radiological equipment used in radiation therapy. Current ISO standards
relevant to radiation
therapy include the following Refs [311 – 313]. It is recommended that the ISO
website is visited to
view the most up-to-date list of standards: http://www.iso.org .
5.53. Guidance on X ray based medical radiological equipment used for
imaging as part of pre-
treatment simulation, IGRT or for follow-up assessment, as described in para
5.51, is given in Section
3, paras. 3.26 – 3.39.
5.54. As the licensee takes responsibility for the radiation safety of medical
radiological equipment to
be used in the radiation therapy facility, he/she should impose purchasing
specifications that include
conditions to meet relevant international standards of the IEC and
ISO and/or equivalent national
standards. In some countries there may be a medical devices agency or similar
organization that gives
type approval to particular makes and models of medical radiological equipment.
Radiation sources,
including radioactive material, equipment and accessories, should be purchased
only from suppliers
who meet national requirements for such dealings.
5.55. Displays, gauges and instructions on operating consoles of medical
radiological equipment, and
accompanying instruction and safety manuals, may be used by staff who may not
understand, or who
have a poor understanding of, the manufacturer’s original language. In such
cases, the accompanying
documents should comply with IEC and ISO standards and should be
translated into the local
language. Software, either used in conjunction with medical
radiological equipment or as part of
treatment planning (see also para. 5.75) should be designed so that it can be
easily converted into the
local language resulting in displays, symbols and instructions that will be
easily understood by the
staff. The translations will require a quality assurance process to
ensure accuracy of the technical
content to avoid operating errors. The same should apply to maintenance
and service manuals and
instructions for maintenance and service engineers and technicians, if these
persons do not have an
adequate understanding of the original language.
5.56. Procedures for the purchase, installation, acceptance,
commissioning, use, maintenance and
quality control of all equipment (hardware and software) should be developed
with the involvement of
a medical physicist together with other radiation therapy professionals as
appropriate (for example,
medical radiological practitioner, medical radiation technologist,
biomedical engineer, information
technology specialist) and the radiation therapy facility’s radiation
protection committee and quality
assurance committee.
5.57. For medical radiological equipment in use, specific criteria of
acceptability should be defined in
order to indicate when remedial action should be taken, including, if
appropriate, taking the
equipment out of service. Examples of remedial and suspension criteria are
given in Ref. [193]. A
strategy or transition period for replacement based on social and economic
factors is helpful. See also
paras. 5.224 – 5.243 on quality assurance programmes for medical exposure.
Design features of medical radiological equipment – general considerations
5.58. The design of medical radiological equipment should be such
that its performance is always
reproducible, accurate and predictable, and that it has features that
facilitate staff in carrying out the
requirements in the BSS for operational considerations of optimization
of patient protection (BSS,
paragraph 3.163 - 3.164). Many design features contribute to the performance of
medical radiological
equipment and should be considered when purchasing such equipment, as
indicated briefly in the
following paragraphs. Further details on design features and
performance standards of medical
radiological equipment used in radiation therapy are given in Refs [294 – 299,
301 - 309, 311]. See
also later paragraphs on quality assurance and acceptance testing, and in
particular para. 5.236.
5.59. Medical radiological equipment should include provisions for selection,
reliable indication and
confirmation (when appropriate and to the extent feasible) of operational
parameters such as type of
radiation, indication of energy, beam modifiers (such as filters and wedges),
treatment distance, field
size, beam orientation and either treatment time or pre-set dose.
5.60. As noted in para. 5.52, radioactive sources for either teletherapy or
brachytherapy should meet
relevant international standards [311- 313].
5.61. Units under software control that are designed to operate within certain
tolerances should have
interruption mechanisms that stop the radiation when the tolerances
are exceeded (for example,
tumour tracking and respiratory gating). The equipment design should include
the ability to override
the software control, but only by appropriate persons that have been
authorized by the radiation
therapy facility’s licensee. When dynamic treatments are interrupted
due to being outside defined
tolerances, there should be a system/method available to resume and complete
the treatment.
5.62. Medical radiological equipment using radioactive sources should be
fail-safe in the sense that
the source will be automatically retracted to its shielded position in the
event of an interruption of
power and will remain shielded until the beam control mechanism is
reactivated from the control
panel.
5.63. Medical radiological equipment used for radiation therapy should
be provided with safety
systems capable of preventing its use by unauthorized personnel. A key should
be required to energize
the system, access to which should be restricted to authorized staff.
5.64. External beam radiation therapy equipment containing radioactive sources
and high dose rate
brachytherapy (HDR/PDR) equipment should be provided with a device to return
sources manually to
the shielded position in the case of a failure of the source to retract. For
SRS and SBRT units it should
be possible to close the shielding door manually.
5.65. The design of safety interlocks should be such that operation
of the medical radiological
equipment during maintenance procedures, if interlocks are bypassed, can be
performed only under
direct control of the maintenance personnel using appropriate devices, codes or
keys.
5.66. Record and verify systems (RVSs) and their related interfaces with
imaging systems, treatment
planning systems, treatment delivery systems, and image and administrative data
storage systems (e.g.
PACS, RIS, OIS) should be systematically verified for all their functionalities
and data integrity. The
RVSs should be able to store complete sets of information, including
the patient’s identification,
prescription, treatment plan, and field parameters, allowing this information
to be entered and called
upon accurately for each treatment. The details about the treatment equipment,
including coordinates,
scales, and angles conventions used, beam energies, available field sizes, and
other parameters and
limitations should be entered, or their entry supervised, by the medical
physicist. The system should
be subject to periodic quality assurance because, if these parameters are
incorrectly introduced into
the RVS, systematic treatment errors will occur. Detailed guidance on RVSs is
given in [309, 314].
5.67. Data transfer and data integrity, including patient information, should
be maintained throughout
the radiation therapy facility’s network. Thus the information technology
specialist should be familiar
with the radiation therapy process and work in close cooperation with the
radiological oncology team
(radiological medical practitioner, medical radiation technologist and medical
physicist).
Design features of medical radiological equipment – external beam therapy
5.68. Medical radiological equipment used for external beam therapy should meet
the specifications
given in relevant IEC standards [294 – 297, 301, 302, 305, 306, 308] and should
follow the guidance
on design specifications and performance in Refs [285, 315 - 317], as
appropriate. In addition to the
recommendations given in paras. 5.58 – 5.67, the following considerations
should also be included:
(a) Safety interlocks or other means designed to prevent the
clinical use of the machine in
conditions other than those selected at the control panel should be provided.
(b) Equipment design should permit interruption of the treatment from
the control panel; after the
interruption, resumption of treatment should be possible only from the control
panel.
(c) Radiation beam control mechanisms should be provided,
including devices that indicate
clearly and in a fail-safe manner whether the beam is ‘on’ or ‘off’. See also
para. 5.19.
(d) The radiation field within the treatment area in the absence of any
radiation beam modifiers
(such as wedges or multileaf collimators) should be as uniform as practicable
and the non-
uniformity be stated by the supplier. The non-uniformity of flattening filter
free (FFF) beams
also should be specified by the supplier.
(e) The design of the unit should lead to exposure rates outside the
treatment area due to radiation
leakage or scattering being kept as low as reasonably achievable.
(f) If primary shielding is incorporated into the equipment,
electrical or mechanical interlocks
should be provided to avoid the beam being directed towards secondary
barriers if the
primary shielding is not intercepting the beam.
5.69. Manufacturers should design accelerators to minimize neutron
production, for example by
avoiding the use of aluminium and other materials which have a high
neutron-capture cross section.
Design features of medical radiological equipment – brachytherapy
5.70. Medical radiological equipment used for brachytherapy should meet the
specifications given in
Refs [298] and should follow the guidance in [285, 318], as appropriate.
5.71. Both LDR and HDR sources should be accompanied by a source certificate
specifying:
(a) The source strength in terms of reference air kerma rate
(RAKR) or the air kerma rate
constant defined by the ICRU [300]. See also para. 5.206(b));
(b) The quality control tests applied to the source.
5.72. Applicators for brachytherapy should be manufactured specifically for the
source to be used or
be compatible with it. Use of reusable LDR radioactive
sources after the working lifetime
recommended by the manufacturer should be continued only after leak
testing by the medical
physicist or RPO and approval by the regulatory body.
5.73. Where manual brachytherapy sources incorporating ²²⁶Ra or encapsulated
¹³⁷Cs are still in use,
efforts should be made to replace them as soon as practicable with modern
afterloading systems. In no
case should sources be left in applicators (pre-loaded applicators) in between
clinical procedures, to
avoid encapsulation and/or applicator rupture due to radiation damage.
5.74. Sources using beta emitters, such as ⁹⁰Sr and ¹⁰⁶Ru in
ophthalmic applicators, should be
provided with low atomic number shielding to minimize bremsstrahlung while they
are in storage and
in preparation for use.
Design features of treatment planning systems
5.75. The capabilities of treatment planning systems (TPS) have evolved in
parallel with advances in
computers and computing. Depending on the TPS, these capabilities may include
complex three- or
four-dimensional image manipulation and dose calculations. The design features
of the TPS should
meet the clinical goals of the radiation therapy facility. TPSs should meet the
standards given in Ref
[307], and should follow the guidance on TPSs, including specifications and
performance given in
[285, 320 - 322].
Design features of simulators and imaging equipment
5.76. The role of radiation therapy simulators, as distinct from imaging
devices, has changed in recent
years with wide bore CT scanners becoming more prevalent and integral to the
treatment planning and
follow-up. Where more conventional simulators are used these should meet the
specifications given in
IEC standards [299, 303, 304] and should follow the recommendations of Refs
[285, 323]. The CT
scanners used as virtual simulators should be designed so that
patients can be simulated in the
treatment position. As noted above in para. 5.53, guidance on medical
radiological equipment used for
imaging as part of radiation therapy, either pre-treatment, during treatment
(IGRT) or for follow-up, is
given in Section 3, paras. 3.26 – 3.39 and Section 4, paras. 4.44 – 4.50, with
specific guidance for
IGRT in Ref [324].
5.77. Guidance applicable to C-arm imaging devices used in
brachytherapy is given in Section 3,
paras. 3.36 – 3.37.
Ancillary equipment
5.78. The radiation therapy facility should have equipment, instruments and
test objects for dosimetry
and quality control which may include ionization chambers (thimble, plane-
parallel and well-type),
solid-state detectors, detectors for small-field dosimetry,
electrometers, thermometers, barometers,
phantoms, geometry and mechanical test tools. Further guidance
on appropriate equipment,
instruments and test objects is given in [285, 298, 310, 317, 325, 326].
5.79. The radiation therapy facility should have a mould room (also known as a
patient preparation
area or workshop) that is equipped to prepare beam modifiers, positioning aids
and immobilization
devices (e.g. blocks, compensators, bolus, etc.). Immobilization devices
are now more commonly
prepared in the simulation area and multileaf collimators remove the
requirement for shielding blocks
in most of the cases. Where blocks are still prepared, electronic transfer of
data from the TPS to the
automatic cutting and milling machines would represent an advantage in terms of
accuracy.
5.80. In addition to laser positioning beams, the radiation therapy
facility may need to have other
positioning devices, including surface optical scanners, RF systems, GPS
transmitters and ultrasound
units.
5.81. For manual brachytherapy, the radiation therapy facility should
be equipped with radiation
safety and source handling equipment including a magnifying glass,
source manipulators (such as
forceps, tweezers or tongs), clippers or wire-cutters, and several shielded
containers.
5.82. For afterloading brachytherapy, the radiation therapy facility
should be equipped for source
handling in the case of a failure of the afterloading unit, including: a
storage container present in the
treatment room, to serve as an emergency source container in case of
failure of the afterloader in
retracting the source; a remote manipulator; and a rod mounted GM detector for
source localization.
5.83. The radiation therapy facility should be equipped with radiation
monitoring instruments (area
detectors and portable/survey meters), including Geiger counters,
ionization chambers with
electrometers or scintillators. For accelerators with energies of 15 MV and
above, access to a neutron
measuring instrument may be needed.
Security of sources
5.84. The objective of source security is to ensure continuity in the control
and accountability of each
source at all times in order to meet the requirement in BBS paragraph 3.53. In
the radiation therapy
facility the sources include sealed sources used in teletherapy and
brachytherapy, and sealed sources
used for calibration or QC tests. Situations that are particularly critical
with respect to source security
in the radiation therapy facility include receipt of sources, storage
of sources, and movement of
sources within the facility. The radiation therapy facility licensee should
develop procedures to ensure
the safe receipt and movement of radioactive sources within the institution and
establish controls to
prevent theft, loss, unauthorized withdrawal of radioactive materials
or entrance of unauthorized
personnel to the controlled areas. An inventory of sources should be
maintained, with procedures in
place to check and confirm the sources are in their assigned locations and are
secure.
Maintenance
5.85. The BSS (BSS, para 3.15(i) and 3.41) gives requirements for maintenance
to ensure that sources
meet their design requirements for protection and safety throughout
their lifetime and to prevent
accidents as far as reasonably practicable. Therefore the licensee of
the radiation therapy facility
should establish the necessary arrangements and coordination with the
manufacturer’s representative
before initial operation and on an on-going basis. This can be
achieved through a maintenance
contract (preventive and corrective) with the manufacturer, or by
in-house staff or third party
contractor only if appropriately trained and authorized (see also Section 2,
para. 2.112).
5.86. Maintenance includes not just the medical radiological equipment
and its hardware, but also
software, networks, data bases and other supporting systems in the
radiation therapy facility, for
example, PACS, HIS and RIS.
5.87. In addition to the guidance in Section 2, paras 2.110 - 2.111, the
licensee of the radiation therapy
facility should ensure that the process of removal from, and return
to, clinical service of radiation
therapy medical radiological equipment for maintenance or source exchange
includes:
(a) A record of maintenance carried out should be kept for each item of
equipment: this should
include information on any defects found by users (a fault log), remedial
actions taken (both
interim and subsequent repairs) and the results of testing before equipment is
reintroduced to
clinical use.
(b) Where maintenance of the therapy and imaging equipment or treatment
planning equipment
may affect the accuracy of the physical or clinical dosimetry or the
safe operation of the
equipment, a radiation therapy medical physicist should perform
specific tests or
measurements in order to determine that the equipment is operating
satisfactorily before it is
used to treat patients (see BSS paragraph 3.167(b)).
5.88. The electrical and mechanical safety aspects of the medical
radiological equipment are an
important part of the maintenance programme, and can have direct or
indirect effects on radiation
safety. This work should be performed by appropriately authorized
persons who understand the
specifications of the medical radiological equipment. Electrical and mechanical
maintenance should
be included in the QA programme at a frequency recommended and preferentially
performed by the
manufacturer of the medical radiological equipment. Servicing should
include a written report
describing the findings. These reports should be archived as part of the QA
programme.
OCCUPATIONAL RADIATION PROTECTION
Introduction
5.89. In radiation therapy radiological procedures, as described in
paras. 5.1 to 5.6, occupationally
exposed individuals are usually the medical radiation
technologists, the radiological medical
practitioners (typically the radiation oncologists) and the medical physicists.
In some radiation therapy
facilities, other health professionals such as nurses may also be considered
occupationally exposed.
5.90. Additional occupationally exposed personnel may include
dosimetrists and biomedical or
service engineers and some contractors, depending on their role.
5.91. Other radiation therapy facility workers such as social workers,
dieticians, physiotherapists,
patient porters, orderlies, assistants, cleaners and other service support
personnel, for whom radiation
sources are not directly related to their work, require the same level of
protection as members of the
public, as stated in the BSS paragraph 3.78.
5.92. This sub-section contains guidance very specific to radiation
therapy. For more general and
comprehensive guidance on occupational radiation protection, including
guidance on radiation
protection programmes, assessment of occupational exposure and providers
of dosimetry services,
applicable to all areas of radiation use (including non-medical uses),
reference should be made to the
IAEA Safety Guide Occupational Radiation Protection [23].
Arrangements under the radiation protection programme
Classification of areas
5.93. Various areas and rooms in a radiation therapy facility should
be classified as controlled or
supervised areas, in line with the requirements given in BSS paragraphs 3.88 to
3.92. All other rooms
and areas, not so-designated, are considered as “public domain” and levels of
radiation in these areas
should be low enough to ensure compliance with the dose limits for public
exposure.
5.94. In a radiation therapy facility, all treatment rooms for
external beam radiation therapy and
remote afterloading brachytherapy, operating theatres used during
brachytherapy procedures with
radioactive sources, brachytherapy patient rooms, radioactive source storage
and handling areas, and
rooms where imaging or simulation procedures are performed meet the criteria
for controlled areas
and should be so designated.
5.95. Supervised areas may include the areas surrounding brachytherapy
patients’ rooms or around
radioactive source storage and handling areas.
5.96. The area around the control panel for all medical radiological
equipment used in radiation
therapy should be classified as either a controlled or a supervised
area, even though the radiation
levels may be very low due the shielding design. In either case, this area
should have restricted access,
among other things to avoid distraction of staff which may lead to
accidental medical exposure of
patients.
5.97. In order to avoid uncertainties about the extent of controlled
and supervised areas, the
boundaries should, when possible, be walls and doors, partitions or
other physical barriers, clearly
marked or identified with ‘radiation area’ signs.
Local rules and procedures – general
5.98. The BSS, in paragraph 3.93, establishes a hierarchy of preventive
measures for protection and
safety with engineered controls, including structured and ancillary
shielding, specific physical
barriers, signs and interlocks, being supported by administrative
controls and personal protective
equipment. To this end, and as required in the BSS paragraph 3.94, written
local rules and procedures
should be established in a radiation therapy facility. Their purpose is to
ensure protection and safety
for workers and other persons. These local rules and procedures should include
measures to minimize
occupational radiation exposure during both normal work and unusual
events. The local rules and
procedures also should cover the wearing, handling and storing of personal
dosimeters, and specify
investigation levels and ensuing follow-up actions (see also paras. 5.155 –
5.167).
5.99. Since all personnel involved in using radiation in radiation therapy
should know and follow the
local rules and procedures, the development and review of these local rules
and procedures should
include representatives of all health professionals involved in radiation
therapy.
5.100. Equipment (hardware and software) should be operated in a manner that
ensures satisfactory
performance at all times with respect to both the tasks to be accomplished and
radiation safety. The
manufacturer’s operating manual is an important resource in this respect, but
additional procedures
should be also considered. The final documented set of operational
procedures should be approved by
the radiation therapy facility’s licensee, and incorporated into the
facility’s quality management
system (see Section 2, paras. 2.136 – 2.147).
5.101. Radiation therapy staff should understand the documented
procedures for operation of the
equipment with which they are working, including the safety features,
and should be trained, with
periodic refresher training, in what to do when things go wrong. Additional
education and training
should occur when new devices or techniques are introduced into radiation
therapy practice.
5.102. Many local rules and procedures address aspects of some or
all of occupational radiation
protection, patient radiation protection and public radiation protection,
either directly or indirectly, as
well as ensuring a successful application of the treatment. This
is the case with the following
paragraphs (5.103 to 5.142) – while placed in this section on
occupational radiation protection
because they are to be followed by workers, the local rules and
procedures often also have
significance for patient and/or public radiation protection. The following
paragraphs (5.103 to 5.142)
give recommendations that should be incorporated into the radiation therapy
facility’s local rules and
procedures.
5.103. No one should be in the treatment room during the delivery of treatment,
except the patient. All
attending personnel should be in appropriately shielded areas.
5.104. Safety features such as interlocks, presence of accessories such as the
T-bar for manual Co-60
source retraction and survey meters functionality should be checked daily prior
to patient treatment.
More detail is given in Ref [285], and see also para. 5.236 on quality control
tests in general.
5.105. Sealed sources should be subject to leak tests prior to their first use
and at regular intervals
thereafter, in conformity with international standards [312]. Leak tests should
be sufficiently sensitive
to be able to detect the presence of very small amounts of removable
contamination, for example 0.2
kBq.
5.106. Area surveys should be performed periodically around all treatment units
and check sources,
including cobalt 60-units, shielded safes and source storage facilities for LDR
and HDR sources.
5.107. Local rules for pregnant workers and persons under 18 should reflect the
guidance given in
paras. 5.178 – 5.181 and 5.182, respectively.
Local rules and procedures – external beam radiation therapy
5.108. Safe operation of external beam radiation therapy units requires
procedures for area surveys,
interlock checks, wipe tests (for sealed sources) and procedures for
contingencies such as a source
becoming stuck in the on or partially on position. Such procedures
require that the necessary
equipment be available, calibrated and in working order, including:
(a) A radiation monitor;
(b) Wipe test capabilities (for radioactive sources);
(c) Personal alarm dosimeters, especially for unplanned exposures.
5.109. The procedures for the use of radiation monitoring equipment
should recognize that some
instruments will “lock up” in a high radiation field and give
erroneous readings, and that this
phenomenon, if it occurs, can be identified by starting the monitoring from
outside the room in which
the source is located, i.e. monitoring from the lower to the higher dose rate
areas.
5.110. The presence of other staff in the area of the control panel
should be kept to the minimum
necessary so as to avoid distraction to the medical radiation technologist, as
stated in para. 5.96.
5.111. As described in para 5.105, regular wipe tests should be
performed for sealed sources. For
external beam therapy the method to be used is the indirect wipe test of the
nearest accessible surface.
5.112. Irradiations that involve long uses of high energy X rays, such as beam
calibration, dosimetry
and quality control measurements, should be scheduled to take place at the end
of the day’s clinical
roster so that neutron activated radionuclides (especially the
longer-lived ones) can decay
significantly overnight.
Local rules and procedures – brachytherapy – general considerations
5.113. Source inventories should be maintained, giving the radionuclide,
location and activity with
reference date of each source at the facility as well as its serial
or batch number, and a unique
identifier. The unique identifier may be either a colour coded or an
alphanumeric identifier.
5.114. Sources should never be left on preparation surfaces. They
should be either in storage, in
transit or in use.
5.115. As described in para 5.105, regular wipe tests should be
performed for sealed sources. For
long-lived LDR brachytherapy sources the typical method is the direct
moist wipe test, while for
remote controlled brachytherapy the method to be used is the indirect
wipe test of the nearest
accessible surface. For an HDR/PDR unit, the wipe tests should be
only carried out on the
afterloading drive assembly and transport containers, since the source itself
has too high a dose rate to
allow this type of test.
5.116. As stated in para 5.106, area surveys should be performed
periodically around the source
storage facilities for LDR and HDR/PDR sources.
5.117. The source storage facilities should be marked to indicate
that they contain radioactive
materials, and instructions given on how to contact the radiation protection
officer, medical physicist
or other responsible radiation safety individual in the event of an emergency.
5.118. Source storage rooms should be kept locked at all times.
5.119. After every brachytherapy treatment, all brachytherapy sources should
be removed from the
patient, except in the case of permanent implants. The patient should be
monitored with a radiation
survey meter to ensure that no radioactive source remains in or on
the patient. Linen, dressings,
clothing, waste and equipment should be kept within the room where the
removal of sources takes
place until all sources are accounted for, and should be monitored with a
radiation detector. Mobile
containers and portable equipment containing radioactive sources should be
removed to storage or to
a secure place when not in use.
5.120. Sterilization processes in brachytherapy should be appropriate
for preventing damage to
sources and applicators that could affect safety.
Local rules and procedures – brachytherapy – additional for LDR sources
5.121. In the case of temporary LDR brachytherapy applications, both
manual as well as remotely
controlled, the following information should be displayed at the
entrance to the treatment room:
identification of the patient, sources, date and time of insertion
and removal, nursing required,
time/distance allowance for nurses and visitors, and concise
instructions for unplanned source and
applicator removal and for dealing with an emergency, including
contact details. A patient with a
removable source in or on his or her body should only leave the room in
exceptional circumstances
and would should be accompanied by an attendant from the radiation therapy
facility at all times.
5.122. Reusable sources should be inspected visually for possible damage after
each use, by means of
magnifying viewers and a leaded viewing window in a shielded work area.
5.123. There should be a diagram at the source storage safe that
shows the exact location of each
source within the safe, thus reducing the time taken to locate and identify a
source.
5.124. Sources should only be handled with long forceps or tongs.
5.125. When transporting sources, a mobile shielded container should be
available and the shortest
route possible should be used. The container should have a long handle and/or a
long handled trolley
should be used.
5.126. Reusable sources that come into direct contact with body
tissues will require cleaning and
sterilization after each use. This can subject the sources to
possible damage from heat, abrasion,
chemicals and mechanical stresses. Therefore, these sources should be inspected
after every use.
5.127. Work surfaces should be continuous, easy to clean and brightly lit to
make it easy to find any
sources that have been dropped.
5.128. If the source storage and preparation room is also the applicator
loading room, there should be
a sink for cleaning the applicators. However, a sink can also lead to a loss of
sources to the sewage
system when a source is left in the applicator or a patient removes a source
and puts it in the sink.
These situations are preventable by placing a filter in the sink’s drain.
Local rules and procedures – brachytherapy – additional for HDR/PDR sources
5.129. The HDR/PDR afterloader should undergo routine quality assurance tests
at the beginning of
each treatment day [327].
5.130. Among other safety checks, the couplings and transfer tubes should be
checked before each
HDR treatment, to ensure that there are no obstacles to prevent motion of the
source. Further details
on safety checks are given in Ref [327].
5.131. Emergency safety precautions require the availability of an
emergency container in the
treatment room, as well as an emergency kit containing surgical clamps and long
handled forceps for
manipulation of the source guide tubes and applicators if the source fails to
return to the safe, or for
other source retrieval actions. The emergency container should be
placed close to the patient and
should be sufficiently large that it can accept the entire applicator
assembly containing the source
removed from any patient.
5.132. Manufacturers provide suggested contingency procedures if the
source fails to return to the
safe. These generally consist of a short single page synopsis, suitable for
posting in an appropriate
place, of the necessary sequential steps involved in the emergency procedure.
They assume that the
physical integrity of the applicator is maintained. These procedures
are specific to the actual
afterloading unit, but, in general, each step assumes that if the
previous action fails to lead to
recovery, then the following actions are required. The general sequence is:
(a) Observation at the console of an error message and emergency indicators
(audible and visible
alarms);
(b) Recovery at the console (e.g. pressing an emergency ‘off’ button);
(c) Entry into the room with a portable radiation survey meter (opening the
door activates the
interlock that retracts the source);
(d) Observation of radiation levels in the room (by mounted monitors or
portable survey meters);
(e) Recovery at the afterloading unit (pressing an emergency ‘off’
button on the remote
afterloading unit);
(f) Manual retraction of the source (using a hand crank);
(g) Patient survey and the afterloader survey (confirming that the source is
in the safe);
(h) Applicator removal and placement in the emergency container;
(i) Patient survey and emergency container survey (to confirm that the source
is not in the patient
and that it is in the emergency container);
(j) Removal of the patient from the vault with subsequent redundant survey
monitoring.
Local rules and procedures – remote control afterloading brachytherapy
5.133. Remote afterloading equipment requires specific contingency
procedures, as these are
especially critical for HDR brachytherapy. These procedures are dealt with in
paras. 5. 310 – 5.313. A
shielded container large enough to accommodate the largest applicator set
should be kept next to the
unit in case the source gets stuck.
Local rules and procedures – manual brachytherapy
5.134. For implants with sources of different activities, after verification of
the source strength, the
source or source holder should be marked with unique identifiers (for
example, a pre-established
colour which cannot be compromised by body fluids), to facilitate visual
recognition and prevent the
possibility of confusion between different sources or batches.
Containers utilized for transport of
radioactive sources should conform to the requirements established in the
IAEA’s Regulations for the
Safe Transport of Radioactive Material [284] – see also paras 5.318 – 5.320.
5.135. The movements of the sources from the time they leave the safe until
their return (if applicable)
should be recorded, with the signature of the person responsible for the move
(using forms or a log
book). A person should be assigned to be in charge of accountability
for the sources. This person
should keep a record of the source order and of issuance from and return to the
safe, with signatures.
See also para. 5.84.
5.136. Reusable sources should be inspected visually for possible damage after
each use by means of
magnifying viewers and a leaded viewing window in a shielded work area.
5.137. Sources should only be handled with long forceps or tongs, never
directly with the fingers.
5.138. A mobile shielded container should be available for transport of sources
and the shortest route
possible should be used. The container should have a long handle and/or a long
handled trolley used.
5.139. Reusable sources which come into direct contact with body
tissues require cleaning and
sterilization after each use; this can subject the sources to
possible damage from heat, abrasion,
chemical attack and mechanical stresses. Therefore, these sources should be
inspected after every use.
5.140. Available safety features listed in para. 5.37 should be effectively
used.
5.141. Precautions to be observed during the cutting and handling of
¹⁹²Ir wires should include
ensuring that:
(a) Appropriate tools and equipment such as forceps, cutting
devices, magnifying glasses and
good illumination of the work surface are available and used and that, if ¹⁹²Ir
wires are cut off
for immediate use, a container to hold cut lengths is provided and labelled;
(b) Radioactive waste is collected and stored in adequate containers,
and properly transferred to
an authorized waste disposal facility (see also paras. 5.288 – 5.289);
(c) Surfaces and tools are properly decontaminated.
Local rules – imaging and simulation
5.142. Local rules and procedures for performing imaging procedures as
part of pre-planning and
simulation should follow the guidance, where appropriate, given in Sections 3
and 4, paras. 3.56 –
3.82 and 4.69 – 4.103, respectively. Additional information relevant to local
rules specific to using
imaging equipment as part of IGRT is given in Ref [324].
Personal protective equipment and in-room protective devices
5.143. The BSS, paragraphs 3.93 and 3.95, requires that personal protective
equipment and in-room
protective equipment are available and used when structural shielding
and administrative controls
alone cannot afford the necessary level of occupational radiation
protection. The need for this
protective equipment should be established by the radiation therapy
facility’s RPO or medical
physicist.
5.144. For current external beam treatment procedures in radiation
therapy, personal protective
equipment is not usually needed. However, during patient preparation, source
implantation or manual
afterloading techniques in brachytherapy, and in the
simulation/preplanning phase when imaging
equipment is in use, e.g. C-arm, CT, PET-CT, the relevant recommendations given
in the Sections 3
and 4 covering these procedures should be applied (see paras. 3.83 – 3.93
and 4.109 – 4.110).
5.145. In the case of manual handling of sources for brachytherapy,
protective equipment such as
shielding blocks on the workbench and lead glass screen should be
used, as well as appropriate
devices for handling sources (see para. 5.138 and 5.141).
5.146. For nursing of brachytherapy patients with either temporary (¹³⁷Cs or
¹⁹²Ir) or permanent
implants (¹²⁵I seeds) consideration should be given to the use of movable
shielding in the ward.
Further advice is given in Ref [328].
5.147. Protective equipment for emergencies in brachytherapy, e.g. a stuck
source in HDR, should
include an emergency container suitable for applicators/sources. See
also the sub-section covering
procedures for contingencies (paras. 5.310 – 5.313).
Monitoring of the workplace
5.148. The BSS, in paragraphs 3.96-3.98, sets out the requirements and
responsibilities for workplace
monitoring. Workplace monitoring comprises measurements made in the working
environment and
the interpretation of such results. Workplace monitoring serves several
purposes, including routine
monitoring, special monitoring for specific occasions, activities or tasks, and
confirmatory monitoring
to check assumptions made about exposure conditions. Workplace monitoring
can be used to verify
the occupational doses of personnel whose work involves exposure to
predictable low levels of
radiation. It is particularly important for staff members who are not
individually monitored. Further
general guidance on workplace monitoring is given in Ref [23].
5.149. Workplace monitoring in areas around each of the medical radiological
equipment (therapy and
imaging) in the radiation therapy facility, when it is being operated, should
be carried out when:
(a) The room and shielding construction has been completed,
either new or renovation, and
before the room is first used clinically;
(b) New or substantially refurbished equipment is commissioned;
(c) Source replacements have taken place in teletherapy or remote
controlled brachytherapy;
(c) New software for the medical radiological equipment is
installed or there is a significant
upgrade;
(d) New techniques are introduced;
(e) Servicing on the medical radiological equipment has been
performed, which may have an
impact on the radiation delivered.
5.150. Initial workplace monitoring includes measurements of radiation leakage
from the equipment
and of scattered radiation using suitable phantoms, and area monitoring of
accessible areas around,
above and below irradiation rooms. This initial monitoring should be performed
as part of acceptance
tests, prior to clinical use of equipment.
5.151. In addition, exposure levels in teletherapy rooms with radioactive
sources and high dose rate
brachytherapy treatment rooms should be continuously monitored through
the use of permanently
installed area monitors. The source storage and handling area should
be monitored with a survey
meter immediately following the removal from, or return to, storage of
brachytherapy sources.
5.152. For treatment rooms where the possibility of induced activity exists,
e.g. with protons, heavy
ions and high energy X ray beams (>10 MV), consideration should be given to the
use of appropriate
area monitors to detect the presence of neutrons and other radiation being from
emitted from induced
radionuclides in the treatment room [329].
5.153. Workplace monitoring should be done in association with
brachytherapy procedures. Soon
after implantation of the sources a survey of exposure rates in the vicinity of
the patient is necessary.
5.154. All survey meters used for workplace monitoring should be calibrated in
terms of ambient dose
equivalent. For radiation therapy procedures, the quantity is H*(10) and the
unit the Sv and its sub-
multiples. The calibration should be traceable to a standards dosimetry
laboratory. The meters should
be subject to regular quality control tests. See also para. 5.241.
Assessment of occupational exposure and workers’ health surveillance
Occupational exposure assessment
5.155. The purpose of monitoring and dose assessment is, inter alia, to provide
information about the
actual exposure of workers and confirmation of good work practices and
regulatory compliance. It
contributes to reassurance and motivation. As stated in the BSS, paragraphs
3.99 to 3.102, individual
monitoring is required for any worker who normally works in a controlled area
and is likely to receive
significant occupational exposure. Workers who may require individual
monitoring include radiation
oncologists, medical physicists, medical radiation technologists, the
radiation protection officer,
biomedical engineers, maintenance and servicing personnel, and any nursing or
other staff who should
spend time with patients who contain sources.
5.156. Monitoring involves more than just measurement. It includes
interpretation, assessment,
investigation and reporting, which may lead to corrective measures, if
needed. Individual external
doses can be assessed by using individual monitoring devices that
include thermoluminescent
dosimeters (TLDs), optical stimulated luminescence dosimeters (OSLDs), film
badges and electronic
dosimeters. Individual monitoring devices must be calibrated, and be
traceable to a standards
dosimetry laboratory. For more detailed guidance see Refs [23].
5.157. Each dosimeter is to be used for monitoring only the person to whom it
is issued, for work
performed at that radiation therapy facility and should not be taken
to other facilities where that
person may work. For example, if a person is issued with a dosimeter at
hospital A it should be worn
only at hospital A and not at any other hospitals or medical
centres where he/she may also work.
Monitoring results can then be interpreted for the person working in
a specific radiation therapy
facility, and allow appropriate review of the effectiveness of the optimization
of protection measures
for that individual in that facility. See also paras. 5.168 – 5.170.
5.158. The monitoring period specified by regulatory bodies in most countries
is typically in the range
of one to three months. A one month monitoring period is usually
used for persons performing
procedures associated with higher occupational exposures. A longer monitoring
period (two or three
months) is more typical for personnel exposed to lower doses, as a one month
cycle would usually
mean that the actual occupational dose was less than the minimum detection
level of the dosimeter
resulting in no detectable doses. With a longer cycle it is more likely to
obtain a reading. Unnecessary
delays in the return, reading and reporting of dosimeters should be avoided –
the radiation therapy
facility should send the dosimeters to the dosimetry service provider
who should then process the
dosimeters and return the dose reports, all in a timely manner. Some regulatory
bodies may specify a
performance criterion for timely reporting.
5.159. The operational dosimetric quantity used is the personal dose
equivalent Hp(d). For weakly
penetrating and strongly penetrating radiation, the recommended depths, d, are
0.07 mm and 10 mm,
respectively. Radiation used in radiation therapy is usually strongly
penetrating and therefore d = 10
mm, except in the case of use of beta sources for brachytherapy. Hp(10) is used
to provide an estimate
of effective dose that avoids both underestimation and excessive overestimation
[18].
5.160. For monitoring the skin and extremities, a depth of 0.07 mm (d = 0.07)
is recommended, and
Hp(0.07) is used to provide an estimate of equivalent dose to the
skin and extremities. When the
possibility of substantial exposure to the hands exists, such as in
the handling of brachytherapy
sources, extremity dosimeters should be worn (if this is compatible with
clinical practice).
5.161. For monitoring the lens of the eye, a depth of 3 mm (d = 3) is
recommended, and Hp(3) is used
to provide an estimate of equivalent dose to the lens of the eye. In practice,
however, the use of Hp(3)
has not been widely implemented for routine individual monitoring. In
radiation therapy it would
generally be expected that the dose to the lens of the eye is not significantly
higher than for the rest of
the body. A possible exception is during the handling of sources for
preparation and insertion, but the
accepted practice of using a workbench provided with L block shielding with a
lead glass viewing
window should adequately protect the eyes. Nonetheless, monitoring of eyes
doses may be considered
in these or similar cases.
5.162. There are three dose limits applicable to workers in radiation
therapy – effective dose; and
equivalent dose to the lens of the eye, and to skin and extremities. The
dosimeter being worn can be
used to estimate one or more of the quantities used for the dose limits.
Depending on the work being
performed by the person being individually monitored, there may be a preferred
position for wearing
the dosimeter, and more than one dosimeter may be used. In radiation therapy,
dosimeters are usually
worn on the front of the upper torso, as occupational exposure arising from
most radiation therapy
procedures results in the whole body being fairly uniformly exposed. If
specialized dosimeters, such
as ring dosimeters for monitoring finger doses, are needed, the
manufacturer’s specific wearing
instructions should be followed.
5.163. When not in use, individual dosimeters should be kept in an established
place and protected
from damage or from irradiation. If an individual’s dosimeter is lost, the RPO
should perform a dose
assessment, record this evaluation of the dose and add it to the worker’s dose
record. Where there is a
national dose registry, information of the dose estimate should be provided in
a timely manner. The
most reliable method for estimating an individual’s dose is to use his or her
recent dose history. In
those cases where the individual performs non-routine types of work, it may be
better to use the doses
of co-workers having similar exposure conditions as the basis for the dose
estimate.
5.164. In some cases occupational doses may be estimated from the results of
workplace monitoring.
The effective dose for personnel can be inferred from the measured ambient dose
equivalent H*(10),
provided the dose gradient in the workplace is relatively low. ICRP
[125] provides conversion
coefficients from ambient dose equivalent to effective dose for
different types of radiation and
energies. The conversion coefficients for photons are close to unity except for
very low energy, such
as the energy of scattered photons from an X ray beam generated at a low
kilovoltage.
5.165. The use of additional direct reading operational dosimeters,
such as electronic dosimeters,
should be considered for use in a radiation therapy facility, e.g.
in a new department or with the
introduction of new modalities or procedures, as these devices can
give the worker an instant
indication of both the cumulative and the current dose rate and also allow
pre-setting of an alarm [23].
Investigation levels for staff exposure
5.166. Investigation levels are a tool used to provide a ‘warning’ of the need
to review procedures and
performance, investigate what is not working as expected and take
timely corrective action. The
following are examples of levels and their related tasks that should not
normally be exceeded and,
therefore, could be suitable as investigation levels. In radiation therapy, for
example, pro rata monthly
values higher than 0.5 mSv (for the dosimeter worn on the torso) should be
investigated. If additional
dosimeters are being used, then values higher than 2 mSv per month for a
dosimeter monitoring the
lens of the eye may indicate that eye doses may be of concern. Values higher
than 15 mSv per month
for hand or finger dosimeters should also be investigated. Abnormal conditions
and/or events should
also trigger an investigation. In all cases the investigation is with a view to
improve implementation of
optimization of occupational protection and results should be recorded.
Investigation levels should
also be set for workplace monitoring.
5.167. The investigation is to be initiated as soon as possible
following the trigger or event and a
written report is to be prepared concerning its cause, including determination
or verification of any
doses received, corrective or mitigating actions, and instructions or
recommendations to avoid
recurrence. Such reports should be reviewed by quality assurance and radiation
safety committees, as
appropriate, and the holder of the licence should be informed. It is
recommended that the report is
submitted to an international or national educational safety reporting
system, for example ROSIS
(Radiation Oncology Safety Information System) [21]. In some cases, the
regulatory body should also
be informed.
Persons who work in more than one place
5.168. Personnel may work in more than one radiation therapy facility. The
facilities may be quite
separate entities in terms of ownership and management, or they may have common
ownership but
separate management, or they may even have common ownership and management, but
be physically
quite separate. No matter which, the occupational radiation protection
requirements for the particular
radiation therapy facility apply when the person is working in that
facility. As described above in
para. 5.157, a dosimeter issued for individual monitoring should be worn only
in the facility for which
it is issued as this facilitates effective implementation of optimization of
protection measures in that
facility. This approach is logistically more easily implemented as
each physical site has its own
dosimeters – there is no need to transport dosimeters between facilities, with
the risk of losing them or
forgetting them. In cases where the facilities are under common
ownership it may be seen as an
unnecessary financial burden to provide more than one set of dosimeters for
staff that work in more
than one of its facilities. However the radiation protection advantages of
having the dosimeter results
linked to a person’s work in only one radiation therapy facility remain. See
also para. 5.170.
5.169. There is however an additional consideration, namely
ensuring compliance with the
occupational dose limits. Any person who works in more than one radiation
therapy facility should
notify the licensee for each of those facilities. Each licensee, through their
RPO, establishes formal
contact with the licensees of the other radiation therapy facilities, and their
RPOs, so that each facility
has an arrangement to ensure that a personal dosimeter is available
and that there is an on-going
record of the occupational doses for that person in all the facilities where
they work.
5.170. Some personnel, such as consultant medical physicists or service
engineers, may perform work
in many radiation therapy facilities and, maybe as well, other medical
radiation facilities. They may
be employed by a company or be self-employed, providing contracted
services to the radiation
therapy and other facilities. In such cases it is simpler for the company or
the self-employed person to
provide the dosimeters for individual monitoring. In other words, in these
cases for each person the
same dosimeter is used for work performed in all radiation therapy
facilities (and other medical
radiation facilities) in the monitoring period.
Records of occupational exposure
5.171. Paragraphs 3.103 to 3.107 of the BSS state the detailed
requirements for records of
occupational exposure, placing obligations on the employer, registrant
and licensee. Apart from
demonstrating compliance with legal requirements, records of occupational
exposure should be used
within the radiation therapy facility for additional purposes, including
assessing the effectiveness of
the facility’s implementation of optimization of protection measures,
and evaluating trends in
exposure. Further general guidance on records of occupational exposure is given
in [23].
Workers’ health surveillance
5.172. The primary purpose of health surveillance is to assess the
initial and continuing fitness of
employees for their intended tasks, and requirements are given in BSS
paragraphs 3.108 to 3.109.
5.173. No specific health surveillance related to exposure to ionizing
radiation is necessary for staff
involved in radiation therapy. Only in cases of overexposed workers, at doses
much higher than the
dose limits (e.g. a few hundred millisieverts or higher), would
special investigations involving
biological dosimetry and further extended diagnosis and medical treatment be
necessary [23]. Under
normal working conditions, the occupational doses incurred in radiation therapy
are low. No specific
radiation related examinations are normally required for persons who are
occupationally exposed to
ionizing radiation as there are no diagnostic tests that yield
information relevant to such normal
exposure. It is, therefore, rare for considerations of occupational exposure
arising from the working
environment of a radiation therapy facility to significantly influence the
decision about the fitness of a
worker to undertake work with radiation or to influence the general conditions
of service [23].
5.174. Counselling should be available to workers who have or may have been
exposed substantially
in excess of dose limits, and information, advice and, if indicated,
counselling should be available to
workers who are concerned about their radiation exposure. In radiation therapy,
the latter group may
include women who are or may be pregnant. Counselling should be
given by appropriately
experienced and qualified practitioners. Further guidance is given in Refs.
[23, 330].
Information, instruction and training
5.175. All staff involved in radiation therapy should fulfil the
respective training and competence
criteria described in Section 2, paras. 2.117 to 2.135. This
will include education, training,
qualification and competence for occupational radiation protection in
radiation therapy. Radiation
oncologists, medical radiation technologists, medical physicists and nurses may
not have been trained
with respect to imaging or pre-planning systems, such as CT, PET-CT, and as
such should undertake
radiation protection and safety training relevant to the additional imaging
modalities in their radiation
therapy facility.
5.176. The BSS, in paragraph 3.110, places responsibilities on
employers to provide, inter alia,
specific instruction and training for protection and safety as it
pertains to their radiation therapy
facilities. This is not only for new staff but also for all staff as part of
their continuing professional
development. Specific instruction and training should be provided when
new medical radiological
equipment, software and techniques are introduced.
Conditions of service and special arrangements
5.177. As required in BSS paragraph 3.111, no special benefits are to be
offered to staff because they
are occupationally exposed. It is simply not acceptable to offer benefits as
substitutes for measures for
protection and safety.
Special arrangements – female workers (pregnant)
5.178. A female worker should notify the licensee that she is pregnant as soon
as she knows of her
condition, so that radiation protection requirements for the embryo/fetus as a
member of the public
can be met.
5.179. Limitation of the dose to the embryo/fetus does not mean that pregnant
women should avoid
working with radiation, but it does imply that the employer should
carefully review the exposure
conditions with regard to both normal exposure and potential exposure. For
example, the dose to the
fetus for workers involved in source handling in manual brachytherapy, under
normal conditions, may
reach the dose limit for members of the public (see Section 2,
Table 2). To prevent this from
happening, rigorous time, shielding and distance restrictions should be
implemented.
5.180. Other possible solutions include reassigning a pregnant worker to duties
where the likelihood
of an accident or incident is unlikely or to a location that has a
lower ambient dose equivalent.
Adequate education and training should accompany such reassignments. A further
consideration is to
avoid using pregnant workers in responding to an incident or emergency such as
those described in
paras. 5.304 – 5.314, for example, with a cobalt unit or an HDR brachytherapy
unit.
5.181. When applying the dose limit to the embryo/fetus, the reading
of the dosimeter may
overestimate embryo/fetal dose by a factor depending on the energy and type of
the incident radiation
(by a factor 10 for low energy X rays and about 2 for Co-60 and MeV X rays).
Information, advice
and, if indicated, counselling for pregnant workers should be available (see
also para. 5.174).
Special arrangements – persons under 18
5.182. In many countries there is the possibility of students aged
16 or more, but under 18,
commencing their studies and training to become a medical radiation
technologist or other health
professional that may involve occupational exposure to ionizing radiation. The
BSS paragraph 3.116
states the requirements for access to controlled areas and the dose limits for
such persons are more
restrictive – see Table 2 in this Safety Guide and Schedule III of the BSS.
Protection of workers responding to incidents in the radiation therapy facility
5.183. The practice of radiation therapy is a planned exposure
situation, and when circumstances
result in incidents that lead to, or could lead to, unintended or accidental
exposures of patients or staff,
they are still within the framework of a planned exposure situation. The
potential occurrence of such
should be considered in advance and contingency plans developed
accordingly – see guidance in
paras. 5.301 – 5.317 in the section on potential exposure, mitigation and
contingency plans.
5.184. Occupational exposure of staff responding to these incidents is still
subject to the occupational
dose limits, and the contingency plans for incidents should include
considerations for the optimization
of protection for the responding workers. The contingency plans also
should include allocation of
responsibilities and provide for the education and training of the
relevant staff in executing the
mitigation measures, which should be periodically rehearsed. Most of these
situations, for example
the retraction of a stuck cobalt-60 source, can be executed in a planned manner
so that doses received
can be kept low.
RADIATION PROTECTION OF PATIENTS, CARERS AND COMFORTERS, AND
VOLUNTEERS IN BIOMEDICAL RESEARCH
Introduction
5.185. This section covers what is more formally called ‘medical exposure’ in
radiation protection.
Medical exposure is defined in the BSS and described in Section 2, para. 2.5.
It concerns radiation
protection of the patient, carers and comforters, and volunteers in
biomedical research. The term
‘patient’, when used in the context of medical exposure, means the patient
undergoing the radiological
procedure. Other patients in the radiation therapy facility or wider medical
facility, including those
who may be waiting for their own radiological procedure, are considered as
members of the public
and their radiation protection is covered in paras. 5.278 – 5.291.
5.186. As described in Section 2 para. 2.8, there are no dose limits for
medical exposure, making it
very important that there is application of the requirements for justification
and optimization.
Justification of medical exposures
5.187. The requirements of the BSS for justification of medical exposure, BSS
paragraphs 3.155 to
3.161, incorporate the ‘3 level approach’ to justification [4, 131, 140].
5.188. The roles of the health authority and professional bodies with respect
to ‘level 2’ or generic
justification of radiological procedures in radiation therapy are described in
Section 2 paras. 2.55 –
2.56 and 2.64, respectively.
Justification for the individual patient
5.189. The BSS invokes a joint approach to justification at the individual
patient level, with a shared
decision involving both the referring medical practitioner (who initiates the
request for a radiological
procedure or procedures) and the radiological medical practitioner. In the case
of radiation therapy,
the requirements for justification are applied more effectively as
part of the medical process of
determining the best approach to treatment. When a patient is
referred by a referring medical
practitioner for treatment, careful consideration should be made by
the multidisciplinary oncology
team regarding whether to treat the patient either by radiation therapy,
another modality, a combined
treatment approach (sequential or concomitant) or not to be treated
at all. Ideally every treatment
decision should be discussed and documented in a tumour or similar board.
5.190. From the radiation protection perspective, not only the radiation
therapy treatment should be
justified, but all the imaging radiological procedures prior to,
during and after the treatment also
should be justified. This includes consideration of the expected
benefits that the imaging brings to
improving the treatment outcome. For example, requesting PET-CT for improved
target delineation or
daily IGRT.
5.191. Two particular groups of patients are identified in the BSS,
paragraph 3.157, for special
consideration with respect to justification – patients who may be pregnant and
paediatrics.
(a) Due to the higher radiosensitivity of the embryo/fetus, it
should be ascertained whether a
female patient is pregnant. Determining pregnancy status is a
requirement in the BSS,
paragraph 3.176, for those radiological procedures that could result in
significant dose to the
embryo or fetus. Pregnancy is a factor in the justification process
and might influence the
timing of the proposed treatment or whether another approach
to treatment is more
appropriate. Confirmation of pregnancy may occur after the initial
justification and before the
treatment commences or during treatment, in which case repeat justification is
required taking
into account the additional sensitivity of the embryo or fetus.
(b) As children may be at greater risk of incurring radiation-induced
stochastic effects, paediatric
treatments require special consideration in the justification process.
5.192. The decision of the multidisciplinary oncology team should be conveyed
to the patient or their
legal authorized representative. The patient, or their legal authorized
representative, also should be
informed about the expected benefits, risks and limitations of the proposed
treatment, as well as the
consequences of not undergoing the treatment. Women of fertile age
should be aware of the
possibility of becoming pregnant during treatment and the risk that this
entails. Consent for treatment
should be obtained before any further patient management action is initiated.
Justification – biomedical research volunteers
5.193. The role of the ethics committee in the justification of medical
exposure of volunteers exposed
as part of a programme of biomedical research is described in Section 2 para
2.98.
Justification – carers and comforters
5.194. The justification of medical exposure incurred by a carer or comforter
is effectively carried out
by the radiological medical practitioner or medical radiation technologist
involved in the treatment or
imaging together with the carer or comforter, prior to the giving of care or
comfort. It depends on the
carer or comforter being correctly informed about the radiation doses and risks
involved, and his or
her understanding of this information and consequent agreement to take
on the role of carer or
comforter.
Optimization of protection and safety
5.195. In medical exposures, optimization of protection and safety has
several components, some
applied directly to the radiological procedure(s) about to be
performed and others providing the
support or framework for the other components. These components of optimization
of protection and
safety are described in the following paras. 5.196 – 5.249.
Design considerations
5.196. The use of appropriate and well-designed medical radiological
equipment and associated
software underpins any treatment in radiation therapy. Linear
accelerators, X ray generators,
radioactive source based equipment (teletherapy and brachytherapy) and their
associated technologies
and accessories (including treatment planning systems) should be designed and
manufactured so as to
facilitate the aim of ensuring that for each patient the exposure of volumes
other than the planning
target volume is kept as low as reasonably achievable consistent with delivery
of the prescribed dose
to the planning target volume within the required tolerances.
Extensive guidance on design
considerations is given in the sub-section on medical radiological
equipment in radiation therapy,
paras. 5.51 to 5.77. Guidance on design considerations for imaging systems,
such as used in radiation
therapy for simulation, patient preparation, image-guidance and follow
up procedures, is given in
section 3, paras 3.26 to 3.39, and Section 4 paras. 4.39 – 4.50. Ultimately, as
stated in the BSS para.
3.162, it is the responsibility of the radiation therapy facility licensee to
ensure that his/her facility
uses only medical radiological equipment and software that meets applicable
international or national
standards.
Operational considerations
5.197. Following justification, the treatment planning and the
treatment delivery are performed in
such a way as to optimize patient protection, as required in the BSS paragraph
3.164. The treatment
goal is to deliver the correct absorbed dose to the correct volume within the
overall prescribed time
while keeping the dose to normal tissue and organs at risk within the
established tolerances and as low
as reasonably achievable. Accurate treatment planning is a crucial
precursor to achieving this
treatment goal.
5.198. Each radiation therapy facility should have written procedures and
protocols for the delivery of
radiation therapy consistent with the above goal. Protocols should be
consistent with current best
radiation therapy practice, published by the relevant professional
bodies – national, regional or
international, for example [331 – 336].
5.199. Advanced radiation therapy techniques (for example, IMRT, SRS,
HDR brachytherapy, and
ion beam therapy) have resulted in the possibility of high conformity
to target volumes or sub-
volumes and therefore dose delivery has very small margins for error.
When delivering radiation
therapy in this way, the radiation therapy facility should utilize
high quality imaging and delivery
equipment and, where applicable, immobilization devices.
5.200. The use of advanced technology has led to the delivery of higher doses
to the target volume,
and frequently uses more complex and unconventional field or source
arrangements. When moving to
more complex modes of delivery, there is a greater risk of error
and the radiation therapy facility
should have all the necessary expertise and resources available before
implementing these techniques.
5.201. Calculation of the dose to the embryo/fetus before the treatment of a
pregnant patient should be
part of the treatment plan. The distance from the field edge to the
embryo/fetus is the most important
factor in embryo/fetal dose, together with other factors such as field size,
angle and radiation energy
[130, 337].
5.202. Specific protocols for the use of imaging equipment (e.g. CT,
PET-CT) in the pre-planning
stage (simulation) of external beam radiation therapy should be
used to ensure appropriate
optimization of protection. In addition to the relevant guidance given
in Section 3 paras. 3.170 -
3.179, these should include the following:
(a) A medical radiation technologist specialized in radiation
therapy should always be present
when images for external beam radiation therapy planning are acquired
in a diagnostic
imaging facility;
(b) Patients should be in the treatment position for all images
acquired for external beam
radiation therapy planning;
(c) The geometry of the imaging modality should be sufficiently
accurate to minimize errors in
dose calculation and target delineation;
(d) When used as a virtual simulator, a CT scanner should have
a large enough bore so that
images can be acquired with the patient in the treatment position;
(e) A flat table top or flat inserts should be used when images
are acquired for external beam
radiation therapy planning;
(f) A reference system consistent with those in the treatment
rooms should be used when
acquiring images for external beam radiation therapy planning. The
treatment planning
system reference point and the patient treatment reference point should be
correlated;
(g) Imaging protocols for radiation therapy should include the
specific technical parameters
required for the simulation. For example, with CT this would include
Hounsfield number
calibration for dose computation accuracy, slice thickness for optimum
planning, the scan
length necessary to encompass the potential volume and other parameters that
may influence
the image quality for radiation therapy planning.
5.203. Specific protocols for the use of imaging equipment in IGRT
should be used to ensure
appropriate optimization of protection. In addition to the relevant guidance
given in Section 3 paras.
3.170 to 3.179, more specific guidance is given in Ref. [324].
Calibration – medical radiological equipment
5.204. The BSS, paragraph 3.167(a), states the requirements for calibration of
sources giving rise to
medical exposure. For radiation therapy, all external beam medical
radiological equipment and
brachytherapy sources used in the radiation therapy facility should be
calibrated, as follows:
(a) External beam radiation therapy medical radiological equipment
should be calibrated in terms
of radiation quality or energy and either absorbed dose or absorbed dose rate
at a predefined
distance under specified conditions; the recommended quantity is
absorbed dose to water
[319, 338]. The calibrations should be for at least the clinically used
energies and qualities.
(b) Sealed sources used for brachytherapy should be calibrated in terms
of reference air kerma
rate in air or equivalent quantity as recommended by the ICRU, at a specified
distance, for a
specified date [319].
(c) Internationally or nationally accepted calibration protocols should
be used. Examples of such
protocols include Refs. [327, 338 – 344].
(d) For brachytherapy, a distinction can be made between
removable and permanent implants.
For removable implants each source should be calibrated individually.
For permanent
implants when a large number of sources are being used, a
representative sample may be
assessed, for example 10% of the sources [341 – 344].
(e) Particular attention should be paid to the calibration of
sources used for special radiation
therapy procedures (e.g. radiosurgery, IORT, stereotactic radiotherapy,
tomotherapy, total
body irradiation) which may require adaptation of the existing international
codes of practice
and introduce additional uncertainties associated with making measurements in
non-reference
conditions. A particular consideration is small field dosimetry –
guidance is given in Ref
[345].
(f) Imaging devices used in the radiation therapy process, such as
conventional simulators, CT
scanners, CBCT, fluoroscopy, radiography and hybrid imaging systems
(PET-CT, SPECT-
CT) should be calibrated following the relevant recommendations in Sections 3
and 4, paras
3.194 – 3.198 and 4.195 – 4.200, respectively. Guidance for MV imaging devices
is given in
Refs [346, 347].
5.205. The BSS, paragraphs 3.154(d) and 3.167, places the responsibility for
calibration in radiation
therapy on the medical physicist, with either direct fulfilment or by
supervision. Correct calibration in
radiation therapy is fundamental and, with increasing complexity in technology
and software, the need
for the direct presence and involvement of the medical physicist is
fundamental. See also sub-sections
on patient dosimetry and the quality assurance programme for medical exposures,
below.
5.206. The BSS, paragraph 3.167(b), specifies when these calibrations should
occur. In addition to the
initial calibration prior to clinical use and calibration after major
maintenance or upgrade, periodic
calibrations should occur. The intervals for these calibrations may differ,
depending on the type of
source and unit. For example, linear accelerators should be calibrated at least
yearly. These intervals
will be specified by the regulatory body in each Member State. Constancy
checks are addressed in
paras. 5.224 – 5.243.
5.207. The BSS, paragraph 3.167(c), requires independent verification of
radiation therapy equipment
calibrations, prior to clinical use, because miscalibration of a radiation
therapy source can result in
inappropriate treatment involving many patients and can lead to
very serious consequences.
Independent verification ideally means verification by a different,
independent medical physicist
using different dosimetry equipment. However, other options, such as
verification by a second
medical physicist or only verification using a second set of equipment, or
using a remote dosimetry
audit (e.g. the IAEA/WHO TLD dosimetry audit) could be acceptable. In checking
for compliance,
the regulatory body should be aware of the limitations of local resources, but
nevertheless some form
of independent verification should take place.
5.208. The licensee of the radiation therapy facility should ensure that
independent verification of all
radiation therapy equipment calibrations is performed at least once
every two years through
participation in a national, regional or international programme. One of the
simplest mechanisms for
independent verifications of external beam calibration or physical
dosimetry is participation in the
IAEA/WHO thermoluminescent dosimetry postal dose quality audit. The
regulatory body should
encourage licensees to participate in this or similar programmes.
5.209. Sealed sources used for external beam and brachytherapy will also have a
calibration certificate
provided by the manufacturer, in accordance with Ref. [311] or its
national equivalent standards.
While important, this does not replace the calibrations required by
the BSS paragraph 3.167 and
described above in paras. 5.204 - 5.208.
5.210. New brachytherapy sources should be calibrated and differences of
more than 5% from the
manufacturer’s certified reference air kerma rate should be investigated. The
sources should not be
used for patient treatment until such differences have been
investigated and resolved. Further
guidance on resolving differences in calibrations is given in Ref [343].
Calibration – dosimetry instrumentation
5.211. Dosimetry instrumentation used at a radiation therapy facility
should be calibrated at
appropriate intervals. Detailed guidance is given in Ref [338]. A period of not
more than three years is
recommended for the reference instruments.
5.212. The BSS in paragraph 3.167(d) requires the calibration of
dosimetry instrumentation to be
traceable to a standards dosimetry laboratory (SDL). Ideally this
would be to the national SDL
(primary or secondary) in the State concerned, with access either directly or
through a duly accredited
calibration facility. However, it may be necessary for dosimetry
instruments to be sent to another
country or state if there is no national SDL in the country or state where the
instruments are used. To
ensure the calibration is maintained, the calibrated dosimeter should
be checked for consistency
periodically in the facility against a reference check source.
5.213. Given the expenses involved in calibrating dosimeters, it is
helpful if the radiation therapy
facility keeps the calibrated dosimeter as its “local standard” and uses it
only for primary calibrations.
Relative calibrations can be done with instruments intercompared with
the “local standard” on a
periodic basis.
5.214. Records of calibration measurements and associated calculations,
including uncertainty
determinations (budgets), should be maintained as described in para. 5.276.
Dosimetry of patients
5.215. The BSS, paragraph 3.168, gives the requirements for dosimetry
of patients in radiation
therapy. Dosimetry is required for each patient undergoing external
beam radiation therapy or
brachytherapy. There are two aspects to the patient dosimetry – absorbed doses
to the planning target
volume(s) and absorbed doses to specific organs and tissues that have been
identified as being at risk
by the radiological medical practitioner (radiation oncologist).
5.216. For external beam radiation therapy, the final doses delivered to a
patient are the result of a
multi-stage process, commencing with the treatment prescription, dated
and signed by the medical
radiological practitioner (radiation oncologist), which should contain the
following information: the
location of the treatment site(s), total dose, dose per fraction,
fractionation and overall treatment
period of each course per site. The treatment prescription should
indicate whether the radiation
therapy will be given alone or in combination, either
concomitantly or sequentially, with
chemotherapy and the timing of other local treatments such as surgery. The
normal tissues or organs
that may receive significant radiation should be identified and the maximum
doses to, and if possible
and necessary the volumetric distribution of dose in, these organs or tissues
at risk should be stated.
Such tissues or organs may be in the irradiated volume or they may receive
doses as a consequence of
leakage or scattered radiation. The treatment prescription is then
used as the basis for treatment
planning, followed by delivery and dose verification. The requirements
of the BSS can be met by
determining the absorbed doses to the planning target volume(s) and the
absorbed doses to specific
tissues and organs that have been identified as being at risk.
5.217. There are many different terms, concepts and approaches in use
for different aspects of
prescribing, recording and reporting of external beam radiation therapy. For
example, there are many
specifications of volumes, including gross tumour volume, clinical
target volume, planning target
volume, organ at risk, planning organ at risk volume. Radiation
therapy facilities should use the
international recommendations of the ICRU for the specification of
volumes and the prescribing,
recording and reporting of doses in external beam radiation therapy [348 –
355]. Further guidance on
dosimetry in external beam radiation therapy is given in Refs. [326, 337 – 340,
354 – 360].
5.218. For brachytherapy, the process also begins with the treatment
prescription, dated and signed by
the radiological medical practitioner (radiation oncologist). The treatment
prescription should contain
the following information: the total dose to a reference point and to
organs at risk, the size of the
reference dose volume, the number of sources and their dose distribution,
the radionuclide and the
source strength at a reference date. As with 5.219, the specification of
volumes and the prescribing,
recording and reporting of doses should follow the recommendations of
the ICRU [361 – 363].
Further guidance on dosimetry in brachytherapy is given in Refs. [359, 364 –
371].
5.219. Absorbed doses to organs as a result of imaging procedures carried out
as part of the radiation
therapy process should be considered both in the irradiated volume and in the
critical organs. While
this estimation does not need to have the accuracy required in the
determination of the doses to the
target volumes and normal tissues or organs at risk, these absorbed doses can
be considerable and they
should then be accounted for and added as appropriate. Guidance specific to
imaging doses during
IGRT is given in Ref [372].
5.220. Absorbed doses arising from neutrons when using high energy
photon beams should be
considered when determining doses to the irradiated volume and to the critical
organs. Methodologies
for this are still being developed [373].
5.221. Whenever appropriate, radiobiological considerations should be
incorporated into the treatment
decisions, for example by calculation of biologically effective doses [374].
Examples are when doses
from external beam radiation therapy and brachytherapy are added, or when the
patient has missed
some fractions due to clinical or technical reasons.
5.222. Treatment planning systems in radiation therapy continue to become more
and more complex
and, at the same time, they are used to predict the doses that the patient will
receive. Therefore, the
radiation therapy facility should have means to verify the dose to selected
points, independent from
the treatment planning systems calculations, for example by performing
by manual calculations or
case-specific QA measurements in a phantom. [321, 375, 376].
5.223. The radiation therapy facility medical physicist should perform
phantom and/or in vivo
measurements as appropriate. An example is to verify lung dose distributions
for total body irradiation
with photons.
Quality assurance for medical exposures
5.224. The BSS paragraph 3.170 requires radiation therapy facilities
to have a comprehensive
programme of quality assurance for medical exposures. General guidance on
management systems is
given in Section 2, paras 2.136 to 2.147, and it is simply
re-iterated here that the programme of
quality assurance for medical exposures should fit in with, and be
part of, the facility’s wider
management system. The paragraphs in this subsection give guidance on the
technical aspects of the
programme of quality assurance for medical exposures.
5.225. When planning and developing an effective quality assurance
programme for medical
exposures, licensees should recognize that it demands strong managerial
commitment and support in
the form of training and allocation of time, personnel and equipment resources.
5.226. The purpose of the programme of quality assurance for medical exposures
is to help ensure
successful implementation of optimization of patient protection in the
radiation therapy facility and to
minimize the occurrence of unintended and accidental medical exposures. The BSS
paragraph 3.171
establishes the elements of the programme.
5.227. By the very nature of radiation therapy, the facility’s
programme of quality assurance for
medical exposures will be complex and should encompass the entire
radiation therapy process,
including the treatment decision, tumour localization, patient positioning and
immobilization, image
acquisition for treatment planning, treatment planning, treatment delivery,
treatment verification and
follow up. With respect to equipment, instrumentation and systems, it
should include both the
hardware and software.
5.228. Measurements on medical radiological equipment used in radiation therapy
are an important
component of the programme. Acceptance tests are required for new or
significantly refurbished or
repaired equipment, or after the installation of new software or modification
of existing software that
could affect protection and safety of patients. The acceptance test
is followed immediately by
commissioning, and then on-going periodic quality control tests,
including constancy tests. The
purpose is to ensure that, at all times, all medical
radiological equipment is performing correctly,
accurately, reproducibly and predictably. Acceptance and commissioning tests
should be performed in
the same way for equipment and/or software that has been donated.
5.229. Acceptance tests and commissioning should not be restricted to radiation
emitting equipment
or sources, but should also be conducted for any system that has
implications for safety, such as
treatment planning systems (TPSs) and other software integral to or
supporting any stage of the
radiation therapy process. Insufficient understanding of TPSs at the
commissioning stage and
thereafter was involved in several accidental medical exposures [377 – 379].
5.230. After equipment or software installation has been completed, acceptance
testing should verify
conformance with the technical specifications given by the manufacturer and
stated in the purchase
agreement, and should verify compliance with relevant safety
requirements from IEC or other
recognized standards [294 – 309]. Depending on the equipment purchase
agreement, acceptance tests
may be performed by the manufacturer’s representative in the presence of the
local medical physicist
representing the user, or, if acceptable to the manufacturer and/or
the purchaser, by a medical
physicist jointly with the manufacturer’s representative. The tests to
be performed as part of the
acceptance testing should be specified in the purchasing conditions, where the
responsibility of the
manufacturer or supplier for resolving issues of non-conformity identified
during acceptance testing
should be clearly established.
5.231. Acceptance tests should ensure that equipment and/or software is
compatible with the other
equipment with which it will have to interface. The accuracy and integrity of
data including during
transfer processes should be verified.
5.232. After acceptance and before starting clinical use, commissioning of
equipment (hardware and
software) is performed – i.e. and radiation sources and radiation beams are
characterized and software
is customized for clinical use. The commissioning process is also a
very important stage for
familiarization of the staff with the equipment (hardware and
software) and for gaining a full
understanding of the equipment’s capabilities and limitations. The process
is critical, and therefore
essential, to safety as shown in reports on unintended and accidental medical
exposures involving a
large number of patients [23, 380, 381]. During commissioning the
medical physicist identifies,
measures and compiles all data required for clinical use. This is followed by
validation of the data
[285, 314, 320].
5.233. During commissioning the quantities and measures including tolerances
and action levels are
defined for the periodic quality control tests, setting the baseline for
subsequent constancy tests (see
also 5.238 below).
5.234. If there has been a major repair or modification or a source replacement
that may affect the
radiation protection and safety of patients, no treatment can take place until
the necessary QC tests
have been completed and checked by the medical physicist who has confirmed that
the equipment is
safe for use. Significant unintended and accidental medical exposure has
occurred because appropriate
tests were not performed following a repair [282, 380, 382].
5.235. As noted above in para. 5.228, the quality assurance
programme, with acceptance,
commissioning and on-going quality control tests should include software –
installation, upgrade, or
modification. A particular case is the software upgrade of a
treatment planning system where the
necessary actions may range from full commissioning to a partial
verification of the relevant
parameters. The medical physicist should be involved in this process.
Where remote software
modifications are possible, a protocol should be in place that
ensures the medical physicist is
informed prior to any modifications being carried out so that appropriate QC
tests can take place prior
to reintroduction of treatment.
5.236. There are many published reports from international and national
organizations and national or
regional professional bodies giving detailed guidance on the range of
acceptance, commissioning and
quality control tests that should be performed on the various
equipment and software used in the
different modalities in, and aspects of, radiation therapy, how they should be
performed, tolerances
and action levels, and recommended frequencies [193, 285, 314 – 316, 318, 320
- 322, 324, 325, 327,
328, 331 - 334, 341, 346, 375, 376, 383 - 401]. In addition, many
of these organizations and
professional bodies have websites where new or updated publications on the
topic can be found. The
regulatory body may have its own specific requirements on the tests that
should be performed and
their frequencies.
5.237. The quality assurance programme for medical exposures should
include testing of sealed
sources for leakage at regular intervals, as required by the
regulatory body. The quality assurance
programme should also include regular inventories of all radiation sources, at
intervals determined by
the regulatory body.
5.238. For guidance with respect to imaging medical radiological
equipment see Sections 3 and 4,
paras. 3.229 and 4.224, respectively. A diagnostic and a radiation therapy
medical physicist should be
present. Radiotherapy specific parameters that should be considered include,
for example, Hounsfield
number calibration for CT and geometric accuracy.
5.239. The results of the quality control tests should be compared with
established tolerance limits.
These limits may have been established to ensure compliance with a regulatory
requirement for the
performance of particular physical parameters or they may be set on the basis
of recommended values
given in published reports, such as referenced in para. 5.236.
As required in the BSS paragraph
3.170(b), if the measured values fall outside the tolerance limits, corrective
actions should take place.
Such corrective actions are likely to include maintenance or servicing of the
equipment, and hence the
radiation therapy facility should have a maintenance programme in
place. In some cases, the
equipment may be outside the tolerance limits by a significant amount and the
equipment should be
immediately taken out of clinical use and not returned until the servicing has
taken place and it has
been ascertained by the medical physicist that the equipment
now meets the performance
requirements for clinical use.
5.240. The programme of quality assurance for medical exposures in radiation
therapy should include
use of “checks and balances” to ensure that the facility’s protocols
and procedures for treatment,
including radiation protection and safety, are being followed – for example,
geometric and dosimetric
verification of the treatment. The periodic review of the protocols and
procedures themselves is part
of the facility’s radiological review (see paras. 5.273 – 5.275).
5.241. The BSS paragraph 3.171(e) specifically requires that dosimetry and
monitoring equipment are
part of the programme of quality assurance for medical exposures. This includes
instrumentation used
for the purposes of calibration and clinical dosimetry, such as ion chambers,
detectors, electrometers,
and beam scanners. The requirement is to ensure that such instrumentation has a
valid calibration (see
paras. 5.211 – 5.214), and that it is functioning correctly. Instrumentation
for calibration and clinical
dosimetry in radiation therapy should undergo acceptance testing and
regular quality control. The
programme of quality assurance for medical exposures should establish a
calibration cycle for each
instrument (see also para. 5.211) and a set of quality control tests on the
operation of each instrument
to be performed at regular intervals, such as
recommended internationally [338]. Preventive
maintenance procedures should be carried out on a regular basis.
5.242. Maintaining records is a crucial aspect of the programme of
quality assurance for medical
exposures. This includes the procedures used in the programme and all ensuing
results. In particular,
all data related to acceptance, commissioning, calibration and
dosimetry should be documented,
including independent verification. Records also should be kept for the results
of the periodic quality
control tests, and corrective actions. The regulatory body, during its
inspections of a radiation therapy
facility, should review the records of the programme of quality assurance for
medical exposures.
5.243. In line with standard practices for quality management, the BSS in
paragraph 3.172 requires
regular and independent audits of the programme of quality assurance for
medical exposures. Because
of the complexity of the radiological procedures being performed in a radiation
therapy facility, such
audits should be performed relatively frequently, perhaps every two
years for a radiation therapy
facility performing complex radiation therapy treatments. Such audits
may be external or internal.
Internal audits are usually logistically simpler to implement, while the
external audit generally has the
advantage of bringing in an outside perspective. The audit of the
radiation therapy facility’s
programme of quality assurance for medical exposures can be incorporated into
more comprehensive
audits that the facility is performing for its management system. Further, the
results of the audit of the
programme of quality assurance for medical exposures will be a major input into
the radiation therapy
facility’s radiological review (see paras. 5.273 - 5.275). If indicated from
the audit, the programme of
quality assurance for medical exposures should be updated or modified,
accordingly. Further,
feedback from operational experience and lessons learned from accidents or near
misses (see also para
5.270) can help identify potential problems and correct deficiencies,
and therefore should be used
systematically in improving the quality assurance programme [17, 19].
Dose constraints – carers and comforters
5.244. In radiation therapy, the potential for persons to act in the
role of a carer or comforter (as
defined in the BSS) is generally limited as accompanying a patient
during external beam radiation
therapy or access to HDR brachytherapy patients during treatment is generally
not allowed. However,
since LDR brachytherapy treatments last 2-3 days, visits by close relatives
could be allowed, provided
dose constraints for these carers or comforters are established
and implemented. Similarly,
brachytherapy treatments that involve permanent implants of sealed
sources may also lead to the
exposure of persons who, in the role of carers or comforters, provide care,
comfort and support to the
patient. This exposure of carers and comforters is defined as medical exposure
(see the BSS) and as
such is not subject to dose limits. However the BSS, paragraphs 3.153 and
3.173, requires that such
carers and comforters are afforded radiation protection through the application
of the requirements for
optimization of protection and safety and, in particular, the use of dose
constraints in this process.
These are the dose constraints established by government, as a result of
consultation with the health
authority, relevant professional bodies and the regulatory body, as
required by the BSS paragraph
3.149(a)(ii). Guidance on setting dose constraints, including considerations
for children and pregnant
women, is given in Section 2, paras, 2.48 - 2.49.
5.245. The radiation therapy facility should have written protocols for
implementing measures for the
optimization of protection for carers and comforters of LDR brachytherapy
patients or patients with
permanent implants. The measures should utilize the basic methods for
radiation protection, i.e.
shielding, time and distance. The protocol should include the following:
(a) Criteria specifying who would be acceptable as acting as a carer or
comforter;
(b) Methods for ensuring that the carer or comforter receives a dose
that is as low as reasonably
acceptable;
(c) The values of the dose constraints to be applied (see Section 2
para. 2.49).
5.246. Licensees should be able to demonstrate that the effective dose to the
carer or comforter, by
applying the protocols, is unlikely to exceed the dose constraints. It is
relatively straightforward to
estimate effective doses to carers and comforters from measurements of the
ambient dose equivalent
rates at the positions where they will be situated with respect to
the patient. These determinations
should be made in advance to verify that dose constraints will not be exceeded.
Therefore, individual
dose monitoring is normally not necessary.
5.247. The BSS paragraph 3.153 states that a carer or comforter must be
informed about the radiation
risks involved in providing care, comfort and support to the patient, and about
the means to be taken
to afford appropriate radiation protection to the carer or comforter.
The carer or comforter should
indicate that he/she has understood the information and is still willing to
care, comfort and support the
patient. Appropriate written instructions should be available and provided to
the carer or comforter.
5.248. Guidance applicable to carers and comforters supporting patients under
imaging radiological
procedures as part of the treatment process in the radiation therapy facility
is given in Section 3 paras.
3.236 to 3.240.
Dose constraints – volunteers in biomedical research
5.249. Participants in a programme of biomedical research may undergo radiation
therapy as part of
the research programme. Guidance on the role of the ethics committee in
approving such programmes
is given in Section 2, para 2.98, and this normally includes the setting of
applicable dose constraints
(Section 2 para. 2.99).
Pregnant females patients
5.250. Female patients who are pregnant form a special sub-group of
patients that should be
particularly considered with respect to radiation protection. These
considerations have been described
in para. 5.191 for justification and para. 5.201 for optimization. None of
these considerations can take
place if it is not known that the patient is pregnant. Therefore it is crucial,
as is required in the BSS
paragraphs 3.175 -3.176, for the radiation therapy facility to have in place
means for ensuring that
pregnancy status is known for female patients.
5.251. The first approach is through posting of clear signs, in languages able
to be understood by the
range of people using the radiation therapy facility, simply asking the
question “Are you pregnant or
possibly pregnant?” and, if so, asking them to inform staff. Such signs should
be posted widely in the
facility, including waiting rooms and cubicles. The second approach is to
directly ask female patients
whether they are or might be pregnant. This may not always be so
easy given social and cultural
sensitivities, but it should be done when needed.
5.252. Neither of the approaches described in para. 5.251 work if the woman
does not know if she is
or may be pregnant. For this reason, the BSS paragraph 3.176 has an
additional requirement on
facilities to have “procedures in place for ascertaining the pregnancy
status of a female patient of
reproductive capacity before the performance of any radiological
procedure that could result in a
significant dose to the embryo or fetus”. In radiation therapy this
situation is likely to occur, in
particular when it includes treatment of the abdomen or pelvis area and
treatment to volumes near the
uterus such that significant leakage and/or scattered radiation
reaches the embryo or fetus.
Cooperation with the referring medical practitioner, through standard requests
for pregnancy status for
specified treatments, is one approach. In case of doubt, a pregnancy
test or hormonal level
determination to assess menopausal status may be carried out.
Release of patients after permanent brachytherapy implants
5.253. As required in the BSS paragraph 3.178, a radiation therapy facility
should have arrangements
in place to manage the release of patients who have permanent
brachytherapy implants. Once the
patient is released, two groups of persons should be afforded appropriate
radiation protection – the
general public with whom the patient may encounter or interact, and members of
the patient’s family
and close friends who may be viewed simply as also being members of the public
or as carers and
comforters. Exposure of members of the public is subject to the
public dose limits (see Section 2,
Table 2), while exposure of carers and comforters is not subject to dose
limits but is instead controlled
through dose constraints (see Section 2 paras. 2.46 – 2.49 and
Section 5 paras. 5.244 – 5.248).
Further, as discussed in para 2.46, public exposure arising from a single
“source”, such as the patient
with the implants, should be subject to dose constraints set at some fraction
of the dose limits.
5.254. The radiation therapy facility’s medical physicist or radiation
protection officer should
establish prior to the release of a patient that the radioactivity of the
implants is such that the doses
that could be received would not exceed public dose limits, for members of the
public, and would be
unlikely to exceed the relevant dose constraints for both members of
the public and carers and
comforters. An acceptable method to estimate the acceptable activity
of permanent implants for
patients being discharged from hospitals is to calculate the time
integral of the ambient dose
equivalent rate, considering the activity, energy and half-life of the
radionuclides. In the case of carers
and comforters, the assumptions made for the calculations should be
consistent with the written
instructions that will be given at the time the patient is discharged from the
facility. Published data
would suggest that systematic dose monitoring, at least in the case
of permanent brachytherapy
implanted sources for prostate cancer, is not necessary [342, 402].
5.255. As indicated in 5.256, the patient or legal guardian of the
patient should be provided with
written instructions on how to keep doses to members of the public and carers
and comforters as low
as reasonably achievable. Areas of particular concern are children and pregnant
partners of patients.
Detailed guidance, including a sample information sheet, has been given by the
ICRP for implanted
sources for prostate cancer [402].
5.256. There is a low probability of an implanted seed being
expelled, for example with prostate
treatment. The written instructions should cover this possibility and give
guidance on what to do and
what not to do. Again detailed advice is given by the ICRP [402].
5.257. The patient with permanent brachytherapy implants should be
informed that if he/she is to
undergo subsequent surgery, then the surgeon should be informed of the presence
of the implants –
for example, a prostate cancer patient undergoing subsequent pelvic or
abdominal surgery. A wallet
card with all relevant information about the implant is useful [402].
5.258. Information also should be provided to the patient on radiation risks,
including guidance with
respect to fertility in the case of implants for prostate cancer [402].
Unintended and accidental medical exposures
Prevention of unintended and accidental medical exposures
5.259. The BSS, paragraphs 3.179 to 3.181, sets out requirements both for
minimizing the likelihood
of unintended and accidental medical exposures and for the ensuing
investigation if such exposures
occur. The pathways identified in the BSS for potentially leading to unintended
or accidental medical
exposures include flaws in the design of medical radiological
equipment, failures of medical
radiological equipment while in operation, failures and errors in software that
control or influence the
delivery of the radiation, and human error. General strategies for addressing
those pathways include
regular maintenance of medical radiological equipment and software, a
comprehensive quality
assurance programme, continuing education and training of staff, and
the promotion of a safety
culture. The lessons learned from events that have occurred should be
used for preventing or
minimizing unintended and accidental medical exposures , as described below in
para. 5.270.
5.260. A reduction in the probability of unintended or accidental
medical exposures in radiation
therapy can be brought about by:
(k) The introduction of safety barriers at identified critical
points in the radiation therapy
pathway, with specific quality control checks at these points. Quality control
is not confined
to radiological equipment physical tests or checks, and can include actions
such as checks of
the treatment plan or dose prescription by independent professionals.
(l) Actively encouraging a culture of always working with awareness and
alertness.
(m) Providing detailed protocols and procedures for each process
in the radiation therapy
pathway.
(n) Education and training, including continuing professional
development, of all staff involved
in the preparation and delivery of radiation therapy.
(o) Clear definitions of the roles, responsibilities and functions of
staff in the radiation therapy
facility that are understood by all staff.
5.261. Unusual and complex treatments should always trigger an extra
warning and each staff
member should be aware and alert in these situations. The use of ‘time-out’
where staff take time to
review what has been planned, prior to delivering treatment, should be
considered.
5.262. As noted in 5.262 above, comprehensive protocols and procedures covering
the various steps
in the process should exist for the major part of the department’s activity
[331 - 336, 396, 403 - 409].
Checklists detailing actions, and signed by the responsible parties at each
step, are very helpful [410].
For the most critical steps, such as commissioning and calibration of
equipment, there should always
be a review, either internally or preferably through an external independent
audit.
5.263. Preventive measures should include incident and near incident reporting,
analysis and feedback
including lessons learnt from international experience [19, 21, 282,
378, 380, 381, 411 – 414].
Preventive measures should also include each radiation therapy facility
checking the robustness of
their safety system against reported incidents. IAEA Safety Reports Series No.
17 [282] and ICRP
Publications 86 and 112 [380, 411] contain reviews of case histories from an
extensive collection of
accidental medical exposures.
5.264. Proactive risk assessment should also be carried out to try to pre-empt
incidents or potential
incidents. The tools used to carry out this type of analysis in radiation
therapy include, for example,
process maps or failure trees to facilitate the identification of possible
failure modes, and then the use
of prospective analyses, such as failure mode and effect analysis (FMEA), root
cause analysis (RCA),
and risk matrix to assess the probability and likely consequences of
such unacceptable events.
Detailed guidance on some of these tools and approaches is given by
the ICRP [411] and the
European Commission [405].
5.265. Before introducing a new technology into a radiation therapy facility,
general lessons obtained
from established technologies may still be useful but there will be no specific
lessons to share and to
apply. In this case, a proactive assessment is even more necessary.
This can be combined with an
early collection and sharing of experience and events by the first users of the
new technology, such as
through participation in SARON, ROSIS or similar [19, 21].
5.266. Building on the guidance from paras 5.259 to 5.265, the following
three-step strategy can help
to prevent unintended and accidental exposures in radiation therapy:
(g) allocate responsibilities to appropriately qualified health
professionals only and ensure that a
quality management system that includes radiation safety is in place;
(h) use the lessons learned from unintended and accidental medical
exposures to test whether the
quality management system, including for radiation safety, is robust
enough against these
types of events;
(i) identify other latent risks by posing the question “what else could
go wrong” or “what other
potential hazards might be present” in a systematic, anticipative
manner to all steps in the
radiation therapy process, using for example the proactive methods briefly
described in para.
5.264.
Investigation of unintended and accidental medical exposures
5.267. The events that constitute unintended or accidental medical exposures
are detailed in the BSS
paragraph 3.180, and for a radiation therapy facility include those associated
with imaging and with
treatment. For the former, reference should also be made to Section
3, paras. 3.251 - 3.255 and
Section 4, paras. 4.250 – 4.251. Unintended and accidental medical exposures
may occur at any stage
in the radiation therapy process. It needs to be stressed that for
treatment in radiation therapy,
unintended or accidental medical exposures may be either underexposures or
overexposures. It also
needs to be stressed that the events in the BSS paragraph 3.180 also include
‘near misses’, and these
nshould be considered in the same way as actual events.
5.268. One of the events identified in the BSS paragraph 3.180 is
when the dose or dose fraction
delivered was “substantially different from (over or under)” the prescribed
dose. Guidance regarding
the level of dose difference that would be considered as
substantially different can be found in
international or regional recommendations [380, 404]. The radiation therapy
facility should have in
place a system with clear procedures for identifying when
this type of event occurs. For example,
unintended or accidental medical exposures involving a total dose 10% or more
over that prescribed
should be detectable in most cases by the radiation oncologist or relevant
health professional, based
on an unusually high incidence of adverse patient reactions [380], and the
radiation therapy facility’s
procedures should include such patient monitoring to act as a trigger for
further investigation. Clinical
identification of situations of under dose is more difficult, but may
become manifest through poor
tumour control – again, such monitoring should be part of the radiation therapy
facility’s procedures.
In addition to the clinically based approaches to identifying doses
that were substantially different
from that prescribed, other approaches should be used in parallel, including
the review processes that
are part of quality assurance.
5.269. Each radiation therapy facility should have a system in place to manage
the investigation of
unintended and accidental medical exposures, and the ensuing actions
and reporting. The BSS,
paragraph 3.181, states what should be done in the course of the
investigation. This includes
calculation or estimation of patient doses, which should be performed
by a medical physicist,
identification and implementation of corrective actions, records of the
investigation and for the
radiological medical practitioner to inform the patient and the patient’s
referring medical practitioner.
A record of the calculation method and results should also be placed
in the patient file. When
required, counselling of the patient should be undertaken by someone with
appropriate experience and
clinical knowledge.
5.270. The investigation of unintended and accidental medical
exposures, as required by the BSS
paragraphs 3.180 – 3.181, has three main purposes. The first is to
assess the consequences for the
patient (s) affected and provide remedial and health care actions if
necessary. The second is to
establish what went wrong and how to prevent or minimize a
recurrence in the radiation therapy
facility – i.e. the investigation is for the facility’s and their patients’
benefit. The second purpose is to
provide information to other persons or other radiation therapy
facilities. Dissemination of
information about unintended and accidental medical exposures and
radiation injuries has greatly
contributed to improving methods for minimizing their occurrence. This
might be through the
regulatory body and/or the health authority for more significant events or as
required by a country’s
regulations, where the regulatory body disseminates information on the
event and the corrective
actions to other facilities that might learn from the event (see also para.
5.271). Another approach,
independent from any legal requirement for reporting to the regulatory
body, is to participate in
voluntary international or national databases designed as educative
tools. Two international such
databases for radiation therapy are the SAFRON and ROSIS reporting
systems [19, 21]. Facilities
performing radiation therapy should be active participants and users of SAFRON,
ROSIS or similar
international databases or equivalent national ones [415, 416].
5.271. As noted in para. 5.270, the BSS requires reporting to the regulatory
body, and to the health
authority if appropriate, for “significant” events. Further, the regulatory
body in a given country may
also specify their own requirements for reporting events to them. It
is difficult to quantify
“significant” – specifying a numerical trigger value immediately
creates an artificial distinction
between values immediately below that value (and hence should not be reported)
and those just above
the value (which should not be reported). However, the attributes of
“significant” events can be
elaborated, and events with one or more of these attributes should be reported
to the regulatory body.
Such attributes would include the occurrence of, or the potential for, serious
unintended or unexpected
health effects due to radiation, the likelihood of a similar event occurring in
other radiation therapy
facilities, a large number of patients were affected, and gross
misconduct or negligence by the
responsible health professionals. As stated in para 5.270 above, one of the
roles of the regulatory body
for such a reported event is to disseminate information on the
event and the lessons learned to all
potentially affected parties – typically other radiation therapy
facilities and relevant professional
bodies, but also in some cases manufacturers, suppliers, and maintenance
companies.
5.272. Irrespective of whether the event is also reported to the
regulatory body, feedback to staff
should be provided in a timely fashion and, where changes are
recommended, all staff should be
involved in bringing about their implementation.
Records and review
Radiological review
5.273. The BSS in paragraph 3.182 requires the performance of a periodic
radiological review at the
radiation therapy facility. This involves considering both justification
and optimization aspects of
radiation protection. For the latter, the results of the programme of
quality assurance for medical
exposures, including the periodic independent audit, would be a significant
input to the process. As
described in Section 2, paras. 2.146 - 2.147, the wider clinical audit could
include the radiological
review with its assessment of the application of the requirements for
justification and optimization in
the facility for the radiation therapy being performed [54, 417].
5.274. To facilitate compliance with the BSS, paragraph 3.182, and to learn
from periodic reviews, the
methodology used, the original physical, technical and clinical
parameters considered, and the
conclusions reached should be documented and taken into account prior to any
new review that may
result in an update of institutional policies.
5.275. In radiation therapy, reviews should consider patient outcome
(survival, control of disease,
acute and late side effects, etc.), the effect of introducing new technologies
on efficiency and cost,
such as, for example, the introduction of hypofractionation either for curative
or palliative intent. The
radiation therapy facility should have a system for the on-going collection of
relevant data to support
such reviews.
Records
5.276. Records should be in place to demonstrate on-going compliance
with radiation protection
requirements. The BSS, paragraphs 3.183 to 3.185, specify requirements
for keeping personnel
records, records of calibration, dosimetry and quality assurance, and
records for medical exposure.
These records should be kept for the period specified by the country’s
regulatory body. In the absence
of such a requirement, a suggested period for keeping records is 10
years. In the case of children,
records should be kept for a longer time.
5.277. In the case of records for a radiation therapy facility, care should be
taken to also retain the
records of the imaging radiological procedures (X ray and nuclear
medicine) performed while
preparing, planning, treating and verifying the treatment.
RADIATION PROTECTION OF THE PUBLIC
Introduction
5.278. Public exposure may arise, from the performance of radiation therapy, to
persons in and around
the radiation therapy facility.
5.279. The requirements of the BSS for public protection, paragraphs 3.117 -
3.123, 3.125 - 3.127,
and 3.135 - 3.137, apply to radiation therapy facilities. This
sub-section contains guidance very
specific to radiation therapy facilities. For more general and
comprehensive guidance on radiation
protection of the public, reference should be made to the IAEA Safety Guide
Radiation Protection of
the Public [24].
Non-occupationally exposed workers and visitors
5.280. Non-occupationally exposed workers are those persons who work
at the radiation therapy
facility but not in a role that is directly involved in the use of radiation,
for example clerical staff, and
cleaning personnel. It also includes those persons who work at the wider
medical facility where the
radiation therapy facility is located. These persons are to be
afforded the same level of radiation
protection as any member of the public, as required by the BSS paragraph 3.78.
5.281. Visitors to the radiation therapy facility include those persons who
will be undergoing radiation
therapy but refers to the time during the visit when the treatment or other
radiological procedure is not
taking place. Similarly for carers and comforters – any exposure other than
during the radiological
procedure in which they are helping will be public exposure. Other
visitors, including persons
delivering goods or supplies, sales personnel, accompanying persons and
escorts, and other patients in
the facility, are also considered members of the public.
External exposure and contamination
5.282. The primary means for protecting the public (non-occupationally exposed
workers and visitors)
is to ensure that the shielding of the radiation therapy facility (see paras.
5.42 – 5.50) is sufficient so
that public exposure resulting from being in any immediate adjacent areas,
including rooms above and
below, accessible by either non-occupationally exposed workers or visitors,
would be in compliance
with the public dose limits, and preferably less than any dose constraint that
the regulatory body may
have applied (see Section 2 paras. 2.16 – 2.17 and 2.46.)
5.283. Patients receiving permanent implants may expose members of the
public in the radiation
therapy facility and upon discharge. Patients receiving temporary implants may
also expose members
of the public in the radiation therapy facility. In the radiation
therapy facility, the RPO should
establish rules to ensure that the exposure of any member of the public will be
less than the public
dose limit and, preferably, lower than any applicable dose constraint.
An acceptable method to
estimate the acceptable retained activity for patients being discharged
is described in para. 5.254.
Assumptions made in these calculations with regard to time and distance should
be consistent with the
instructions given to patients at the time of discharge of the patient from the
radiation therapy facility.
Results of the calculations should be recorded. Examples of such calculations
are given in Ref. [418].
5.284. When deciding on the appropriate activity at discharge for a particular
patient, the licensee and
RPO should take into account the transport and the living conditions of the
patient, such as the extent
to which the patient can be isolated from other family members and the need to
manage safely the
patient’s excreta and body fluids that may contain a migrating
source. In some cases, such as for
elderly or child patients, it may be necessary to discuss the precautions to be
taken with other family
members.
5.285. Radiation protection precautions may be required after the death of a
patient with permanent
implants, for autopsy, embalming, burial or cremation. These precautions should
be determined by the
RPO, based on a generic safety assessment of the need for monitoring personnel
who carry out these
procedures, the need for monitoring the premises and the need for
minimizing external radiation
exposure and the potential for contamination. Whole body monitoring and finger
monitoring may be
required for autopsy and embalming personnel, as contamination and radioactive
waste are likely to
be generated [383]. Other considerations, such as cultural or ethical concerns,
should be taken into
account. A particular example is the cremation of patients with
permanent implants, where strict
radiation protection considerations would indicate the need to store the ashes
until adequate decay has
been achieved before releasing them to the family.
Control of access
5.286. Access to areas where radiation is being used should be controlled to
provide for controlling
doses to visitors and non-occupationally exposed workers. Visitors
should not be allowed to enter
treatment rooms or other controlled areas while in use. Exceptionally, a
visitor may be permitted to
enter a controlled or supervised area, accompanied at all times
by a staff member who knows the
protection and safety measures for the area. The radiation therapy
facility should have written
procedures specifying when such exceptions can take place and who may
do the accompanying.
Similarly, the facility should have established the rules regarding
non-occupationally exposed
workers, especially regarding access to controlled and supervised
areas. Particular attention, in all
cases, should be made with respect to potentially pregnant women.
5.287. Controlled and supervised areas should be clearly identified to help
prevent inadvertent entry
to areas where treatment or other radiological procedures are being performed.
See also para. 5.19.
Further control can be afforded by the use of keys (or passwords)
to restrict access to the control
panels of medical radiological equipment to only authorized persons.
Radioactive sources no longer in use
5.288. When radioactive sources in the radiation therapy facility become
surplus to requirements or
are no longer viable for their medical purpose, the licensee should ensure that
the sources are either
transferred or disposed of, appropriately. The licensee retains responsibility
for the sources until the
time of their transfer to another appropriate licensee or to an
authorized waste disposal facility.
Detailed guidance on the management of radioactive waste, applicable to
radiation therapy facilities,
is given in Ref [283].
5.289. Specifically for radioactive source teletherapy equipment, the licensee
should:
(a) Notify the regulatory body of any intention to transfer or
decommission ⁶⁰Co teletherapy
equipment prior to initiating an action. Depleted uranium used as shielding
material should also be
treated as radioactive waste. For example, a ⁶⁰Co teletherapy head may contain
depleted uranium and
is to be managed appropriately.
(b) Ensure that resources for the disposal of the sources will be made
available when the teletherapy
equipment is to be decommissioned.
5.290. Regulatory bodies may need to require applicants for licences to have in
place a programme for
safe disposal or return of the radioactive sources when their use is
discontinued, before authorization
for the import or purchase of equipment or radiation sources is
given. A contract with the
manufacturer or representative for the return of sources is acceptable evidence
of such a programme.
Monitoring and reporting
5.291. The programme for monitoring public exposure arising from radiation
therapy should include
dose assessment in the areas in and surrounding the radiation therapy facility
which are accessible to
the public. This can be achieved from the shielding calculations in the
planning stage, combined by
area monitoring at the initial operation of the facility and
periodically thereafter. Records of these
assessments should be kept for typically 7-10 years, but in any case
for a period that meets any
relevant regulatory requirements.
PREVENTION AND MITIGATION OF ACCIDENTS
Safety assessments
5.292. To comply with the BSS requirements for safety assessment (paragraphs
3.29 to 3.36), , the
registrant or licensee should conduct a safety assessment applied to
all stages of the design and
operation of the radiotherapy facility. The safety assessment report
should to be submitted to the
regulatory body if required. Basically, the safety assessment deals
with determining ‘what can go
wrong’ and how it can be prevented and, in case it occurs, how it can be
mitigated. Section 2, paras
2.148 – 2.152, describes general considerations for facilities using
ionizing radiation for medical
purposes.
5.293. The safety assessment should be systematic and contain
information on identification of
possible events leading to accidental exposure. Information on events, causes
and contributing factors
identified from reported accidents is available in Refs. [282, 378, 380, 381,
405, 406, 411 – 414] (see
also Appendix I for a summary of typical causes and contributing factors to
accidental exposures in
radiation therapy). The safety assessment should not only cover these
events, but also aim at
anticipating other events that have not previously been reported. Clearly the
safety assessment should
be documented.
5.294. The safety assessment should be revised when:
(a) New or modified radiation sources are introduced - including equipment
and new or renovated
facilities
(b) Operational changes occur, including workload;
(c) Operational experience or information on accidents or errors indicates that
the safety assessment is
to be reviewed.
5.295. Safety assessments in radiation therapy facilities performing
brachytherapy or teletherapy with
sealed sources should consider additional steps associated with sealed
sources, including ordering,
transport and receipt of sealed sources, unpacking, storage, preparation and
handling of the sources
prior to the use in the treatment of the patient, care of patients with
high amounts of activity, and
storage and handling of sources after removal and the management of unused
radioactive seeds.
5.296. The performance of safety assessments in radiation therapy can
be complemented by
participation in international networks to share information, such as SAFRON
and ROSIS [18, 20] or
national ones such as ROILS and NRLS [415, 416]. In order to ensure that the
safety assessment is
comprehensive and is not restricted to past events but also
anticipates other possible events,
consideration should also be given to using systematic techniques,
e.g. fault and event trees and
probabilistic safety assessment technique, such as described for
unintended or accidental medical
exposure of patients in paras 5.264.
5.297. For radiation therapy, as discussed in para 5.259, possible
scenarios for potential exposure
include flaws in the design of medical radiological equipment,
failures of medical radiological
equipment while in operation, failures and errors in software that control or
influence the delivery of
the radiation, and human error. Potential exposure can also arise
from imaging during patient
preparation, simulation in treatment planning and guidance during.
Accident prevention
5.298. Accident prevention is clearly the best means for avoiding
potential exposure and the BSS,
paragraphs 3.39 to 3.41, set out requirements based on good engineering
practice, defence in depth,
and facility-based arrangements to achieve this. Design considerations
for medical radiological
equipment, ancillary equipment and the radiation therapy facility are described
in paras 5.8 to 5.83.
5.299. The licensee should incorporate:
(a) Defence in depth measures to cope with identified events, and evaluation of
the reliability of the
safety systems (including administrative and operational procedures, equipment
and facility design).
(b) Operational experience and lessons learned from accidents and
errors [282, 380, 411]. This
information should be incorporated into the education and training, maintenance
and quality assurance
programmes.
Mitigation and contingency plans
5.300. On the basis of events identified by the safety assessment
for the radiotherapy facility,
contingency plans should be prepared for events associated with
potential exposure, including
allocation of responsibilities and resources, the development and
implementation of procedures, and
the provision of training and periodic retraining of the relevant
staff in executing the mitigation
measures. Such contingency plans should be based on the events identified by
the safety assessment
for the radiation therapy facility.
5.301. The exposure of workers involved in such radiation therapy events cannot
be considered an
unexpected exposure and whether deliberate or not should be
controlled, and the dose limits for
workers in planned exposure situations should apply.
5.302. For medical exposure, potential exposure when it does occur is manifest
as an unintended or
accidental medical exposure. Means for preventing or minimizing unintended and
accidental medical
exposures in radiation therapy are described in paras. 5.259 – 5.266, and the
ensuing investigation and
corrective actions are described in paras. 5.267 – 5.272.
5.303. The BSS, paragraphs 3.43 to 3.44, sets out the requirements for
emergency preparedness and
response. Situations that may be treated as emergency should be identified in
the safety assessment,
and emergency plan should be prepared according to the requirements
in the GSR Part 7 and
guidelines by the regulatory authority [7]. Potential situations that
can lead to emergency in
radiotherapy department settings are loss of control over a dangerous
radiation therapy source in
result of unauthorized or malicious act, or conventional emergencies such as
fires and earthquakes.
Stuck sources, general
5.304. Contingency procedures should be short, concise, unambiguous and, if
necessary, illustrated
with drawings without explanatory text. They should be able to be read at
‘first sight’ and followed. It
should be made clear that the first sight procedures refer to actions to be
taken immediately to prevent
or limit serious overexposures, or take other lifesaving actions [396]. Further
actions to recover the
sources, to repair and test the equipment for returning it to use, are not
directly part of the contingency
response. The occupational protection requirements, including dose
limits, will apply to these
subsequent actions.
5.305. In radiation therapy, however, the patient is directly in the
radiation beam or brachytherapy
sources are placed inside the patient; for this reason some of the
contingency response actions
coincide with source recovery actions, for example the retrieval of
remote control brachytherapy
sources from the patient to the safe, either manually, electrically or using
the manual crank.
Stuck sources, cobalt 60
5.306. Contingency procedures should be posted at the treatment unit.
These procedures should
ensure that the patient is removed from the primary beam as quickly and
efficiently as possible whilst
minimizing unnecessary exposure to the involved personnel.
5.307. In the case of an event, the first step is to note the time, and
immediately use the source driving
mechanism to return the source to the shielded position. If there is a patient
on the treatment couch,
the patient should be removed from the area and the area should be
secured from further entry.
Emphasis should be placed on avoiding exposure of personnel to the
primary beam. The medical
physicist or radiation protection officer should notified and take
control of the situation, including
deciding when it is safe to re-enter the room. Before resuming patient
treatment, the medical physicist
should check the calibration of the radiation therapy and verify that it has
not changed, particularly the
timer error in Co-60 teletherapy units.
5.308. Actions are to be performed only by personnel that have been
educated and trained in the
contingency actions understand them and have regularly rehearsed the
procedures. The actions
required should be performed within seconds. Therefore personnel present in all
procedures, i.e. the
staff on the unit, should be educated and trained in these actions.
5.309. After the contingency actions, the following should be done:
(a) The maintenance or service engineer should be contacted to perform an
inspection of the machine;
(b) The medical physicist should assess the patient doses and clear the use of
the machine after the
ensuing maintenance;
(c) The radiation protection officer should assess the doses to personnel as a
result of the contingency
actions and recovery operations;
(d) A record should be kept of all actions;
(e) The regulatory body may need to be notified, depending on the country’s
regulations;
(f) Information should be sent to an international safety learning system such
as SAFRON or ROSIS
[20, 22] or national ones [415, 416].
Stuck sources, remote control brachytherapy units
5.310. Contingency plans require having an emergency container available in the
treatment room, as
well as an emergency kit containing long handled forceps for manipulation of
the source guide tubes,
and applicators if the source fails to return to the safe, as
stated in paras. 5.129 and 5.131. The
emergency container should be placed close to the patient and should be
sufficiently large to accept
the entire applicator assembly containing the source which has been removed
from a patient.
5.311. In HDR applications the short response time required for
contingency actions (minutes)
imposes the need for the immediate availability of a radiological
medical practitioner, a medical
physicist and a medical radiation technologist during all applications. Each
one of these professionals
should be educated and trained in contingency procedures and actions.
5.312. Manufacturers usually provide suggested contingency procedures if the
source fails to return to
the safe. They assume that the physical integrity of the applicator is
maintained. These procedures are
specific to the actual afterloading unit but generally involve a standard
sequence, as detailed in para.
5.132.
5.313. After the contingency actions have been successful, the following should
be done:
(a) The maintenance or service engineer should be contacted to
perform an inspection and, if
necessary, repair the machine;
(b) The medical physicist should make an assessment of the patient doses
arising from the incident,
and clear the re-use of the machine after the ensuing maintenance;
(c) The radiation protection officer should make an assessment of the dose to
personnel involved in
the contingency actions and recovery operations;
(d) The assessments should be recorded;
(e) The regulatory body may need to be notified, depending on the country’s
regulations.
(f) Information should be sent to an international safety learning system such
as SAFRON or ROSIS
[19, 21] or national ones [415, 416].
Incidents during source replacement
5.314. Only trained and authorized maintenance or servicing personnel should
handle incidents during
a source change of external beam therapy and remote control brachytherapy
units. If the participation
of radiation therapy personnel is necessary for any of these actions,
the scope of this participation
should be limited to operating the equipment. The respective
responsibilities of radiation therapy
personnel and maintenance or servicing personnel for these specific
situations should be clearly
defined.
Contamination
5.315. Although ²²⁶Ra has been removed from most radiation therapy
facilities, encapsulated ¹³⁷Cs
sources used in manual afterloading still exist, and there is always a
possibility that the encapsulation
may rupture. In the case of a contamination event the area should be closed to
further entry and that
all who were in the area remain to be surveyed and decontaminated if necessary.
If there are windows
or other ventilation systems, these should be closed or turned off.
The radiation protection officer
should be contacted immediately once the possibility of contamination is
suspected. Contact details
for the radiation protection officer should be posted throughout the radiation
therapy facility.
Lost radiation therapy sources
5.316. A detailed, up to date inventory of all sources should be
maintained by the RPO of the
radiation therapy facility so that it can be determined immediately which
sources are missing, their
type and activity, their last known location, and who last had possession of
them. The area where the
sources were last known to be should be closed to entry and exit
until after a survey. This search
should be performed with the most sensitive radiation detection survey meter
available.
5.317. When the source cannot be located and it is suspected that it is off
site, emergency plan should
be followed, according to the requirements in the BSS ([3], Section
4) for emergency exposure
situations. Thee regulatory body should be informed immediately. These
situations typically require
actions by national intervention organizations [7].
SAFETY IN THE TRANSPORT OF RADIOACTIVE MATERIALS
5.318. The BSS paragraph 2.25 sets out the requirements for the transport of
radioactive materials,
invoking in particular the IAEA Regulations for the Safe Transport of
Radioactive material [266]. The
IAEA Regulations for the Safe Transport of Radioactive material use the defined
terms “consignor” to
mean any person, organization or government that prepares a
consignment for transport, and
“consignee” to mean any person, organization or government that is
entitled to take delivery of a
consignment. “Consignment” is also a defined term, meaning any package or
packages, or load of
radioactive material, presented by a consignor for transport.
5.319. The licensee of a radiation therapy facility may be both a consignee and
a consignor, and hence
may have responsibilities for both receipt and shipment of radioactive sources
– for example, external
beam sources and brachytherapy sources.
5.320. The detailed requirements for the safe transport of radioactive
material, including general
provisions, activity limits and classification, requirements and controls for
transport, requirements for
radioactive material and for packagings and packages, test
procedures, and approval and
administrative requirements, are given in the IAEA Regulations for the Safe
Transport of Radioactive
material [283]. Emergency arrangements during the transport of
radioactive material should be in
place, in line with the requirements in the GSR Part 7 and guidelines by the
regulatory authority [7].
The licensee and the RPO of the radiation therapy facility should be familiar
with these regulations to
ensure that their transport of radioactive materials will be in compliance.
APPENDIX I
SUMMARY OF TYPICAL CAUSES OF AND CONTRIBUTING FACTORS TO
ACCIDENTAL EXPOSURES IN MEDICAL USES OF IONIZING RADIATION
DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES
I.1. Problems leading to accidental exposures associated with using radiation
in diagnostic radiology
and image guided interventional procedures that have been identified from
reported events include:
• Equipment not meeting IEC or equivalent national standards;
• Maintenance errors;
• Errors in the identification of patients and examination sites;
• Inappropriate or lack of examination protocols.
I.2. Factors that may influence the frequency and severity of accidental
exposures include:
• Insufficient training and expertise of radiological medical practitioners
(in particular
interventionists), medical physicists or medical radiation technologists;
o Lack of knowledge about the equipment being used and its features and
options;
o Lack of knowledge about optimization of protection for patients;
o Lack of knowledge about optimization of protection for staff;
• No reassessment of staffing requirements after purchasing new equipment
or increasing
workload;
• Inadequate quality assurance and lack of defence in depth;
o Dose rates for interventional equipment set too high;
o AEC malfunction;
• Lack of a programme for acceptance tests and commissioning of equipment;
• Lack of a maintenance programme;
• Poor, misunderstood or violated procedures;
• Lack of operating documents in a language understandable to users;
• Lack of dose or dose rate display during a procedure;
• Lack of dose alerts if selected factors seem inappropriate;
• Lack of radiation protection tools and devices in the examination room;
• Misunderstanding of displays or software messages;
• Inattention;
• Inconsistent use of different quantities and units.
I.3. In most accidental exposures there was a combination of several
contributing factors, which can
be summarized as:
• Lack of commitment of the licensee (medical facility and/or radiology
facility administrators
and managers);
• Staff insufficiently trained;
• Insufficient quality assurance.
NUCLEAR MEDICINE
I.4. Problems leading to accidental exposures associated with using radiation
in nuclear medicine that
have been identified from reported events include:
• Communication errors, faulty transmission of information,
misunderstanding of prescriptions
and protocols, or use of obsolete protocols;
• Errors in the identification of the patient;
• Use of the wrong source, the wrong radiopharmaceutical or the wrong
activity;
• Calibration errors;
• Maintenance errors.
I.5. Factors that may influence the frequency and severity of accidental
exposures include:
• Insufficient training and expertise of radiological medical practitioners
(nuclear medicine
physicians), medical physicists or medical radiation technologists (nuclear
medicine
technologists);
• No reassessment of staffing requirements after purchasing new equipment,
hiring new
medical radiation technologists or increasing workload;
• Inadequate quality assurance and lack of defence in depth;
• Lack of a programme for acceptance tests and commissioning of equipment;
• Lack of a maintenance programme;
• Poor, misunderstood or violated procedures;
• Lack of operating documents in a language understandable to users;
• Misunderstanding of displays or software messages;
• Inattention;
• Inconsistent use of different quantities and units.
I.6. In most accidental exposures there was a combination of several
contributing factors, which can
be summarized as:
• Lack of commitment of the licensee (medical facility and/or nuclear
medicine facility
administrators and managers);
• Staff insufficiently briefed or trained;
• Insufficient quality assurance.
RADIATION THERAPY
I.7. Problems leading to accidental exposures associated with using radiation
in radiation therapy that
have been identified from reported events include:
Common to external beam radiation therapy and brachytherapy:
• Equipment not meeting IEC or equivalent national standards;
• Maintenance errors;
• Errors in the identification of patients and treatment sites;
• Conflicting signals and displays misinterpreted or not followed up;
• Communication errors, transmission of information and misunderstanding of
prescriptions
and protocols, or use of obsolete protocols;
• Use of obsolete files and forms which were still accessible.
External beam radiation therapy:
• Errors in acceptance tests and commissioning or lack of tests of both
radiation equipment and
sources and TPSs;
• Errors in the calibration of radiotherapy beams;
• Errors in the preparation of tables and curves from which the treatment
time is calculated;
• Errors in the use of TPSs for individual patients.
Brachytherapy:
• Using an incorrect source or incorrect units of source strength;
• Dislodging of HDR brachytherapy sources;
• Mistakes in source handling by nurses during brachytherapy treatment;
• Leakage of sealed sources;
• Sources left in patients and loss of radiation sources.
I.8. The following contributing factors allowed these errors to remain
undetected until they became
accidental medical exposures:
• Insufficient education of the radiological medical practitioner
(radiation oncologist), medical
physicist, medical radiation technologist (radiotherapy technologist),
maintenance engineers
and brachytherapy nurses;
• Overloaded staff when new equipment was purchased or workload increased;
• Insufficient quality assurance and lack of independent checks for safety
critical activities,
such as beam calibration;
• Lack of a programme for acceptance testing and commissioning;
• Lack of a maintenance programme;
• Poor, misunderstood or violated procedures;
• Lack of operating documents in a language understandable to the users;
• Inattention (environment prone to distraction);
• Inconsistent use of quantities and units.
I.9. In a number of the reported accidents there was a combination of several
of the above contributing
factors. Concurrent occurrence of several contributing factors may be
indicative of a more general
problem involving:
• Lack of commitment of the licensee (medical facility and/or radiation
therapy facility
administrators and managers);
• Insufficiently educated or trained staff;
• Insufficient quality assurance and defence in depth.
APPENDIX II
AVOIDANCE OF PREGNANCY FOLLOWING RADIOPHARMACEUTICAL
THERAPY
II.1 The periods for which it is recommended to avoid pregnancy
following radiopharmaceutical
therapy with long-lived radionuclides are given in Table 3, adapted from Ref
[251].
TABLE 3. RECOMMENDATIONS FOR AVOIDANCE OF PREGNANCY FOLLOWING
RADIOPHARMACEUTICAL THERAPY
Nuclide and form Disease
All activities up to*
(MBq)
Avoid pregnancy
(months)
³²P-phosphate Polycythaemia and related disorders
200 3
⁸⁹Sr-chloride Bone metastases
150 24
⁹⁰Y-colloid Arthritic joints
400 0
⁹⁰Y-colloid Cancer
4000 1
¹³¹I-iodide Thyrotoxicosis/non-toxic goitre
800 6 (at least)
¹³¹I-iodide Thyroid cancer
6000 6 (at least)
¹³¹I MIBG
Phaeochromocytoma 7500
3
¹⁵³Sm-colloid Bone metastases
2600 1
¹⁶⁹Er-colloid Arthritic joints
400 0
* Note: The administration of activities smaller than those indicated in column
3 does not imply that
the advisory period specified in column 4 may be reduced.
.
APPENDIX III
CESSATION OF BREAST FEEDING
III.1. Recommendations for cessation of breast feeding following
administration of various
radiopharmaceuticals for diagnostic procedures are given in Table
4, adapted from the
recommendations of Refs. [214, 247, 251]. In addition, it is recommended that
there is a cessation
period of 4 hours for breast-feeding patients undergoing a PET procedure with
¹⁸F-FDG, taking into
account both exposure from breast milk and external exposure to the baby while
being fed [65].
TABLE 4. RECOMMENDATIONS FOR CESSATION OF BREAST FEEDING FOR
RADIOPHARMACEUTICALS IN DIAGNOSTIC EXAMINATIONS
Radiopharmaceutical Administered activity
to the mother, MBq
(mCi)
Need for
counselling
Advice
³²P-phosphate Any
Yes Cessation
⁵¹Cr-EDTA 1.85 (0.05)
No
⁶⁷Ga-citrate 185 (5.0)
Yes Cessation
⁸¹ᵐKr-gas 6000 (160)
No
⁹⁹ᵐTc-DISIDA 300 (8)
No
⁹⁹ᵐTc-DMSA 80 (2)
No
⁹⁹ᵐTc-DTPA 740 (20)
No
⁹⁹ᵐTc-exametazime 500 (14) No
⁹⁹ᵐTc-glucoheptonate 740 (20) No
⁹⁹ᵐTc-HAM 300 (8)
No
⁹⁹ᵐTc-MAA 148 (4)
Yes Cessation for 12 hours
⁹⁹ᵐTc-MDP 740 (20)
No
⁹⁹ᵐTc-MIBI 1110 (30)
No
⁹⁹ᵐTc-pertechnetate 800 (20)
Yes Cessation for 40 hours
⁹⁹ᵐTc-PYP 740 (20)
No
⁹⁹ᵐTc-RBCs in vitro 740 (20)
No
⁹⁹ᵐTc-RBCs in vivo 740 (20) Yes
Cessation for 12 hours
⁹⁹ᵐTc-sulphur colloid 444 (12)
No
⁹⁹ᵐTc-DTPA aerosol 37 (1)
No
⁹⁹ᵐTc-MAG3 370 (10)
No
⁹⁹ᵐTc-WBCs 185 (5)
Yes Cessation for 48 hours
¹¹¹In-WBCs 18.5 (0.5)
No
¹²³I-MIBG 370 (10)
Yes Cessation for 48 hours
¹²³I-NaI 14.8 (0.4)
Yes Cessation
¹²³I-OIH 74 (2)
No
¹²⁵I-fibrinogen Any
Yes Cessation
¹²⁵I-HSA Any
Yes Cessation
¹²⁵I-OIH 0.37 (0.01)
No
¹³¹I-NaI 5550 (150)
Yes Cessation
¹³¹I-OIH 11.1 (0.3)
No
²⁰¹Tl 111 (3)
Yes Cessation for 96 hours
Note: The abbreviations used in this table are as follows:
DISIDA, di-isopropyl-iminodiacetic acid;
DMSA, dimercaptosuccinic acid;
DTPA, diethylene-triamine-penta-acetic acid;
EDTA, ethylene-diamine-tetra-acetic acid;
HAM, human albumin microsphere;
HAS, human serum albumin;
MAA, macro-aggregated albumin;
MAG3, mercapto-acetyl-triglycine;
MDP, methylene diphosphonate;
MIBG, meta-iodo-benzyl-guanidine;
MIBI, methoxy-isobutyl-isonitrile;
OIH, ortho-iodo-hippurate;
PYP, pyrophosphate;
RBCs, red blood cells;
WBCs, white blood cells;
REFERENCES
[1] UNSCEAR, Sources and Effects of Ionizing Radiation, UNSCEAR 2008 Report
to the General
Assembly with Scientific Annexes, Vol. 1, United Nations, New York (2010)
[2] EUROPEAN ATOMIC ENERGY COMMUNITY, FOOD AND AGRICULTURE
ORGANIZATION OF THE UNITED NATIONS, INTERNATIONAL ATOMIC ENERGY
AGENCY, INTERNATIONAL LABOUR ORGANIZATION, INTERNATIONAL
MARITIME ORGANIZATION, OECD NUCLEAR ENERGY AGENCY, PAN AMERICAN
HEALTH ORGANIZATION, UNITED NATIONS ENVIRONMENT PROGRAMME,
WORLD HEALTH ORGANIZATION, Fundamental Safety Principles, IAEA Safety Standards
Series No. SF-1, IAEA, Vienna (2006).
[3] EUROPEAN COMMISSION, FOOD AND AGRICULTURE ORGANIZATION OF THE
UNITED NATIONS, INTERNATIONAL ATOMIC ENERGY
AGENCY,
INTERNATIONAL LABOUR ORGANIZATION, OECD NUCLEAR ENERGY AGENCY,
PAN AMERICAN HEALTH ORGANIZATION, UNITED NATIONS ENVIRONMENT
PROGRAMME, WORLD HEALTH ORGANIZATION, Radiation Protection and Safety of
Radiation Sources: International Basic Safety Standards, IAEA Safety Series No.
GSR Part 3,
IAEA, Vienna (2014).
[4] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, The
2007
Recommendations of the International Commission on Radiological
Protection, Publication
103, Ann. ICRP 37 2–4, Elsevier Science, Oxford (2007).
[5] INTERNATIONAL ATOMIC ENERGY AGENCY, IAEA Safety Glossary:
Terminology
Used in Nuclear Safety and Radiation Protection, 2007 Edition, IAEA, Vienna
(2007).
[6] INTERNATIONAL ATOMIC ENERGY AGENCY. Safety Guide on Non-medical
human
imaging. DS471 (new under development).
[7] FOOD AND AGRICULTURE ORGANIZATION OF THE UNITED
NATIONS,
INTERNATIONAL ATOMIC ENERGY AGENCY, INTERNATIONAL LABOUR
ORGANIZATION, OECD NUCLEAR ENERGY AGENCY, PAN AMERICAN HEALTH
ORGANIZATION, UNITED NATIONS OFFICE FOR THE COORDINATION OF
HUMANITARIAN AFFAIRS, WORLD HEALTH ORGANIZATION, Preparedness and
Response for a Nuclear or Radiological Emergency, IAEA Safety Standards Series
No. GSR
Part 7, IAEA, Vienna (2014)(to be updated).
[8] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Patient Dosimetry for X Rays Used in Medical Imaging, ICRU Rep. 74, ICRU,
Bethesda, MD
(2006).
[9] INTERNATIONAL ATOMIC ENERGY AGENCY, Dosimetry in Diagnostic Radiology:
An
International Code of Practice, Technical Reports Series No. 457, IAEA, Vienna
(2007)
[10] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Radiation Dosimetry and Image Quality Assessment in Computed Tomography, ICRU
Rep. 87,
ICRU, Bethesda, MD (2013).
[11] INTERNATIONAL ATOMIC ENERGY AGENCY, Governmental, Legal and
Regulatory
Framework for Safety, IAEA Safety Standards Series No. GSR Part 1, IAEA, Vienna
(2010).
[12] EUROPEAN COMMISSION. Radiation protection 109. Guidance on
diagnostic reference
levels (DRLs) for medical exposures. Directorate-General, Environment,
Nuclear Safety and
Civil Protection (1999).
[13] IPEM, BIR, RCR, NRPB and CoR, Guidance on the Establishment
and Use of Diagnostic
Reference Levels for Medical X-Ray Examinations, Produced by a joint
Working Party of
IPEM, BIR, RCR, NRPB and CoR (2004).
[14] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS,
Reference Levels and Achievable Doses in Medical and Dental Imaging:
Recommendations for
the United States, NCRP Rep. 172, NCRP, Bethesda, MD (2012).
[15] EUROPEAN COMMISSION, European ALARA Network, 2007.
http://www.eu-
alara.net/index.php/surveys-mainmenu-53/36-ean-surveys/156-drls.html
[16] EUROPEAN COMMISSION. DDM2 Project Report Part 2: Diagnostic
Reference Levels
(DRLs) in Europe Contract ENER/2010/NUCL/SI2.581237 Study on European
Population
Doses from Medical Exposure (Dose
Datamed 2, DDM2).
http://ddmed.eu/_media/news:ddm2_project_report_part_2_final_rb_16052014.pdf
(to be
updated once the final report is published)
[17] AUSTRALIAN RADIATION PROTECTION AND NUCLEAR SAFETY
AGENCY
(ARPANSA), National Diagnostic Reference Level Database (NDRLD) [Online],
ARPANSA,
Yallambie, (2011). http://www.arpansa.gov.au/services/ndrl/index.cfm.
[18] INTERNATIONAL ATOMIC ENERGY AGENCY, Release of patients after
radionuclide
therapy, Safety Report Series 63, IAEA, Vienna (2009).
[19] INTERNATIONAL ATOMIC ENERGY AGENCY, Safety in
Radiation Oncology
(SAFRON), https://rpop.iaea.org/SAFRON
[20] INTERNATIONAL ATOMIC ENERGY AGENCY, Safety in Radiological
Procedures
(SAFRAD), https://rpop.iaea.org/safrad/
[21] ROSIS. Radiation Oncology Safety Information System,
http://www.rosis.info/
[22] INTERNATIONAL ATOMIC ENERGY AGENCY, Regulatory control of radiation
sources,
Safety Standards Series No. GS-G-1.5, IAEA, Vienna (2004).
[23] INTERNATIONAL ATOMIC ENERGY AGENCY, Occupational Radiation Protection,
Safety
Standards Series No. DS453
[24] INTERNATIONAL ATOMIC ENERGY AGENCY, Radiation Protection of
the Public,
Standards Series No. DS432
[25] INTERNATIONAL ATOMIC ENERGY AGENCY, Application of the management
system
for facilities and activities, Safety Standards Series No.GS-G-3.1, IAEA,
Vienna (2006).
[26] INTERNATIONAL ATOMIC ENERGY AGENCY, Roles and Responsibilities, and
Education
and Training Requirements for Clinically Qualified Medical Physicists, Human
Health Series
no. 25, IAEA, Vienna (2013).
[27] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION,
Radiological
Protection in Biomedical Research, Publication 62. Ann. ICRP 22 (3), 1992.
[28] AUSTRALIAN RADIATION PROTECTION AND NUCLEAR SAFETY AGENCY, Code of
Practice for the Exposure of Humans to Ionizing Radiation for Medical
Research Purposes.
Radiation Protection Series RPS 8, ARPANSA, Yallambie (2005).
[29] INSPECTION GÉNÉRALE DES AFFAIRES SOCIALES. Summary of ASN report n° 2006
ENSTR 019 - IGAS n° RM 2007-015P on the Epinal radiotherapy accident (2007).
[30] WORLD HEALTH ORGANIZATION, Procurement process resource guide,
WHO Medical
device technical series, ISBN 978 92 4 150137 8, World Health
Organization (2011).
http://whqlibdoc.who.int/publications/2011/9789241501378_eng.pdf
[31] EUROPEAN COORDINATION COMMITTEE OF THE
RADIOLOGICAL,
ELECTROMEDICAL AND HEALTHCARE IT INDUSTRY (COCIR), Medical Electrical
Equipment - Good refurbishment practice,
Brussels, COCIR, 2009.
http://www.cocir.org/site/fileadmin/6.1_Initiatives_Refurbishment/COCIR_Industry
_standard_
on_Good_Refurbishment_Practice_June_2009.pdf
[32] EUROPEAN COORDINATION COMMITTEE OF THE
RADIOLOGICAL,
ELECTROMEDICAL AND HEALTHCARE IT INDUSTRY (COCIR), Medical Electrical
Equipment - Good refurbishment practice,
Brussels, COCIR, 2009.
http://www.cocir.org/site/fileadmin/6.1_Initiatives_Refurbishment/COCIR_Industry
_standard_
on_Good_Refurbishment_Practice_June_2009.pdf
[33] WORLD HEALTH ORGANIZATION, Guidelines for healthcare equipment
donations, WHO,
Geneva (2000).
http://www.who.int/hac/techguidance/pht/1_equipment%20donationbuletin82WHO.pdf
[34] WORLD HEALTH ORGANIZATION, Medical device donations:
considerations for
solicitation and provision, WHO Medical device technical series, WHO,
Geneva (2011).
http://whqlibdoc.who.int/publications/2011/9789241501408_eng.pdf
[35] EUROPEAN COORDINATION COMMITTEE OF THE
RADIOLOGICAL,
ELECTROMEDICAL AND HEALTHCARE IT INDUSTRY (COCIR). Good Maintenance
Services Practice Guide - Optimising the Equipment Life Cycle,
Brussels, COCIR (2013).
http://www.cocir.org/site/fileadmin/Publications_2013/13004.COC.Good_Maintenance
_WEB.
[36] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A Guide for
Establishing a
Credentialing and Privileging Program for Users of Fluoroscopic
Equipment in Healthcare
Organizations, Report of Task Group 124, AAPM Report No. 124, New York (2012).
[37] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Education and
Training in Radiological Protection for Diagnostic and Interventional
Procedures, Publication
113, Ann. ICRP 39(5), Elsevier Science, Oxford (2009).
[38] EUROPEAN COMMISSION. Guidelines on Radiation Protection Education and
Training of
Medical Professionals in the European Union. Radiation Protection No.
175, Luxembourg:
Publications Office of the European Union, 2014
[39] INTERNATIONAL LABOUR ORGANIZATION, International Labour
Organization
International standard classification of occupations. ISCO-480 08 Group
definitions – final
draft (word), Page 57. http://www.ilo.org/public/481
english/bureau/stat/isco/isco08/index.htm.
[40] INTERNATIONAL ATOMIC ENERGY AGENCY, Postgraduate Medical Physics Academic
Programmes, Training Course Series No. 56, IAEA, Vienna (2013).
[41] INTERNATIONAL ATOMIC ENERGY AGENCY, Clinical Training of Medical
Physicists
Specializing in Radiation Oncology, Training Course Series No. 37, IAEA, Vienna
(2009).
[42] INTERNATIONAL ATOMIC ENERGY AGENCY, Clinical Training of Medical
Physicists
Specializing in Diagnostic Radiology, Training Course Series No. 47, IAEA,
Vienna (2010).
[43] INTERNATIONAL ATOMIC ENERGY AGENCY, Clinical Training of Medical
Physicists
Specializing in Nuclear Medicine, Training Course Series No. 50, IAEA, Vienna
(2011).
[44] EUROPEAN COMMISSION. European Guidelines on Medical Physics
Expert. Radiation
Protection No. 174, Luxembourg: Publications Office of the European Union
(2014)
[45] INTERNATIONAL ATOMIC ENERGY AGENCY, Competency Based
Hospital
Radiopharmacy Training, Training Course Series No. 39, IAEA, Vienna (2010).
[46] INTERNATIONAL ATOMIC ENERGY AGENCY, A Syllabus for the
Education and
Training of Radiation Oncology Nurses, Training Course Series No. 28, IAEA,
Vienna (2008).
[47] INTERNATIONAL ATOMIC ENERGY AGENCY. The management system for facilities
and
activities, Safety Standards Series No. GS-R-3, IAEA, Vienna (2006).
[48] NATIONAL ELECTRICAL MANUFACTURERS ASSOCIATION (NEMA), Digital Imaging
and Communication in Medicine (DICOM). http://dicom.nema.org/ .
[49] HEALTH LEVEL SEVEN INTERNATIONAL, Introduction to
HL7 Standards.
http://www.hl7.org/implement/standards/index.cfm?ref=nav
[50] INTEGRATING THE HEALTHCARE ENTERPRISE (IHE), Home Page,
http://www.ihe.net/
[51] THE SCOTTISH OFFICE, The Interface between Clinical Audit and Management
— A Report
of a Working Group Set Up by the Clinical Resource and Audit Group,
Scottish Office,
Edinburgh (1993).
[52] INTERNATIONAL ATOMIC ENERGY AGENCY, Comprehensive Clinical
Audits of
Diagnostic Radiology Practices: A Tool for Quality Improvement, Human Health
Series No. 4,
IAEA, Vienna (2010).
[53] INTERNATIONAL ATOMIC ENERGY AGENCY, Quality management audits in
nuclear
medicine practices, IAEA, Vienna (2008).
[54] INTERNATIONAL ATOMIC ENERGY AGENCY, Comprehensive audits of
radiotherapy
practices: a tool for quality improvement: Quality Assurance Team for
Radiation Oncology
(QUATRO), IAEA, Vienna (2007).
[55] INTERNATIONAL ATOMIC ENERGY AGENCY, Safety Assessment for
Facilities and
Activities, Safety Standards Series No. GSR Part 4, IAEA, Vienna (2009).
[56] PAN AMERICAN HEALTH ORGANIZATION, THE ROTARY INTERNATIONAL
DISTRICT 6440. Diagnostic Imaging in the Community A Manual for
Clinics and Small
Hospitals Diagnostic Imaging in the Community- A Manual for Clinics and Small
Hospitals.
Washington DC (2011).
[57] NATIONAL HEALTH SERVICE NHS, Facilities for Diagnostic Imaging and
Interventional
Radiology, HBN 6 Volume 1, NHS Estate (2001).
[58] RADIOLOGICAL PROTECTION INSTITUTE OF IRELAND, The design of
diagnostic
medical facilities where ionizing radiation is used. RPII, Dublin, (2009).
[59] INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Basic Ionizing Radiation
Symbol, ISO 361, ISO, Geneva (1975).
[60] Sutton DG, Martin CJ, Williams, JR, Peet DJ. Radiation Shielding in
Diagnostic Radiology. 2
ed. London: BIR (2012).
[61] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS,
Structural shielding design for medical X ray imaging facilities, NCRP Report
147, Bethesda,
USA (2005).
[62] DEUTSCHES INSTITUT FÜR NORMUNG, DIN 6812:2013-06.
Medizinische
Röntgenanlagen bis 300 kV - Regeln für die Auslegung des baulichen
Strahlenschutzes
[63] Hanson Gerald P., and Palmer Philip E.S. Radiation Shielding for Clinics
or Small Hospitals
with a WHIS-RAD. Pan American Health Organization (PAHO) and the Rotary
International
District 6440, Washington DC (2013)
[64] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, AAPM Task Group
108:
PET and PET/CT Shielding Requirements, Med. Phys. 33, 1, 4-15 (2006).
[65] INTERNATIONAL ATOMIC ENERGY AGENCY, Radiation protection in newer
medical
imaging techniques: PET/CT, Safety reports series no. 58, IAEA, Vienna (2008).
[66] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60336, Edition 4.0 (2005-
04-20),Medical electrical equipment - X-ray tube assemblies for
medical diagnosis -
Characteristics of focal spots
[67] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60336, Edition 4.0 (2006-
05-18),Corrigendum 1 - Medical electrical equipment - X-ray tube
assemblies for medical
diagnosis - Characteristics of focal spots
[68] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60522, Edition 2.0 (1999-
02-10), Determination of the permanent filtration of X-ray tube assemblies
[69] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60526, Edition 2.0 (1978-
01-01), High-voltage cable plug and socket connections for medical X-ray
equipment
[70] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60526, Edition 2.0 (2010-
04-22), Corrigendum 1 - High-voltage cable plug and socket connections
for medical X-ray
equipment
[71] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60580, Edition 2.0 (2000-
01-14), Medical electrical equipment - Dose area product meters
[72] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-1-3, Edition
2.1
(2013-04-19), Medical electrical equipment - Part 1-3: General
requirements for basic safety
and essential performance - Collateral Standard: Radiation protection
in diagnostic X-ray
equipment
[73] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-28, Edition
2.0
(2010-03-10), Medical electrical equipment - Part 2-28: Particular
requirements for the basic
safety and essential performance of X-ray tube assemblies for medical diagnosis
[74] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-43, Edition
2.0
(2010-03-25), Medical electrical equipment - Part 2-43: Particular
requirements for the basic
safety and essential performance of X-ray equipment for interventional
procedures
[75] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-44, Edition
3.1
(2012-09-18), Medical electrical equipment - Part 2-44: Particular
requirements for the basic
safety and essential performance of X-ray equipment for computed tomography
[76] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-45, Edition
3.0
(2011-02-10), Medical electrical equipment - Part 2-45: Particular requirements
for basic safety
and essential performance of mammographic X-ray equipment and mammographic
stereotactic
devices
[77] INTERNATIONAL ELECTROTECHNICAL COMMISSION,
INTERNATIONAL
ELECTROTECHNICAL COMMISSION, IEC 60601-2-54, Edition 1.0 (2009-06-29), Medical
electrical equipment - Part 2-54: Particular requirements for the
basic safety and essential
performance of X-ray equipment for radiography and radioscopy
[78] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-54, Edition
1.0
(2010-03-10), Corrigendum 1 - Medical electrical equipment -
Part 2-54: Particular
requirements for the basic safety and essential performance of X-ray equipment
for radiography
and radioscopy
[79] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-54, Edition
1.0
(2011-06-10), Corrigendum 2 - Medical electrical equipment -
Part 2-54: Particular
requirements for the basic safety and essential performance of X-ray equipment
for radiography
and radioscopy
[80] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-63, Edition
1.0
(2012-09-19), Medical electrical equipment - Part 2-63: Particular
requirements for the basic
safety and essential performance of dental extra-oral X-ray equipment
[81] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60601-2-65, Edition
1.0
(2012-09-19), Medical electrical equipment - Part 2-65: Particular
requirements for the basic
safety and essential performance of dental intra-oral X-ray equipment
[82] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60613, Edition 3.0 (2010-
01-27), Electrical and loading characteristics of X-ray tube assemblies for
medical diagnosis
[83] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60627, Edition 3.0 (2013-
07-09), Diagnostic X-ray imaging equipment - Characteristics of
general purpose and
mammographic anti-scatter grids
[84] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60806, Edition 1.0 (1984-
01-01), Determination of the maximum symmetrical radiation field from a
rotating anode X-ray
tube for medical diagnosis
[85] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TS 61223-1, Edition
1.0
(1993-07-20), Evaluation and routine testing in medical imaging departments -
Part 1: General
aspects
[86] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TS 61223-2-1, Edition 1.0
(1993-07-20), Evaluation and routing testing in medical imaging
departments - Part 2-1:
Constancy tests - Film processors
[87] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61223-2-6, Edition
2.0
(2006-11-28), Evaluation and routine testing in medical imaging
departments • Part 2•6:
Constancy tests • Imaging performance of computed tomography X-ray equipment
[88] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61223-3-2, Edition
2.0
(2007-07-10), Evaluation and routine testing in medical imaging
departments - Part 3-2:
Acceptance tests - Imaging performance of mammographic X-ray equipment
[89] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61223-3-4, Edition
1.0
(2000-03-21), Evaluation and routine testing in medical imaging
departments - Part 3-4:
Acceptance tests - Imaging performance of dental X-ray equipment
[90] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61223-3-5, Edition
1.0
(2004-08-23), Evaluation and routine testing in medical imaging
departments - Part 3-5:
Acceptance tests - Imaging performance of computed tomography X-ray equipment
[91] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61223-3-5, Edition
1.0
(2006-03-24), Corrigendum 1 - Evaluation and routine testing in medical imaging
departments
- Part 3-5: Acceptance tests - Imaging performance of computed tomography X-ray
equipment
[92] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61262-1,
Edition 1.0
(1994-07-14), Medical electrical equipment - Characteristics of
electro-optical X-ray image
intensifiers - Part 1: Determination of the entrance field size
[93] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61262-2,
Edition 1.0
(1994-07-14), Medical electrical equipment - Characteristics of
electro-optical X-ray image
intensifiers - Part 2: Determination of the conversion factor
[94] INTERNATIONAL ELECTROTECHNICAL COMMISSION , IEC 61262-3, Edition
1.0
(1994-07-21), Medical electrical equipment - Characteristics of
electro-optical X-ray image
intensifiers - Part 3: Determination of the luminance distribution and
luminance non-uniformity
[95] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61262-4,
Edition 1.0
(1994-07-29), Medical electrical equipment - Characteristics of
electro-optical X-ray image
intensifiers - Part 4: Determination of the image distortion
[96] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61262-5,
Edition 1.0
(1994-08-05), Medical electrical equipment - Characteristics of
electro-optical X-ray image
intensifiers - Part 5: Determination of the detective quantum efficiency
[97] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61262-6,
Edition 1.0
(1994-07-21), Medical electrical equipment - Characteristics of
electro-optical X-ray image
intensifiers - Part 6: Determination of the contrast ratio and veiling glare
index
[98] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61262-7,
Edition 1.0
(1995-09-13), Medical electrical equipment - Characteristics of
electro-optical X-ray image
intensifiers - Part 7: Determination of the modulation transfer function
[99] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61267, Edition 2.0 (2005-
11-09), Medical diagnostic X-ray equipment - Radiation conditions for use in
the determination
of characteristics
[100] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61331-1, Edition
1.0
(1994-10-21), Protective devices against diagnostic medical X-radiation - Part
1: Determination
of attenuation properties of materials
[101] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61331-2, Edition
1.0
(1994-10-21), Protective devices against diagnostic medical X-radiation
- Part 2: Protective
glass plates
[102] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61331-3, Edition
1.0
(1998-11-06), Protective devices against diagnostic medical X-radiation
- Part 3: Protective
clothing and protective devices for gonads
[103] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61674, Edition 2.0 (2012-
11-29), Medical electrical equipment - Dosimeters with ionization
chambers and/or
semiconductor detectors as used in X-ray diagnostic imaging
[104] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61676, Edition 1.1 (2009-
01-26), Medical electrical equipment - Dosimetric instruments
used for non-invasive
measurement of X-ray tube voltage in diagnostic radiology
[105] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/PAS 61910-1, Edition 1.0
(2007-07-11), Medical electrical equipment - Radiation dose
documentation - Part 1:
Equipment for radiography and radioscopy
[106] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 62220-1, Edition
1.0
(2003-10-23), Medical electrical equipment - Characteristics of digital X-ray
imaging devices -
Part 1: Determination of the detective quantum efficiency
[107] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 62220-1-2, Edition
1.0
(2007-06-05), Medical electrical equipment - Characteristics of digital X-ray
imaging devices -
Part 1-2: Determination of the detective quantum efficiency - Detectors used in
mammography
[108] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 62220-1-3, Edition
1.0
(2008-06-11), Medical electrical equipment - Characteristics of digital X-ray
imaging devices -
Part 1-3: Determination of the detective quantum efficiency -
Detectors used in dynamic
imaging
[109] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 62494-1, Edition
1.0
(2008-08-13), Medical electrical equipment - Exposure index of digital X-ray
imaging systems
- Part 1: Definitions and requirements for general radiography
[110] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 62563-1, Edition
1.0
(2009-12-10), Medical electrical equipment - Medical image display
systems - Part 1:
Evaluation methods
[111] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Cardiac
Catheterization
Equipment Performance, AAPM Report No. 70, AAPM (2001)
[112] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, The Critical Examination
of X-ray Generating Equipment in Diagnostic Radiology, IPEM Report 107 (2012)
[113] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS, Medical
X-Ray, Electron Beam and Gamma-Ray Protection for Energies up to 50
Mev (Equipment
Design, Performance and Use), NCRP Report No. 102, NCRP (1989).
[114] NATIONAL ELECTRICAL MANUFACTURERS ASSOCIATION, Computed Tomography
Dose Check, NEMA Standards Publication XR 25-2010, NEMA, Rosslyn (2010).
[115] WHO, Technical specifications for the World Health Imaging System
for radiography the
WHIS-RAD, WHO/RAD/TS/95.1,
WHO, 1995.
http://apps.who.int/iris/bitstream/10665/60643/1/WHO_RAD_TS_95. 1.pdf?ua=1
[116] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Assessment of
Display
Performance for Medical Imaging Systems, Report of the American Association of
Physicists
in Medicine (AAPM) Task Group 18, Medical Physics Publishing, Madison, WI, AAPM
On-
Line Report No. 03, April 2005. http://www.aapm.org/pubs/reports/OR_03.pdf
[117] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Quality
Control in
Diagnostic Radiology, AAPM Report No. 74, AAPM (2002)
[118] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Recommended Standards
for the Routine Performance Testing of Diagnostic X-Ray Systems, IPEM Report 91
(2005)
[119] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, The
Commissioning and
Routine Testing of Mammographic X-Ray Systems: A Technical Quality
Control Protocol,
IPEM Report 89, IPEM (2005)
[120] INTERNATIONAL ATOMIC ENERGY AGENCY, Quality Assurance Programme for Screen
Film Mammography, Human Health Series no. 2, IAEA, Vienna (2009).
[121] AMERICAN COLLEGE OF RADIOLOGY, Mammography quality control manual,
American
College of Radiology, Reston, VA (1999).
[122] EUROPEAN COMMISSION, European guidelines for quality assurance in
breast cancer
screening and diagnosis, 4th Edition, EC (2006).
[123] Miller DL, Vano E, Bartal G, et al, Occupational radiation
protection in interventional
radiology: A Joint Guideline of the Cardiovascular and Interventional
Radiology Society of
Europe and the Society of Interventional Radiology. Cardiovasc Interv
Radiol 33:230–239
(2010)
[124] Duran A, Sim K-H, Miller DL, et al, Recommendations for Occupational
Radiation Protection
in Interventional Cardiology, Catheterization and Cardiovascular
Interventions 82:29–42
(2013).
[125] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Conversion
Coefficients for Radiological Protection Quantities for External
Radiation Exposures. ICRP
Publication 116. Ann. ICRP 40(2–5) (2010).
[126] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS, Use of
Personal Monitors to Estimate Effective Dose Equivalent and Effective Dose to
Workers for
External Exposure to Low-Let Radiation, NCRP Report No. 122, NCRP (1995).
[127] INTERNATIONAL ATOMIC ENERGY AGENCY. Implications for Occupational
Radiation
Protection of the New Dose Limit for the Lens of the Eye. IAEA. TECDOC series,
No. 1732,
Vienna (2013).
[128] Martin CJ, A review of radiology staff doses and dose
monitoring requirements, Radiation
Protection Dosimetry Vol. 136, No. 3, pp. 140–157 (2009)
[129] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Avoidance of
Radiation Injuries from Medical Interventional Procedures. ICRP Publication 85.
Ann. ICRP 30
(2) (2000).
[130] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Pregnancy and
Medical Radiation. ICRP Publication 84. Ann. ICRP 30 (1), (2000).
[131] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION.
Radiological
Protection and Safety in Medicine. ICRP Publication 73. Ann. ICRP 26(2) (1996).
[132] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION.
Radiological
Protection in Medicine. ICRP Publication 105. Ann. ICRP 37(6) (2007).
[133] WORLD HEALTH ORGANIZATION, Effective Choices for Diagnostic Imaging in
Clinical
Practices, Technical Report Series No. 795, WHO, Geneva (1990).
[134] WORLD HEALTH ORGANIZATION, Rational Use of Diagnostic Imaging in
Paediatrics,
Technical Report Series No. 757, WHO, Geneva (1987).
[135] EUROPEAN COMMISSION. Referral Guidelines for Imaging. Radiation
Protection No. 118,
2000.
http://ec.europa.eu/energy/nuclear/radioprotection/publication/doc/118_en.pdf
[136] EUROPEAN COMMISSION. Referral Guidelines for Medical Imaging Availability
and Use in
the European Union. Radiation Protection
No. 178, 2014.
http://ec.europa.eu/energy/nuclear/radiation_protection/doc/publication/178.pdf
[137] AMERICAN COLLEGE OF RADIOLOGY, ACR
Appropriateness Criteria®.
http://www.acr.org/Quality-Safety/Appropriateness-Criteria
[138] ROYAL COLLEGE OF RADIOLOGISTS. iRefer: Making the best use of clinical
radiology,
7th edition. London: The Royal College of
Radiologists (2012).
http://www.rcr.ac.uk/content.aspx?PageID=995
[139] SOCIÉTÉ FRANÇAISE DE RADIOLOGIE. Guide du bon usage des examens
d'imagerie
médicale. http://gbu.radiologie.fr
[140] GOVERNMENT OF WESTERN AUSTRALIA, Department of Health. Diagnostic
Imaging
Pathways. 2014. http://www.imagingpathways.health.wa.gov.au/
[141] CANADIAN ASSOCIATION OF RADIOLOGISTS, 2012 CAR Diagnostic Imaging Referral
Guidelines, CAR 2012. http://www.car.ca/en/standards-guidelines/guidelines.aspx
[142] AMERICAN DENTAL ASSOCIATION, Dental Radiographic
Examinations:
Recommendations for Patient Selection and Limiting Radiation Exposure,
ADA and FDA,
2012.
http://www.ada.org/sections/professionalResources/pdfs/Dental_Radiographic_Exami
nations_2
012.pdf
[143] COMMITTEE ON MEDICAL ASPECTS OF RADIATION IN THE
ENVIRONMENT
(COMARE), TWELFTH REPORT: The impact of personally initiated X-ray
computed
tomography scanning for the health assessment of asymptomatic individuals. HPA,
UK, 2007.
[144] UK Statutory Instruments, 2011 No. 1567, Health And Safety, The Ionising
Radiation (Medical
Exposure) (Amendment)
Regulations 2011.
http://www.legislation.gov.uk/uksi/2011/1567/made
[145] AMERICAN COLLEGE OF RADIOLOGY, Practice Guidelines by Modality,
Diagnostic
Radiology: Adult and Pediatric Guidelines.
http://www.acr.org/Quality-Safety/Standards-
Guidelines/Practice-Guidelines-by-Modality/General-Diagnostic
[146] AMERICAN COLLEGE OF RADIOLOGY, Practice Guidelines by Modality,
Computed
Tomography (CT) Guidelines.
http://www.acr.org/Quality-Safety/Standards-
Guidelines/Practice-Guidelines-by-Modality/CT
[147] AMERICAN COLLEGE OF RADIOLOGY, Practice Guidelines by Modality, Breast
Imaging
and Intervention Guidelines.
http://www.acr.org/Quality-Safety/Standards-Guidelines/Practice-
Guidelines-by-Modality/Breast-Imaging
[148] AMERICAN COLLEGE OF RADIOLOGY, Practice Guidelines by Modality,
Interventional
Radiology Guidelines.
http://www.acr.org/Quality-Safety/Standards-Guidelines/Practice-
Guidelines-by-Modality/Interventional
[149] AMERICAN COLLEGE OF RADIOLOGY, Practice Guidelines by
Modality, Pediatric
Guidelines.
http://www.acr.org/Quality-Safety/Standards-Guidelines/Practice-Guidelines-by-
Modality/Pediatric
[150] EUROPEAN COMMISSION. European Guidelines on Quality Criteria
for Diagnostic
Radiographic Images. EUR 16260 EN, Brussels, Luxembourg (1996).
[151] EUROPEAN COMMISSION. European Guidelines on Quality Criteria
for Diagnostic
Radiographic Images in Paediatrics. EUR 16261 EN, Brussels, Luxembourg (1996).
[152] EUROPEAN COMMISSION, European Guidelines for Quality Criteria
for Computed
Tomography, Rep. EUR 16262, EC, Luxembourg (2000).
[153] IMAGE GENTLY. How to develop CT protocols
for Children. 2007.
http://pedrad.org/Portals/6/Procedures/Protocols.pdf
[154] IMAGE GENTLY, Protocols for
Imaging Gently.
http://imagegently.dnnstaging.com/Procedures/InterventionalRadiology/Protocols.a
spx#198976
9-protocol-recommendations
[155] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, CT Scan
Protocols.
http://www.aapm.org/pubs/CTProtocols/
[156] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Protection of the
Patient in Diagnostic Radiology. ICRP Publication 34. Ann. ICRP 9 (2-3) (1982).
[157] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Managing Patient
Dose in Digital Radiology. ICRP Publication 93. Ann. ICRP 34 (1) (2004).
[158] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION.
Radiological
protection in paediatric diagnostic and interventional radiology. ICRP
Publication 121. Ann.
ICRP 42(2) (2013).
[159] ASRT, Best Practices in Digital
Radiography, White Paper
http://www.asrt.org/docs/whitepapers/asrt12_bstpracdigradwhp_final.pdf
[160] INTERNATIONAL ATOMIC ENERGY AGENCY, Radiation protection in
Paediatric
Radiology, Safety reports series no. 71, IAEA, Vienna (2012).
[161] INTERNATIONAL ATOMIC ENERGY AGENCY, Dosimetry in Diagnostic
Radiology for
Paediatric Patients, Human Health Series No. 24, IAEA, Vienna (2013).
[162] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Cost-Effective Methods
of
Patient Dose Reduction in Diagnostic Radiology, Prepared by a Joint Working
Party: IPEM,
BIR & NRPB, IPEM Report 82 (2001).
[163] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, An Exposure
Indicator for
Digital Radiography: Report of AAPM Task Group 116, AAPM Rep. 116, New York
(2009).
[164] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS, Report
No. 149 - A Guide to Mammography and Other Breast Imaging Procedures (2004)
[165] ACR–AAPM–SIIM Practice Guideline for Determinants of Image
Quality in Digital
Mammography,
2007.
http://www.acr.org/~/media/ACR/Documents/PGTS/guidelines/Image_Quality_Digital_M
amm
o.pdf
[166] INTERNATIONAL ATOMIC ENERGY AGENCY, Quality Assurance Programme for
Digital
Mammography, Human Health Series no. 17, IAEA, Vienna (2011).
[167] SECHOPOULOS, I, A review of breast tomosynthesis. Part I. The image
acquisition process,
Med Phys, 40, 014301 (2013);
doi: 10.1118/1.4770279
http://scitation.aip.org/docserver/fulltext/aapm/journal/medphys/40/1/1.
4770279.pdf?expires=1
395361418&id=id&accname=guest&checksum=573F0407C8A3909CE84F4504FD7CDD15
[168] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Managing Patient
Dose in Computed Tomography. ICRP Publication 87. Ann. ICRP 30(4) (2000).
[169] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Managing Patient
Dose in Multi-Detector Computed Tomography (MDCT). ICRP Publication 102. Ann.
ICRP 37
(1) (2007).
[170] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, The
Measurement,
Reporting, and Management of Radiation Dose in CT, AAPM Rep. 96, AAPM, College
Park,
MD (2008), http://www.aapm.org/pubs/reports/RPT_96.pdf
[171] NAGEL, HD (Ed.), Radiation Exposure in Computed Tomography: Fundamentals,
Influencing
Parameters, Dose Assessment, Optimisation, Scanner Data, Terminology,
4th edn, COCIR,
Hamburg (2002).
[172] McCOLLOUGH, CH, BRUESEWITZ, MR, KOFLER, JM, CT dose reduction
and dose
management tools: Overview of available options, Radiographics 26: 503–512
(2006).
[173] EUROPEAN COMMISSION, Cone Beam CT for Dental and
Maxillofacial Radiology:
Evidence Based Guidelines, Radiation Protection No. 172 (2012).
[174] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION.
Radiological
Protection in Cone Beam Computed Tomography (CBCT). ICRP Publication (in
preparation).
[175] EUROPEAN COMMISSION, European guidelines on radiation protection in
dental radiology:
The safe use of radiographs in dental practice, Radiation Protection No. 136
(2004).
[176] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS, Report
No. 145 - Radiation Protection in Dentistry (2003)
[177] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION.
Radiological
protection in cardiology. ICRP Publication 120. Ann. ICRP 42(1) (2013).
[178] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION.
Radiological
Protection in Fluoroscopically Guided Procedures outside the Imaging
Department. ICRP
Publication 117, Ann. ICRP 40(6) (2010).
[179] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Functionality and
Operation
of Fluoroscopic Automatic Brightness Control/Automatic Dose Rate Control Logic
in Modern
Cardiovascular and Interventional Angiography Systems, AAPM Report No.
125, AAPM
(2012).
[180] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS,
Radiation Dose Management for Fluoroscopically-guided Interventional
Medical Procedures,
NCRP Rep. 168, NCRP, Bethesda, MD (2010).
[181] INTERNATIONAL ATOMIC ENERGY AGENCY, Dual Energy X Ray Absorptiometry for
Bone Mineral Density and Body Composition Assessment, IAEA Human Health Series
No. 15,
(2011).
[182] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Radiation Dosimetry and Image Quality Assessment in Computed Tomography, ICRU
Rep. 87,
ICRU, Bethesda, MD (2013).
[183] INTERNATIONAL ATOMIC ENERGY AGENCY, Diagnostic Radiology
Physics: A
Handbook for Teachers and Students, IAEA, Vienna (2014).
[184] INTERNATIONAL ATOMIC ENERGY AGENCY, Implementation of the International
Code
of Practice on Dosimetry in Diagnostic Radiology (TRS 457): Review of Test
Results, IAEA
Human Health Reports No. 4, IAEA, Vienna (2011).
[185] INTERNATIONAL ATOMIC ENERGY AGENCY, Status of Computed
Tomography
Dosimetry for Wide Cone Beam Scanners, IAEA Human Health Reports No. 5, IAEA,
Vienna
(2011).
[186] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE,
Comprehensive
Methodology for the Evaluation of Radiation Dose in X-Ray Computed Tomography,
AAPM
Report No. 111, AAPM (2010).
[187] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Size-specific Dose
Estimates
(SSDE) in Paediatric and Adult Body CT Examinations, Rep. 204, AAPM, New York
(2011).
[188] Balter, S., Hopewell, J.W., Miller, D.L., Wagner, L.K., Zelefsky,
M.J., Fluoroscopically
Guided Interventional Procedures: A Review of Radiation Effects on Patients’
Skin and Hair,
Radiology Vol 254(2), 326-341 (2010).
[189] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. ICRP Statement
on Tissue Reactions / Early and Late Effects of Radiation in Normal
Tissues and Organs –
Threshold Doses for Tissue Reactions in a Radiation Protection Context. ICRP
Publication 118.
Ann. ICRP 41(1/2) (2012).
[190] HEGGIE J.C.P., I.D. McLEAN, J. HERLEY, F.J. THOMSON, R.K. GREWAL, J.
DIFFEY,
AND L.T. COLLINS, ACPSEM Position Paper: RECOMMENDATIONS FOR A DIGITAL
MAMMOGRAPHY QUALITY ASSURANCE PROGRAM V3.0
(2012).
http://www.acpsem.org.nz/index.php/radiologydocuments/cat_view/334-groups/187-sp
ecialty-
groups/189-radiology/201-reports
[191] INTERNATIONAL ATOMIC ENERGY AGENCY, Quality Assurance
Programme for
Computed Tomography: Diagnostic and Therapy Applications, Human Health
Series no. 19,
IAEA, Vienna (2012).
[192] INTERNATIONAL ATOMIC ENERGY AGENCY, Quality assurance for PET and PET/CT
systems, Human Health Series no. 1, IAEA, Vienna (2009).
[193] EUROPEAN COMMISSION. Radiation Protection 162: Criteria for Acceptability
of Medical
Radiological Equipment used in Diagnostic Radiology, Nuclear Medicine
and Radiotherapy.
Luxembourg: Office for Official Publications of the European Communities
(2012).
[194] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Validating
Automatic Film
Processor Performance, AAPM Report No. 94, AAPM (2008).
[195] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Acceptance Testing
and
Quality Control of Photostimulable Storage Phosphor Imaging Systems, AAPM
Report No. 93,
AAPM (2006)
[196] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Mammography - Assessment of Image Quality, Report 82, ICRU (2009)
[197] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Quality Assurance of PET
and PET/CT Systems, IPEM Report 108, IPEM (2013)
[198] BRITISH INSTITUTE OF RADIOLOGY, Assurance of Quality in the
Diagnostic X Ray
Department, 2nd edn, QA Working Group of the Radiation Protection Committee of
the BIR,
London (2001).
[199] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Measurement of
the
Performance Characteristics of Diagnostic X-Ray Systems: Part I: 2nd
Edition X-Ray Tubes
and Generators, Report 32 series, HPA (1996).
[200] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Measurement of
the
Performance Characteristics of Diagnostic X-Ray Systems: Part II:
2nd Edition Image
Intensifier TV Systems, Report 32 series, IPSM (1996).
[201] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Measurement of
the
Performance Characteristics of Diagnostic X-Ray Systems: Part III: 2nd
Edition Computed
Tomography X-Ray Scanners, Report 32 series, IPEM (2003).
[202] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Measurement of
the
Performance Characteristics of Diagnostic X-Ray Systems: Part IV: 2nd Edition
Intensifying
Screens, Films, Processors & Auto Exposure Control, Report 32 series, IPSM
(1997).
[203] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Measurement of
the
Performance Characteristics of Diagnostic X-Ray Systems: Part V: 2nd Edition
Conventional
Tomographic Equipment, Report 32 series, HPA (1996).
[204] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Measurement of
the
Performance Characteristics of Diagnostic X-Ray Systems: Part VI: 2nd
Edition Image
Intensifier Fluorography Systems, Report 32 series, HPA (1996).
[205] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Measurement of
the
Performance Characteristics of Diagnostic X-Ray Systems: Part VII: 1st
Edition Digital
Imaging Systems, Report 32 series, IPEM (2010).
[206] NATIONAL COUNCIL ON RADIATION
PROTECTION AND
MEASUREMENTS, Report No. 099 - Quality Assurance for Diagnostic Imaging.
Bethesda,
Maryland (1988)
[207] US MQSA Mammography Facility Surveys, Mammography Equipment
Evaluations, and
Medical Physicist Qualification Requirements under MQSA, Guidance for Industry
and FDA
staff,
2005.
http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/Guidance
Docu
ments/ucm094411.pdf
[208] NATIONAL HEALTH SERVICE, NHS BREAST SCREENING
PROGRAMME,
Commissioning and routine testing of full field digital mammography
systems, NHS Cancer
Screening Programme Rep. NHSBSP Equipment report 0604, NHSBSP, Sheffield
(2006).
[209] NHS BREAST SCREENING PROGRAMME, Routine control tests for full
field digital
mammography systems, National Breast Screening Programme Rep. NHSBSP
Equipment
Report 0702, NHSBSP, Sheffield (2007), www.cancerscreening.nhs.uk
[210] AMERICAN COLLEGE OF RADIOLOGY, ACR Technical Standard for Medical
Physics
Performance Monitoring of SPECT-CT Imaging Equipment,
ACR (2009).
http://www.acr.org/Quality-Safety/Standards-Guidelines/Technical-Standards-by-
Modality/Medical-Physics .
[211] AMERICAN COLLEGE OF RADIOLOGY, ACR–AAPM Technical Standard for
Medical
Physics Performance Monitoring of PET/CT Imaging Equipment,
ACR (2013). .
http://www.acr.org/Quality-Safety/Standards-Guidelines/Technical-Standards-by-
Modality/Medical-Physics
[212] SHOPE, TB, Radiation-induced skin injuries from fluoroscopy,
Radiographics 16: 1195–99
(1996).
[213] FOOD AND DRUG ADMINISTRATION, Public Health Advisory: Avoidance of
serious X-
ray induced skin injuries to patients during fluoroscopy guided procedures,
Rockville, Md: US
Food and Drug
Administration, 1994.
http://www.fda.gov/MedicalDevices/Safety/AlertandNotices/PublicHealthNotificatio
ns/ucm06
3084.htm. Published September 30, 1994. Updated April 30 (2009).
[214] INTERNATIONAL ATOMIC ENERGY AGENCY, Nuclear Medicine Resources
Manual,
IAEA, Vienna (2006).
[215] THE INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE Report No.
xx,
Radiation Protection in Nuclear Medicine, IPEM, (2014).
[216] INTERNATIONAL ATOMIC ENERGY AGENCY, Planning a clinical PET
centre, Human
Health Series no. 11, IAEA, Vienna (2010).
[217] INTERNATIONAL ATOMIC ENERGY AGENCY, Operational Guidance on
Hospital
Radiopharmacy: A Safe and Effective Approach, IAEA, Vienna (2008).
[218] EUROPEAN ASSOCIATION OF NUCLEAR MEDICINE, Guidelines on current good
radiopharmacy practices (cGRPP) for radiopharmaceuticals in nuclear medicine.
Eur J Nucl
Med Mol Imaging (2010).
http://www.eanm.org/publications/guidelines/5_EJNMMI_Guidance_cGRPPfulltext_05_2
010.
[219] INTERNATIONAL ATOMIC ENERGY AGENCY, A Handbook of physics for
nuclear
medicine, IAEA, Vienna (2014).
[220] WORLD HEALTH ORGANIZATION, QA of Pharmaceuticals: A Compendium of
Guidelines
and Related Materials, Vol. 2, WHO, Geneva (1999).
[221] COUNCIL OF EUROPE DIRECTORATE FOR THE QUALITY OF MEDICINES, European
Pharmacopoeia 1997, 3rd edn, Council of Europe, Strasbourg (1996), as
amended by Suppl.
2000, Council of Europe, Strasbourg (1999).
[222] BRITISH PHARMACOPOEIA COMMISSION, British Pharmacopoeia 2000, The
Stationery
Office, London (2000).
[223] UNITED STATES PHARMACOPOEIA, NATIONAL FORMULARY, United
States
Pharmacopoeia 24/National Formulary 19, The United States Pharmacopoeial
Convention Inc.,
Rockville, MD (1999).
[224] INTERNATIONAL ATOMIC ENERGY AGENCY, ARCAL XV: Producción y
Control de
Radiofármacos, Manual de Protocolos de Calidad de Radiofármacos,
ARCAL/IAEA, IAEA,
Vienna (1999).
[225] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60789 (2005-10) Ed.
3.0
Medical electrical equipment - Characteristics and test conditions of
radionuclide imaging
devices - Anger type gamma cameras.
[226] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60789 Corr.1
(2009-10)
Ed. 3.0 Corrigendum 1 - Medical electrical equipment - Characteristics and test
conditions of
radionuclide imaging devices - Anger type gamma cameras.
[227] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61303 (1994-10) Ed.
1.0
Medical electrical equipment - Radionuclide calibrators - Particular
methods for describing
performance.
[228] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61675-1 (2013-09) Ed. 2.0
Radionuclide imaging devices - Characteristics and test conditions - Part 1:
Positron emission
tomographs.
[229] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61675-2 (2005-02) Ed. 1.1
Radionuclide imaging devices - Characteristics and test conditions -
Part 2: Single photon
emission computed tomographs.
[230] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61675-3 (1998-02) Ed. 1.0
Radionuclide imaging devices - Characteristics and test conditions -
Part 3: Gamma camera
based whole body imaging systems.
[231] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 61948-1 (2001-02) Ed.
1.0 Nuclear medicine instrumentation - Routine tests - Part 1: Radiation
counting systems.
[232] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 61948-2 (2001-02) Ed.
1.0 Nuclear medicine instrumentation - Routine tests - Part 2: Scintillation
cameras and single
photon emission computed tomography imaging.
[233] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 61948-3 (2005-07) Ed.
1.0 Nuclear medicine instrumentation - Routine tests - Part 3: Positron
emission tomographs.
[234] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 61948-4 (2006-11) Ed.
1.0 Nuclear medicine instrumentation - Routine tests - Part 4: Radionuclide
calibrators.
[235] NATIONAL ELECTRICAL MANUFACTURERS ASSOCIATION,
Performance
Measurements of Gamma Cameras, NEMA Standards Publication NU
1-2012, NEMA,
Rosslyn (2013).
[236] NATIONAL ELECTRICAL MANUFACTURERS ASSOCIATION,
Performance
Measurements of Positron Emission Tomographs, NEMA Standards Publication
NU 2-2012,
NEMA, Rosslyn (2013).
[237] NATIONAL ELECTRICAL MANUFACTURERS ASSOCIATION,
Performance
Measurements and Quality Control Guidelines for Non-Imaging Intraoperative
Gamma Probes,
NEMA Standards Publication NU 3-2004, NEMA, Rosslyn (2004).
[238] AMERICAN COLLEGE OF RADIOLOGY, ACR Technical Standard for Medical
Nuclear
Physics Performance Monitoring of PET Imaging
Equipment, ACR 2011.
http://www.acr.org/Quality-Safety/Standards-Guidelines/Technical-Standards-by-
Modality/Medical-Physics
[239] AMERICAN COLLEGE OF RADIOLOGY, ACR–AAPM Technical Standard for
Medical
Nuclear Physics Performance Monitoring of Gamma
Cameras, ACR 2013. .
http://www.acr.org/Quality-Safety/Standards-Guidelines/Technical-Standards-by-
Modality/Medical-Physics
[240] EUROPEAN ASSOCIATION FOR NUCLEAR MEDICINE, Acceptance testing for
nuclear
medicine instrumentation, Eur J Nucl Med Mol Imaging
(2010) 37:672–681.
http://www.eanm.org/publications/guidelines/3_EJNMMI_Physics_GL_AcceptTesting_fu
lltext
_03_2010.pdf
[241] INTERNATIONAL ATOMIC ENERGY AGENCY, Quality assurance for SPECT
systems,
Human Health Series no. 6, IAEA, Vienna (2009).
[242] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, DICOM Image and Data
Management for Nuclear Medicine, Physiological Measurements,
Radiotherapy and
Ultrasound, Report 99, IPEM (2010).
[243] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, The
Selection, Use,
Calibration, and Quality Assurance of Radionuclide Calibrators Used in
Nuclear Medicine,
Report of AAPM Task Group 181, June 2012
[244] ISO 3925:2014, Unsealed radioactive substances -- Identification and
documentation
[245] THE ROYAL COLLEGE OF RADIOLOGISTS. Guidelines for the use of PET-CT in
children,
Second edition. London: The Royal College of Radiologists (2014).
[246] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Doses to the
Embryo and Fetus from Intakes of Radionuclides by the Mother. ICRP Publication
88. Ann.
ICRP 31(1-3) (2001).
[247] STABIN, M., BREITZ, H., Breast milk excretion of
radiopharmaceuticals: Mechanisms,
findings, and radiation dosimetry, J. Nucl. Med. 41: 863–873 (2000).
[248] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Doses to Infants
from Ingestion of Radionuclides in Mothers' Milk. ICRP Publication 95. Ann.
ICRP 34(3-4),
(2004).
[249] RADAR, Internal Sources – Nuclear Medicine.
http://www.doseinfo-radar.com/RADAR-INT-
NM.html
[250] ACR, Practice Guidelines by Modality, Nuclear Medicine.
http://www.acr.org/Quality-
Safety/Standards-Guidelines/Practice-Guidelines-by-Modality/Nuclear-Medicine
[251] ADMINISTRATION OF RADIOACTIVE SUBSTANCES ADVISORY COMMITTEE
(ARSAC), Notes for Guidance on the Clinical Administration of
Radiopharmaceuticals and
Use of Sealed Radioactive Sources, Rev 2011. Health Protection Agency, UK
(2006).
[252] EUROPEAN ASSOCIATION OF NUCLEAR MEDICINE (EANM),
Guidelines.
http://www.eanm.org/publications/guidelines/index.php?navId=37
[253] IMAGE WISELY, Imaging modalities,
nuclear medicine.
http://www.imagewisely.org/Imaging-Modalities/Nuclear-Medicine
[254] INTERNATIONAL ATOMIC ENERGY AGENCY, Quantitative nuclear medicine
imaging:
concepts, requirements and methods, Human Health Reports no. 9, IAEA, Vienna
(2014).
[255] SNMMI/ASNC/SCCT Guideline for Cardiac SPECT/CT and PET/CT 1.0 ,
J Nucl Med.
54:1485-1507 (2013).
[256] GELFAND MJ., MT. PARISI, AND ST TREVES, Pediatric
Radiopharmaceutical
Administered Doses: 2010 North American Consensus Guidelines,
J Nucl Med.
2011;52(2):318–322. Accessed through
Image Gently:
http://pedrad.org/associations/5364/ig/Procedures/InterventionalRadiology/Protoc
ols.aspx#198
9769-protocol-recommendations
[257] EUROPEAN ASSOCIATION OF NUCLEAR MEDICINE (EANM), Paediatric Dosage Card
(2014). http://www.eanm.org/publications/dosage_calculator.php?navId=285
[258] LASSMANN M, TREVES ST; EANM/SNMMI Paediatric Dosage Harmonization
Working
Group. Paediatric radiopharmaceutical administration: harmonization of
the 2007 EANM
paediatric dosage card (version 1.5.2008) and the 2010 North American consensus
guidelines.
Eur J Nucl Med Mol Imaging 41(5):1036-41 (2014).
[259] AMERICAN COLLEGE OF RADIOLOGY, ACR–ASTRO Practice Guideline
for The
Performance of Therapy with Unsealed Radiopharmaceutical Sources,
ACR (2010).
http://www.acr.org/~/media/ACR/Documents/PGTS/guidelines/Unsealed_Radiopharms.pd
f
[260] SILBERSTEIN EB, ALAVI A, BALON HR, CLARKE SE, DIVGI C, etc. The SNM
Practice
Guideline for Therapy of Thyroid Disease with ¹³¹I 3.0. J Nucl Med.
Oct;53(10):1633-51
(2012).
[261] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Release
of
patients after therapy with unsealed radionuclides. ICRP Publication
94. Ann. ICRP 34(2),
(2004).
[262] AMERICAN THYROID ASSOCIATION TASKFORCE ON RADIOIODINE SAFETY.
Radiation Safety in the Treatment of Patients with Thyroid Diseases
by Radioiodine 131I:
Practice Recommendations of the American Thyroid Association, Thyroid
Apr;21(4):335-46
(2011)
[263] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Absorbed-dose Specification in Nuclear Medicine, ICRU Publication 67, 2002.
[264] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS,
Management of Radionuclide Therapy Patients, NCRP Report 155, NCRP, Bethesda,
2006.
[265] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A
Primer for
Radioimmunotherapy and Radionuclide Therapy, AAPM Report No. 71, AAPM, 2001.
[266] EUROPEAN ASSOCIATION OF NUCLEAR MEDICINE (EANM), EANM
Dosimetry
Committee guidance document: good practice of clinical dosimetry reporting, Eur
J Nucl Med
Mol Imaging, DOI 10.1007/s00259-010-1549-3, 2010
[267] HINDORF C, G GLATTING, C CHIESA, O LINDÉN, G FLUX. EANM Dosimetry
Committee guidelines for bone marrow and whole-body dosimetry, Eur J Nucl Med
Mol
Imaging, DOI 10.1007/s00259-010-1422-4, 2010.
http://www.eanm.org/publications/guidelines/EJNMMI_Bone_Marrow_Dosimetry_06_2010
.p
df
[268] HÄNSCHEID H, CANZI C, ESCHNER W, FLUX G, LUSTER M, STRIGARI L,
LASSMANN M. EANM Dosimetry Committee Series on Standard Operational Procedures
for
Pre-Therapeutic Dosimetry II. Dosimetry prior to radioiodine therapy of benign
thyroid
diseases, Eur J Nucl Med Mol Imaging. Jul;40(7):1126-34 (2013).
[269] ROYAL COLLEGE OF PHYSICIANS. Radioiodine in the management of benign
thyroid
disease: clinical guidelines. Report of a Working Party. London: RCP (2007).
[270] Colletti, P.M., PET-CT in the Pregnant
Patient, Image Wisely, 2012.
http://www.imagewisely.org/~/media/ImageWisely%20Files/NucMed/PETCT%20in%20the%
20Pregnant%20Patient
[271] INTERNATIONAL ATOMIC ENERGY AGENCY, Quality Assurance for
Radioactivity
Measurement in Nuclear Medicine, Technical Report Series No. 454, IAEA, Vienna
(2006).
[272] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Radiation Dose to
Patients from Radiopharmaceuticals (Addendum to ICRP Publication 53). ICRP
Publication 80.
Ann. ICRP 28(3) (1998).
[273] NUCLEAR REGULATORY COMMISSION, Radiation Dose
Estimates for
Radiopharmaceuticals, Rep. NUREG/CR-6345, Oak Ridge Institute for Science and
Education,
(1996).
[274] STABIN, M, Internal Dosimetry in Pediatric Nuclear
Medicine, In: Pediatric Nuclear
Medicine, S. Treves, ed., Springer-Verlag (1995).
[275] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Radiation Dose to
Patients from Radiopharmaceuticals - Addendum 3 to ICRP Publication 53. ICRP
Publication
106. Ann. ICRP 38 (1-2) (2008)
[276] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION. Radiation Dose to
Patients from Radiopharmaceuticals. ICRP Publication 53. Ann. ICRP 18 (1-4)
(1988).
[277] DEZARN WA., JT CESSNA, LA. DEWERD, W FENG, VL. GATES, J
HALAMA, AS.
KENNEDY, S NAG, M SARFARAZ, V SEHGAL, R SELWYN, MG. STABIN, BR.
THOMADSEN, LE. WILLIAMS, R SALEM, Recommendations of the American Association
of Physicists in Medicine on dosimetry, imaging, and quality assurance
procedures for 90Y
microsphere brachytherapy in the treatment of hepatic malignancies, AAPM Report
No. 144,
Med. Phys. 38 (8), (2011).
[278] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE,
Dosimetry for
Radionuclide Therapy, Report 104, IPEM (2011).
[279] The UK Gamma Probe Working Group, Guidelines on the Quality Assurance of
Intraoperative
Gamma Probes, BNMS, UK NHS (2005).
[280] EUROPEAN ASSOCIATION FOR NUCLEAR MEDICINE, Routine quality
control
recommendations for nuclear medicine instrumentation, Eur J Nucl Med Mol
Imaging (2010)
37:662–671. http://www.eanm.org/publications/guidelines/index.php?navId=37
[281] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Quality
Assurance in
Gamma Camera Systems, Report 86, IPEM (2003).
[282] INTERNATIONAL ATOMIC ENERGY AGENCY, Lessons Learned from
Accidental
Exposures in Radiotherapy, Safety Reports Series No. 17, IAEA, Vienna (2000).
[283] INTERNATIONAL ATOMIC ENERGY AGENCY, Predisposal Management of Waste from
the Use of Radioactive Materials in Medicine, Industry, Research, Agriculture
and Education,
Safety Standards Series, DS454
[284] INTERNATIONAL ATOMIC ENERGY AGENCY, Regulations for the Safe
Transport of
Radioactive Material, 2012 Edition, IAEA Safety Standards Series No. SSR-6,
IAEA, Vienna
(2012).
[285] INTERNATIONAL ATOMIC ENERGY AGENCY, Setting Up a Radiotherapy
Programme:
Clinical, Medical Physics, Radiation Protection and Safety Aspects, IAEA,
Vienna (2008).
[286] INTERNATIONAL ATOMIC ENERGY AGENCY, Radiation Protection in the
Design of
Radiotherapy Facilities, Safety Report Series No. 47, IAEA, Vienna (2006).
[287] INTERNATIONAL ATOMIC ENERGY AGENCY, Radiotherapy Facilities: Master
Planning
and Concept Design Considerations, IAEA Human Health Reports No. 10,
IAEA, Vienna
(2014).
[288] INTERNATIONAL ATOMIC ENERGY AGENCY, Planning National Radiotherapy
Services:
A Practical Tool, Human Health Series No. 14, IAEA, Vienna (2010).
[289] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 61859, Edition 1.0
(1997-05-09), Guidelines for radiotherapy treatment rooms design.
[290] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS,
Structural Shielding Design and Evaluation for Megavoltage X- and Gamma-Ray
Radiotherapy
Facilities, NCRP Report No. 151, NCRP, Bethesda, Maryland (2005).
[291] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, The
Design of
Radiotherapy Treatment Room Facilities, Rep. 75, IPEM, York (1997).
[292] Jack Venselaar, Ali S. Meigooni, Dimos Baltas, Peter J. Hoskin,
editors. Comprehensive
Brachytherapy: Physical and Clinical Aspects. CRC Press (2013).
[293] CONFERENCE OF RADIATION CONTROL PROGRAM DIRECTORS. CRCPD Publication
E-11-2. Technical White Paper: Guidance for State Programs that Regulate the
New Therapy
Modality Electronic Brachytherapy (2011).
http://www.crcpd.org/Pubs/H-
36BrachytherapyWhitePaper.pdf
[294] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical Electrical Equipment,
Part 1: General Requirements for Safety, Rep. IEC 601-1, IEC, Geneva (1988).
[295] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical Electrical Equipment,
Part 2-1: Particular requirements for the basic safety and essential
performance of electron
accelerators in the range 1 MeV to 50 MeV, Edition 3.0, Rep. IEC 60601-2-1,
IEC, Geneva
(2009).
[296] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical Electrical Equipment,
Part 2-8: Particular requirements for basic safety and essential performance of
therapeutic X-
ray equipment operating in the range 10 kV to 1 MV, Edition 2.0, Rep. IEC
60601-2-8, IEC,
Geneva (1999).
[297] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical Electrical Equipment,
Part 2-11: Particular requirements for the basic safety and essential
performance of gamma
beam therapy equipment, Edition 3.0, Rep. IEC 60601-2-11, IEC, Geneva (2013).
[298] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical Electrical Equipment,
Part 2-17: Particular requirements for the basic safety and
essential performance of
automatically-controlled brachytherapy afterloading equipment, Edition 3.0, IEC
60601-2-17,
IEC, Geneva (2013).
[299] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical Electrical Equipment,
Part 2-29, Particular requirements for the basic safety and
essential performance of
radiotherapy simulators, Edition 3.0, Rep. IEC 60601-2-29, IEC, Geneva (2008).
[300] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60731, Edition
3.0,
Medical electrical equipment - Dosimeters with ionization chambers as used in
radiotherapy,
IEC, Geneva (2011).
[301] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 60976, Edition 2.0 (2007-
10-16), Medical electrical equipment - Medical electron accelerators -
Functional performance
characteristics
[302] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 60977, Edition 2.0
(2008-07-09), Medical electrical equipment - Medical electron
accelerators - Guidelines for
functional performance characteristics.
[303] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61168, Edition 1.0 (1993-
12-22), Radiotherapy simulators - Functional performance characteristics.
[304] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Radiotherapy
Simulators:
Guidelines for Functional Performance Characteristics, Rep. IEC 61170, IEC,
Geneva (1993).
[305] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 61217, Edition 2.0 (2011-
12-07), Radiotherapy equipment - Coordinates, movements and scales.
[306] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 61852, Edition 1.0
(1998-04-15), Medical electrical equipment - Digital imaging and communications
in medicine
(DICOM) - Radiotherapy objects,
[307] INTERNATIONAL ELECTROTECHNICAL COMMISSION, Medical Electrical Equipment:
Requirements for the Safety of Radiotherapy Treatment Planning Systems,
Edition 2.0, Rep.
IEC 62083, IEC, Geneva (2009).
[308] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC/TR 62266, Edition 1.0
(2002-03-15), Medical electrical equipment - Guidelines for
implementation of DICOM in
radiotherapy.
[309] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 62274, Edition 1.0 (2005-
05-27), Medical electrical equipment - Safety of radiotherapy record and verify
systems.
[310] INTERNATIONAL ELECTROTECHNICAL COMMISSION, IEC 62467-1, Edition
1.0
(2009-06-09), Medical electrical equipment - Dosimetric instruments as used in
brachytherapy -
Part 1: Instruments based on well-type ionization chambers.
[311] INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Radiological protection -
Sealed Radioactive Sources - General requirements and
classification, Standard ISO-
2919:2012, ISO, Geneva (2012).
[312] INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Radiation protection
-
Sealed radioactive sources - Leakage test methods, ISO 9978:1992, ISO, Geneva
(1992)
[313] INTERNATIONAL ORGANIZATION FOR STANDARDIZATION, Clinical dosimetry -
Beta radiation sources for brachytherapy, ISO 21439:2009, ISO, Geneva (2009).
[314] INTERNATIONAL ATOMIC ENERGY AGENCY, Record and Verify Systems for
Radiation
Treatment of Cancer: Acceptance Testing, Commissioning and Quality Control,
Human Health
Reports No. 7, IAEA, Vienna (2013).
[315] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, AAPM Code of Practice for
Radiotherapy Accelerators, Report of AAPM Radiation Therapy Task Group
45, AAPM
Report No. 47, AAPM (1994).
[316] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE Report 94
Acceptance
Testing and Commissioning of Linear Accelerators, 2007 ISBN 978 1 903613 30 6.
[317] INTERNATIONAL ATOMIC ENERGY AGENCY, Radiation Oncology
Physics: A
Handbook for Teachers and Students, IAEA, Vienna (2005).
[318] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Remote
Afterloading
Technology, AAPM Report No. 41, published for the AAPM by the
American Institute of
Physics, New York (1993).
[319] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Fundamental Quantities and Units for Ionizing Radiation (ICRU Report
85a-Revised) (2011)
[320] INTERNATIONAL ATOMIC ENERGY AGENCY, Commissioning and Quality Assurance of
Computerized Planning Systems for Radiation Treatment of Cancer, Technical
Reports Series
No. 430, IAEA, Vienna (2004).
[321] Benedick Fraass, Karen Doppke, Margie Hunt, Gerald Kutcher,
George Starkschall, Robin
Stern, Jake Van Dyke, American Association of Physicists in Medicine
Radiation Therapy
Committee Task Group 53: Quality assurance for clinical radiotherapy
treatment planning,
AAPM Report No. 62, Med. Phys. 25 (10), (1998).
[322] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, A Guide to Commissioning
and Quality Control of Treatment Planning Systems, Rep. 68, IPEM, York (1996).
[323] BRITISH INSTITUTE OF RADIOLOGY, Treatment Simulators, BJR Supplement 23
(1989)
[324] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, The Role of In-Room kV X-
Ray Imaging for Patient Setup and Target Localization, Report of AAPM
Task Group 104,
AAPM Report No. 104, MD (2009).
[325] DAS I, CHENG CW, WATTS R, AHNESJÖ A, GIBBONS J, LI XA,
LOWENSTEIN J,
MITRA RK, SIMON W, ZHU T. Accelerator beam data commissioning
equipment and
procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM,
Medical
Physics, Vol. 35, No. 9 (2008).
[326] LOW D, MORAN J, DEMPSEY J, DONG L, OLDHAM M. Dosimetry tools and
techniques
for IMRT, Report of AAPM Task Group 120, Med. Phys. 38 (3) (2011)
[327] NATH, R, ANDERSON, LL, MELI, JA, OLCH, AJ, STITT, JA, WILLIAMSON, JF,
Code of
practice for brachytherapy physics: Report of the AAPM Radiation Therapy
Committee Task
Group No. 56, AAPM Report No. 59, Medical Physics, vol. 24, no. 10, pp. 1557-98
(1997).
[328] The GEC ESTRO Handbook of Brachytherapy, Edited by Alain
Gerbaulet, Richard Pötter,
Jean-Jacques Mazeron, Harm Meertens, Erik Van
Limbergen, ESTRO, 2002.
http://www.estro.org/about/governance-organisation/committees-activities/gec-est
ro-handbook-
of-brachytherapy
[329] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Neutron
Measurements
Around High Energy X-ray Radiotherapy Machines, A report of Task
Group 27 Radiation
Therapy Committee, AAPM Report No. 19, AAPM (1986).
[330] NATIONAL COUNCIL ON RADIATION PROTECTION AND MEASUREMENTS,
Preconception and Prenatal Radiation Exposure: Health Effects and
Protective Guidance,
NCRP Report No. 174, NCRP, Bethesda, Maryland (2013).
[331] THOMADSEN BR, ERICKSON BA, EIFEL PJ, HSU IC, PATEL RR, PETEREIT
DG,
FRAASS BA, RIVARD MJ. A review of safety, quality management, and practice
guidelines
for high-dose-rate brachytherapy: executive summary. Supplemental
material (2014).
http://www.practicalradonc.org/cms/attachment/2014676581/2036190596/mmc1.pdf
[332] JAFFRAY DA, LANGEN KM, MAGERAS G, DAWSON LA, YAN D, ADAMS R, MUNDT
AJ, FRAASS B. Assuring Safety and Quality in Image Guided Delivery of Radiation
Therapy.
Pract Radiat Oncol. Supplemental material (2013).
http://www.practicalradonc.org/cms/attachment/2009137979/2031372254/mmc1.pdf
[333] MORAN JM, DEMPSEY M, EISBRUCH A, FRAASS BA, GALVIN JM, IBBOTT GS,
MARKS LB. Safety considerations for IMRT, Pract Radiat Oncol. Supplemental
material
(2011). http://download.journals.elsevierhealth.com/mmcs/journals/1879-
8500/PIIS1879850011001627.mmc1.pdf
[334] SOLBERG TD, BALTER JM, BENEDICT SH, FRAASS BA, KAVANAGH B, MIYAMOTO
C, PAWLICKI T, POTTERS L, YAMADA Y. Quality and safety considerations in
stereotactic
radiosurgery and stereotactic body radiation therapy, Pract Radiat Oncol.
Supplemental
material (2012).
http://download.journals.elsevierhealth.com/mmcs/journals/1879-
8500/PIIS1879850011002165.mmc1.pdf
[335] THE ROYAL COLLEGE OF RADIOLOGISTS, SOCIETY AND
COLLEGE OF
RADIOGRAPHERS, INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE,
NATIONAL PATIENT SAFETY AGENCY, BRITISH INSTITUTE OF RADIOLOGY.
Towards safer radiotherapy, The Royal College of
Radiologists, London (2008).
https://www.rcr.ac.uk/docs/oncology/pdf/Towards_saferRT_final.pdf
[336] THE ROYAL COLLEGE OF RADIOLOGISTS, SOCIETY AND
COLLEGE OF
RADIOGRAPHERS, Children’s Cancer and Leukaemia Group, Good practice
guide for
paediatric radiotherapy, The Royal College of
Radiologists, London (2012).
https://www.rcr.ac.uk/docs/oncology/pdf/BFCO%2812%295_Good_practice.pdf
[337] Stoval, M., et al., Fetal dose from radiotherapy with photon
beams: Report of the AAPM
Radiation Therapy Task Group No. 36, Med.
Phys. 22 (1995).
http://www.aapm.org/pubs/reports/RPT_50.pdf
[338] INTERNATIONAL ATOMIC ENERGY AGENCY, Absorbed Dose Determination in
External
Beam Radiotherapy, Technical Reports Series No. 398, IAEA, Vienna (2000).
[339] Peter R. Almond, Peter J. Biggs, B. M. Coursey, W. F. Hanson, M. Saiful
Huq, Ravinder Nath,
D. W. O. Rogers, AAPM’s TG-51 protocol for clinical reference
dosimetry of high-energy
photon and electron beams, AAPM Task Group 51, AAPM Report 67, Med.
Phys. 26 (9)
(1999).
[340] C.-M. Ma, C. W. Coffey, L. A. DeWerd, C. Liu, R. Nath, S. M. Seltzer, J.
P. Seuntjens, AAPM
protocol for 40–300 kV x-ray beam dosimetry in radiotherapy and radiobiology,
AAPM Task
Group 61, AAPM Report 76, Med. Phys. 28(6) (2001).
[341] Kutcher, G.J., Comprehensive QA for radiation oncology: Report of AAPM
Radiation Therapy
Committee Task Group 40, AAPM Report No. 46. Med. Phys. 21 (4): 581–618 (1994).
[342] Yan Yub, Lowell L. Anderson, Zuofeng Li, David E. Mellenberg, Ravinder
Nath, M. C. Schell,
Frank M. Waterman, Andrew Wu, John C. Blasko, Permanent
prostate seed implant
brachytherapy: Report of the American Association of Physicists in Medicine
Task Group No.
64, AAPM Report 68, Med. Phys. 26 (10) (1999).
[343] Wayne M. Butler, William S. Bice, Jr., Larry A. DeWerd, James M. Hevezi,
M. Saiful Huq,
Geoffrey S. Ibbott, Jatinder R. Palta, Mark J. Rivarda, Jan P. Seuntjens, Bruce
R. Thomadsen,
Third-party brachytherapy source calibrations and physicist
responsibilities: Report of the
AAPM Low Energy Brachytherapy Source Calibration Working Group, AAPM Report No.
98,
Med. Phys. 35 (9) (2008).
[344] INTERNATIONAL ATOMIC ENERGY AGENCY, Calibration of Photon and
Beta Ray
Sources Used in Brachytherapy: Guidelines on standardized procedures at
Secondary Standards
Dosimetry Laboratories (SSDLs) and hospitals, IAEA TECDOC 1274, IAEA, Vienna
(2002).
[345] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Small Field MV
Photon
Dosimetry, Report 103, IPEM (2010).
[346] Michael G. Herman, James M. Balter, David A. Jaffray, Kiarin P.
McGee, Peter Munro,
Shlomo Shalev, Marcel Van Herk, John W. Wong, Clinical use of electronic portal
imaging:
Report of AAPM Radiation Therapy Committee Task Group 58, AAPM Report No. 75,
Med.
Phys. 28(5) (2001).
[347] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Radiotherapy Portal
Imaging
Quality, A report of Task Group 28, AAPM Report No.
24, AAPM (1987).
http://www.aapm.org/pubs/reports/RPT_24.pdf
[348] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Prescribing, Recording and Reporting Photon Beam Therapy, ICRU
Report 50, ICRU
Publications, Bethesda, Maryland (1993).
[349] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Prescribing, Recording and Reporting Photon Beam Therapy (Supplement to ICRU
Report 50),
ICRU Report 62, ICRU Publications, Bethesda, Maryland (1999).
[350] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Prescribing, Recording, and Reporting Intensity-Modulated Photon-Beam
Therapy (IMRT),
ICRU Report 83, ICRU Publications, Bethesda, Maryland (2010)
[351] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Prescribing, Recording and Reporting Electron Beam Therapy, ICRU Report
71, ICRU
Publications, Bethesda, Maryland (2004).
[352] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Prescribing, Recording, and Reporting Proton-Beam Therapy, ICRU
Report 78, ICRU
Publications, Bethesda, Maryland (2007).
[353] INTERNATIONAL ATOMIC ENERGY AGENCY, Dose Reporting in Ion Beam
Therapy,
IAEA TECDOC 1560, IAEA, Vienna (2007).
[354] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE,
Recommendations for
clinical electron beam dosimetry: Supplement to the recommendations of
Task Group 25,
AAPM Report 99, Med. Phys., Vol 36, (7) (2009).
[355] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Diode In Vivo Dosimetry
for
Patients Receiving External Beam Radiation Therapy, Report of Task
Group 62 of the
Radiation Therapy Committee, AAPM Report No. 87 (2005).
[356] ESTRO Booklet No. 1 Methods for in vivo Dosimetry in
external Radiotherapy.
http://www.estro.org/school/articles/publications/publications
[357] ESTRO Booklet No. 5 Practical Guidelines for the Implementation of in
vivo dosimetry with
diodes in external radiotherapy with photon
beams (entrance dose),
http://www.estro.org/school/articles/publications/publications
[358] ESTRO Booklet No. 10 Independent Dose Calculations
Concepts and Models,
http://www.estro.org/school/articles/publications/publications
[359] INTERNATIONAL ATOMIC ENERGY AGENCY, Development of Procedures for In Vivo
Dosimetry in Radiotherapy, Human Health Reports No. 8, IAEA, Vienna (2013).
[360] BRITISH INSTITUTE OF RADIOLOGY, Central Axis Depth Dose
Data for Use in
Radiotherapy, BJR Supplement 25 (1996)
[361] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Dose and Volume Specification for Reporting Intracavitary Therapy in
Gynecology, ICRU
Report 38, ICRU Publications, Bethesda, Maryland (1985).
[362] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Dose and volume specification for reporting interstitial therapy, ICRU
Report 58, ICRU
Publications, Bethesda, Maryland (1997).
[363] INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS,
Dosimetry of Beta Rays and Low-Energy Photons for Brachytherapy with
Sealed Sources
ICRU Report 72 (2004).
[364] PEREZ-CALATAYUD J, BALLESTER F, DAS RK, DEWERD LA, IBBOTT
GS,
MEIGOONI AS, OUHIB Z, RIVARD MJ, SLOBODA RS, WILLIAMSON JF.
Dose
calculation for photon-emitting brachytherapy sources with average energy
higher than 50 keV:
Report of the AAPM and ESTRO, AAPM Rep No. 229, Med. Phys. 39 (5) (2012).
[365] BEAULIEU L, CARLSSON TEDGREN A, CARRIER JF, DAVIS SD, MOURTADA
F,
RIVARD MJ, THOMSON RM, VERHAEGEN F, WAREING TA, WILLIAMSON JF. Report
of the Task Group 186 on model-based dose calculation methods in brachytherapy
beyond the
TG-43 formalism: Current status and recommendations for clinical
implementation, Med. Phys.
39 (10) (2012).
[366] CHIU-TSAO ST, SCHAART DR, SOARES CG, NATH R. AAPM
Therapy Physics
Committee Task Group No 149. Dose calculation formalisms and consensus
dosimetry
parameters for intravascular brachytherapy dosimetry: Recommendations
of the AAPM
Therapy Physics Committee Task Group No. 149, Med. Phys. 34 (11) (2007).
[367] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, AAPM Recommendations on
Dose Prescription and Reporting Methods for Permanent Interstitial
Brachytherapy for Prostate
Cancer, Report of AAPM Task Group 137, AAPM Report No. 137 (2009).
[368] CHIU-TSAO ST, ASTRAHAN MA, FINGER PT, FOLLOWILL DS, MEIGOONI AS,
MELHUS CS, MOURTADA F, NAPOLITANO ME, NATH R, RIVARD MJ, ROGERS DW,
THOMSON RM. Dosimetry of ¹²⁵I and ¹⁰³Pd COMS eye plaques for intraocular
tumors: Report
of Task Group 129 by the AAPM and ABS, Med. Phys. 39 (10) (2012).
[369] RIVARD MJ, COURSEY BM, DEWERD LA, HANSON WF, HUQ MS, IBBOTT GS,
MITCH MG, NATH R, WILLIAMSON JF. Update of AAPM Task Group No. 43 Report: A
revised AAPM protocol for brachytherapy dose calculations, AAPM Report
No. 84, Med.
Phys. 31 (3) (2004).
[370] Nath R, Amols H, Coffey C, Duggan D, Jani S, Li Z, Schell M, Soares C,
Whiting J, Cole PE,
Crocker I, Schwartz R. Intravascular brachytherapy physics: Report of
the AAPM Radiation
Therapy Committee Task Group No. 60, AAPM Report No. 66, Med. Phys. 26 (2)
(1999).
[371] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Dosimetry of
Interstitial
Brachytherapy Sources: Recommendations of the AAPM Radiation Therapy Committee
Task
Group No. 43, Med. Phys. 22: 209–234 (1995).
[372] MURPHY MJ1, BALTER J, BALTER S, BENCOMO JA JR, DAS IJ, JIANG SB, MA CM,
OLIVERA GH, RODEBAUGH RF, RUCHALA KJ, SHIRATO H, YIN FF. The management
of imaging dose during image-guided radiotherapy: Report of the AAPM
Task Group 75,
AAPM Report No. 95, Med. Phys. 34 (10) (2007).
[373] F SANCHEZ-DOBLADO, C DOMINO, F GOMEZ, ET AL, Estimation of
neutron-equivalent
dose in organs of patients undergoing radiotherapy by the use of a novel online
digital detector,
Phys. Med. Biol. 57: 6167–6191 (2012).
[374] FOWLER J. F., 21 years of Biologically Effective Dose, The British
Journal of Radiology, 83
(2010), 554–568.
[375] INTERNATIONAL ATOMIC ENERGY AGENCY, Specification and Acceptance Testing
of
Radiotherapy Treatment Planning Systems, IAEA TECDOC 1540, IAEA, Vienna (2007).
[376] INTERNATIONAL ATOMIC ENERGY AGENCY, Commissioning of
Radiotherapy
Treatment Planning Systems: Testing for Typical External Beam Treatment
Techniques, IAEA
TECDOC 1583, IAEA, Vienna (2008).
[377] DEPARTMENT OF HEALTH. Report of the independent inquiry commissioned by
the West
Midlands Regional Health Authority into the conduct of isocentric
radiotherapy at the North
Staffordshire Royal Infirmary between 1982 and 1991. London: Department
of Health and
Social Security. Library and Information Service, (1992).
[378] INTERNATIONAL ATOMIC ENERGY AGENCY, Investigation of an Accidental
Exposure
of Radiotherapy Patients in Panama, IAEA, Vienna (2001).
[379] SCOTTISH EXECUTIVE. Unintended overexposure of patient Lisa Norris during
radiotherapy
treatment at the Beatson Oncology Centre, Glasgow in January 2006, Report of an
investigation
by the Inspector appointed by the Scottish Ministers for The
Ionising Radiation (Medical
Exposures) Regulations 2000,
Scotland (2006).
http://www.scotland.gov.uk/Resource/Doc/153082/0041158.pdf
[380] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Prevention of
Accidental Exposures to Patients Undergoing Radiation Therapy, ICRP
Publication 86, Ann.
ICRP 30 (3) (2000).
[381] INTERNATIONAL ATOMIC ENERGY AGENCY, Accidental Overexposure of
Radiotherapy
Patients in San José, Costa Rica, IAEA, Vienna (1998).
[382] SOCIEDAD ESPAÑOLA DE FÍSICA MÉDICA, The Accident of the Linear
Accelerator in
the “Hospital Clínico de Zaragoza”, SEFM, Madrid (1991).
[383] EUROPEAN SOCIETY FOR THERAPEUTIC RADIOLOGY AND ONCOLOGY, Practical
Guidelines for the Implementation of a Quality System in Radiotherapy,
ESTRO, Brussels
(1999).
[384] EUROPEAN SOCIETY FOR THERAPEUTIC RADIOLOGY AND ONCOLOGY, Quality
assurance in radiotherapy, Radiother. Oncol. 35: 61–73 (1995).
[385] ESTRO Booklet No. 2 Recommendations for a Quality Assurance
Programme in External
Radiotherapy, http://www.estro.org/school/articles/publications/publications
[386] ESTRO Booklet No. 4 Practical Guidelines for the Implementation
of Quality System in
Radiotherapy, http://www.estro.org/school/articles/publications/publications
[387] ESTRO Booklet No. 6 Monitor Unit Calculation For High Energy
Photon Beams, Practical
Examples, http://www.estro.org/school/articles/publications/publications
[388] ESTRO Booklet No. 7 Quality Assurance of Treatment Planning Systems –
Practical Examples
for non-IMRT Photon Beams,
http://www.estro.org/school/articles/publications/publications
[389] ESTRO Booklet No. 8 A Practical Guide to Quality Control of
Brachytherapy Equipment,
http://www.estro.org/school/articles/publications/publications
[390] [ESTRO Booklet No. 9 Guidelines for
the verification of IMRT,
http://www.estro.org/school/articles/publications/publications
[391] GIBBONS JP, ANTOLAK JA, FOLLOWILL DS, HUQ MS, KLEIN EE, LAM KL, PALTA
JR, ROBACK DM, REID M, KHAN FM. Monitor unit calculations for external
photon and
electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71,
AAPM
Report No. 258, Med. Phys. 41 (3) (2014).
[392] Katja M. Langenb, Niko Papanikolaou, John Balog, Richard Crilly, David
Followill, S. Murty
Goddu, Walter Grant III, Gustavo Olivera, Chester R. Ramsey, Chengyu Shi, QA
for helical
tomotherapy: Report of the AAPM Task Group 148, Med. Phys. 37 (9), (2010).
[393] LANGEN KM, PAPANIKOLAOU N, BALOG J, CRILLY R, FOLLOWILL D, GODDU SM,
GRANT W 3RD, OLIVERA G, RAMSEY CR, SHI C; AAPM TASK GROUP 148.Quality
assurance of medical accelerators, Med. Phys. 36 (9), (2009)
[394] EZZELL GA, BURMEISTER JW, DOGAN N, LOSASSO TJ, MECHALAKOS
JG,
MIHAILIDIS D, MOLINEU A, PALTA JR, RAMSEY CR, SALTER BJ, SHI J, XIA P, YUE
NJ, XIAO Y. IMRT commissioning: Multiple institution planning and dosimetry
comparisons,
a report from AAPM Task Group 119, Med. Phys. 36 (11) (2009).
[395] MUTIC S, PALTA JR, BUTKER EK, DAS IJ, HUQ MS, LOO
LN, SALTER BJ,
MCCOLLOUGH CH, VAN DYK J; AAPM RADIATION THERAPY COMMITTEE TASK
GROUP NO. 66. Quality assurance for computed-tomography simulators and
the computed
tomography- simulation process: Report of the AAPM Radiation Therapy
Committee Task
Group No. 66, AAPM Report No. 83, Med. Phys. 30 (10), (2003).
[396] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Report of
the AAPM
Radiation Therapy Committee Task Group No. 59: High Dose Rate Brachytherapy
Treatment
Delivery, AAPM Report No. 61, AAPM, New York (1998).
[397] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Physical Aspects of
Quality
Assurance in Radiation Therapy, Rep. No. 13, American Institute of
Physics, New York
(1984).
[398] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE Report 93
Guidance for
Commissioning and QA of a Networked Radiotherapy Department, 2006 ISBN 1 903613
28 0
[399] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, Physics Aspects of
Quality
Control in Radiotherapy, Rep. 81, IPEM, York (1999).
[400] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE,
Commissioning and
Quality Assurance of Linear Accelerators, Rep. 54, IPEM, York (1990).
[401] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Commissioning and quality
assurance of X-ray–based image-guided radiotherapy systems, AAPM
Medical Physics
Practice Guideline 2.a, Journal of Applied Clinical Medical Physics,
Volume 15, Number 1,
2014. http://www.aapm.org/pubs/MPPG/documents/MPPG2a.pdf
[402] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Radiation Safety
Aspects of Brachytherapy for Prostate Cancer using Permanently
Implanted Sources. ICRP
Publication 98. Ann. ICRP 35 (3) (2005).
[403] Gary A. Ezzell, James M. Galvin, Daniel Low, Jatinder R. Palta,
Isaac Rosen, Michael B.
Sharpe, Ping Xia, Ying Xiao, Lei Xing, Cedric X. Yu, Guidance
document on delivery,
treatment planning, and clinical implementation of IMRT: Report of the IMRT
subcommittee
of the AAPM radiation therapy committee, AAPM Report No. 82, Med. Phys. 30 (8)
(2003).
[404] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, Medical Accelerator
Safety
Considerations: Report of AAPM Nuclear Medicine Committee Task Group 35, AAPM
Report
No. 56, AAPM (1996).
[405] Patient safety in external beam radiotherapy – 11 Guidelines on risk
assessment and analysis of
adverse events and near misses, ACCIRAD, http://www.accirad.eu
[406] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Prevention of
High-dose-rate Brachytherapy Accidents. ICRP Publication 97. Ann. ICRP 35 (2)
(2005).
[407] INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE Report 96 Guidance for
the
Clinical Implementation of Intensity Modulated Radiation Therapy (2008).
[408] THE ROYAL COLLEGE OF RADIOLOGISTS, Quality assurance practice
guidelines for
transperineal LDR permanent seed brachytherapy of prostate cancer, The
Royal College of
Radiologists, London,
(2012).
https://www.rcr.ac.uk/docs/oncology/pdf/BFCO%2812%294_QA_prostate.pdf
[409] THE ROYAL COLLEGE OF RADIOLOGISTS, SOCIETY AND
COLLEGE OF
RADIOGRAPHERS, INSTITUTE OF PHYSICS AND ENGINEERING IN MEDICINE, On
target: ensuring geometric accuracy in radiotherapy, The Royal College
of Radiologists,
London, 2008.
https://www.rcr.ac.uk/docs/oncology/pdf/BFCO%2808%295_On_target.pdf
[410] AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, AAPM TG230.
Medical
Physics Practice Guideline Task Group #3: The Development,
Implementation, Use and
Maintenance of Safety Checklists, new, not yet published
[411] INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Preventing
Accidental Exposures from New External Beam Radiation Therapy
Technologies. ICRP
Publication 112. Ann. ICRP 39 (4) (2009).
[412] INTERNATIONAL ATOMIC ENERGY AGENCY, The Radiological Accident in
Istanbul,
IAEA, Vienna (2000).
[413] INTERNATIONAL ATOMIC ENERGY AGENCY, The Radiological Accident in
Samut
Prakarn, IAEA, Vienna (2002).
[414] INTERNATIONAL ATOMIC ENERGY AGENCY, Accidental Overexposure of
Radiotherapy
Patients in Bialystok, IAEA, Vienna (2004).
[415] ASTRO. RO-ILS: Radiation Oncology Incident
Learning System™.
https://www.astro.org/Clinical-Practice/Patient-Safety/ROILS/Index.aspx
[416] NATIONAL PATIENT SAFETY AGENCY. National Reporting and
Learning Systems
(NRLS). https://report.nrls.nhs.uk/nrlsreporting/
[417] INTERNATIONAL ATOMIC ENERGY AGENCY, On-site Visits to Radiotherapy
Centres:
Medical Physics Procedures, Quality Assurance Team for Radiation
Oncology (QUATRO),
IAEA TECDOC 1543, IAEA, Vienna (2007).
[418] NUCLEAR REGULATORY COMMISSION, Consolidated Guidance about
Materials
Licenses: Program-Specific Guidance About Medical Use Licenses,
NUREG-1556, Vol. 9,
Revision 2, Office of Federal and State Materials and Environmental Management
Programs,
Washington, DC (2008).
[419] INTERNATIONAL ATOMIC ENERGY AGENCY, Method for Developing Arrangements for
Response to a Nuclear or Radiological Emergency, EPR-METHOD, IAEA, Vienna
(2003).